DOC-016 Iss 1 3
3) The illumination appears weak.
Check that the intensity settings on the Control Pod are set at an appropriate level.
Has the optical setup procedure been carried out (shown in Section 7 of the User
Manual)?
Has the microscope been setup correctly?
Check that apertures are open.
Check for any ND filters in the optical path.
Check that an appropriate filter cube set is being used with the selected
excitation wavelength.
4) Illumination is not flat and even at the sample
(homogenous).
Has the optical setup procedure been carried out (shown in Section 7 of the User
Manual)?
5) Illumination is not centralized over the field of view.
Rotate the Light Source in the epi-port whilst observing the field of view.
If the offset moves with the rotation of the Light Source then the system has lost
its alignment. This will need to be returned to CoolLED to be reset.
If the offset does not move with the rotation of the system then the issue is with
the optical path of the microscope. A person that is competent in microscope
servicing will need to investigate this cause.
6) Image of the LEDs on the sample.
Has the optical setup procedure been carried out (shown in Section 7 of the User
Manual)?
Is the pE-100 being used with any additional parts that may increase the optical path
length? For example an external shutter or a pE-Combiner.