5Bualo Filter® VisiClear™ Operator’s Manual
Standard Accessories:
• Operator’s Manual
• Power Cord
• Pneumatic Footswitch
Please contact BUFFALO FILTER® Customer Service to purchase the following accessories:
• Replacement Filters
• EZLink™ Automatic Activation Device
• Hoses, Tubing, Laparoscopic Kits, Adapters, Wands & Other Accessories
1.3 Operational Information
e operational information contained in this section is intended for the customer review of
regulatory issues. e information pertains to the use of the products both domestically and
internationally:
1. e BUFFALO FILTER® VisiClear™ Surgical Smoke Plume Evacuator(s) complies
with IEC60601-1:2005 electrical specications in the following systems:
100/120 VAC, 50/60 Hz, 220/240 VAC, 50/60 Hz
2. Type of protection against electrical shock: Class I
3. Degree of protection against electric shock: Type CF Applied Part
4. Degree of protection against ingress of water: IPX1
5. Method of sterilization or disinfection recommended by BUFFALO FILTER®:
Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution
or soapy water. Wipe dry with a clean cloth. Do not steam sterilize.
6. Degree of safety of application in the presence of ammable anesthetic mixture with air or
with oxygen or nitrous oxide: Not Suitable
7. Mode of operation: Continuous
8. e fuses on the circuit board are (F1, F20 to be serviced by an authorized BUFFALO
FILTER® technician as follows:
100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo)
with Interruption Rating: 1500A@250V
220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-Blo)
with Interruption Rating: 1500A@250V
9. e fuse on the transformer circuit (F6) to be serviced by an authorized BUFFALO FILTER®
technician as follows:
100/120 VAC, 50/60 Hz use 100 Amp 250 Volt Fuse (Fast-Act)
with interruption rating: 1500A@250V
220/240 VAC, 50/60 Hz use 50Amp 250 Volt Fuse (Fast-Act)
with interruption rating: 1500A@250V
10. e fuses on the motor circuit are to be serviced by an authorized BUFFALO FILTER®
technician as follows: 220/240 VAC, 50/60 Hz use 3,15 Amp 250 Volt Fuse (Fast-Acting), (F3)
with Interruption Rating: 1500A@250V
11. is equipment needs special precautions regarding Electro Magnetic Compatibility and
needs to be installed according to EMC information found in this manual.