
8. Press and release the Scan button.
A scanning symbol appears in the upper right-hand corner of the LCD screen during the scan.
Scan button
9. Hold the probe steady until the scan is nished.
When you hear the end-scan tone, the scan is complete.
10. Using the Results screen, verify that the aim was accurate.
If no arrow is displayed, then the
bladder was completely contained
within the scanning cone. The
measurement is accurate.
If a solid arrow is displayed, then
the bladder was mostly inside the
scanning cone. The results are
satisfactory, but in order to ensure
accuracy, Verathon recommends
that you re-aim in the direction of
the arrow and perform the scan
again.
If a ashing arrow is displayed,
then the scan is off-target, and the
bladder was not enclosed within
the scanning cone. Re-aim the
probe in the direction indicated by
the ashing arrow and perform the
scan again.
11. Note the patient’s bladder volume.
The BVI 6100 instrument will not save this measurement after a new scan is begun.
FLASHING ARROWSOLID ARROWNO ARROW
EC Representative:
Verathon Medical (Europe) B.V.
Willem Fenengastraat 13
1096 BL Amsterdam, The Netherlands
Tel: +31 (0) 20 210 30 91
Fax : +31 (0) 20 210 30 92
Corporate Headquarters:
Verathon Inc.
20001 North Creek Parkway
Bothell, WA 98011, USA
Tel: 800.331.2313 (USA & Canada only)
Tel: 425.867.1348 Fax: 425.883.2896
BladderScan, the BladderScan symbol, Verathon, and the Verathon Torch symbol are
trademarks of Verathon Inc. © 2007, 2010, 2015 Verathon Inc. 0900-0589-10-60
verathon.com
https://my.scanpoint.com0123
800.331.2313 | verathon.com