Storage
0483
Finger, hand or body is moving.
People with a low oxygen saturation value are more likely to experience the following symptoms: shortness of
breath, increased heart rate, weakness, nervousness and outbreaks of sweating. If oxygen saturation is known
to be chronically diminished, it requires monitoring using the pulse oximeter under medical supervision. If you
have acutely diminished oxygen saturation, with or without the accompanying symptoms, you must consult a
doctor immediately as it could lead to a life-threatening situation. The pulse oximeter is particularly suitable for
patients at risk such as people with heart disease or asthma, but also for athletes and healthy people who
exercise at high altitude (e.g. mountaineers, skiers or amateur pilots).
evaluated at the measurement site. Rather, they are exclusively used to display the current visual signal varia
tion at the measurement site and do not enable reliable diagnostics for the pulse.
Operating
98 65
98 65
Pulse oximeter
ENGLISH
Notes on electromagnetic compatibility • The device is
suitable for use in all environments listed in these instructions for use, including domestic
• The pulse oximeter displays an instantaneous measurement but cannot be used for continuous monitoring. •
The pulse oximeter does not have an alarm function and is therefore not suitable for evaluating medical
• There must not be any nail varnish, artificial nails or other cosmetics on the finger to be measured. •
Ensure that the finger nail on the finger to be measured is short enough that the fingertip covers the sensor
elements in the housing.
• Avoid using this device directly next to other devices or stacked on top of other devices, as this could lead to
faulty operation. If, however, it is necessary to use the device in the manner stated, this device as well as
the other devices must be monitored to ensure they are working properly.
• Keep your hand, finger and body steady during the measurement.
• For people with cardiac arrhythmia, the measurement values of SpOÿ and the heart rate may be incorrect
or the measurement may not be possible at all.
• Protect the pulse oximeter from dust, shocks, moisture, extreme temperatures and explosive materials.
• Do not self-diagnose or self-medicate on the basis of the measurements without consulting your doctor.
• The use of accessories other than those specifed or provided by the manufacturer of this device can lead to
an increase in electromagnetic emissions or a decrease in the device’s electromagnetic immunity; this can
result in faulty operation.
• The displays for the pulse wave and pulse bar do not allow the strength of the pulse or circulation to be
• In cases of carbon monoxide poisoning, the pulse oximeter displays a measurement value that is too high.
• Keep portable RF communication devices (including peripheral equipment, such as antenna cables or ex
ternal antennas) at least 30 cm away from all device parts, including all cables included in delivery. Failure
to comply with the above can impair the performance of the device.
• To avoid falsifying the measurement, there should not be any strong light sources (e.g. fluorescent lamps or
• Failure to comply with the above can impair the performance of the device.
• Check to ensure that the package contains all the parts that should be included in the delivery.
• People with low blood pressure, who suffer from jaundice or take medication for vascular contraction, may
experience incorrect or falsified measurements.
• The use of the device may be limited in the presence of electromagnetic disturbances. This could result in
• Check the pulse oximeter regularly before use to ensure that there is no visible damage to the device and
the batteries are still sufficiently charged. In case of doubt, do not use the device and contact Beurer
customer services or an authorised retailer.
• This device conforms with the European standards EN60601-1 and EN60601-1-2 (in accordance with
CISPR 11, IEC 61000-4-2, IEC 61000-4-3, IEC 61000-4-8) and is subject to particular precautions with
regard to electromagnetic compatibility. Please note that portable and mobile HF communication systems
may interfere with this device. For more details, please contact our Customer Services at the address
indicated.
• Do not look directly inside the housing during the measurement. The red light and the invisible infra-red
light in the pulse oximeter are harmful to your eyes.
• Do not use any additional parts that are not recommended by the manufacturer or offered as equipment.
• Incorrect measurements are likely for patients who have been administered medical dye in the past or for
those who have abnormal haemoglobin levels. This applies in particular for cases of carbon monoxide
poisoning and methaemoglobin poisoning, which can occur for example from the administration of local
anaesthetics or from an existing methaemoglobin reductase deficiency.
• Using the pulse oximeter for long periods may cause pain for people with circulatory disorders. Therefore
• This device is not intended for use by people (including children) with restricted physical, sensory or mental
skills or a lack of experience and/or a lack of knowledge, unless they are supervised by a person who has
responsibility for their safety or they receive instructions from this person on how to use the device.
• Under no circumstances should you open or repair the device yourself, as faultless functionality could no
longer be guaranteed thereafter. Failure to comply will result in voiding of the warranty. For repairs, please
contact Beurer customer services or an authorised retailer.
Source: Adapted to “Windisch W et al. Guidelines for Non-Invasive and Invasive Home Mechanical Ventila
tion for Treatment of Chronic Respiratory Failure Update 2017; Pneumologie 2017; 71: 722795”
%SpO2 PRbpm
7.2 Attaching the lanyard
Function button
Note
Dear customer,
Note
7.1 Inserting the batteries
Notes on handling batteries • If
your skin or eyes come into contact with battery fluid, rinse the affected areas with water and seek medi
Note
IMPORTANT:
Decline in oxygen saturation depending on altitude
WARNING
www.beurer-healthguide.com
4
www.beurer.com • www.beurer-gesundheitsratgeber.com
5
Beurer GmbH • Soeflinger Strasse 218 • 89077 Ulm, Germany
%SpOÿ
PR bpm Pulse rate (beats per minute)
- from Bluetooth® 4.0
IP 22
- iOS from 10.0, AndroidTM from 5.0
1 2
3
AFTER 60
%SpO2 PRbpm
ÿ on large fingers that do not fit into the device easily (fingertip: width approx. > 20 mm, thickness approx.
>15 mm).
The „beurer HealthManager“ app must be activated if you wish to transfer data.
• Option 1: When the pulse oximeter is switched off, press and hold the function button for 5 seconds.
„SYNC“ will flash on the display. The device will now attempt to connect to the app for approx. 10 seconds.
• Option 2: After measurements are taken, the data will automatically be sent to the app. „SYNC“ will flash
on the display. The device will now attempt to connect to the app for approx. 10 seconds. „SYNC“ stops
flashing as soon as a connection is established. All measurement data in the memory is automatically trans
ferred to the app. The pulse oximeter will then switch off automatically. If Bluetooth® is deactivated (OFF),
the message OFF will appear after you have taken the measurement.
Note: Bluetooth® must be activated (ON) if you wish to transfer data.
„SYNC“ stops flashing as soon as a connection is established. All measurement data in the memory is
automatically transferred to the app. The pulse oximeter will then switch off automatically.
9. System requirements for the app „beurer HealthManager“
16. Technical data
5. Warnings and safety notes
14. Disposal
Display description
8. Operation
2. Intended use
12. Maintenance/cleaning
6. Unit description
3. Getting to know your device
1. Included in delivery
15. Warranty/service
13. Storage
7. Initial use
11. Evaluating measurement results
15. What if there are problems?
10. Transfer of measured values via Bluetooth® low energy technology
4. Signs and symbols
WARNING
454.20-PO60_2020-04-24_04_IM1a_BEU_DE-EN
Read these instructions for use carefully! Non-observance of the following information may result in personal injury or material
damage. Store these instructions for use and make them accessible to other users. Make sure you include these instructions for
use when handing over the device to third parties.
- products purchased as seconds or as used goods;
Google Play and the Google Play logo are trademarks of Google LLC.
The Beurer PO 60 pulse oximeter provides a non-invasive measurement of the arterial oxygen saturation (SpO2 ) and the
heart rate (pulse rate). Oxygen saturation indicates the percentage of haemoglobin in arterial blood that is loaded with oxygen.
Therefore it is an important parameter for assessing the respiratory func tion. To take a measurement, the pulse oximeter uses
two rays of light with differing wavelengths, which strike the finger inserted inside the housing. A low oxygen saturation value
generally indicates underlying illnesses (respiratory diseases, asthma, heart failure etc.).
The serial number is located on the device or in the battery compartment.
Replace the batteries.
Compatible with Bluetooth® 4.0 smartphones/tablets
switch it on.
The following are explicitly excluded from this warranty: -
deterioration due to normal use or consumption of the product;
Safety note regarding potential for
damage to the device/accessories
The warranty only applies to products purchased by the buyer as a consumer and used exclusively for perso nal purposes in the
context of domestic use.
3. Close the battery compart ment
cover again.
Solution
ÿ near flammable or explosive gas mixtures. issues such as error messages or the failure of the display/device.
2. Pulse rate (value in beats per minute)
The function button on the pulse oximeter has four functions in total:
• Protect batteries from excessive heat.
The following symbols are used in these instructions for use, on the packaging and on the type plate for the device:
Cardiac arrhythmia.
ÿ during an MRI or CT scan.
measurements using Bluetooth® low energy technology“.
pulse 30 – 250 bpm, ±2beats/minute
Activate Bluetooth® on the smart phone
and start the app.
93-90
IMPORTANT
2. Insert the two supplied bat
– To switch off the pulse oximeter, wait for 10 seconds. The pulse oximeter switches itself off automatically.
customer;
Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of
Apple Inc., registered in the U.S. and other countries.
direct sunlight) in the immediate vicinity of the pulse oximeter.
• If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
The pulse oximeter is not
displaying measurement
values.
cedeclarationofconformity.php.
2402 MHz – 2480 MHz frequency band Max.
2.87 dBm transmission power,
1. Oxygen saturation (value in percent)
• Do not disassemble, open or crush the batteries.
Source: Hackett PH, Roach RC: High-Altitude Medicine. In: Auerbach PS (ed): Wilderness Medicine, 3rd edition; Mosby,
St.Louis, MO 1995; 1-37.
SpOÿ 70 –100%, ±2%,
„beurer HealthManager“ app is not
activated or Bluetooth® is switched off in
the smartphone settings.
Possible cause
• Brightness function: To select your desired display brightness, hold down the function button for slightly longer during
operation.
ÿ on patients who are not steady at the site of application (e.g. trembling).
Lanyard holder
Alarm suppression
3. Your measurement values will appear on the screen
after a few seconds.
The buyer will then receive further information about the processing of the warranty claim, e.g. where they can send the product
and what documentation is required.
In particular, do not start taking any new medication or change the type and/or dosage of any existing medication without
prior approval.
ÿ on small children or babies.
• Transferring measurements to the app and synchronising time and date: See point 10 „Transferring
The pulse oximeter uses Bluetooth® low energy technology,
Accuracy
• Do not use rechargeable batteries.
Directive – WEEE
The worldwide warranty period is 5 years, commencing from the purchase of the new, unused product from the seller.
environments.
1. Slide the battery compartment lid
open.
ÿ on fingers that are too small, as with small children for example (width approx. < 10 mm, thickness ap prox. < 5 mm).
Problem
AAA batteries
Serial number
• Observe the plus (+) and minus (-) polarity signs.
Replace the batteries.
The batteries in the pulse oximeter are
empty.
Declaration of Conformity for this product can be found at: www.beurer.com/web/we-landingpages/de/
For environmental reasons, do not dispose of the device in the household waste at the end of its useful life. Dispose
of the device at a suitable local collection or recycling point. Dispose of the device in accordance with EC Directive –
WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local authorities
responsible for waste disposal.
Keep your finger, hand and body still during
the measurement.
ÿ if the device or the finger you are using is damp.
Non-observance of the following instructions can lead to incorrect or failed measurements.
Repairs or an exchange in full do not extend the warranty period under any circumstances.
SpOÿ 0 – 100%,
pulse 30 – 250 beats/minute
Supervise children around the device to ensure they do not play with it.
1. Insert one finger into the finger opening of the pulse
oximeter as shown and hold it steady.
• Clean the housing and the interior rubber surface with a soft cloth dampened with medical alcohol after
List of supported smartphones/tablets
Permissible operating temperature and
humidity
Altitude sickness, acclimatisation recommended
AFTER 60
If the buyer wishes to make a warranty claim, they should approach their local retailer in the first instance: see the
attached “International Service” list of service addresses.
Bluetooth® is not activated (OFF) at the
device.
- Each time data is transferred, the time and date are synchronised with the smartphone.
1x PO 60 pulse oximeter, 2x 1,5 V AAA batteries, 1x lanyard, 1x belt bag, 1x these instructions for use
IP22, application part, type BF
• We hereby confirm that this product complies with the European RED Directive 2014/53/EU. The CE
of the applicable European and national
directives.
Electrical and Electronic Equipment EC
Measurement method Non-invasive measurement of arterial oxygen saturation of haemoglobin and pulse rate in finger
• Use identical or equivalent battery types only. • Always
replace all batteries at the same time.
• If you are not going to use the pulse oximeter for more than one month, remove both batteries from the device to avoid
possible leaking.
2. Draw the other end of the lanyard through the loop at
the narrow end and tighten.
ÿ on fingers with anatomical changes, oedemas, scars or burns.
2 x 1.5V
Under no circumstances should you hold the pulse oximeter under water, as this can cause liquid to enter and damage the
pulse oximeter.
• Choking hazard! Small children may swallow and choke on batteries. Store the batteries out of the reach of small children.
Permissible storage temperature and
humidity
Model No.
Measurement finger is too large or too
small.
results.
Cd = Battery contains cadmium, Hg =
Battery contains mercury
ÿ if you are allergic to rubber products.
Normal range
+10°C to +40°C, ÿ 75% relative humidity, 700 – 1060 hPa ambient pressure
No altitude sickness (normally)
– To return to the measurement, press and hold the function button for 5 seconds.
- damage that arises during transport between manufacturer and customer, or between service centre and
The word mark Bluetooth® and accompanying logo are registered trademarks of Bluetooth SIG, Inc. Any and all use of these
marks by Beurer GmbH is done so under licence. Other trademarks and trade names are the property of the relevant holder.
If you have a pre-existing condition, always consult your doctor to evaluate your measurements.
Observe the warnings and safety notes
in section 5.
Function button
-40°C to +60°C, ÿ 95% relative humidity, 500 – 1060 hPa ambient pressure
IMPORTANT: Do
not use high-pressure sterilisation on the pulse oximeter!
Measurement range
The batteries in the pulse oximeter are
too low or are empty.
- The „beurer HealthManager“ app must be activated to allow data transfers.
cal assistance.
Pb = Battery contains lead,
Device protected against foreign objects ÿ12.5
mm and against water dripping at an angle
Do NOT use the pulse oximeter
Application part, type BF
Seek medical attention urgently
~90
holder as shown.
Insufficient circulation in the meas
urement finger.
• The empty, completely flat batteries must be disposed of through specially designated collection boxes,
recycling points or electronics retailers. You are legally required to dispose of the batteries.
Impact on human body
Classification/measures to be taken
The pulse oximeter is
displaying measurement
interruptions or high meas
urement value jumps.
Only use the Beurer PO 60 pulse oximeter on humans to measure the arterial oxygen saturation (SpOÿ) of haemoglobin and
the heart rate (pulse rate). The pulse oximeter is suitable for private use (at home) as well as for use in the medical sector (hospitals,
medical establishments).
must be activated on your smartphone and on the pulse oximeter (ON). You must also ensure that the „beurer HealthManager“ app
is open on your smartphone.
Beurer guarantees the perfect functionality and completeness of this product.
The display orients automatically (vertical format, horizontal format). This ensures that the values are easy to read on the display
at all times, regardless of how you hold the pulse oximeter.
– Briefly press the function button, in order to activate (ON) or deactivate (OFF) Bluetooth®.
Activate Bluetooth® on the device by
following the instruction of section 8 (ON).
The following table for evaluating your measurements does NOT apply to people with certain pre-existing conditions (e.g. asthma,
heart failure, respiratory diseases) or whilst staying at altitudes above 1500 metres.
1. Insert the narrow end of the lanyard through the
Critical range:
The following table informs you of the effects of various altitudes on oxygen saturation value and its impact on the human body.
The following table does NOT apply to people with certain pre-existing conditions (e.g. asthma, heart failure, respiratory diseases
etc.). People with pre-existing conditions can show signs of ill ness (e.g. hypoxia) at lower altitudes.
ÿ on fingers that have nail varnish on, are dirty or have a plaster or other dressing on them.
Power supply
Arterial oxygen saturation of
haemoglobin (in percent)
A warranty claim shall only be considered if the buyer can provide Beurer, or an authorised Beurer partner, with - a copy of the
invoice/purchase receipt, and - the original product.
Finger opening
Manufacturer
The warranty shall apply without prejudice to any mandatory statutory provisions on liability.
Thickness between 5 and 15 mm.
99-94
• The codes below are printed on batteries containing harmful substances:
> 90
Observe the instructions for use
– After switching on the pulse oximeter, press and hold the function button for 5 seconds, in order to access the
Bluetooth® settings. „ON“ or „OFF“ will appear in the display.
Note on important information
To transport the pulse oximeter more easily you can attach a lanyard to the device.
Store the pulse oximeter in a dry place (relative humidity ÿ 95%). If the humidity is too high it may shorten the service life of the
pulse oximeter or damage it. Store the pulse oximeter in a place where the ambient temperature is between -40°C and 60°C.
Approx. 27 g (without batteries)
2 AAA batteries last for approx. 2 years of operation at 3 measurements per day (each of 60
seconds).
This product satisfies the requirements
• If the device is not to be used for a relatively long period, take the batteries out of the battery compartment.
Disposal in accordance with the Waste
Altitude
There are two ways in which you can synchronise data with your smartphone. In both instances, Bluetooth®
• If a low battery status appears on the display of the pulse oximeter, change the batteries.
Data transfer
5. Battery level indicator
Permissible operating
conditions
2500-3000 m
Visit to the doctor recommended
Seek medical attention.
Thank you for choosing one of our products. Our name stands for high-quality, thoroughly tested products for applications in the
areas of heat, weight, blood pressure, body temperature, pulse, gentle therapy, mas sage, beauty, baby and air. Please read
these instructions for use carefully and keep them for later use, be sure to make them accessible to other users and observe the
information they contain.
ÿ whilst taking a blood pressure measurement on the same arm using a cuff.
Weight
2. Press the function button. The pulse oximeter begins
its measurement. Do not move during the measurement.
SpOÿ (oxygen saturation) measurement in %
The warranty conditions below shall not affect the seller’s statutory warranty obligations which ensue from the sales
agreement with the buyer.
Fingertip must have the following
measurements: Width between 10 and
20 mm.
Expected SpOÿ value
(oxygenÿsaturation) in %
Batteries not inserted correctly.
Classification
To transfer the measurements to your smartphone via Bluetooth®, proceed as follows: Activate Bluetooth® in your smartphone
settings, open the „beurer HealthManager“ app and follow the instructions. In the settings menu of the app, select and connect the
PO 60. A randomly generated six-digit PIN code is displayed on the pulse oximeter, and at the same time, an input field appears
on the smartphone in which you must enter this six-digit PIN code.
– Briefly press the function button to switch on the pulse oximeter.
Note
During the warranty period, should this product prove to be incomplete or defective in functionality in accordance with the
following provisions, Beurer shall carry out a repair or a replacement delivery free of charge, in accordance with these warranty
conditions.
do not use the pulse oximeter for longer than approx. 2 hours on one finger.
- Bluetooth® is activated automatically.
No data transfer possible for
measured values.
4. Pulse bar
Your Beurer team
Battery life
CE labelling
• Do not charge or short-circuit batteries.
Warning instruction indicating a risk
of injury or damage to health
Android is a trademark of Google LLC.
each use.
Beurer GmbH, Söflinger Straße 218, 89077 Ulm, Germany (hereinafter referred to as “Beurer”) provides a warranty for this
product, subject to the requirements below and to the extent described as follows.
Technical information is subject to change without notification to allow for updates.
Sensor to measure SpOÿ Red light (wave length 660 nm); infra-red (wave length 905 nm); silicon receiver diode
1500-2500 m
Decreased range:
- accessories supplied with this product which are worn out or used up through proper use (e.g. batteries, rechargeable batteries,
cuffs, seals, electrodes, light sources, attachments and nebuliser accessories); - products that are used, cleaned, stored or
maintained improperly and/or contrary to the provisions of the instructions for use, as well as products that have been opened,
repaired or modified by the buyer or by a service centre not authorised by Beurer;
Do not dispose of batteries containing
hazardous substances with household waste.
teries into the pulse oximeter
battery compartment following the
correct polarity as shown.
German law shall apply.
Dimensions
- consequential damage arising from a fault in this product (however, in this case, claims may exist arising from product liability
or other compulsory statutory liability provisions).
• Switch-on function: When the pulse oximeter is switched off you can press the function button briefly to
WARNING
• Risk of explosion! Never throw batteries into a fire.
Each time data is transferred, the time and date are synchronised with the smartphone. To save all of your measurements with
the correct date, we recommend connecting your PO 60 to your smartphone before taking the first measurement.
Reinsert the batteries. If after
reinserting the batteries correctly there
are still no measurement values displayed,
contact customer services.
3. Pulse wave (plethysmographic wave)
• Activating and deactivating Bluetooth®:
Permissible storage and
transport conditions
Subject to errors and changes
L 58.5 mm x W 31 mm x H 32 mm
< 90