BEYES AirLight M800 User manual

Instructions For Use
M800 High Speed Collection
Table of Contents
• Before You Begin
• Safety Information
• Technical Specifications
• Assembly & Removal
• Friction Grip Bur Specifications
• Bur Insertion & Removal
• Cleaning & Lubrication
• Warranty
WARNING!
(Risk of injury)
ATTENTION!
(To prevent damage)
Thermal washer
disinfectable
General
explanations,
without risk to
persons or objects
Sterilizable
up to the stated
temperature
Fig.1 Fig.2
Fig. 2.1
Intended Use
AirLight Dental Handpiece, models AirLight are intended to be used for removing carious
material, reducing hard tooth structure, cavity preparation, finishing tooth preparations
and restorations and polishing teeth.
Before You Begin
Please read this Manual prior to using your AirLight Highspeed Handpiece. The use of
this product is only intended for use by trained dental personnel. We recommend that
you retain this manual for future reference.
Safety Information
• High speed handpieces should be sterilized after each patient to prevent any
transmission of infective organisms.
• Suitable hygienic measures must be taken to prevent cross contamination between
patients, operators and other personnel.
• Only operate this highspeed handpiece when the bur is properly inserted in the chuck.
Do not attempt to extend the length of the bur by improper positioning in the chuck.
Technical Specifications
Assembly & Removal
This handpiece should only be used on dental delivery systems that provide filtered
and regulated air and water. Filtration of 25 microns is recommended.
Security Ring
Non-Swivel Handpiece (Fig.1)
Connection: M4= 4 hole, B2 = 2 hole
1. Insert the non-swivel handpiece into the tubing
2. Firmly screw the tubing nut of the handpiece tubing by hand in a clockwise
3.
Unscrew the tubing nut of the handpiece tubing by hand in a counterclockwise for removal
Swivel Handpiece
Ez-Swivel QD Coupling (Fig.2)
Connection: M4= 4 hole, B2 = 2 hole
1. Ensure QD coupling is firmly attached to the handpiece
tubing
2. Align QD coupling into handpiece and push together by
pulling back the security ring
3. Ensure handpiece is firmly attached to coupling (Note: the
security ring must return to original position)
4. Remove QD coupling by pulling back the security ring
Ez-Swivel PD Coupling (Fig. 2.1)
Connection: M4= 4 hole, B2 = 2 hole
1. Ensure PD coupling is firmly attached to the handpiece
tubing
2. Align PD coupling into handpiece and push together
3. Ensure handpiece is firmly attached to coupling
4. Pull the retention sleeve of the PD coupling back and
remove the turbine handpiece by pulling in an axial direction
Other Approved Manufacturer’s Couplings (Fig. 2.2)
Connection: Refer to original manufacturer's instruction for use. (Do NOT
use unapproved couplings)
1. Attach the turbine handpiece to the Kavo®, NSK®, W&H®
and Star®coupling as described by the coupling manufac-
turer
2. Check that the connection to the coupling is secure.
3. Remove the Kavo®, NSK®, W&H® and Star® coupling as
described by the coupling manufacturer.
Water & Air Adjustment
Coolant Water:
Adjust the water volume at the delivery unit to obtain a satisfactory spray mist.
Air Pressure: The operating pressure Must be 2.62 bar to 2.90 bar, and recommendation
is 2.76 bar. It is highly recommended that the operating pressure initially be set using a
handpiece pressure gauge due to the pressure loss caused by the handpiece tubing
material and length.
Warning: Do not exceed 3.52 bar operating pressure as this will shorten turbine life
and may cause additional damage to the handpiece.
Handpiece Operation
Use standard length concentric diamond and carbide burs with 1.59mm nominal diameter
friction grip shanks designed for high speed handpieces.
Use only burs with hardened tempered steel shanks. Before use, verify concentric rotation
by operating handpiece outside the or alcavity in a safe place.
Warning: Do not use burs that produce excessive noise or do not run concentrically. Do
not use burs that are modified, eccentrically designed, or have bent or ‘soft’ (nonheat
treated) shanks.
Friction Grip Bur Specifications
For application details and operating data of the different rotary instruments please refer to
the information provided by the manufacturer.The push button autochuck turbines have the
following bur requirements:
Use only instruments whose shanks comply with the requirements of ISO 1797-1.
• Shank diameter of 1.59—1.60mm
• Overall length up to max. 25mm
• Overall length up to max. 21mm (mini)
• Working diameter up to max. 2.1mm (ISO 021)
Instruments that differ from the specified data must not be used.
Otherwise the mechanism of the clamping system could be damaged.
Fig. 2.2
KavoR
NSKR
W&HR
StarR
Do not assemble/remove during operation!
Rotary instruments
Idle mode speed
Coolant supply volume
Water setting range
Chip air setting range
Return air pressure
Operating pressure
Air consumption
Chip air consumption at 2 bar
Intensity of illumination
Standard Head (S)
350,000-360,000
1.6-0.01
>50
0.7-2
1.5-3
<0.5
30-45
>1.5
22,000 22,000
>1.5
30-45
0.7-2
1.5-3
<0.5
>50
1.6-0.01
380,000-410,000
2.62-2.90 2.62-2.90
Mini Head (M)
EN ISO 1797-1:1995 (0mm)
(rpm)
(ml/min)
(bar)
(bar)
(bar)
(bar)
(NL/min)
(NL/min)
(Lux)
25 22Power/Torque (watt)
Dia. Amm Dia. 1.59 ●1.60mm
Controler/Racer max.25mm, Mini max. 21mm
Fig.3 Bur Specifications

Fig.5
Document #: ENI007
Rev.4 / 23.07.19
Fig.4
Fig.8
Fig.6.2
Fig.7
Fig.6.1
0197
Bur Insertion (Fig.4)
1. Insert bur into autochuck until resistance is felt.
2. Depress the autochuck end cap button and continue to feed-in the
bur until it clicks and stops.
3. Release the autochuck end cap button. (Always pull on the bur after
it has been installed to confirm that it is seated properly).
4. Never depress the autochuck end cap when the bur is rotating.
Use only faultless burs and diamonds. These must be clean to prevent
dirt from being transferred into the chuck.
Bur Removal (Fig.5)
1. Grasp handpiece firmly so your thumb touches the end cap of the
autochuck and push until a slight ‘click’ is noticed.
2. Pull bur straight out.
3. Release end cap.
Never depress the autochuck end cap button or rest the
end cap against any part of the oral cavity during operation.
Depressing the end cap while a bur is rotating can activate
the chuck mechanism and result in the end cap button with
excessive heat and / or loss of bur retention.
External Cleaning
LED Light
Clean the LED light using isopropyl alcohol or other disinfection agents.
Super soft cleaning cloth must be used. Never use a cleaning agent
that has an ammonium, chlorine, or acid base. (Fig.6.1)
External Surface
Clean debris from the external surface by using isopropyl alcohol
or other disinfecting agents. Never use a cleaning agent that has
an ammonium, chlorine, or acid base. Dry thoroughly.
Never immerse the handpiece in disinfection baths.
Improper care and maintenance may lead to a premature
wear and failure of the handpiece. (Fig.6.2)
Beyes Dental Canada Inc.
23-595 Middlefield Road
Toronto, Ontario, M1V 3S2
Canada
Tel: 1-416-281-1888
Email: [email protected]
Web: www.beyes.ca
Lotus NL B.V.
Koningin Julianaplein 10, Ie Verd,
2595AA, The Hague,
Netherlands
Tel: +31645171879 (English)
+31626669008 (Dutch)
Operating conditions for high-tempearture washer-disinfectors
* When using a high-temperature washer-disinfector to clean the handpiece,
strictly adhere to the conditions specified below.
High-temperature cleaning conditions
Unit Name
Miele G7881 Vario TD
Mode Detergent
(concentration)
neodisher Mediclean
(0.3-0.5%)
Neutralizer*
(concentration)
neodisher Z
(0.1 - 0.2%)
neodisher mieclear
(0.02-0.04%)
Rinse
(concentration)
* After cleaning there may be streaks or white spots on the instrument.
Use a neutralizer only if there are streaks or white spots.
Operating Precautions
• Always use a handpiece holder when washing the handpiece, making sure to rinse the
inside of the handpiece thoroughly.
• If any medical agent remains inside the handpiece, it may corrode or clog the air or
water pipe, resulting in a malfunction of the handpiece.
• For details on handling medical agents or adjusting their concentration, refer to the user
manual for the washing device.
• After washing is complete, check to see if the handpiece, including its inside, is
completely dry. If any water remains inside the handpiece, expel it with an air gun or
another such tool. Failure to do so could result in the remaining water coming out during
use and cause poor lubrication or sterilization.
• Always lubricate the handpiece after washing.
Inappropriate cleaning methods and solutions will damage the handpiece
Do not clean the handpiece using strong acidic or alkaline solutions that could
cause the metal to corrode.
Do not leave a handpiece inside the high-temperature washer-disinfector.
Lubrication
Chuck System Lubrication (Fig.7)
1. Fit the spray nozzle onto the spray can
2. Grip the handpiece firmly
3. Press the tip of the spray nozzle firmly into the chuck system
4. Spray about 1 second
5. Push the button three times
Handpiece Lubrication (Turbine)
• After internal cleaning
• Prior Sterilization
•
The Handpiece must be cleaned and sterilized between patient uses, or at least once daily.
The handpiece must be cleaned and sterilized between patient uses.
Purging
Connect handpiece to tubing and run handpiece for 20-30 seconds to purge remaining lubricant.
Automatic Handpiece Lubrication
After cleaning the handpiece simply connect the handpiece to the correct adaptor and
activate the Lubrication System according to the manufacturer’s system instructions.
Manual Lubrication (Fig.8)
After cleaning the handpiece apply lubricant
BEFORE sterilization
1. Use only a high quality lubricant such as Beyes Lube.
2. Ensure the correct application nozzle is connected to
the plastic valve at the top of the spray can.
3. Firmly insert the application nozzle into the rear
of the handpiece
4. Spray for 2-3 seconds
5. Connect it to the tubing and run it with bur for 25 seconds.
6. Wipe any excess lubricant from the handpiece.
Packing
Approved Sterilization Procedure
Following your country specific directives, standard and guidelines
• Steam sterilization class B (pre and post vaccum) with sterilizers in accordance with
EN 13060. Sterilization holding time a minimium of 3 minutes at 134°C (273°F)
OR
• Steam sterilization class S with sterilizers in accordance with EN 13060
The sterilizer manufacturer must give the express approval for the sterilzation of turbine
handpieces. Sterilization holding time a minimium of 3 minutes at 134°C (273°F)
An unwrapped handpiece needs to be used immediately or sterility is not maintained.
Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
Use steam sterilization exclusively for complete sterility. Never dry heat sterilize,
ultrasonically clean, or expose handpiece or swivel to chemical disinfection or cold
sterilization.
Limited Warranty
Beyes Dental Canada hereby warrants against defect in material and workmanship that for a
period of 12 months from the date of purchase, this product will perform satisfactorily under
normal use and service. No warranty claim can be inferred here, as wear may occur earlier or
later than indicated above depending on use, frequency of sterilization and frequency of
maintenance. Beyes shall under no circumstances be liable for incidental or consequential
damages or damage due to improper use or maintenance.
This Warranty is subject to the following conditions.
Beyes will, without charge repair or replace (at Beyes sole discretion) any defective part/s
covered in this warranty.If the conditions are met, the defective item must be returned to an
authorized Beyes service center or distributor. Accompanied by the original invoice or packing
list. This warranty is in lieu of all other warranties, expressed or implied .
* This device is not field repairable * All brands are holders of their respective trademarks.
!
!
!
Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner licensed by the law of the states in
which he or she practices to use or order the use of this device. Beyes Dental Canada Inc. is not responsible for any typographical errors.
Subjects to technical changes without notice.
Place the handpiece in a sterilization pouch.
Table of contents
Other BEYES Dental Equipment manuals

BEYES
BEYES APEXPILOT ALL-IN-ONE User manual

BEYES
BEYES ApexPilot G2 User manual

BEYES
BEYES Accurator A2 User manual

BEYES
BEYES EndoPilot User manual

BEYES
BEYES AirLight M800 User manual

BEYES
BEYES Maxso X99 Series User manual

BEYES
BEYES Comfort-Sonic P9 User manual

BEYES
BEYES ApexPilot G1 User manual

BEYES
BEYES Maxso E600P User manual

BEYES
BEYES Comfortek User manual

BEYES
BEYES Comfort-Sonic P6 Plus User manual

BEYES
BEYES GPRO P1 User manual

BEYES
BEYES Maxso X200 User manual

BEYES
BEYES SLimax-C User manual

BEYES
BEYES Maxso E600 User manual

BEYES
BEYES AirLight Highspeed Handpiece Series User manual

BEYES
BEYES AirLight M800 User manual

BEYES
BEYES Maxso High Speed M200 User manual