CAUTION
・Before use, check that the nozzle will not come off from the syringe. Do not use if the syringe nozzle comes loose.
・Make sure that the nozzle is tightened securely and will not come off when pressing the spray button. Do not use if the
syringe nozzle comes loose.
・Do not subject the unit to shock. Doing so check that the nozzle will not come off from the syringe.
Cleaning must be done within 1 hour after use.
Replace with a new syringe nozzle and saliva ejector tip and do not take autoclave for following cases.
・Any waste material can not be removed by clogged hole.
・Contamination and solid material attached to vacuum handpiece and saliva ejector can not be removed.
・Sterilization by class B cycles.
・Skip the drying process if the temperature is to exceed 135°C.
・The cap, filter and body are made of resin. They may become deteriorated if they are sterilized in an autoclave many times.
・After autoclave sterilization, the cap, filter, body and valve are subject to discoloration, which does not have a negative
effect on performance.
・The slide knob can be autoclave 100 times and is expendable supplies.
・Sterilization temperature is 135 degrees or less.
・If damage occurs to the sterilization pouch, discard and sterilize again using a new pouch.
・For cleaning, do not use cleaning agents containing solvent or abrasives, thinners or oil-based alcohol
(butanol and isopropyl alcohol), which may cause cracks. Clean with a wet soft cloth or use a mild detergent diluted to an
approximately 10% solution. Wipe off water and residual disinfectant immediately.
・This could cause corrosion, damage or incorrect operation of the unit. They may facilitate rust generation.
Should drug solution or water come into contact with this unit, immediately wipe it off with a dry soft
towel, etc. This could result in defective function or electric leakage as well as spotting or rusting.
・Flow rate adjustment knobs for water and air are intended to increase/decrease the flow rate. Caution should be exercised
as tuning the knob excessively may cause it to turn free.
・There is cupfiller flow rate adjusting screw (pinch valve) for maintenance. The flow rate is adjusted when installing the unit.
・Do not spray liquids directly onto the surfaces of the holder. In order to prevent damage to electrical components and system.
Wipe off the holder surface with soft cloth moistened with cleaner or disinfectant. Wipe off with dry soft cloth to dried up
of the holder after cleaning and disinfection.
・Completely dried up the cleaning solution before use the product.
23. Precautions for after using of the syringe
Clean the filter with running water. Do not use brush to clean the filter which may damage to the filter.
・After use, wipe off syringe nozzle, body contaminated with blood or saliva and clean it with cleaning agent as necessary.
・Sterilization with autoclave, be sure to use sterilization pouch.
・For cleaning of other syringes, observe the instructions shown in the package inserts and instruction Manuals included with
the syringes. Before use, be sure to carefully read the package inserts and instruction manuals accompanying the various
pieces of equipment (including optional articles) to ensure proper use.
24. Precautions after the cleaning of syringe
25. Precautions for cleaning of the solid corrector
26. Precautions for cleaning the syringe nozzle, saliva ejector tip
27. Use of flow rate adjustment knobs
28. Precautions for sterilization of syringe nozzle, saliva ejector tip
29. Precautions for cleaning the product exterior
30. Immediately wipe off drug solution when it comes into contact with this unit
31. Precautions for cleaning the Dr. instrument holder and the assistant holder