1 of 4
LB-0266 Rev DRAFT 1.00
Instruction Procedure to Replace Fuzz Button NeuroPort Plug with a Filament
Film NeuroPort Plug
The instructions below is a procedure to replace the existing fuzz button NeuroPort Plug with
the beta sample of Filament Film NeuroPort Plug on the Patient Cable (P/N: 4463).
1) Gently pull the existing NeuroPort Plug (P/N: 4152) from the head stage housing of the
Patient Cable Assembly (P/N: 4463). This is done by applying slight pressure to pull the
edges around the stainless steel wheel of the NeuroPort plug as shown in the figure
below. Gently rock the plug back and forth using care not to bend the internal ICS 96
pins.
2) Unplug the whole NeuroPort plug from the Patient Cable as shown in the figure below
and place it in an ESD safe bag.