TSV-1000 Operation
By default, the device will start in TACHOMETER mode
and will pulse at RPM preset # 1 of 10.
To switch to STROBOSCOPE mode, press Swhile in
TACHOMETER mode.
To switch to TACHOMETER mode, press Twhile in
STROBOSCOPE mode.
• TACHOMETER MODE
To test a non-contact tachometer, hold the tachometer
so that its sensor is 0.5-2 inches / 12-50mm from the
TSV-1000 pulse sensor.
When the tachometer is in measuring mode its display
should be in-sync or close to in-sync with TSV-1000’s
display. The allowable tolerance will depend on the ta-
chometer’s specifications.
While in TACHOMETER mode, press SP+ or SP- to
toggle between presets (10).
Press ADJ+ or ADJ- to adjust the pulse rate.
To adjust the presets press DEFAULT, the display will
show Def X: Y where X is the preset number (0—9) and
Y is the preset value in RPM.
Press SP+ or SP- to toggle between presets (10)
Press ADJ+ or ADJ- to adjust a preset.
Press Default to save changes and exit the preset ad-
justment mode returning to normal operating mode.