Fig.5
Document #: ENI015
Rev.4/ 18.09.19
Fig.4
Fig.8
Fig.6
Fig.7
0197
Bur Insertion (Fig.4)
1. Insert bur into autochuck until resistance is felt.
2. Depress the autochuck end cap button and continue to feed-in
the bur until it clicks and stops.
3. Release the autochuck end cap button. (Always pull on the bur
after it has been installed to confirm that it is seated properly).
4. Never depress the autochuck end cap when the bur is rotating.
Use only faultless burs and diamonds.
These must be clean to prevent dirt from being transferred into
the chuck.
Bur Removal (Fig.5)
1. Grasp handpiece firmly so your thumb touches the end cap of
the autochuck and push until a slight ‘click’ is noticed.
2. Pull bur straight out.
3. Release end cap.
Never depress the autochuck end cap button or rest the
end cap against any part of the oral cavity during operation.
Depressing the end cap while a bur is rotating can activate
the chuck mechanism and result in the end cap button with
excessive heat and / or loss of bur retention.
External Cleaning
Clean debris from the external surface by using isopropyl alcohol
or other disinfecting agents. Never use a cleaning agent that has
an ammonium, chlorine, or acid base. Dry thoroughly.
Never immerse the handpiece in disinfection baths.
Improper care and maintenance may lead to a premature
wear and failure of the handpiece. (Fig.6)
Operating conditions for high-tempearture washer-disinfectors
* When using a high-temperature washer-disinfector to clean the handpiece,
strictly adhere to the conditions specified below.
High-temperature cleaning conditions
Unit Name
Miele G7881 Vario TD
Mode Detergent
(concentration)
neodisher Mediclean
(0.3-0.5%)
Neutralizer*
(concentration)
neodisher Z
(0.1 - 0.2%)
neodisher mieclear
(0.02-0.04%)
Rinse
(concentration)
* After cleaning there may be streaks or white spots on the instrument.
Use a neutralizer only if there are streaks or white spots.
Operating Precautions
• Always use a handpiece holder when washing the handpiece, making sure to rinse the
inside of the handpiece thoroughly.
• If any medical agent remains inside the handpiece, it may corrode or clog the air or
water pipe, resulting in a malfunction of the handpiece.
• For details on handling medical agents or adjusting their concentration, refer to the user
manual for the washing device.
• After washing is complete, check to see if the handpiece, including its inside, is
completely dry. If any water remains inside the handpiece, expel it with an air gun or
another such tool. Failure to do so could result in the remaining water coming out during
use and cause poor lubrication or sterilization.
• Always lubricate the handpiece after washing.
Inappropriate cleaning methods and solutions will damage the handpiece
Do not clean the handpiece using strong acidic or alkaline solutions that could
cause the metal to corrode.
Do not leave a handpiece inside the high-temperature washer-disinfector.
Lubrication
Chuck System Lubrication (Fig.7)
1. Fit the spray nozzle onto the spray can
2. Grip the handpiece firmly
3. Press the tip of the spray nozzle firmly into the chuck system
4. Spray about 1 second
5. Push the button three times
Handpiece Lubrication (Turbine)
• After internal cleaning
• Prior Sterilization
• The Handpiece must be cleaned and sterilized between patient uses.
Purging
Connect handpiece to tubing and run handpiece for 20-30 seconds to purge remaining
lubricant.
Automatic Handpiece Lubrication
After cleaning the handpiece simply connect the handpiece to the correct adaptor and
activate the Lubrication System according to the manufacturer’s system instructions.
Manual Lubrication (Fig.8)
After cleaning the handpiece apply lubricant
BEFORE sterilization
1. Use only a high quality lubricant such as Beyes Lube.
2. Ensure the correct application nozzle is connected to
the plastic valve at the top of the spray can.
3. Firmly insert the application nozzle into the rear
of the handpiece
4. Spray for 2-3 seconds
5. Wipe any excess lubricant from the handpiece.
Packing
Approved Sterilization Procedure
Following your country specific directives, standard and guidelines
• Steam sterilization class B (pre and post vaccum) with sterilizers in accordance with
EN 13060. Sterilization holding time a minimium of 3 minutes at 134°C (273°F)
Minimum drying time after sterilization: 10 minutes
OR
• Steam sterilization class S with sterilizers in accordance with EN 13060
The sterilizer manufacturer must give the express approval for the sterilzation of turbine
handpieces. Sterilization holding time a minimium of 3 minutes at 134°C (273°F)
Minimum drying time after sterilization: 10 minutes
An unwrapped handpiece needs to be used immediately or sterility is not maintained.
Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
Use steam sterilization exclusively for complete sterility. Never dry heat sterilize,
ultrasonically clean, or expose handpiece or swivel to chemical disinfection or cold
sterilization.
Limited Warranty
Beyes Dental Canada hereby warrants against defect in material and workmanship that
for a period of 12 months from the date of purchase, this product will perform satisfacto-
rily under normal use and service. No warranty claim can be inferred here, as wear may
occur earlier or later than indicated above depending on use, frequency of sterilization
and frequency of maintenance. Beyes shall under no circumstances be liable for
incidental or consequential damages or damage due to improper use or maintenance.
This Warranty is subject to the following conditions: Beyes will, without charge repair or
replace (at Beyes sole discretion) any defective part/s covered in this warranty.If the
conditions are met, the defective item must be returned to an authorized Beyes service
center or distributor. Accompanied by the original invoice or packing list. This warranty is
in lieu of all other warranties, expressed or implied .
* This device is not field repairable
* All brands are holders of their respective trademarks.
!
!
!
Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner
licensed by the law of the states in which he or she practices to use or order the use of this device.
Place the handpiece in a sterilization pouch.