Calibration Instruction 1Calibration Instruction for Salwico ST400 H2S
5103028_Salwico ST400 H2S_Calibration Instruction_I0_EN_2019_C 1
1 Calibration Instruction for Salwico ST400
H2S
1.1 Preliminaries
This instructions applies to the following products:
Figure 1. Salwico ST400 H2S, 0-30 ppm 4-20 mA. Part no. 4462
The H2S sensor is Electrochemical and measures H2S content by means of
chemical reaction.
DANGER!
H2S – Hydrogen Sulphide – is toxic and dangerous also in rather low concentrations.
H2S is a sticky gas and smells very distinctly from rotten eggs.
Consilium Replacement article number for ST400 H2S detectors is no. 004462-S.
A new sensor is supplied without the metal housing and must be fitted in the
housing. This applies to both H2S and O2sensors of ST400 type.
Calibration equipment
Normal measuring range is 0-30 ppm, our standard calibration bottle no.
5201005-00A contains 20 ppm of gas in air.
For SW2020 a mix-bottle is often used, with H2S 20 ppm as well as N299% and
C3H850% LEL, article no. 5201010-01A.
Test points (A) for the multi-meter as well as “SPAN” (B) and “ZERO” (C)
potentiometers are indicated below.
CAUTION!
Do not adjust the other potentiometers.