
Biotage®Extrahera™Installation and Safety | © Biotage 2020
Safety
Operation
»
Use the system only for its intended purpose, as described
in the user documentation delivered with the system and
user documentation available at www.biotage.com.
»
To avoid leakage, check the following
before operating the system:
»
Ensure that all connections are properly connected and
tightened (see the "Connections" section on page 5).
»
Ensure that the extraction waste collector, waste
reservoir, and solvent rack with five reservoirs (in
position 5 on the working area) are in place, and that the
waste reservoir and the pipette tip waste bin are not full.
»
Ensure that pipette tips of the correct size have been
loaded into the correct positions in the pipette tip racks.
»
To avoid damaging the system, use care to ensure that all
solvents used with the system are free of particulates.
»
Never operate the system when damaged.
Chemical and Biological Safety
»
Check regularly for leaks and spills. If leakage is
observed, follow the instructions in the “Troubleshooting”
section in the “Biotage®Extrahera™User Manual”
document. If spillage has occurred, follow the
instructions for cleaning the interior of the system in
the “Biotage®Extrahera™User Manual” document.
»
Always place the solvent bottles and waste
reservoir on the side of the system.
»
Use appropriate caps on the solvent bottles to
prevent harmful solvent vapors from escaping
and the contents from being spilled.
»
The system operates using electricity, which can
introduce additional hazards with certain solvents
if not properly connected, vented, or set up with
recommended manufacturer approved settings.
»
Never leave solvents, samples, or waste inside
the system when the ventilation is turned off.
»
Follow all generally-accepted lab safety procedures
and applicable laws and regulations.
»
All samples and waste should be treated
as potentially biohazardous.
»
Always follow local and national safety regulations
related to storage, handling and disposal of
chemicals, biological samples and waste.
»
Read and understand the safety data sheet (SDS)
provided by the chemical manufacturer before
storing, handling, working with, or disposing
of any chemical or hazardous substance.
»
Personnel working with or near the system must wear
protective clothing, safety gear, and eye protection that
comply with local and national safety regulations.
Maintenance and Troubleshooting
»
If the system is found with the door closed and the
power off, ensure to ventilate the system properly before
turning the system back on. Follow the instructions
in the "Explosion Risk" section on page 9.
»
If the fan stops working, shut down the system,
remove any solvents, samples, and waste inside the
system, and contact Biotage 1-Point Support.
»
If the ventilation is too low, potentially harmful gases will
escape into the surrounding environment. Ensure to take the
necessary precautions to avoid exposure to harmful gases.
»
Follow all maintenance instructions in the “Maintenance”
chapter of the “Biotage®Extrahera™User Manual” document.
»
Clean the waste tubing regularly to avoid leakage
caused by the tubing getting clogged. See instructions
in the “Biotage®Extrahera™User Manual” document.
»
There are no user serviceable parts inside the system.
Covers and safety shields may only be removed by an
authorized Biotage service engineer. Potential electrical
hazard exists due to high voltage circuits inside the system.
»
The power cord should be inspected periodically
and replaced if damaged or altered. Use only
a power cord supplied by Biotage.
»
The system uses double pole fusing. Use only
exact replacement fuses supplied by Biotage.
Incorrect fuses create a potential fire hazard.
»
Use only tubing, nuts, and ferrules supplied by Biotage.
»
Use caution when finger-tightening fittings to
prevent stripped threads or crushed ferrules.