Brooks CoolCell SV10 User manual

CoolCell®SV10 CoolCell®SV10
Controlled-rate cell freezing container
Controlled-rate cell freezing container
Quick Start
• The six chambers and 10ml serum vials should be dry to avoid vial sticking upon freezing.
• Make sure the core (black ring) is at room temperature and seated in the bottom of the central cavity.
• Place sample vials, each containing 5ml of cell suspension, in each well. For reproducible freezing
profiles, each well should contain a filled vial; “blank” vials containing 5ml of freezing medium should be
placed in empty chambers if freezing less than six vials of cell suspension.
• Check that the tubes slide in and out freely.
• Fully seat the lid on CoolCell SV10 container.
• Place CoolCell SV10 container into a -8ooCfreezer or dry ice locker. Ensure that there is at least one inch
of free space clearance around the CoolCell SV10 module.
• Freeze for four hours before transferring samples to archive storage.
Transferring frozen samples to archive storage
• Prepare an insulated pan or container with a one inch (2.5cm) layer of pulverized or pellet dry ice.
• Remove CoolCell SV10 container from the freezer and gently remove the lid using a twisting and rocking
motion.
• Vial tops will be exposed once lid is removed and vials should be quickly placed onto the dry ice.
Special Notes:
• Always use dry ice to transfer frozen vials containing cells to permanent storage to avoid temperature
rise and cell damage. Vial contents can rise from -800Cto over -500Cin less than one minute if exposed to
room temperature air.
• It is strongly recommended that all frozen cell cultures be checked for viability before the stock culture is
terminated.
Recycling CoolCell SV10 to room temperature
The CoolCell SV10 container is ready to freeze again as soon as the core (black ring) is at room temperature.
To rapidly recycle CoolCell SV10 container to room temperature, remove the center solid core ring by invert-
ing and tapping on a surface. The CoolCell SV10 body and lid will return to room temperature in 10 to 15
minutes. Check that all chambers are dry. Dry the core ring before re-inserting into the central chamber.
About CoolCell SV10
The CoolCell SV10 container, in combination with a -800Cfreezer or dry ice locker, will provide the freezing
rate of -10Cper minute that is ideal for cryo-preservation of most cultured cell lines. The CoolCell SV10 design
uses a combination of insulation foam, radial symmetry, and a heat transfer core to regulate heat loss rather
than using a large thermal mass (alcohol-based freezing container).
Quick Start
• The six chambers and 10ml serum vials should be dry to avoid vial sticking upon freezing.
• Make sure the core (black ring) is at room temperature and seated in the bottom of the central cavity.
• Place sample vials, each containing 5ml of cell suspension, in each well. For reproducible freezing
profiles, each well should contain a filled vial; “blank” vials containing 5ml of freezing medium should be
placed in empty chambers if freezing less than six vials of cell suspension.
• Check that the tubes slide in and out freely.
• Fully seat the lid on CoolCell SV10 container.
• Place CoolCell SV10 container into a -8ooCfreezer or dry ice locker. Ensure that there is at least one inch
of free space clearance around the CoolCell SV10 module.
• Freeze for four hours before transferring samples to archive storage.
Transferring frozen samples to archive storage
• Prepare an insulated pan or container with a one inch (2.5cm) layer of pulverized or pellet dry ice.
• Remove CoolCell SV10 container from the freezer and gently remove the lid using a twisting and rocking
motion.
• Vial tops will be exposed once lid is removed and vials should be quickly placed onto the dry ice.
Special Notes:
• Always use dry ice to transfer frozen vials containing cells to permanent storage to avoid temperature
rise and cell damage. Vial contents can rise from -800Cto over -500Cin less than one minute if exposed to
room temperature air.
• It is strongly recommended that all frozen cell cultures be checked for viability before the stock culture is
terminated.
Recycling CoolCell SV10 to room temperature
The CoolCell SV10 container is ready to freeze again as soon as the core (black ring) is at room temperature.
To rapidly recycle CoolCell SV10 container to room temperature, remove the center solid core ring by invert-
ing and tapping on a surface. The CoolCell SV10 body and lid will return to room temperature in 10 to 15
minutes. Check that all chambers are dry. Dry the core ring before re-inserting into the central chamber.
About CoolCell SV10
The CoolCell SV10 container, in combination with a -800Cfreezer or dry ice locker, will provide the freezing
rate of -10Cper minute that is ideal for cryo-preservation of most cultured cell lines. The CoolCell SV10 design
uses a combination of insulation foam, radial symmetry, and a heat transfer core to regulate heat loss rather
than using a large thermal mass (alcohol-based freezing container).
Instructions for Use
Instructions for Use
As a result, freezing profiles are extremely consistent from one run to the next. Also, because of this low ther-
mal mass, the CoolCell SV10 container will not cause a rise in local freezer temperature and will protect nearby
samples already stored in the freezer. Low thermal mass also means CoolCell SV10 container will rapidly return
to room temperature for another freezing cycle (see fast recycle instructions above).
CoolCell SV10 freezing performance
The CoolCell SV10 container will freeze six serum vials
each containing 5ml of cell suspension at -1oCper minute
when placed in a -8ooCenvironment (mechanical freezer
or dry ice locker). The graph (left) displays a typical
freezing profile obtained using these conditions.
Note: “CC-12” refers to CoolCell 12-well 1ml cryovial
freezer, #BCS-136.
Troubleshooting
Care and Cleaning
The CoolCell SV10 container is constructed of closed cell crosslinked polyethylene foam and a solid thermo-
conductive core. CoolCell SV10 container is compatible with prolonged cryogenic temperature exposure. The
foam may be cleaned by water and mild soap. Rinse and dry thoroughly. CoolCell SV10 container is resistant to
alcohols and 10% bleach solutions. Do not autoclave. Maximum temperature exposure: 60oC. Avoid prolonged
exposure to UV light sources.
As a result, freezing profiles are extremely consistent from one run to the next. Also, because of this low ther-
mal mass, the CoolCell SV10 container will not cause a rise in local freezer temperature and will protect nearby
samples already stored in the freezer. Low thermal mass also means CoolCell SV10 container will rapidly return
to room temperature for another freezing cycle (see fast recycle instructions above).
CoolCell SV10 freezing performance
The CoolCell SV10 container will freeze six serum vials
each containing 5ml of cell suspension at -1oCper minute
when placed in a -8ooCenvironment (mechanical freezer
or dry ice locker). The graph (left) displays a typical
freezing profile obtained using these conditions.
Note: “CC-12” refers to CoolCell 12-well 1ml cryovial
freezer, #BCS-136.
Troubleshooting
Care and Cleaning
The CoolCell SV10 container is constructed of closed cell crosslinked polyethylene foam and a solid thermo-
conductive core. CoolCell SV10 container is compatible with prolonged cryogenic temperature exposure. The
foam may be cleaned by water and mild soap. Rinse and dry thoroughly. CoolCell SV10 container is resistant to
alcohols and 10% bleach solutions. Do not autoclave. Maximum temperature exposure: 60oC. Avoid prolonged
exposure to UV light sources.
Problem Solution
Vials do not freely fit in the
chambers
The CoolCell SV10 container is designed to fit standard serum vial
shaped vessels up to 23mm in diameter and up to 47 mm in height.
Check that flag style labels, if used, will not bind and hinder insertion
or removal.
Vials are stuck in CoolCell SV10
after freezing
It is likely moisture was in the vial chambers or on the sample vial prior
to freezing. Remove the core (black ring) and tap the CoolCell SV10 to
dislodge vials.
The lid does not fully seat Ensure that sample vials are properly seated. The maximum height of
the tube is 47mm.
Problem Solution
Vials do not freely fit in the
chambers
The CoolCell SV10 container is designed to fit standard serum vial
shaped vessels up to 23mm in diameter and up to 47 mm in height.
Check that flag style labels, if used, will not bind and hinder insertion
or removal.
Vials are stuck in CoolCell SV10
after freezing
It is likely moisture was in the vial chambers or on the sample vial prior
to freezing. Remove the core (black ring) and tap the CoolCell SV10 to
dislodge vials.
The lid does not fully seat Ensure that sample vials are properly seated. The maximum height of
the tube is 47mm.
15 Elizabeth Dr., Chelmsford, MA 01824 USA brookslifesciences.com/contact-us brookslifesciences.com
CoolCell SV10 Instructions, May 2018
CoolCell SV10 ©Brooks Automation, Inc. All rights reserved.
15 Elizabeth Dr., Chelmsford, MA 01824 USA brookslifesciences.com/contact-us brookslifesciences.com
CoolCell SV10 Instructions, May 2018
CoolCell SV10 ©Brooks Automation, Inc. All rights reserved.

Other Brooks Laboratory Equipment manuals

Brooks FluidX Aperio Series User manual

Brooks

Brooks FluidX Aperio Series User manual

Brooks MagnaTran LEAP User manual

Brooks

Brooks MagnaTran LEAP User manual

Brooks CryoPod Carrier User manual

Brooks

Brooks CryoPod Carrier User manual

Brooks IntelliXcap AcoustiX User manual

Brooks

Brooks IntelliXcap AcoustiX User manual

Brooks IntelliXcap User manual

Brooks

Brooks IntelliXcap User manual

Brooks MARATHON LEAP AX User manual

Brooks

Brooks MARATHON LEAP AX User manual

Brooks VX TopCooler User manual

Brooks

Brooks VX TopCooler User manual

Brooks Marathon Express User manual

Brooks

Brooks Marathon Express User manual

Brooks CryoPod User manual

Brooks

Brooks CryoPod User manual

Brooks FluidX XSD-1 User manual

Brooks

Brooks FluidX XSD-1 User manual

Brooks CoolBox 2XT User manual

Brooks

Brooks CoolBox 2XT User manual

Popular Laboratory Equipment manuals by other brands

Logicube TALON ULTIMATE user manual

Logicube

Logicube TALON ULTIMATE user manual

Siemens Hematek operating manual

Siemens

Siemens Hematek operating manual

Metrohm 814 manual

Metrohm

Metrohm 814 manual

Rainin Pipetman P-2 manual

Rainin

Rainin Pipetman P-2 manual

COBI PLATEC X-PLA quick guide

COBI PLATEC

COBI PLATEC X-PLA quick guide

Teledyne ACCQPrep HP125 Installation and operation guide

Teledyne

Teledyne ACCQPrep HP125 Installation and operation guide

Data Translation DT2802 user manual

Data Translation

Data Translation DT2802 user manual

cytiva Biacore 8K operating instructions

cytiva

cytiva Biacore 8K operating instructions

Mecmesin MultiTest-i Series Assembly and installation

Mecmesin

Mecmesin MultiTest-i Series Assembly and installation

Labconco Protector XL 1124 Series user manual

Labconco

Labconco Protector XL 1124 Series user manual

NuAire LabGard NU-640-400 Operation and maintenance manual

NuAire

NuAire LabGard NU-640-400 Operation and maintenance manual

Sigma SGP-1000-2 instruction manual

Sigma

Sigma SGP-1000-2 instruction manual

IKA C-MAG HS 10 Service manual

IKA

IKA C-MAG HS 10 Service manual

Labnet Spectrafuge 16M instruction manual

Labnet

Labnet Spectrafuge 16M instruction manual

GeneReach Taco 24 user manual

GeneReach

GeneReach Taco 24 user manual

IKA Algaemaster 10 operating instructions

IKA

IKA Algaemaster 10 operating instructions

Phcbi MBR-305GR operating instructions

Phcbi

Phcbi MBR-305GR operating instructions

Prior PureFocus850 user manual

Prior

Prior PureFocus850 user manual