PlumeSafe® Whisper602 Service Manual
Section B.
Regulations and Compliance
The operational information contained in this section is intended for the customer review of
compliance issues. The information pertains to the use of Buffalo Filter products both
domestically and internationally:
1. The PlumeSafe Whisper 602 complies with IEC601.1 electrical specifications in the following
systems: PlumeSafe Whisper 602 -120V/60Hz, 100V/50Hz & 60Hz , 220V/50 Hz,
240V/50Hz
2. The PlumeSafe Whisper 602 Smoke Evacuation System(s) has a classification under the
UL2601-1, Clause 5.1 of : CLASS 1 Equipment
3. The PlumeSafe Whisper 602 Smoke Evacuation System(s) has an identification under the
UL2601-1, Clause 5.2 according to the degree of protection against electric shock: Type B
Equipment
4. The PlumeSafe Whisper 602 Smoke Evacuation System(s) has an identification under the
UL2601-1, Clause 5.3 according to the degree of protection against harmful ingress of water:
Ordinary Equipment (enclosed equipment without protection against ingress of
water)
5. The PlumeSafe Whisper 602 Smoke Evacuation System(s) has a classification under the
UL2601-1, Clause 5.4 according to the method of sterilization or disinfection recommended by
BUFFALO FILTER: Unplug System. Wipe system with a damp cloth containing mild
surface disinfectant solution, or soapy water. Wipe dry with a clean cloth. Do not
steam sterilize or autoclave.
6. The PlumeSafe Whisper 602 Smoke Evacuation System(s) has a classification under the
UL2601-1, Clause 5.5 according to the degree of safety of application in the presence of
flammable anesthetic mixture with air or with oxygen or nitrous oxide: Equipment not
suitable for use in the presence of Flammable Anesthetic mixture with air or with
oxygen or nitrous oxide.
7. The PlumeSafe Whisper 602 Smoke Evacuation System has a classification under the UL2601-
1, Clause 5.6 according to the mode of operation: Continuous Operation
8. The fuses used on the circuit board are to be serviced by an authorized Buffalo Filter
technician as follows: 120V/60Hz, 100V/50 Hz & 60 Hz, 220V/50Hz and 240V/50Hz
use a 0.0625A/250V fuse
9. The PlumeSafe Whisper 602 Smoke Evacuation System and all filters and accessories are not
intended for contact with patients.
10. The PlumeSafe Whisper 602 Smoke Evacuation System is in full compliance with the Medical
Devices Directive II, Council Directive 93/42/EEC of 14 June 1993
11. The PlumeSafe Whisper 602 Smoke Evacuation System complies with the City of Los
Angeles Electrical Codes.