Bestcare PL350CT User manual

BESTLIFT®
FULL BODY PATIENT LIFT
Model: PL350CT
Ensure the product has been assembled according to the instructions in this
manual. All operators should read and understand the instructions for safe and
proper operation of the patient lift.

1
Date of Manufacture
BestLift® PL350CT
MAX LOAD: 350LB
S/N: PL350CTYYMM1XXX
2920 Pacific Dr,
Norcross, GA 30071
1-678-679-6690
MASS OF LIFT:95LB(43KG)
Thank you for choosing Bestcare!
To better serve you, please record the following information for
future use:
Supplier Name: ___________________________________________
Supplier Telephone: ________________________________________
Product Serial Number: _____________________________________
Date of Purchase: _________________________________________
This product has been tested by SGS, an accredited testing institute to comply with
the requirements of:
ISO 10535, IEC 60601-1, ANSI/AAMI ES60601-1, CAN/CSA-C22.2 No.60601-1,
IEC 60601-1-11, ANSI/AAMI HA60601-1-11, CAN/CSA-C22.2 No.60601-1-11,
IEC 60601-1-2
800519
The manufacture date can be found on the serial number
label that is located at the top of the mast. The YY in the
serial number is the manufacture year, and the MM is the
manufacture month.

2
i
i
Product Description
Intended Use
The Bestlift® PL350CT allows the patient to be lifted and
transferred safely with minimal physical effort provided by the
operator. This is due to it being a full electric lift . The PL350CT
uses a special clip sling that will make transferring the patient
more comfortable. The PL 350CT is also foldable and can be
stored away in small spaces and ts in the trunk of your car.
With the PL350CT patients can be lifted from oor to bed with
ease. Additionally, patients can be moved in and out of the car
effortlessly as a result of our special spreader bar. With the
addition of the optional battery pack you will never have to be
worried about your lift not being charged and ready for use.
The PL350CT is intended to be used with patients who are
completely immobile or totally dependent on the patient lift for
transfers. The PL350CT is intended to be used for transfers
of the patient in and out of car, bed, their wheelchair, to and
from the commode, or any other type of surface. The special
spreader bar is intended to make transferring patients in and
out of the car easier. The PL350CT is foldable and can t in
small spaces as well as the trunks of most cars. The PL350CT
is intended to be operated by one caregiver who will perform
the lift and transfer. Transfers should always be done on a at
dry surface and are not intended to be used in high humidity
environments for extended amounts of time.

3
DEFINITIONS & SYMBOLS
In this manual the user refers to the patient or resident and may be used
interchangeably at different times. Caregiver refers to the operator or person
who is assisting with the transfer.
Symbols used in this manual and on the product and their meanings:
Warning! Failure to heed this warning may
result in damage to the product or serious
injury to the operator and/or user.
Do Not Bleach. No bleach product may be
used. The garment is not colorfast or
structurally able to withstand any bleach.
Important instructions follow. Read and
understand the instructions in the manual
before using the product.
Do Not Dryclean. Garment may not be
commercially drycleaned.
Type BF applied part. Refer to the part of
the medical device which come into
physical contact with the patient in order
for the device to carry out its intended
Do Not Iron. Item may not be smooted or
finished with an iron.
Class II symbol. To identify equipment is
designed that a connection to electrical
earth (ground) is not required.
Tumble Dry, Normal, Low Heat. A machine
dryer may be regularly used at a maximum
of Low Heat setting.
Note! Important information concerning
the product and/or its correct and proper
usage follows.
Do Not Dry. A machine dryer may not be
used.
Advise the operator to consult the
accompanying documents before use.
Machine Wash, Hot. Initial water
temperature should not exceed 82°C or
180°F.
Right side up Do Not Wash
Fragile, handle with care Keep dry
Carton storage temperature limit of -25°C
to 70°C Carton storage humidity limit of 0% to 90%
European Conformity mark, The product
conforms to European Medical Directive
93/42/EEC and meets applicable health,
safety and environmental requirements.
Handle with care

4
iSafety Instructions
A BestLift® will provide years of service if it is properly maintained as
any electric and/or mechanical equipment requires. Please pay careful
attention to the following important information regarding the care,
maintenance, and operation of the patient lift. Carefully read these
instructions before assembling the lift, or attempting to lift a user with
the device.
• Before attempting to lift any patient, please practice operating the
lift and explain the lift procedures to the patient
• Please provide special assistance under the caretakers’ discretion.
(i.e. patients who are comatose, spastic, agitated, or otherwise
severely handicapped.)
• The patient lift should be used solely for transferring a user/patient
from one utility (beds, chairs, toilets, etc.) to another. The patient lift
should not be used for transporting or moving any patient from one
location to another location.
• During lifting or lowering, whenever possible, always keep the base
of the lift in the widest position.
• The base of the lift should be closed before moving the lift.
• Do not roll casters over any object while the user/patient is in the
sling.
• Do not lock casters during lifting.
• While being lifted in a sling, always keep the user/patient centered
over the base and facing the caregiver operating the lifter.
• Never leave the user/patient unattended during lifting.
Service personnel may contact Bestcare for any information,
instructions, or certain parts for servicing purpose.

WARNING!
5
• NO modication of this equipment is allowed. It is unsafe to modify the
equipment from its original state. This product is deemed safe to use with
only the original parts as delivered. Unauthorized use of any modied
equipment may result in serious injury or death.
• DO NOT replace any components of the lift without consulting with
Bestcare and must follow proper instruction from Bestcare when replacing
any components.
• DO NOT service any parts of the lift while in use with a patient.
• It is NOT safe to load the equipment more than the indicated maximum safe
working load.
• IF any accessories with lower load capacity than the equipment itself is used,
DO NOT load the equipment more than the lowest maximum safe
working load.
• REMOVE the battery pack if the lift will not be used for some time, to prevent
battery leakage. PERIODICALLY check the battery pack and the charger
• This equipment DOES NOT consist of any materials or ingredients to
which patients are exposed, that may cause bodily harm or injury.
• Prolonged exposure or physical contact with any part of the equipment WILL
NOT cause skin irritation, allergic reaction, etc.
• Using accessories, detachable parts, or materials not described in the
instruction manual MAY RESULT IN SERIOUS INJURIES.
• Keep the equipment away from children and pets when in use. IT MAY
CAUSE SERIOUS INJURY OR DEATH.
• Keep the equipment away from any pests or pest infested area. Diseases
transferred from pests to the equipment MAY CAUSE SERIOUS ILLNESS
to patients upon contact.
• If the lift is exposed to the sun for long periods of time, be CAUTIOUS as the
surface of the lift may be hot.
• There are no risk if the lift is exposed to reasonably foreseeable
environmental conditions
• CAUTION! The controller contains parts and assemblies that could be
susceptible to damage by electrostatic discharge (ESD). To reduce
the likelihood of this, one should be grounded prior to using the controller.
• DO NOT use the lift on a inclined surface that is greater than 5°.

6
FEATURES & OVERVIEW
iPrior to assembly, unpack and inspect all parts from the shipping carton.
Contact you dealer immediately if any parts are damaged or missing.
5
3
8
12
9
2
6
7
1
10
11
4
UNLESS OTHERWISE SPECIFIED:
SHEET:
SCALE:
PART NUMBER:
DESCRIPTION:
DATE:
DRAWN BY:
MATERIAL:
FINISH:
2920 PACIFIC DRIVE
NORCROSS, GA 30071
PROPRIETARY AND CONFIDENTIAL
THE INFORMATION CONTAINED IN THIS
DRAWING IS THE SOLE PROPERTY OF
BESTCARE LLC. ANY REPRODUCTION IN PART
OR AS A WHOLE WITHOUT THE WRITTEN
PERMISSION OF BESTCARE LLC IS PROHIBITED.
9/8/2017
1 of 1
DaTong Wrench
ALL DIMENSIONS IN MM
1-6
6-30
30-120
120-400
0.1
0.2
0.3
0.5
Easy Tools Screw Driver
8.4 lb 0.7 lb
(3.8 kg) (0.3 kg)
7.5 lb
(3.4 kg)
11 lb
(5 kg)
Emergency Lowering
2Actuator
5Spreader Bar
6Rear Caster 10
11
4
Hand Control
8
Mast
12 Emergency Stop Button
1
Combined
Weight of
66 lb
(30 kg)
**Base
3
Control Unit
7Front Caster
Leg Opening Pedal
9
Boom
** Heaviest detachable part
Class II Equipment

mm (in)
A1729.2 (68.1)
B1070.8 (42.2)
C658.4 (25.9)
D492.2 (19.4)
E823.2 (32.4)
F766.9 (30.2)
G576.7 (22.7)
H333.3 (13.1)
I496.9 (19.6)
J275 (10.8)
K442.7 (17.4)
L118.5 (4.7)
M20.3 (0.8)
PL350CT
A
B
C
600
D
E
700
G
H
F
M
L
I
J
K
7

8
SPECIFICATIONS & OPTIONS
C
B
H
E
F
G
D
A
Standard Features
1
Foldable Lift Frame
2
4-Points Dynamic Spreader Bar
3
4" Rear Caster with Brake
4
3" Front Caster
5
Power Control Unit
Optional Features
6
Charging Cradle
7
Second Battery pack
Dimensions & Specifications
Model
PL350CT
Safe Working Load
350 lb
159 kg
ABase Length
40.7"
1033.8 mm
BBase Ext. Width - Closed
25.5"
648 mm
CBase Int. Width - Opened
32.6"
828.8 mm
DMast Height
47"
1192.2 mm
EFront Caster Height
4.7"
118.5 mm
FLifting Height - Lowest
21.2"
537.22 mm
GLifting Height - Highest
63.3"
1608 mm
H Turning Diameter
55.7”
1414 mm
Specications of Control Unit
Input Voltage 220-240VAC 50Hz / 110-120VAC 60Hz
Output Voltage 24 VDC
Battery Pack 24 VDC 5AH
Lift Control Handset and Panel
Protection Class IP54 (Protection against dust and splash of water)
Sound Level Under Full Load 16db
Sound Level Under No Load 11db
Lift Cycles Per Charge 25 full cycles under full load
Overload Warning by LCD
Battery Status by LCD
Duty Cycle 10% Duty cycle (Max 2 minute ON,18 minute OFF)
PL350CT
Lift Weight Unloaded (lb) 95
Lift Weight Unloaded (kg) 43
* Bestcare is committed to continuous improvements of our products therefore
the specication, dimensions, and features listed above are for guidance only
and are subject to change without prior notice.

9
Step 2: Secure the stand to the mast. Align the hole on the stand
bracket with the lower hole on the mast. Make sure the tab on the
stand bracket insert into top hole on the mast, and insert the screw to
secure the stand bracket as follows.
Step 3: Remove the quick release pin that is securing the mast.
Step 4: Lift the mast upright.
Setup Your PL350CT for First Use
Step 1: Remove folded lift from box.

10
Step 5: Insert the pin to secure the mast.
Step 6: Remove the pin that is securing the boom.
Step 7: Remove lower actuator pin on the mast.
Step 8: Move the boom out of the way, and install the lower
end of the actuator to the mast using a pin and retaining ring
as follows.

Step 9: Install the upper end of actuator to the boom by
reinsert the quick release pin to secure it.
Step 10: Insert the spreader bar into the tube on the boom as
follows. Align the holes and secure the spreader bar using a
pin and retainer ring.
WARNING! The tube on the boom and the
spreader bar are intended to be hard to adjust.
Attached the spreader bar and use the handle on the
spreader to adjust if needed.
11

11 12
Power Control Unit Installation
4
3
21
5
M5 SCREWS
M4 SCREWS
UNLESS OTHERWISE SPECIFIED:
SHEET:
SCALE:
PART NUMBER:
DESCRIPTION:
DATE:
DRAWN BY:
MATERIAL:
FINISH:
2920 PACIFIC DRIVE
NORCROSS, GA 30071
PROPRIETARY AND CONFIDENTIAL
THE INFORMATION CONTAINED IN THIS
DRAWING IS THE SOLE PROPERTY OF
BESTCARE LLC. ANY REPRODUCTION IN PART
OR AS A WHOLE WITHOUT THE WRITTEN
PERMISSION OF BESTCARE LLC IS PROHIBITED.
9/27/2017
1 of 1
M4x0.7x10mm Machine Screw
ALL DIMENSIONS IN MM
1-6
6-30
30-120
120-400
0.1
0.2
0.3
0.5

Fold Up The Lift For Storage or Transport
Step 1: Remove The Spreader Bar
WARNING: actuator should be fully retracted (boom in
lowest position) prior to getting ready for storage
Step 2: Detach The Actuator
13
1 2
1 2

13
Step 3: Fold Up The Boom
Step 4: Fold Up The Mast
14
1 2
3
12
3

Standing Up
Lying Down
15
Warning - The lift may fall and cause bodily injury if it is not secured against the wall.

16
Operating Instructions
i
Double check all assemblies for tightness and read operating
instructions carefully prior to use.
PREPARATION BEFORE LIFTING
• Twist the RED EMERGENCY STOP BUTTON clockwise.
• Verify the Lifting and Lowering Functions by pressing either the UP or DOWN
button on the hand control once.
• Position the base of the lift around or under the object.
• Widen the base and disengage the caster brakes.
Do not attempt to transfer a patient without prior approval of the person’s nurse.
Also, do not transfer without having studied the instructions and performed
several practices in operating the product. Together (with the patients doctor,
nurse, or medical attendant) select a sling that is both practical and comfortable.
The sling selected should be one that serves the needs of the patient, while
providing the patient with optimal safety. Never interfere with the lift, unless
instructed by the attendant. Have a doctor, nurse, or medical attendant
(experienced in the use of the BestStand) present during the rst few times the
lift is used to transfer a new user.
i
For optimum performance the lift should be transported and stored in
following condition range:
• -25°C to +5°C (-13°F to 41°F), and
• +5°C to +35°C (41°F to 95°F) at a non-condensing relative humidity 0% to 90%
• +35°C to +70°C at a water vapor pressure up to 5 kPa
For optimum performance the lift should be operated in following condition
range:
• at a temperature range of +5°C to +40°C (+41°F to +104°F)
• at a non-condensing relative humidity range of 15 % to 90 %, but not requiring a
water vapor partial pressure greater than 5 kPa
• 70 kPa to 106 kPa atmospheric pressure
The lift will not have any potential electromagnetic or other interference with
other devices.

Power Control Unit Overview
17
1. Connect Actuator as shown above.
2. Connect Hand Control as shown above.
Actuator
Battery Pack
Release Handle
Emergency Stop Button
Up / Down Button
LCD Display Panel
Battery Pack
Charger
DC In
Not Used
Hand Control
Charging Cradle
(Optional)
Additional Battery Pack
(Optional)
WARNING!
• Contains a long charger cord that may cause strangulation if
used improperly
• Keep long charger cord out of walkway where it poses a
tripping hazard
• Have the charger port on the controller facing away from the
wall for easy disconnection

DC Charger
(included)
LED Indicator
* Amber = charging
* Green = charging completed
Charger Plug-in
LED Indicator
* Green = power connected
* Blank = power disconnected
Charging Cradle
(Optional)
LCD Display Panel Signs - Charging
During charging, the LCD Display Panel will show one of the four signs below.
• The DC Charger must be connected.
• The emergency stop button must be released.
• The battery will not be charged if the emergency button is pressed in.
LCD Display Panel Signs
When the emergency stop button is released, the LCD Display Panel will show one of the four signs
below.
• The sign will display for 5 seconds.
• Then the lift will go into standby mode and the sign disappears.
• The lift is ready for use.
Charging: DC Charger or Charging Cradle
Release Pressed in
18

19
LED Indicator
* Green = In Use
* Blank = Standby
Lifting
Lowering
Operating Lift: Using Hand Control
Operating Lift: Using LCD Display Panel
Other manuals for PL350CT
1
Table of contents
Other Bestcare Medical Equipment manuals

Bestcare
Bestcare BESTLIFT PL400 User manual

Bestcare
Bestcare BESTLIFT PL400 User manual

Bestcare
Bestcare BestLift 400 Series User manual

Bestcare
Bestcare BestLift Luna User manual

Bestcare
Bestcare SA400HE User manual

Bestcare
Bestcare 33232 User manual

Bestcare
Bestcare BESTLIFT PL600 User manual

Bestcare
Bestcare BESTSTAND SA500 User manual

Bestcare
Bestcare BestStand 600 Series Programming manual

Bestcare
Bestcare BESTSTAND SA228 User manual