ThermoScan®
PRO 6000
Cleaning and
maintenance
guide
What aects accuracy?
• Damage, modication or adjustment to the lens window.
Do not touch lens directly.
• Use of cleaners other than alcohol.
• Fingerprints, cerumen or dust on window.
(if placed in ear without a probe cover, clean immediately)
• Reuse of probe covers
Cleaning lens window & probe
Cleaning body
Cleaning accessories
Only use cleaning solutions approved for the thermometer body such as:
Alcohol, 10% bleach solution, Oxivir, Virox, CaviWipes™, Clinell or SaniCloth.
Remove probe cover box carrier
from the cradle by rotating the probe
cover box carrier forward
Wipe the cradle and probe cover carrier with a
damp cloth or cleaning wipe with cleaning an
approved cleaning solution.
About Braun ThermoScan®
disposable probe covers:
• To ensure accurate
measurements always
use Braun ThermoScan®
Probe Covers.
Intended for single use only!
Always use a new probe cover
for each measurement to
ensure accuracy and hygiene.
• Never use a damaged,
perforated or soiled
probe cover. This can
lead to inaccurate
measurements.
This product is manufactured by Kaz USA, Inc. under
license to the‘Braun’trademark. ‘Braun’is a registered
trademark of Braun GmbH, Kronberg, Germany.
Welch Allyn Technical Support
visit locations (www.welchallyn.com/about/company/locations.html)
®
1
2
1
1
1
3
2
2
Gently wipe surface of probe lens window with cotton swab or
cloth slightly moistened with alcohol. To prevent damage to the
sensor only use gentle pressure when cleaning. Gently wipe dry.
With probe facing down, wipe the probe
with a cleaning wipe moistened with alcohol.
Gently wipe dry.
Allow at least 5 minutes drying time before taking
temperature. Make sure probe lens window is clean
and dry prior to use.
For additional protection, we recommend placing a new
probe cover on the thermometer probe to protect this area
when cleaning the body of the thermometer.
Use a damp cloth or cleaning wipe with moistened with an
approved cleaner. To clean the body, ensure that the wipe is
moist, not saturated. Wipe the body with the display facing up.
Only use cleaning solutions approved for the thermometer body such as:
Alcohol, 10% bleach solution, Oxivir, Virox, CaviWipes™, Clinell or SaniCloth.
Cradle
Tether To clean the tether, ensure that the wipe is moist,
not saturated. Wipe the tether with a damp cloth or
cleaning wipe with an approved cleaning solution.
Accuracy check:
It is recommended that accuracy should
be checked on an annual basis or whenever
clinical accuracy of the thermometer is in
question using the 9600 Plus Calibration
Tester (REF 01802‑110).
17”
11”
Spot Colors
Brand:
Braun
Category: Thermometers
Model: PRO 6000
Oracle Desc. : POSTER, CLEANING AND
MAINTENANCE, PRO6000, ENG, 2014
Subject: PRO 6000 Cleaning &
Maintenance Poster
Artwork Part #: 31BR6000190
Die Line Part #: NA
Region: US
Flat Size:
MM: 431.80 x 279.40
IN: 17 x 11
Folded Size: NA
Material:
Page count: 1
Revision:
1
Release Date:
Date: 23MAR15
Scale: 1/1
Colors:
Special Instructions:
Dielines (Do not print)
Cyan
Coating:
Overall Gloss
Varnish:
Spot UV
100%
Magenta
100%
Yellow
100%
Black
100%
PMS
XXXX
PMS
XXXX
Kaz USA, Inc.
Creative Services, 250 Turnpike Road
Southborough, 01772 MA, USA
+1 508 490 7000
Quality Requirement of Artwork and Quality
Clarification Process of Artwork Printing:
Meet Eng-QS-06&02
CREATIVE DEPARTMENT
ARTWORK SPECIFICATIONS