Bosean Electronic Technology K-600 User manual

K-600
Portable Multi-gas Detector
with Built-in Pump
User Manual
Ver:BSA20180501001
Bosean Electronic Technology Co.,Ltd
ISO9001-2008

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 1 -
Brief Introduction
K-600 portable multi-gas detector with built-in pump is a new
intelligent gas detector, which adopts an advanced integrated
circuit, with standard intelligent level design technology, and
proprietary digital analog hybrid communication technology. The
detector is of excellent sensitivity and repeatability, which makes it
very easy to use and maintain. Thus greatly meets the safety
requirements of industrial sites with high reliability. The detector is
made of high-strength engineering plastics, compound non-slip
rubber, which is of high strength and good hand feeling. what’s
more, the detector is water-proof, dust-proof and explosion-proof.
It is widely used in petroleum, chemical, environmental, metallurgy,
refining, gas transmission and distribution, biochemical medicine,
agricultural and other industries.
1. Main features
*Gas pump sampling method and high-sensitivity sensor, with
high sensitivity and repeatability.
* 32-bit built-in MCU, high reliability and self-adaptation ability.
* Full functions, easy operation.
* CSTN colorful LCD, more intuitionistic, abundant and clear
indication.
* Compact design, easy carrying.
* High strength engineering plastics and compound anti-slippery
rubber; high strength, water-proof, dust-proof and explosion-proof.
*1200pcs data records, can be contacted with computer via
software. Can print the data records, can store data records, can
make the data analysis.
*Gas curve display.

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 2 -
2. Structure & Function
2.1 Appearance
2.2 Detector structure: the main shell, circuit boards, batteries,
display, sensors, chargers of the components.
2.3 Principle: Electrochemical or Catalytic or PID’s or infrared
sensor or VOC’s sensor.
3. Technical Data
Target Gas
Range
Low alarm
High alarm
Resolution
Ex
0~100%LEL
20%LEL
50%LEL
1%LEL
H2S
0~100ppm
10ppm
35ppm
1ppm
CO
0~1000ppm
50ppm
150ppm
1ppm
O2
0~30%vol
19.5%vol
23.5%vol
0.1%vol
Other gases needed, please contact supplier
1
Built-in pump gas inlet
2
LED
3
Button
4
Buzzer
5
LCD screen

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 3 -
Gas sampling method: Gas pump sampling
Detecting gas: Combustible gas, H2S, CO, O2
Accuracy: ≤±5% F.S.
Response Time: ≤30s
Indication: LCD displays real-time and system status;
LED, audio and vibration alert for gas leakage, fault and low
voltage, pump working state.
Working environment: -20℃~50℃,<95%RH (no dew)
Power Source: DC3.7V Li-on battery, 3200mAh
Charging time: 6h~8h
Working time: ≥ 8h continuously (without alarming)
Gas Sensor Life: 2 years
Explosion-proof grade: Ex ib IIB T4 Gb
Protection Grade: IP65
Weight: Appr. 400g (with battery)
Dimensions: 130mm×67mm×30mm(L × W × H)
4. Operation & Function
4.1 Turn on self check and preheating process
Press the button for 4s and then release it. the LCD
screen of the detector is turned on and the welcome interface is
displayed. The detector automatically performs self-test of the
sound, light and vibration alarm signals, displays the system
information and performs the preheating countdown. The above
information can be used to confirm the integrity of the detector
performance.
4.1.1 After the detector is warmed up and
enters the power-on state, the normal detection
state of the detector is shown in Figure 1. (The
display mode varies according to the position of
the sensor, and the gas in the figure represents
methane.) Fig.1

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 4 -
4.2 Turn off
Under the detector is in the normal detection state, press
button, the LCD screen will prompt the user to shut down the
interface as shown in Figure 2. The user selects whether to shut
down through button. If select sure, the screen no longer
displays any information, and the detector enters Off state.
Fig. 2
4.3 Button function
Detector in the normal detection state as shown in Figure 3:
Fig. 3
Press any key to turn on the backlight (user can set the backlight
time)
Press to enter the function menu;
Press the button to turn the gas sampling pump on or off.
Press to enter the gas concentration curve display interface;
meanwhile press to view the concentration curve of different
gases.
Shut down?
BACK SURE

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 5 -
4.3.1when gas detector detects the gas less than the low alarm (
Mark: for oxygen, between the low alarm and high alarm value), g
as detector is under normal detecting state, no alarm at this time.
4.3.2 When the detected gas concentration is higher than low
alarm value and lower than the high alarm value (Note: When the
oxygen concentration is lower than the low alarm value), the
detector is in a low alarm state. The buzzer emits an alarm sound
of “beep,beep...”every 0.5s, and the red alarm light flashes
synchronously. At the same time, the gas concentration value on
the screen turns yellow and the backlight and vibrator are also
turned on remind of low alarm; press button to cancel the
audible alarm, but still display alarm information. The buzzer will
resume ringing until the new alarm is triggered, and the alarm light
and vibrator will resume operation. When the gas concentration
value detected by the detector returns below the low limit alarm
value, the gas concentration value turns green and the alarm
signal is automatically released. When the alarm is detected, the
detector is shown in Figure 4.
Fig.4
4.3.3 When the detected gas concentration is higher than the set
high alarm value, the detector is in the high alarm state, at this

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 6 -
time the buzzer emits “beep,beep,…every 0.25s”. the red
indicator light flashes synchronously, and the color of the gas
concentration value changes to red on the screen. The backlight
and the vibrator also open at the same time, indicating the high
alarm; long press the button to release the audible alarm. but
there is still an alarm message. The buzzer will resume ringing
until the new alarm is triggered, and the alarm light and vibrator
will resume operation. When the gas concentration value detected
by the detector returns below the low alarm value, the gas
concentration value turns green and the alarm signal is
automatically released. When the alarming, the detector is shown
in Figure 5.
Fig 5
4.3.4 When the detected gas concentration is higher than the test
range, the buzzer of the detector will beep normally, the LCD
screen will be on constantly, the vibrator will be normally open, and
“HHHH” will be displayed on the screen, indicating over-range. at
that time, you can press the button to release the audible
alarm.
4.3.5 Display time, gas sampling pump flag , buzzer indication
(signal when no mute is , when mute is ), the
corresponding concentration information display of different gas

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 7 -
types, battery power, current date.
Note: The above alarm sounds can be manually cleared by
pressing the key. After clearing, the alarm information is still
displayed normally; if the detector triggers the alarm again, the
corresponding alarm sound information can be sent again.
4.4 Use and set functions
The detector has a total of alarm record viewing function, gas
concentration curve display, setting gas parameters, calibration,
zero calibration, language setting, information viewing, time setting,
etc. in the normal detection interface, press button to enter the
function menu, as shown in Figure 6. Then press button
to enter the function selection. When the cursor moves to the
return type on the function selection interface, continue to
return to the normal monitoring interface.
Fig.6
Note: The following menus and function keys exit to return to the
monitoring interface. This operation is no longer repeated.
4.4.1 Alarm Record Gas curve display
1. Alarm record function:
Press the button in the normal detection main interface to enter
the function menu, select the , press button to enter the

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 8 -
alarm record interface, the record content includes the type of
alarm sensor, enter the corresponding sensor to
display the alarm value, alarm time, as shown in
Figure 7: This function can view the gas alarm
history.
Fig.7
The maximum alarm record is up to 1200 groups.
2. Curve display function:
Press in the normal detection main interface to enter the curve
display interface.
Then press button to view the gas concentration curve of each
channel sensor. Each channel sensor has a curved display. As
shown in Figure 9:
Oxygen as an example:
Fig.8
06/08
21:18
O2 20.9% Low 19.5 % High 23.5%
Description
The abscissa X axis is the time,
the current record within 10s of
the curve
The vertical axis Y is the sensor
value, data of the sensor can be
seen from the Y axis
Note: the axis and the data is green, the data curve is blue, the
low alarm line is yellow, high alarm line is red.

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 9 -
4.4.2 Detector menu settings
In the normal detection mode, press to enter the menu, then
press to select the shut down,setting,alarm point, alarm record
calibration zero, calibration, time setting and system reset. These
function menus can be displayed cyclically and can be selected by
cursor movement.
The menu function is as follows:
Description
Setting content
Main
interface
Press to move the cursor to the return
function, press to return to the main interface.
(The cursor moves and selects the entry according to
the above method, and will not be described later)
Shut down
Press to move the cursor to the return
function, press to enter the shutdown selection
interface.
System
Settings
Enter the system settings interface to set the detector
channel, delete alarm records, view system
information, language settings, backlight settings,
and air pump settings.
Alarm
point
Set the low and high value
Alarm
record
Alarm record information of the corresponding
channel gas
Gas Zero
Enter the zero calibration interface to zero the four
sensors.

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 10 -
The menu function is as follows:
fig. 9
fig10
Alarm Settings
Select alarm settings under the menu interface,
press the button like this choose the gas you
need to detect. Take the natural gas for
example, as the picture 10 shows,press the
button to edit, after finishing it,press the to
save it.
Gas
Calibration
Enter the calibration interface to calibrate the four
sensors.
Date time
Set the current date and time setting
System
reset
Enter the password "1111" for factory reset

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 11 -
fig11
fig12
Zero calibration and calibration
Select Zero calibration under the menu
interface, press the button ,Take the natural
gas for example, as the picture 11 shows,press
the button to edit, after finishing it, press the
to save it.press the button return to the
upper level menu.
Attention:1. Zero calibration should be carried
out in clean air, otherwise the degree of gas
concentration in the environment will affect the
accuracy of the portable pumping gas detector.
2. The detector is calibrated passing the
standard gas before leaving factory, and users
should not operate this item.
fig13
The Date
Select The Date under the menu interface,enter
the time setting interface, as the picture 19
shows , press the button to move the
cursor, press the to set the system time,
after the modification is completed, the cursor
moves to the last press of the , return to the
main menu.
4.4.3 the system setting of the detector
Display screen parameters
Detail Features

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 12 -
fig14
fig15
1. Press the button enter the feature
menu, moves the cursor to system
settings by pressing , as the
picture 8 shows. Enter system settings by
pressing , as the picture 14 shows
fig16
fig17
2. Channel settings include channel
switch, gas type, unit selection, accuracy,
and range settings. We take the natural
gas for example, as the picture 10
shows,moves the cursor to select gas type
(natural gas) by pressing .While
the cursor moves to natural gas, you can
enter the natural gas channel settings by
pressing the ,like picture 11 shows.
Press the to adjust the feature and
return to channel settings .
Don’t change the alarm value parameter,if
there is no special requirement.

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 13 -
fig18
fig19
3. Clear record
Use the to clear record, press
to enter the input password interface, like
picture 18 shows, the password is “1111”,
enter the clear password judgment, as
picture 19 shows.
fig20
4. System information
Press the to select the system
information, press the enter the system
information interface. As the picture 20
shows.
fig21
5. Language
Press the to select language,
then press the enter the language
settings interface.As the picture 21 shows.

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 14 -
fig 22
6. backlight settings
press the select backlight
settings, and press the enter the
backlight settings interface. As the
picture 22 shows.
fig23
7.Air pump setting
4.4.4. Alarm information
The following table shows the details of each alarm:
Alarm type
Low alarm:
Slowly tweak the alarm tone
The alarm indication is yellow
The red alarm light flashes
vibrating
High alarm:
Abnormal harsh tone of the drop alarm sound
The alarm indication is red
The red alarm light flashes
vibrating

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 15 -
4.5. Charging
Please charge the detector when it shows low battery or the
detector can’t be turned on due to low battery. Before charging,
please turn off the detector. After you connect the charger correctly
between the detector and AC power source, the detector will be
turned on automatically. When the battery mark on the screen is
full and doesn’t change any more, it means the charging is
completed. Please pull off the charger.
Warning: During charging status, the detector can’t detect the gas
leakage. Please do not try to charge it at testing places to avoid
fire or explosion. Please do not charge it when the detector is
working to avoid potential damage.
Note: Make sure full charge for at least once within 3 months
since production date.
5. Possible fault and corresponding solution
Possible fault
Possible reason
Corresponding solution
The detector
can’t be turned
on
Too low battery
Please charge it in time.
The detector dies
Please contact the
manufacturer of dealer
Fault of electric
circuit
Please contact the
manufacturer of dealer
No response to
the gas
Warm up is not
finished
Wait till warm up is
finished
Fault of electric
circuit
Please contact the
manufacturer of dealer

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 16 -
Inaccurate
indication
Sensor is overdue
Please contact the
manufacturer or dealer to
replace the gas senor
Uncalibrated for long
time
Please calibrate it in time
Fault indication
of time
Battery voltage is
used up
Please charge it and reset
time
Strong
electromagnetism
disturb
Please reset time
Zero calibration
is unavailable
Too much zero drift
of gas sensor
Please calibrate or
replace the gas sensor
Minus gas level
displayed
Gas sensor drift
Calibrate zero point
Sensor fault
indication
Sensor fault
Please contact the
manufacturer or dealer to
replace the gas senor
6. Notices
6.1 Falling down from high places or strong shake is prohibited.
6.2 The detector may not work properly at interferential
high-concentration gas.
6.3 To avoid incorrect result or possible damage to the detector,
please operate and handle the detector in accordance with the
manual.
6.4 The detector should be not stored or used neither under the
circumstance with caustic gas (such as Cl2), nor under the other
rugged circumstances, including excessive high or low
temperature, high humidity, electromagnetic field and strong

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 17 -
sunshine.
6.5 If there is dust on the surface of the detector after a long-term
use, please clean it lightly with clean soft cloth. The surface may
be scraped or destroyed with caustic solvent or hard things.
6.6 To assure the testing accuracy, the detector should be
calibrated periodically. And the calibration period should be less
than one year.
6.7 Please put the used Lithium batteries to the appointed places
or send to our company. Don’t discard them into the dustbin at
random.
7. Standard accessories:
Suit case packaging 1pc
Charger 1pc
Calibration cap 1pc
Communications cable 1pc
User manual 1pc

Bosean Electronic Technology Co.,Ltd www.bosean.com
- 18 -
Appendix
Gas
Range
L-alarm
H-alarm
CH4
0-100%LEL
20%LEL
50%LEL
C3H8
0-100%LEL
20%LEL
50%LEL
H2
0-100%LEL
20%LEL
50%LEL
H2
0-1000ppm
35ppm
250 ppm
H2S
0-100ppm
10ppm
15ppm
H2S
0-100ppm
10ppm
20ppm
CO
0-1000ppm
35ppm
200ppm
CO
0-1000ppm
30ppm
60ppm
C2H4O
0-20ppm
10ppm
15ppm
C2H4
0-100%LEL
20%LEL
50%LEL
C2H4
0-20ppm
5ppm
10ppm
O2
0-30%vol
19.5%vol
23.5%vol
C2H5OH
0-100%LEL
20%LEL
50%LEL
NH3
0-100ppm
25ppm
50ppm
CL2
0-20ppm
5ppm
10ppm
O3
0-20ppm
5ppm
10ppm
SO2
0-20ppm
2ppm
5ppm
SO2
0-100ppm
2ppm
5ppm
PH3
0-20ppm
0.3ppm
5ppm
PH3
0-5ppm
0.3ppm
2ppm
NO
0-250ppm
20ppm
50ppm
NO2
0-20ppm
5ppm
10ppm
HCN
0-500ppm
10ppm
20ppm
HCN
0-50ppm
10ppm
20ppm
HCL
0-50ppm
10ppm
20ppm
CH2O
0-10ppm
2ppm
5ppm
VOC
0-100ppm
20ppm
50ppm
C6H6
0-100ppm
20ppm
50ppm
CO2
0-5000ppm
1000ppm
2000ppm
CO2
0-50000ppm
1000ppm
2000ppm
Other Bosean Electronic Technology Gas Detector manuals