Benchmark Orbi-Shaker CO2 User manual

Operations Manual
Index:
I. General Information
II-1. Introduction
II-2. Installation
II-3. Specifications
II-4. Features
III-1, Operation
III-2. Platform Installation
III-3. Error Messages
III-4. Maintenance
IV-1. Service and Contact
I. General Information
NOTE : Before operating the Orbi-Shaker CO2 please take a few minutes to
familiarize yourself with the contents of this manual.
Thank you for purchasing the Orbi-Shaker CO2. This operations manual
contains instructions for the care of this Instrument. In order to use the
instrument properly, please read this manual carefully before operation.
Please keep it available for future use.
Prior to the initial operation:
Please check the Instrument and the accessories against the packing list when
you first open the shipping carton. If you find there is damage, a discrepancy,
or missing items, please contact your supplier immediately.
II-1. Introduction
The Orbi-Shaker CO2 is designed for use in extreme environments, including CO2
incubators. The product is ideal for many applications used in general research
laboratories, including cell culture, bacteria suspension, etc.
The Microprocessor’s constant monitoring system ensures accurate shaking speed.
The beltless drive system minimizes noise and vibration while increasing the life span
of the unit. Optional accessories are available for compatibility with a wide variety of
laboratory vessels, including flasks, tubes, dishes and trays.
II-2. Installation
-Choose a location for the shaking mechanism.
-To assure proper ventilation, allow a minimum of 2 inches (5 Cm) of the clearance
between the rear and side of the unit and the adjacent wall.
-Always ensure that the machine is installed on a level surface.
-Once the machine is placed in the desired location, attach the flat cable to the
remote controller.
-Place the remote controller in the desired location. (Ensure that the remote
controller is located in a dry area.
-Before plugging the remote controller into the outlet, ensure that the electrical
specifications (listed on the serial number label on the rear side of the controller)
matches the electrical specifications of your outlet.
-ALWAYS: Connect the flat cable from the remote controller to the shaking
mechanism before connecting the controller to the electrical outlet.
II-3. Specification
1) Dimension (Exterior) : 30 (W) X 34 (D) X 9 (H) cm / 11.8 (W) x 13.5 (D) x 3.5 (H) in.
(Platform): 30 (W) x 33 (D) cm / 12(W) x 13(D) in.
2) Speed range : 30 - 300 RPM
3) Shaking Orbit: 19mm (3/4”)
4) Timer : Digital - 1 min. to 47hr 59min. (1 min. increments)
Or continuous
5) Motor : Brushless DC - Beltless
6) Weight: Shaker: 25 lbs / 11.5 kg
Controller: 5 lbs / 2.3 kg
7) Electrical: 120V or 230V, 50-60Hz / 200W
8) Operating Environment: 0 to 60℃
II-4. Features
1) Constant Monitoring System (CMS) ensures precise control over speed
2) Sealed shaking mechanism for use in extreme environments (humidity up to 100%)
3) Controller allows for speed/time adjustment without opening door
4) MAGic Clamp platform available as optional accessory.
5
) Blushless DC motor
and beltless drive system minimizes noise and
vibration
III-1. Operation
1) Install the Shaker on a level surface. (If installing in an incubator, it is
recommended to start on the lowest available section of the incubator.)
2) Place the power switch (located on the rear of the controller) in the "ON" position
(the display will illuminate).
3) To set the desired time, Press and hold the "RPM/TIME" key for approximately 1
second until LED displays "t” followed by the current set time. Press the "SET"
key located underneath the time/speed display. The set time is displayed and will
blink continuously. It can be adjusted by using the UP (▲) and DOWN (▼) keys.
Once the desired time is displayed, press the “SET” key to confirm.
* If you do not press the "SET" key to store the desired time the new time will not
be saved and the set time will return to the prior setting.
3
) To set the desired mixing speed Press and hold the
"
RPM/TIME
"
key for
Running
Pilot Lamp
A
d
j
ustment Button Run or Stop
RPM/TIME
Select Button
Setting/Save
III-2. Platform Installation
The Orbi-Shaker includes a flat non-slip platform for low speed operation with many
common vessels, including trays and dishes. For installation of the optional MAGic
Clamp platform, please follow the instructions below:
1.) Unscrew the four metal thumb screws from the corners of the internal platform
and remove the green rubber mat.
2.) Unscrew the four center screws and two additional screws (using a Phillips
screw driver.
3.) Remove the included platform and apply the MAGic Clamp platform.
4.) Screw in the six screws that were previously removed from the included
platform.
When installing the microplate platform (H1000-P-MP) or universal spring platform
(H1000-P-SP), please follow the simple instructions below:
1.) Unscrew the four metal thumb screws from the corners of the included platform.
2.)Apply the desired platform lining up the four corner holes and screw in the 4
thumb screws.
III-3. Error Messages
1) This message “rErrL” represents a low speed speed error
-Check the platform to ensure that nothing is obstructing it’s movement
2) This message “rErrH” represents a high speed error.
- Removal of load or partial load.
Solution : If a vessel has been removed, replace the vessel or restart the machine.
3) This message, “Error” represents an electrical error.
- Machine has lost power during the run.
Solution : Check that the power cord is properly installed and restart or check that
the circuit has not been interrupted and restart.
4.) Display will not illuminate
(1) Electrical connection - Please disconnect main power cord and connect again.
(2) Voltage supply - Please check to voltage.
(3) Check the fuse
(4) Verify voltage on the serial number label of the machine
(5) Defective power switch (check voltage at switch)
III-4. Maintenance:
The platform of the Orbi-Shaker CO2 should be cleaned regularly with a soft
cloth dampened with water or alcohol. (Always ensure that the machine has
been powered down prior to cleaning.
In the event of a spillage: Power down the unit, remove the platform and dry the
internal chamber with a soft cloth. DO NOT attempt to operate the unit until the
spillage had been properly dried.
IV-4. Service and Contact
Service on the Orbi-Shaker CO2 should only be performed by qualified service
personnel. To request service or technical support, please contact Benchmark
Scientific or your local Benchmark Scientific representative.

Other manuals for Orbi-Shaker CO2

1

Other Benchmark Laboratory Equipment manuals

Benchmark H2020 User manual

Benchmark

Benchmark H2020 User manual

Benchmark Orbi-Shaker User manual

Benchmark

Benchmark Orbi-Shaker User manual

Benchmark Micro Bead Sterilizer B1201 User manual

Benchmark

Benchmark Micro Bead Sterilizer B1201 User manual

Benchmark R4040 User manual

Benchmark

Benchmark R4040 User manual

Benchmark MyBath B2000-2 User manual

Benchmark

Benchmark MyBath B2000-2 User manual

Benchmark Orbi-Shaker User manual

Benchmark

Benchmark Orbi-Shaker User manual

Benchmark TC 9639 User manual

Benchmark

Benchmark TC 9639 User manual

Benchmark R2020 User manual

Benchmark

Benchmark R2020 User manual

Benchmark BSH1001 User manual

Benchmark

Benchmark BSH1001 User manual

Benchmark Orbi-Shaker CO2 User manual

Benchmark

Benchmark Orbi-Shaker CO2 User manual

Benchmark MyBath B2000-2 User manual

Benchmark

Benchmark MyBath B2000-2 User manual

Popular Laboratory Equipment manuals by other brands

Hettich ROTOFIX 46 operating instructions

Hettich

Hettich ROTOFIX 46 operating instructions

Hettich ROTINA 35 R Repair instructions

Hettich

Hettich ROTINA 35 R Repair instructions

Selecta PRESOCLAVE III 50L manual

Selecta

Selecta PRESOCLAVE III 50L manual

Biomeme Franklin three9 user manual

Biomeme

Biomeme Franklin three9 user manual

LW Scientific BioVID instruction manual

LW Scientific

LW Scientific BioVID instruction manual

Thermo Scientific iCAP RQ ICP-MS Pre-installation requirements guide

Thermo Scientific

Thermo Scientific iCAP RQ ICP-MS Pre-installation requirements guide

Diesse MINI-CUBE user manual

Diesse

Diesse MINI-CUBE user manual

Agilent Technologies Varian 325-MS LC/MS Pre-installation instructions

Agilent Technologies

Agilent Technologies Varian 325-MS LC/MS Pre-installation instructions

Thermo Scientific Labofuge 200 Operating lnstructions

Thermo Scientific

Thermo Scientific Labofuge 200 Operating lnstructions

BioLAB BCFL-203 Operation manual

BioLAB

BioLAB BCFL-203 Operation manual

ZINEXTS ZiXpress 64 system instruction manual

ZINEXTS

ZINEXTS ZiXpress 64 system instruction manual

Carestream INDUSTREX M43ic user guide

Carestream

Carestream INDUSTREX M43ic user guide

4titude 4s3 Operation manual

4titude

4titude 4s3 Operation manual

Sony FilmStation UP-DF750 Setup and user's manual

Sony

Sony FilmStation UP-DF750 Setup and user's manual

VERDER Carbolite Gero GP 450 Installation, operation and maintenance instructions

VERDER

VERDER Carbolite Gero GP 450 Installation, operation and maintenance instructions

Elmi CM-7S Plus user manual

Elmi

Elmi CM-7S Plus user manual

MRC AM-4214SD operating manual

MRC

MRC AM-4214SD operating manual

Micromeritics AquaPrep II 056 Operator's manual

Micromeritics

Micromeritics AquaPrep II 056 Operator's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.