Direct Fit
Attach the Light Source directly to the epi-illumination port of a microscope using a suitable CoolLED
adaptor.
Note:
•The CoolLED adaptor is specified on ordering and will already be installed on your Light Source.
•The exact method of attachment depends on the microscope configuration –please follow
microscope manufacturer’s instructions.
•The Light Source may be oriented horizontally or vertically to fit the microscope, provided the
minimum clearance distances are maintained.
A pE-400max Light Source fitted to a microscope in a horizontal orientation.
Liquid Light Guide
Liquid Light Guide configurations accept a 3 mm liquid light guide.
1. Insert a free end of a 3 mm liquid light guide into the Light Source, ensuring that the light guide is
fully seated in the output barrel.
2. While holding the liquid light guide securely in place, use the 1.5 mm Hex Key to gently fasten
the grub screws found at the output of the Light Source.
Warning: do not over-tighten the adaptor retention grub screws. Over-tightening may
damage the liquid Light guide or the Light Source and result in an unsafe or poorly
performing installation.