14 109118-1 Rev. A, 10/96
3.2 Operating Hints
This section gives some brief operating hints and good laboratory
practices to help assure trouble free pH measurements.
1. New electrodes (or electrodes that have been in storage)
should be conditioned in one inch (25 mm) of pH 7.0 or 4.0
buffer for several hours prior to their use. Make sure that the
wetting cap has been removed from the tip of the electrode.
2. Always remove the wetting cap and fill hole plug during
calibration and measurements. Replace the fill hole plug
when done.
3. Calibrate the electrode daily. Though a one point
calibration may be suitable for some applications, we
recommend that a two point calibration (that brackets the
expected pH range) be performed.
4. Be aware that if you are measuring hot or cold samples
without the use of an ATC probe, the values displayed are
not accurate. You should be calibrating and measuring at
room temperature.
5. When transferring from one sample to another, always
rinse the electrode with distilled water and blot dry. Handle
the electrode carefully and do not use it as a stirring rod.
6. Do not use calibration buffers after the expiration date
printed on the package. Keep the bottles tightly capped and
stored according to the manufacturer's instructions. Never
put used buffer back into the bottle of new buffer.
7. For small sample volumes, make sure that both the pH bulb
and the reference junction are in contact with the sample.
8. Response time is a function of both the electrode and the
solution. Some solutions have very fast response while
others, particularly those with low ionic strength, may take
several minutes.
9. Samples must be in solution (water). You cannot measure
the pH of a dry sample.
Refer to the electrode manufacturer's instructions for proper
electrode care and maintenance. See Section 3.3.2 for general
electrode maintenance and Section 3.4.3 for problem solving.
3