BioLAB BTUR-303 User manual

Thank you for Choosing Biolab products. Please read the “Operating Instructions” and
“Warranty” before operating this unit to assure proper operation.
Operation Manual
3D Rotator
BTUR-303

Index
. SAFETY FEATURES…………….…………………………....................................................................03
2. PACKING ………………………..……………. ……….……...................................................................03
3. STRUCTURE DESCRIPTION ……………….…….……...................................................................05
4. TECHNICAL SPECIFICATIONS………….…. …………..................................................................05
5. F -F9 FIXED PROGRAM OPERATING INSTRUCTIONS …………………… ………………...09
6. PROGRAM DESCRIPTION……………………. …………...............................................................
7. CUSTOMIZATION PROGARM C -C2……………………........................................................... 3
8. C ,C2 CUSTOMIZATION MODE SETTING ……………........................................................... 3
www.biolabscientific.com
3D Rotator BTUR-303
2

0 Safety Features
Please read this manual carefully before operate this machine.
. It must use the original adapter.
2. The machine should be placed with a flat and steady surface for operating.
3. The machine which have large shake extent, please cover the tube then prevent liquid
samples overflowing.
4. Please cut off the power before replacing the tube.
5. Please do not move machine while working.
6. The liquid samples loading should not over the heaviest loads.
02 Packing
Main Body Unit
Standard Rack Unit
Adapter Unit
Wrench for Rack Unit
Packing Box Unit
Operation Manual Unit
Warranty Information
We provide one-year warranty against any manufacturing fault.
If BTUR-303 fails to function according to specifications at any time then please contact the
manufacturer or local representative immediately. The warranty will not cover defects caused
by normal wear or by a breach of the manufacture’s recommendations .Each BTUR-303 is
tested before shipping by the manufacturer. The quality assurance procedure guarantees that
the BTUR-303 you have purchased is ready for use.
Warranty repair could be performed only if the equipment is delivered in the original
manufacturers packaging or equally safe packaging. Therefore, please keep then
transportation packaging after unpacking the device. If the above warranty requirements are
disturbed, repair charges are applied to the customer. For all further questions and
maintenance please contact manufacturer or local representative.
3
www.biolabscientific.com
3D Rotator BTUR-303

MODEL DIMENSION
(CM)
WEIGHT (KG)
BTUR-303 with BX- 03945 3 6x 25x 68 .5
BTUR-303 with BX- 03944 344X 25x 68 .6
BTUR-303 with BX- 03943 420x 25x 68 .9
05 F -F9 Fixed Program Operating Instructions
A. Press the power button 3 seconds and on/off the machine after plug into a
power.
B. Program mode select button. Program mode from F to F9 .In customization
mode for “ rotation angles”,” shaking range”, ” pause time”and “No action” selecting.
The display showing
C. Speed and mode adjustable button. Press this button can be adjust the
speed when the machine is operating. And it can be adjust mode setting in customization
mode.
D. Rotating and shaking showing.
8. Press button for sixth stage setting.
www.biolabscientific.com
3D Rotator BTUR-303
8

Press button for selecting “No rotation ”action
9. Press button for seventh stage setting
Flashing on display,press button for adjusting shaking time parameters
0. Press button for eighth stage setting Flashing on display,press
button for setting the action time for complete program
. Press button ,then press button for adjusting rotation and shaking speed
2. Press button and save the complete program
3. Press button for starting the complete program .
Press button for adjusting angle parameters
4. Press button for second stage setting
Press button for selecting shaking action
Press button for adjusting “shaking” amplitude
5. Press button for third stage setting
Press button for selecting “ clockwise rotation ” action
Press button for adjusting rotation angle
6. Press button for fourth stage setting
Press button for selecting “Pause time ”action
Press button for adjusting “Pause time ”
7. Press button for fifth stage setting
9
www.biolabscientific.com
3D Rotator BTUR-303

Press button for selecting “clockwise rotation ”action
Press button for adjusting rotation angle
E. Operating mode showing.
F. Shaking time and Operating time showing.
G. Rotating angle and shaking frequency showing.
H. Press this button the rack can be 80 ° rotating when the machine stopped for
exchange tubes.
I. Select fixed mode F to F9 or customization mode C ,C2 ,Press this button for
confirming and setting the “rotating angles”, “ shaking range” , “ shaking time” “, pause time”
“ operating time” when the machine stopped.
J. Start and stop button.
06 Program Description
MODE display Rack movement
overview
Mixing program description
F Variable rotation speed continuous clockwise rotation
and complete program can be timing .
F2 Variable rotation speed for clockwise
360° rotation and hold for 2 secs
In initial point ; complete program can be timing .
F3 Variable rotation speed ; 80°clockwise
rotation , 2 sec hold, 80°clockwise rotation,
hold in initial point 2 sec ; the complete
program can be timing.
F4 Variable shaking amplitude & rotation speed;
35°Clockwise rotation,2 sec shaking; 225°
clockwise rotation ,2 sec hold in initial point;
the complete program can be timing.
www.biolabscientific.com
3D Rotator BTUR-303
10

F5 Variable shaking amplitude & rotation speed ;
35°clockwise rotation , 2 sec shaking ; 90°
clockwise rotation, 2 sec shaking; 35°
clockwise rotation ;2 sec hold in initial point;
the complete program can be timing.
F6
Variable shaking amplitude & rotation speed ;
35°Clockwise rotation, 2 sec shaking ,90°
clockwise rotation, 2 sec shaking ,
35°clockwise rotation , 2 sec shaking ,
the complete program can be timing.
F7 Variable rotation speed ; 0°clockwise rotation , 2 sec
hold , , 0°counter clockwise rotation 2 sec hold in initial
point, 0°counter clockwise rotation, 2 sec hold ,
0°clockwise rotation, 2sec hold in initial point,;
the complete program can be timing.
F8
Variable shaking amplitude & rotation speed ;
0°clockwise rotation , 2sec shaking , 0°
counter clockwise rotation , hold in initial point
2sec, 0°counter clockwise rotation,2 sec
shaking ; 0°clockwise rotation, hold in initial
point 2sec ; the complete program can be timing.
3. Press button for parameters increasing or decreasing.
4. Press button select the “rotation angles”, “ shaking range” and “No action” in each
paragraph of customization program C -C2. Display show
5. Press button confirming the customization program
6. Press button for adjusting speed.
11
www.biolabscientific.com
3D Rotator BTUR-303

07 C , C2 Customization Mode Setting
The the required actions as below,
. Counter clockwise rotation 70°
2.Shaking amplitude”3”
3. Clockwise rotation 90°
4. Pause time 8 sec
5. Counter clockwise rotation 80° 6. No need any actions
7. Shaking time 0 seconds
8. The complete set action for one hour.
. Press button for selecting C or C2 customization mode setting
2. Press button for first stage setting.
3. Press button for selecting counter clockwise angle.
4.Press button for confirming the setting
5.Press button for adjusting the speed
08 Customization Program C -C2
.Press button select customization program C -C2
www.biolabscientific.com
3D Rotator BTUR-303
12

2.Press button select ”clockwise rotation “,”counter clockwise rotation “, “ shaking ” ,
“pause ” “ No action” setting.
F9 Variable amplitude & speed ; 0°
clockwise rotation ,2 sec shaking ; 0°
counter clockwise rotation , Initial point 2sec shaking ,
0°counter clockwise
rotation,2sec shaking ; 0°clockwise rotation , Initial point
2sec shaking ;
the complete program can be timing.
_ clockwise rotation and counter clockwise rotation (rotating
angles can be adjustable , speed can be adjustable )
C C2 Customization mode for the “rotating angles”,"shaking
amplitude” , “pause time” “ shaking time” , “ operating time”
setting.
FIXED PRO RAM FROM F1 TO F9
13
www.biolabscientific.com
3D Rotator BTUR-303
Table of contents
Other BioLAB Laboratory Equipment manuals

BioLAB
BioLAB BODR 100 Series User manual

BioLAB
BioLAB BCFL-203 User manual

BioLAB
BioLAB BIDP 100 Series User manual

BioLAB
BioLAB BCMI-502 User manual

BioLAB
BioLAB BDIB-103 User manual

BioLAB
BioLAB BCLT-202 User manual

BioLAB
BioLAB BIFL-208 User manual

BioLAB
BioLAB BCLT-204 User manual

BioLAB
BioLAB BBBF-106 User manual

BioLAB
BioLAB BMRW-104 User manual

BioLAB
BioLAB BBWA-102 User manual

BioLAB
BioLAB BTUR-101 User manual

BioLAB
BioLAB BCBL-301 User manual

BioLAB
BioLAB BVMX-102 User manual

BioLAB
BioLAB BCMI-505 User manual

BioLAB
BioLAB BLVR-202 User manual

BioLAB
BioLAB BIGP-502 User manual

BioLAB
BioLAB BBWA-100 User manual

BioLAB
BioLAB BVMX-201 User manual

BioLAB
BioLAB BAHZ 400 B Series User manual