BioLAB BFSD-502 User manual

Operation Manual
Ductless Fume Hood
BFSD-502
Thank you for Choosing Biolab products. Please read the “Operating Instructions” and
“Warranty” before operating this unit to assure proper operation.

Preface
Dear respected user
Welcome to select and purchase Ductless Fume Hood BFSD-502, here please accept our
sincere thanks!
We sincerely hope that our products can bring the greatest help for your work.
In order to let you be more familiar with the Ductless Fume Hood BFSD-502, please carefully
read the manual firstly. Being familiar with the contents of this manual can ensure you operate
the equipment safely and correctly! Please keep this manual properly.
After having been familiar with the contents, please keep this manual in a place which is
convenient to use for easy access.
www.biolabscientific.com
2
Ductless Fume Hood BFSD-502

Index
1. Unpacking, Installation, Debugging...............................................................................04
2. User Instructions....................................................................................................................11
3. Trouble Shooting and Solution........................................................................................26
4. Warranty....................................................................................................................................30
3
Ductless Fume Hood BFSD-502
www.biolabscientific.com

01 Unpacking, Installation, Debugging
Please firstly check if the packing box is in good condition. If the packing box is damaged,
please take photos.
1.1 Unpacking of Upper Cabinet
box 1 Method 1 Use M8 Wrench to unpack (Figure1)
Figure 1
box 1 Method 2 Necessary tools for unpacking Electric drill with hexagon dead M8(Figure 2)
www.biolabscientific.com
Ductless Fume Hood BFSD-502

Figure 2
Quick unpacking diagram (Remove the screws, as are shown below
, and then take the
wooden box to the left. )
5
Ductless Fume Hood BFSD-502
www.biolabscientific.com

Figure 3
box2 Quick unpacking
Cut the packing tape down with a scissors, remove the upper cover of the carton, and then
move carton upwards.
Figure 4
1.2 Unpacking of Lower Cabinet
box 1 Unpack with M8 wrench or electric drill
www.biolabscientific.com
6
Ductless Fume Hood BFSD-502

Figure 5
box 2 Cut the packing tape down with a scissors, remove the upper cover of the carton, and
then move carton upwards.
Figure 6
1.3 Accessories Check
Please refer to the packing list, carefully check whether the accessories and information are
complete.
7
Ductless Fume Hood BFSD-502
www.biolabscientific.com

BFSD 500 Series Fume Hood Packing List
Upper cabinet :
No. Item Quantity
1 Upper Cabinet 1set
2 User Manual 1copy
3 Quality Certificate 1pc
4 Test Report 1copy
5 Fuse 10A 1pc
6 Power Cord 1pc
7 Inner Hexagon Wrench 1pc
Lower cabinet(optional) :
No. Item Quantity
1 Lower Cabinet 1set
2 Base 1set
1.4 Installation Conditions and Operating nvironment
Installation position of fume hood should be avoided to face road and sites people frequently
go through, avoided to clog windows and places where ventilation and lighting are not good,
avoided to block the entrance and exit and place where the open or close of the door is
affected ,and avoided to place oppositely or in the corner of the wall.
Working environment
(1) Only applicable for indoor operation
(2) Ambient temperature 15° C ~35° C
(3) Relative Humidity ≤75%;
(4) Atmospheric pressure range 70 kPa ~106 kPa;
(5) Electrical parameters Adequate power supply to the fume hood (See 2.1.4 Technical
Parameters);
(6) The power should be grounded (Use a multimeter to test the voltage of the live line to
ground and the voltage of the null line to ground respectively, among which, the voltage of
the live line to ground should be the grid power, and the voltage of the null line to ground
should be 0, otherwise, the power is not grounded well).
1.5 Installation
a) Remove all packaging components
www.biolabscientific.com
8
Ductless Fume Hood BFSD-502

b) Check whether the outer surface of the host has scratches, deformations or foreign
objects;
c) Check carefully the accessories and information by referring to the packing list in the
operating manual;
d) Move the entire equipment to a site which is as close as possible to the position where
the equipment is placed finally and convenient for installation;
e) Check if the environment voltage frequency is the same with the voltage frequency
shown in the label;
Method to connect the upper cabinet and the lower cabinet
(1) Place the lower cabinet in the proper installation position, brake the caster brakes, then
place the upper cabinet on the lower cabinet, and let the left, right, rear side of the upper
cabinet align with those of the lower cabinet;
(2) Fix the upper and lower cabinets to prevent skidding and open the front acrylic door on
the upper cabinet, and then take the accessories out including stainless steel hexagon socket
head cap bolts M10 * 20 and stainless steel large flat washer 5, and fasten well the lower
cabinet and the upper cabinet, as is shown in Figure 7
9
Ductless Fume Hood BFSD-502
www.biolabscientific.com

(3
) Take the activated carbon filter from the paper package, and put it in the filter locating
slot on the filter mounting panel, and then place the fan assembly above the activated carbon
filter located in the bottom, and finally place the second activated carbon filter above the fan
assembly.
(4
) Connect the butt plug on the fan assembly with the butt plug on the operating panel,
and fasten them well; connect the butt plug of the fluorescent lamp with the butt plug on the
operating panel, and fasten them well; connect the butt plug of the VOC sensor with the butt
plug on the operating panel, and fasten them well; plug the air velocity sensor plug on the
corresponding position of the operating panel; plug the humidity sensor plug on the
corresponding position of the operating panel.
1.6 Inspection after Installation
After the equipment is powered on, check the following items according to 2.3.2
Checking Items Normal Working Status
Equipment power status Connect the ambient power, the equipment
can be powered on
Fan running situation
Press the fan button, the fan runs normally,
press the fan air velocity button, the speed
can be adjusted
Fluorescent lamp Press the fluorescent lamp button, the lamp
can be lighted up.
If you have any questions, please contact our agent to debug. Debugging method
is in the After-sale service manual.
www.biolabscientific.com
10
Ductless Fume Hood BFSD-502

02 User Instructions
2.1 Functions
2.1.1 Product Concept
In chemical laboratories, a lot of odors, moisture and corrosive substances will be generated
during the experiment. To protect the safety of users and prevent the spread of pollutants in
laboratories, fume hoods are used.
Note: Filters of BFSD 500 series non-ducted fume hood have poor adsorption effect to
hydrogen, helium and inert gases, methane, ethane, ethylene oxide, carbon monoxide,
carbon dioxide, nitrogen monoxide, propylene, propyne, acetylene, organophosphorus
compounds Mercury, hydrogen cyanide, radioisotope. Generally speaking, the
adsorption effect to gas with a molecular weight of less than 36 is not good, in such
case, it is not recommended to use such a fume hood. In addition, in experiments with
large amounts of volatile chemicals, experiments with large amounts of smoke,
experiments with strong acid and alkali, experiments with flammable,explosive, toxic,
such kind of fume hood cannot be used.
2.1.2 Working theory/Air flow Pattern and Protected Area Figure
11
Ductless Fume Hood BFSD-502
www.biolabscientific.com

Figure 8
2.1.3 Protected Objects
Safety is the biggest mission pursued by fume hoods. The use of fume hoods in laboratories is
to ensure the safety of users and to prevent contamination of the laboratory environment.
2.1.4 Technical Parameters
M
ode
Parameters
BFSD-501 BFSD-502 BFSD-503 BFSD-504
Rated Voltage
AC 220V±10% □ 110V±10% □
Rated
Frequency 50 Hz □ 60Hz □
External Size 1000*830*214
0 (mm )
1200*830*2140
(mm )
1500*940*2140
(mm )
1800*940*2140
(mm )
Internal Size 910*690*740
(mm )
1110*690*740
(mm )
1410*690*740
(mm )
1710*690*740
(mm )
Rated Power 400 W 400 W 500 W 500 W
Airflow
Velocity 0.4 ~0.6m/s
Fluorescent
Lamp Power 220V 16W*2 220V 16W*2 220V 16W*3 220V 16W*3
UV
Lamp (optio
T8 20W T8 20W T8 30W T8 30W
www.biolabscientific.com
12
Ductless Fume Hood BFSD-502

n)
Noise ≤54dB (A)
Note: (1) Power consumption does not include power consumption loaded by cabinet socket
(load can not exceed 500W);
(2) Biolab reserves all the rights to changes in product design, if there are any design change,
we will not inform in advance.
1)Vibration amplitude
The vibration amplitude between frequency 10Hz and 10KHz should not exceed 5μm(rms).
2) Voltage resistance
The voltage increases to 1390V (AC) in 5s and keep for another 5s without breakdown.
3) Grounding resistance ≤0.1Ω
2.2 Product Structure
2.2.1 Structural Composition of BFSD 500 Series Fume Hood
13
Ductless Fume Hood BFSD-502
www.biolabscientific.com

Figure 9
Figure 10
1. Castor 8.Wind Speed Sensor 15.External Circuit Hole
2. Lower Cabinet (Optional) 9.Filter Mounting Panel 16. Hand Rack
3.Socket (Optional) 10. Chemical Filter 17. Rocker Switch
4.Control Valve for Water, Gas
Tap (Optional) 11. Fan Assembly 18. Tail Plug
5. Side Panel 12. Operating Panel 19. Fluorescent Lamp
6.Work Table 13.Fluorescent Lamp 20. Fan
www.biolabscientific.com
1
Ductless Fume Hood BFSD-502

7.Front Window Acrylic 14. Side Panel Acrylic 21. VOC Sensor
2.2.2 Structure Composition
Lighting Source
Three anti-fluorescent lamp is used to ensure that the average illumination in the operating
area is in line with standard requirements.
Control Panel
Main Interface Description
15
Ductless Fume Hood BFSD-502
www.biolabscientific.com

1.Power button Under the power-on state, the icon color is green, at this time, the UV lamp,
fluorescent lamp, fan, socket can be turned on and off; click the power button, and icon color
turns to be red, indicting the equipment is powered off, at this time, the UV lamp, fluorescent
lamp, fan, outlet can not be turned on, all components are close and the buttons have no
response.
2. Fan button When the power button is turned on, click the button, and the icon turns to be
green, the fan is turned on; when the icon turns to be gray, the fan is turned off.
3. Socket button When the power button is turned on, click the button, and the icon turns to
be green and the socket is opened; when the icon turns to be gray, the socket is closed.
4 . Fluorescent lamp button when the power button is turned on, click the button, and the
icon turns to be green, the fluorescent lamp is turned on; when the icon turns to be gray, the
fluorescent lamp is turned off.
5. UV lamp button when the power button is turned on, click the button, and the icon turns
to be green, the fluorescent lamp is turned on; when the icon turns to be gray, the fluorescent
lamp is turned off.
Temperature and humidity sensor display Tem temperature, Hum humidity.
Air velocity display The white area shows the air velocity value of the air velocity sensor at this
moment. One minute after the fan is turned on, if the air velocity value is lower than the low
air velocity alarm value set by the user, it will generate an alarm and prompt "Warning! Air
www.biolabscientific.com
16
Ductless Fume Hood BFSD-502

velocity is too low"
Filter time display The white area shows the fan working time, when the time is greater than
3000 hours, it will generate an alarm and prompt "please replace the filter."
Filter time grid As the filter time increases, the grid will slowly increase, when the grid is full,
the cumulative filter time reaches 3000 hours.
Alarm window Here shows the alarm information.
VOC sensor pollution coefficient display The larger the value is, the more serious the air
pollution is.
Setting button Click the button, the setting interface is reached.
Setting Interface Description
17
Ductless Fume Hood BFSD-502
www.biolabscientific.com

1. User setting User can set low air velocity alarm value and turn on/off the fan at specified
time intervals
2. System setting User can modify the fan gear, debug value, clear filter, UV lamp time.
3. Use instruction Briefly introduce the use method of the product.
4. VOC Alarm setting :
VOC setting used to adjust the display value and alarm value of VOC.
User Setting Interface
www.biolabscientific.com
18
Ductless Fume Hood BFSD-502

1. Users can set low air velocity alarm value independently, one minute after the fan is turned
on, if the air velocity is lower than the user-set low air velocity alarm value, the equipment will
alarm.
2. Start time of regularly start/end fan Users can set fan start time independently, h hour; m
minute.
3. End time of regularly start/end fan ( fan running time) Users can set fan end (running) time
independently, h hour; m minute.
4. Ok button Click this button to memorize the low air velocity alarm value set by the user,
otherwise the value can not be memorized, click the button, switch into the main interface.
19
Ductless Fume Hood BFSD-502
www.biolabscientific.com

5. Start button Click this button to memorize the start/end time of fan set by the user,
otherwise the value can not be memorized
6. Return button Click this button to return to the main interface, and does not memorize the
start/end time of fan set by the user.
System Setting Interface:
1. Velocity display The white area shows the air velocity value of the air velocity sensor at this
time.
2. Debug adjust Click the , to increase/decrease value, the middle area shows
the debug value at this time, by which, the shown air velocity can be adjusted.
3. Fan adjust velocity button The fan gear can be adjusted by clicking this button.
4. Fan gear adjust The fan gear can be adjusted by clicking the , buttons. All in
all, there are 70 gears.
www.biolabscientific.com
20
Ductless Fume Hood BFSD-502
Table of contents
Other BioLAB Laboratory Equipment manuals

BioLAB
BioLAB BIFL-201 User manual

BioLAB
BioLAB BCBL-301 User manual

BioLAB
BioLAB BCLT-201 User manual

BioLAB
BioLAB BBAL-108 User manual

BioLAB
BioLAB BIFL-203 User manual

BioLAB
BioLAB BIFL-205 User manual

BioLAB
BioLAB BIFL-207 User manual

BioLAB
BioLAB BSFLS-101 User manual

BioLAB
BioLAB BLVR-202 User manual

BioLAB
BioLAB BCLT-203 User manual

BioLAB
BioLAB BAVT 500 Series User manual

BioLAB
BioLAB BVMX-102 User manual

BioLAB
BioLAB BBBF-106 User manual

BioLAB
BioLAB BCBHR-304 User manual

BioLAB
BioLAB BHMS-104 User manual

BioLAB
BioLAB BVMX-201 User manual

BioLAB
BioLAB BODR 100 Series User manual

BioLAB
BioLAB BIFL-202 User manual

BioLAB
BioLAB BCLT-204 User manual

BioLAB
BioLAB BBWA-102 User manual