3. INSTALLATION INSTRUCTIONS
3-1. PRE-INSTALLATION REQUIREMENTS
A. General Requirements
(1) The contractor is to supply the necessary service and materials to complete the installation of
the satisfaction of the dentists and the installation engineer.
(2) This includes the supply and installation of the electric power supply cables with main isolating
switch and fuses, air supply piping, water supply piping, suction piping including vacuum pump
and its control wires and drain piping as noted on the installation
diagrams.
B. Setting Requirements
(1) The VOYAGER-II dental unit comprises a Chair section, Cuspidor section, Doctor unit section
and Light section.
(2) TheVOYAGER-II should be mounted taking the opening end of drain pipe into due consideration.
Refer to Page 5, Fig.3-1-1.
(3) The place on which the VOYAGER-II (approx. 210 kg) is set must have endurance force of
250kg/m .
(4) The setting position of the VOYAGER-II chair is shown in Fig.3-1-1 as recommended example.
C. Piping and Plumbing Requirements
(1) All pipings and conduits for cables are to be laid down under the floor and to come out from
the floor.
(2) The installation position and height from the floor of each pipe and cable conduit are shown
in Fig.3-1-1.
(3) The recommended sizes, materials and end piece are shown in Table 1.
-3-
Note : The suction pipe and drain pipe should be laid under the floor with an inclination of
1/200 - 1/400.
Air vacuum type does not require suction pipe, vacuum control wire and its conduit.
Item Material Size End Piece
Compressed Air Supply Pipe Shock Resistance Out. Dia.18mm PT1/2
P.V.C. Pipe HI-13 In. Dia. 13mm
Water Supply Pipe Shock Resistance Out. Dia. 18mm PT1/2
P.V.C. Pipe HI-13 In. Dia. 13mm
Suction Pipe P.V.C. Pipe VP-20 Out Dia. 26mm
IN. Dia. 20mm
Drain Pipe P.V.C. Pipe VP-50 Out. Dia. 58mm
In. Dia. 50mm
Power Supply Cable Conduit P.V.C. VE-16 In. Dia. 16mm
Vacuum Control Wire Coduit P.V.C. VE-16 In. Dia. 16mm
Table 1 The Recommended Sizes, Materials and End Piece of Pipes
2
(5) Do not install on wet floor.