Handpiece Lubrication (Turbine)
• After internal cleaning
• Prior Sterilization
• After 30 minutes of use or at least once daily
Purging
Connect handpiece to tubing and run handpiece for 20-30 seconds to purge remaining
lubricant.
Automatic Handpiece Lubrication
After cleaning the handpiece simply connect the handpiece to the correct adaptor and
activate the Lubrication System according to the manufacturer’s system instructions.
Manual Lubrication (Fig.7)
After cleaning the handpiece apply lubricant
BEFORE sterilization
1. Use only a high quality lubricant such as Beyes Lube.
2. Ensure the correct application nozzle is connected to
the plastic valve at the top of the spray can.
3. Firmly insert the application nozzle into the rear
of the handpiece
4. Spray for 2-3 seconds
5. Wipe any excess lubricant from the handpiece.
Sterilization
Following your country specific directives, standard and guidelines
• Steam sterilization class B (pre and post vaccum) with sterilizers in accordance with
EN 13060.
OR
• Steam sterilization class S with sterilizers in accordance with EN 13060
The sterilizer manufacturer must give the express approval for the sterilzation of turbine
handpieces.
An unwrapped handpiece needs to be used immediately or sterility is not maintained.
Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
Use steam sterilization exclusively for complete sterility. Never dry heat sterilize,
ultrasonically clean, or expose handpiece or swivel to chemical disinfection or cold
sterilization.
Limited Warranty
Beyes Dental Canada hereby warrants against defect in material and workmanship that
for a period of 12 months (six months for turbine) from the date of purchase, this product
will perform satisfactorily under normal use and service. No warranty claim can be inferred
here, as wear may occur earlier or later than indicated above depending on use,
frequency of sterilization and frequency of maintenance. Beyes shall under no
circumstances be liable for incidental or consequential damages or damage due to
improper use or maintenance.
This Warranty is subject to the following conditions
Beyes will, without charge repair or replace (at Beyes sole discretion) any defective part/s
covered in this warranty.If the conditions are met, the defective item must be returned to
an authorized Beyes service center or distributor. Accompanied by the original invoice or
packing list. This warranty is in lieu of all other warranties, expressed or implied .
* This device is not field repairable
* All brands are holders of their respective trademarks.
Fig.4
Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner licensed by
the law of the states in which he or she practices to use or order the use of this device.
I123EN.REV0Z1721
Bur Insertion (Fig.3)
1. Insert bur into autochuck until resistance is felt.
2. Depress the autochuck end cap button and continue to feed-in
the bur until it clicks and stops.
3. Release the autochuck end cap button. (Always pull on the bur
after it has been installed to confirm that it is seated properly).
4. Never depress the autochuck end cap when the bur is rotating.
Use only faultless burs and diamonds.
These must be clean to prevent dirt from being transferred into
the chuck.
Fig.3
Bur Removal (Fig.4)
1. Grasp handpiece firmly so your thumb touches the end
cap of the autochuck and push until a slight ‘click’ is noticed.
2. Pull bur straight out.
3. Release end cap.
Never depress the autochuck end cap button or rest the
end cap against any part of the oral cavity during operation.
Depressing the end cap while a bur is rotating can activate
the chuck mechanism and result in the end cap button with
excessive heat and / or loss of bur retention.
Chuck System Lubrication (Fig.6)
1. Fit the spray nozzle onto the spray can
2. Grip the handpiece firmly
3. Press the tip of the spray nozzle firmly
into the chuck system
4. Spray about 1 second
5. Push the button three times
Fig.7
Fig.6
Fig.5.2
Fig.5.1
Lubrication
Beyes Dental Canada Inc.
30 Casebridge Court
Toronto, Ontario, M1B 3M5
Canada
Tel: 1-416-281-1888
Beyes Europe MEDevice Union ltd.
26 York Street
London, W1U 6PZ
United Kingdom
Tel: 44-20-3514-0949
External Cleaning
Fiber Optic Rod
Clean the Fiber Optic Rod using isopropyl alcohol or other disinfection agents. Super soft
cleaning cloth must be used. Never use a cleaning agent that has an ammonium,
chlorine, or acid base. (Fig.5.1)
External Surface
Clean debris from the external surface by using isopropyl alcohol
or other disinfecting agents. Never use a cleaning agent that has
an ammonium, chlorine, or acid base. Dry thoroughly.
Never immerse the handpiece in disinfection baths.
Improper care and maintenance may lead to a premature
wear and failure of the handpiece. (Fig.5.2)