
Walk200b
2. SAFETY INFORMATION
4
specific materials may cause additional risks to operators, patients or bystanders or may damage the
device. Do not sterilise the device or the patient cable with ethylene oxide gas (EO). Refer to Section
7 for instructions on proper cleaning and disinfection.
Do not leave the air tube unattended in the presence of children as they could be accidentally
strangled.
The Walk200b device is designed to be connected to a PC. The system that consists of the Walk200b
device and the non-medical PC equipment is defined as an electromedical system. The adoption of
an electromedical system requires compliance with the requirements of the Directive.
The use of a medical separation transformer must be used with the electromedical system consisting
of the Walk200b, a PC and any other non-medical equipment (such as printer and monitor). Consult
a CARDIOLINE® Service Centre for more detailed information. Alternatively, a PC compliant with
medical standard EN60601-1 can be used. A separation transformer cannot be used unless
compliance with the following conditions has been confirmed with absolute certainty:
After measurements taken according to the EN60601-1 standard, the dispersion currents in
the case of the PC and of any other non-medical devices that are connected (monitor,
printer, etc.) are less than 0.1 mA.
The non-medical devices in the system are connected outside the patient area, that is, not
within 1.5 m of the patient and/or his or her bed.
If a separation transformer is not used, install the system in such a way that the operator
cannot touch the patient and the case of the any non-medical equipment outside the
patient area at the same time.
If the PC and any other non-medical devices that are connected (monitor, printer, etc…) are
powered by the medical separation transformer, they may be located inside the “patient
area”, i.e. within 1.5 m of the patient and/or his or her bed.
The PC and any other devices to be used in the system (monitor, printer, etc.) must only be
connected to the separation transformer using the cables and any other accessories
supplied with the system. Extension cords, socket adapters and other types of connections
may not be used unless supplied with the system.
It is categorically forbidden to connect other equipment that is not part of the system to the
separation transformer or to a multiple extension cord supplied with the system.
If a non-medical device that is part of the system is connected to a non-prescribed type of
socket (a wall socket, for example), the system no longer complies with medical standards
and the patient may be endangered.
If a non-standard device is connected to the socket of the separation transformer specified
for the system devices, this could create situations of danger for the patient, operators or
environment (overloading or overheating of the sockets and separation transformer.
The separation transformer and any multiple sockets connected to it must not be placed
directly on the floor or any area that may be flooded, or in which dirt may accumulate. The
transformer and sockets must be placed in an area that is easy to inspect, both for regular
cleaning and for the required maintenance.
The installer must install the system in the optimum way to satisfy the safety regulations and
provide the greatest facility and ease of use by both the user and the patient.