CORNING Lambda Plus User manual

Corning®Lambda™
Plus Pipettors
Single-channel, 8-channel,
and 12-channel Pipettors
Instruction Manual

Table of Contents
1. Product Description ..............................................................1
2. Packaging.................................................................................4
3. Setting the Volume................................................................5
4. Aspiration and Dispensing Instructions...........................6
5. Pre-rinsing................................................................................ 8
6. Dense and Viscous Liquids...................................................8
7. Corning® Pipet Tips................................................................ 9
8. Recommendations.................................................................9
9. Recalibration .........................................................................10
10. Troubleshooting ...................................................................13
11. Cleaning and Sterilization..................................................15
12. Spare Parts.............................................................................16
13. Limited Warranty .................................................................17

Instruction Manual | 1
1. Product Description
The Corning® Lambda™ Plus single-channel pipettor is a
volumetric instrument designed to measure and transfer
liquids precisely and safely. It can accommodate volumes
from 0.1 μL to 1,000 μL depending on the model.
Corning Lambda Plus 8-channel and 12-channel pipettors
have been designed for microplate lling. The pipettors
enable precise and simultaneous delivery of 8 or 12 preset
volume doses of liquid. These pipettors are produced in
four volume ranges: 0.5 to 10 μL, 5 to 50 μL, 20 to 200 μL,
50 to 300 μL.
Model Volume Range (μL) Application
2, 10
Measurement and transfer of
micro-volumes, DNA sequencing,
and enzyme assay applications.
20, 50, 100, 200,
300, 1,000
Measurement and transfer of general
aqueous solution, acids, and bases.
The single-, 8-, and 12-channel pipettors feature a digital
counter which shows the pipetting volume. The set
volume is visible in the window on the handle. Volumes
are set by either turning the pipetting pushbutton (A1) or
the adjustment knob (B) in the correct direction.
A1. Pushbutton
A2. Knob
B. Adjustment knob
P. Identication ring
C. Shaft
D. Ejector
E. Tip

2 | CORNING LAMBDA PLUS PIPETTORS
The pipettor design allows the user to lock the volume
setting by sliding the locking ring upwards.
SS
The symbols printed on the handle show locked and
unlocked positions. The volume range of each pipettor
model is shown on the pipetting pushbutton (A1).
Liquids are aspirated into disposable pipet tips attached
to the pipettor shaft. Disposable pipet tips ensure
maximum safety and eliminate the possibility of cross-
contamination of the liquid samples. To protect the user
from contaminated pipet tips, the pipettor is tted with a
built-in tip ejector (D).
Single-channel Pipettor Ejector
The ejector length can be adjusted to accommodate all
common brands of pipet tips. When using narrow tubes,
it may be necessary to remove the tip ejector. This is done
simply by pulling down.
Adjusting the Tip Ejector Length
2 to 1,000 μL Pipettors
The “H“ spacers included with
each pipettor, allow the ejector
length to be adjusted by +1 mm
or +2 mm.
An ”H0” spacer is tted as
standard. The outside shape
of the spacer identies the
adjustment.
2 and 10 μL Pipettors
The 2 μL to 10 μL pipettors may
require the use of an ejector cap
“M” to eciently eject certain
brands of pipet tips. Simply
place the cap (supplied with the
pipettor) on the bottom of the
pipettor shaft and slide the cap
upwards until it surrounds the
bottom of the tip ejector.
M
H

Instruction Manual | 3
Nominal
Volume
(μL) Cat. No.
Volume
(μL)
Accuracy
(%)
Precision
(%)
Fit to Tips
Without
Filter
(μL)
Single-channel Pipettors
2 4070
Min. 0.1 ±40.0 ≤12.0
101 ±2.7 ≤1.30
Max. 2 ±1.5 ≤0.70
10 4071
Min. 0.5 ±4.0 ≤2.80
105 ±1.0 ≤0.60
Max. 10 ±0.5 ≤0.40
20 4072
Min. 2 ±3.0 ≤1.50
20010 ±1.0 ≤0.50
Max. 20 ±0.8 ≤0.30
100 4073
Min. 10 ±1.6 ≤0.80
20050 ±0.8 ≤0.24
Max. 100 ±0.8 ≤0.20
200 4074
Min. 20 ±1.2 ≤0.60
200100 ±0.8 ≤0.25
Max. 200 ±0.6 ≤0.20
1,000 4075
Min. 100 ±1.6 ≤0.40
1,000500 ±0.7 ≤0.20
Max. 1,000 ±0.6 ≤0.15
The accuracy and precision are obtained gravimetrically using
Corning® pipet tips, performing at least 10 measurements
of distilled water at a reference temperature of 20°C ± 1°C
according to EN ISO 8655 standard.
8-channel and 12-channel Pipettor Ejector
The 8- and 12-channel pipettor construction allows the
lower 8- and 12-channel modules to be rotated 360°.
This allows the user to disable the ejector to avoid
accidental use and to change its position to make lling
of microplates more comfortable and convenient. It is
recommended to turn the module clockwise.
The pipettor is a high quality instrument which oers
excellent accuracy and precision.
The accuracy and precision (repeatability) of the liquid
volume depends on the quality of pipet tips used. The
values for accuracy and precision shown in the table below
were obtained using Corning pipet tips.

4 | CORNING LAMBDA PLUS PIPETTORS
Nominal
Volume
(μL) Cat. No.
Volume
(μL)
Accuracy
(%)
Precision
(%)
Fit to Tips
Without
Filter
(μL)
8-channel and 12-channel Pipettors
8 - 10
12 - 10
4080
4084
Min. 0.5 ±10.0 ≤8.0
105 ±4.0 ≤2.0
Max. 10 ±2.0 ≤1.2
8 - 50
12 - 50
4081
4085
Min. 5 ±4.0 ≤2.5
20025 ±3.0 ≤1.2
Max. 50 ±1.6 ≤0.6
8 - 200
12 - 200
4082
4086
Min. 20 ±3.0 ≤1.5
200100 ±1.5 ≤0.8
Max. 200 ±1.0 ≤0.6
8 - 300
12 - 300
4083
4087
Min. 50 ±1.6 ≤1.5
300150 ±1.2 ≤1.0
Max. 300 ±1.0 ≤0.6
◗ Accuracy is the dierence between the volume
dispensed and the test volume selected.
◗ Precision is the degree to which repeated measurements
under unchanged conditions show the same results.
The accuracy and precision are obtained gravimetrically
using Corning® pipet tips, performing at least 10
measurements of distilled water at a reference temperature
of 20°C ± 1°C according to EN ISO 8655 standard.
2. Packaging
The following components are supplied with the pipettor:
◗ Instruction manual
◗ Calibration key
◗ Ejector length adjustment spacers (for single-channel
pipettors)
◗ Ejector cap (for single-channel pipettor models 2 and 10)
◗ Identication labels
◗ Lubricant (for single-channel pipettors)
◗ Color identication rings (for single-channel pipettors).

Instruction Manual | 5
3. Setting the Volume
The volume of the pipettor is set by turning the pipetting
pushbutton (A2) or the black adjustment knob (B).
The volume display consists of three numbers and is
read from top to bottom. The three numbers indicate the
volume selected and are colored black and red.
Red numbers represent decimal fractions for microvolume
pipettors or milliliters for macrovolume pipettors.
An example for each of these pipettors is given on the
next page.
Nominal Volume
(μL)
Counter
Readings
Set
Volume Increment
Single-channel Pipettors
2
1
2
5
1.25 μL 0.002 μL
10
0
7
5
7.5 μL 0.02 μL
20
1
2
5
12.5 μL 0.02 μL
100
0
7
5
75.0 μL 0.2 μL
200
1
2
5
125.0 μL 0.2 μL
1,000
0
7
5
750.0 μL 2 μL
8-channel and 12-channel Pipettors
8 - 10
12 - 10
0
3
5
3.5 μL 0.02 μL
8 - 50
12 - 50
0
6
5
6.5 μL 0.10 μL

6 | CORNING LAMBDA PLUS PIPETTORS
Nominal Volume
(μL)
Counter
Readings
Set
Volume Increment
8-channel and 12-channel Pipettors (continued)
8 - 200
12 - 200
0
8
5
85.0 μL 0.20 μL
8 - 300
12 - 300
2
5
0
250.0 μL 1.0 μL
To attain the maximum accuracy, the set volume must be
approached from a higher value.
If the desired value is lower than the previous value, adjust
the value down to a one-third turn above the required
setting. If the desired value is higher than the previous
value, adjust the value to a one-third turn above the
required value. The setting should then be performed in
the following manner:
◗ Slowly turn the pipetting pushbutton or the black
adjustment knob to the required volume.
◗ If the knob is accidentally turned too far below the
required value, the process must be repeated.
◗ Set the locking ring into the upper position following
the volume adjustment, thus locking the knob and
preventing accidental volume change.
4. Aspiration and Dispensing Instructions
Single-channel Pipettors
Place a pipet tip on the shaft of the pipettor. See Section 6
for the appropriate tip. Press the pipet tip on rmly using a
slight twisting motion. This will ensure an airtight seal.
8-channel and 12-channel Pipettors
To put pipet tips on the shafts: hold the pipettor vertically
and press it against the tips in the rack box until the shafts
retract approximately 1.5 mm into the manifold. The
suspension system of the shafts ensures even and tight
sealing of the pipet tips. The rocking movement does not
have to be performed to seal the pipet tips tightly.
NOTE: Never handle a liquid with a pipettor which has not
been tted with a pipet tip.

Instruction Manual | 7
1234
56
Rest position
First stop Second stop
Aspiration
1.Press the pushbutton to the rst positive stop. Holding
the pipettor vertically, immerse the pipet tip into the
sample liquid. The depth to which the pipet tip is
immersed in the sample liquid depends on the model.
Model Volume Range (μL) Immersion Depth (mm)
2, 10 ≤1
20, 50, 100 2 - 3
200, 1,000 2 - 4
2.Release the pipetting pushbutton slowly and smoothly
to aspirate the sample. Wait one second and then
withdraw the pipet tip from the liquid. If the pipet tip
is not immersed to the recommended depth or if the
pipetting pushbutton is rapidly released, air may enter
the disposable pipet tip.
NOTE: Avoid touching the orice of the pipet tip.
Dispensing
3.Place the end of the pipet tip against the inside wall
of the vessel at an angle of 10° to 40°. Press the
pushbutton smoothly to the rst stop. Wait one second.
4.Press the pushbutton to the second stop to expel any
remaining liquid. Keep the pushbutton depressed to the
very end, and remove the pipettor by drawing the pipet
tip against the inside surface of the receiving vessel.
5.Release the pushbutton to its starting position.
6.Eject the pipet tip by pressing the tip ejector button.
Remember to change the pipet tip whenever a dierent
type of liquid is to be sampled.

8 | CORNING LAMBDA PLUS PIPETTORS
5. Pre-rinsing
When pipetting liquids of higher viscosity or lower surface
tension than water (e.g., sera or organic solvents), a lm
of liquid is formed on the inside wall of the pipet tip. This
lm can create an error. As the lm remains relatively
constant in successive pipetting operations with the same
tip, this error can be eliminated by forming the lm before
transferring the rst sample. This is done by aspirating a
sample and dispensing it back into the same vessel. Now,
as the lm is already formed, all of the following samples
will have better accuracy and repeatability.
This pre-rinsing operation should be repeated when the
volume to be aspirated is changed or when a new pipet tip
is used.
6. Dense and Viscous Liquids
The pipettor’s specications of accuracy and precision are
based on pipetting distilled water. Handling of liquids with
physical qualities of density, viscosity, and surface tension
which dier substantially from water may need to be
gravimetrically checked for compensation of the volume
setting. Normally the degree of error resulting from heavy
or viscous liquids is negligible if pipetting is done slowly
and carefully. It is most important to give the liquids
some time to react to the change of pressure by holding
the pipet tip in its position for at least 2 seconds after the
aspiration and the blow out stroke.
If in extreme cases, this method of operation does not
result in accurate values, compensation may be achieved
as follows:
◗ Weigh the pipetted liquid when the pipettor is set to the
nominal value. Then calculate the set-o from the
nominal value:
Correction, value = 2 x nominal val. —
m = weight of the sample
γ= density of liquid
◗ Check this operation once again and correct if necessary.
Note the corrected value for further pipetting of the
same kind of liquid.
γ
m

Instruction Manual | 9
7. Corning® Pipet Tips
Corning DeckWorks pipet tips are made from high
performance polypropylene and their quality ensures the
precision and accuracy associated with the pipettor. Strict
control is maintained throughout the manufacturing
process to ensure the highest quality.
The accuracy and precision gures for the pipettors are
only guaranteed when Corning pipet tips are used. The
use of inferior quality tips will seriously degrade the
performance of the pipettor.
Refer to Pages 3 and 4 for information on proper pipet tips
sizes, accuracy, and precision values.
8. Recommendations
Observing the following recommendations will ensure
maximum possible accuracy and precision of liquid sampling.
◗ Make sure to operate the pipettor slowly and smoothly.
◗ The depth of immersion in the sample liquid should be
the minimum necessary and should remain constant
during aspiration.
◗ The pipettor should be held in a vertical position.
◗ Change the pipet tip when volume setting is changed or
when a dierent liquid is to be aspirated.
◗ Change the pipet tip if a droplet remains on the end of
the tip from the previous pipetting operation.
◗ Each new pipet tip should be pre-rinsed with the liquid
to be pipetted.
◗ Liquid should never enter the pipettor shaft. To prevent this:
- Press and release the pushbutton slowly and smoothly.
- Never turn the pipettor upside down.
- Never lay the pipettor on its side when there is liquid
in the tip.
◗ Never force the volume setting beyond its recommended
limits.
◗ When pipetting liquids with temperatures dierent
from the ambient temperature, it is recommended to
pre-rinse the tip several times before use.
◗ Do not pipette liquids with temperatures above 70°C.
◗ When pipetting acids or corrosive solutions which emit
vapors, it is recommended to disassemble the shaft
and rinse the piston and seal with distilled water after
nishing the pipetting operation.

10 | CORNING LAMBDA PLUS PIPETTORS
9. Recalibration
The accuracy and precision of the pipettors are obtained
gravimetrically using Corning® pipet tips, performing at
least 10 measurements of distilled water according to
EN ISO 8655 standard.
If during pipettor operation you nd that the accuracy
error (the dierence between the real aspirated volume
and the preset volume) exceeds the permissible value
given in the table in Section 1, the pipettor recalibration
procedure should be carried out.
Before starting the recalibration it is necessary to check
whether the following requirements have been met
during error determination:
◗ Ambient temperature and the temperature of the
pipettor, pipet tips, and water were identical.
◗ Density of the liquid used is close to that of distilled water.
◗ Balance with appropriate sensitivity has been used.
◗ Conversion factor mg/μL has been taken into account.
◗ The requirements shown in Sections 3 and 7 have
been met.
Volume Checked
(μL)
Balance Sensitivity
(mg)
0.1 - 10 ≤0.001
10 - 100 ≤0.01
>100 ≤0.1
If the above conditions are satised and the accuracy
error for selected volume given in Section 1 exceeds the
permissible value, the pipettor recalibration procedure
should be continued.
The recalibration can be performed within one full turn of
the calibration key to the right or to the left only.
Recalibration Conditions
Ambient temperature and the temperature of the pipettor,
pipet tips, and liquid should be within the range of 20°C to
25°C and stabilized during weighing within ±0.5°C.
◗ Measurements should be conducted using distilled water.
◗ Balance sensitivity should be suitable for the volume to
be measured.

Instruction Manual | 11
Recalibration Procedure
◗ Set the dose volume depending on the pipettor volume
according to the following tables:
Nominal
Volume
(μL)
Range of
Pipettor
Volumes (μL)
Preset
Volume
(μL)
Permissible
Volumes
(μL)
Volume Change
ΔV for Full Turn of
the Calibration Key
(24 Increments)
(μL)
Single-channel Pipettors
2 0.1 - 2 0.2 0.176 - 0.224 0.06
10 0.5 - 10 0.5 0.48 - 0.52 0.33
20 2 - 20 2 1.94 - 2.06 0.63
100 10 - 100 10 9.84 - 10.16 2.50
200 20 - 200 20 19.76 - 20.24 6.30
1,000 100 - 1,000 100 98.4 - 101.6 25.00
8-channel and 12-channel Pipettors
8 - 10
12 - 10 0.5 - 10 1 0.92 - 1.08 0.33
8 - 50
12 - 50 5 - 50 5 4.8 - 5.2 1.67
8 - 200
12 - 200 20 - 200 20 19.4 - 20.6 6.30
8 - 300
12 - 300 50 - 300 50 49.2 - 50.8 10.00
◗ Perform 5 aspirations, weigh each one, and calculate the
average value of the aspirations.
◗ Perform 3 aspiration series (each series should include
the aspirations from all channels), weigh each time, and
calculate the average value of the aspirations.
◗ Calculate average aspirated volume in μL multiplying the
average aspiration amount [mg] by the distilled water
density coecient [μL/mg], which depends on temperature
and pressure according to the following table.
Temperature
(°C)
Pressure (kPa)
95.0 101.3 105.0
20 1.0028 1.0029 1.0029
21 1.0030 1.0031 1.0031
22 1.0032 1.0033 1.0033
23 1.0034 1.0035 1.0036
24 1.0037 1.0038 1.0038
25 1.0039 1.0040 1.0040

12 | CORNING LAMBDA PLUS PIPETTORS
If the average aspirated volume exceeds the permissible
value, the following should be performed:
12
34
K
1.Remove the pipetting pushbutton.
WARNING: The pipetting pushbutton consists of 2 parts:
the lower volume-setting section (A2) and the upper
button (A1). After removal of the pushbutton, both parts
will separate.
2.Holding the volume setting knob to protect it against
rotation, insert the calibration key (K) into the cuts of the
calibration screw.
3.Turn the calibration key clockwise to reduce the aspirated
volume, or counterclockwise to increase the volume.
One full turn of the calibration key changes the pipettor
aspiration volume by the amount given in the table.
4.Take out the calibration key and x the pipetting
pushbutton. The pipetting pushbutton should be xed
by placing the lower section on the arbor rst (A2) and
then the button (A1).
Determine the average aspirated volume. The average
volume should be within the permissible range shown
in the table. If the volume exceeds the values stated, the
recalibration procedure should be repeated.
When pipetting liquids with physical properties that are
considerably dierent from those of water, follow the
instructions as described in Section 5.

Instruction Manual | 13
10. Troubleshooting
If you notice an improper pipettor operation, identify the
cause and eliminate the fault. Follow the instructions in
the sequence provided. Replacement of parts should be
required only occasionally and should not occur under
normal pipettor use.
Problem Cause Solution
Droplets of
liquid remain
in the pipet
tip.
The pipet tip is
emptied too fast.
Decrease the speed
of pressing the pipettor
pushbutton.
Press the pushbutton to the
second stop to expel all the
droplets.
The pipet tip
wettability has
increased due to
extensive use.
Replace the tip with
a new one.
Droplets
of air appear
in the liquid
aspirated
into the pipet
tip.
The pipet tip
immersion
is too shallow.
Immerse the tip
deeper according to the
instructions.
The pipet tip is
incorrectly pressed
onto the pipettor
shaft.
Press the pipettor rmly.
The tip is damaged
or worn out due to
extensive use.
Replace the tip with
a new one.
The pipettor
incorrectly
aspirates
the liquid
or the liquid
drops out
from the tip.
The pipet tip is
incorrectly pressed
onto the pipettor
shaft.
Press the pipet tip rmly.
The pipettor
incorrectly
aspirates
the liquid
or the liquid
drops out
from the tip.
The shaft nut is
loose in the single-
channel pipettors
2 μL to 1,000 μL.
Tighten the shaft nut.
The sealing surface
of the shaft is
cracked or scored.
Remove the tip ejector. Unscrew
the shaft nut and inspect the
shaft and the piston assembly.
Replace the damaged parts.
When reassembling the
pipettor, the nut should be
hand-tightened. In the single-
channel pipettor model, the
damage of the shaft may also
cause a damage of the piston
assembly. Replace the damaged
parts. When reassembling the
pipettor, the nut should be
hand-tightened.

14 | CORNING LAMBDA PLUS PIPETTORS
Problem Cause Solution
The pipettor
incorrectly
aspirates
the liquid or
liquid drops
out from the
pipet tip.
Damage to the
piston or seal due
to prolonged use
with chemically
aggressive liquids.
Replace the piston, seal, and
O-ring. Rinse the inside of
the shaft in distilled water
and dry. Lubricate the seal
and O-ring with the lubricant
which has been included
with each pipettor. The
replacement of the piston
requires conducting the
calibration procedure.
NOTE: The parts of the single-
channel (2 μL and 10 μL)
pipettors should be lubricated
evenly with a minimum
amount of lubricant.
The pipettor is
reassembled
improperly.
Disassemble the pipettor and
reassemble it, observing the
proper sequence of steps.
No lubricant on the
sealing elements.
Remove the tip ejector.
Unscrew the shaft nut and
remove the shaft, piston
assembly, seal, and O-ring.
Rinse the removed parts
in distilled water and dry
thoroughly. Lightly lubricate
the inside surfaces of the
seal and the O-ring with
the included lubricant.
Reassemble the pipettor in
the reverse order.
Contamination of
the inside of the
pipettor caused by
extensive aspiration
of chemically
aggressive liquids or
because liquid got
inside the pipettor.
Contamination of the inside
of the pipettor caused by
extensive aspiration of
chemically aggressive liquids
or because liquid got inside
the pipettor.
If the problem continues after carrying out the above
steps, contact your Corning representative.
Before returning the pipettor, please ensure the pipettor
is completely free of any chemical, radioactive, or
microbiological contamination which could pose a threat
during transport and repair.

Instruction Manual | 15
11. Cleaning and Sterilization
Cleaning
External surfaces of the pipetting pushbutton, the
ejector pushbutton, the handgrip, the shaft nut, and the
adjustment knob may be cleaned using a cloth dampened
in isopropyl alcohol. The remaining parts removed from
the pipettor during pipettor disassembly may be washed
with distilled water or isopropyl alcohol.
WARNING: Before using cleaning agents other than
those recommended by the manufacturer, check the
compatibility charts and consider chemical resistance of
the following plastics: PEI, PP, PC, POM, PA, PPS, PVDF
used to produce the pipettor parts.
Sterilization
The pipettor can be sterilized in an autoclave at 121°C for
20 minutes. After sterilization, the pipettor should be dried
and cooled to room temperature. It is recommended to:
◗ Unscrew the shaft nut slightly prior to sterilization and
re-tighten after autoclaving.
◗ Set the locking ring in lower (unlocked) position prior to
sterilization.
◗ Sterilize the pipettors in autoclave with an initial
vacuum and drying cycle.
The precision of the results should not alter if the
pipetting process and autoclaving are carried out as
described in this manual. Because a slight change in the
accuracy of the dosage may occur, it is recommended to:
◗ Check the calibration of the pipettor after the initial rst,
third, and fth autoclaving cycles and then after every
10 autoclaving cycles.
Ultra Violet (UV) Sterilization
The pipettors are UV-resistant. However, the recommended
distance from the radiation source to the exposed element
should be not less than 50 cm. Prolonged and very intense
UV exposure can cause decoloration of pipettor parts
without eecting their performance.

16 | CORNING LAMBDA PLUS PIPETTORS
12. Spare Parts
A
N
B
S
F
G
J
I
C
D
A1
A2
K
H
P
M
A: Pipetting pushbutton A1: Pushbutton A2: Knob
B: Adjustment knob
C: Shaft
D: Ejector
F: Shaft nut
G: Piston assembly
H: Spacer
I: O-ring
J: Seal
K: Calibration key
M: Ejector cap
N: Ejector pushbutton
P: Identication ring
S: Locking ring
Please contact Corning Customer Service to inquire about
availability of spare parts. Pipettor model and name of the
part required should be specied.
The replacement of the plunger requires conducting the
calibration procedure according to Section 9.
To nd your nearest service center and for scientic
support, visit www.corning.com/lifesciences.

Instruction Manual | 17
13. Limited Warranty
Corning Incorporated (Corning) warrants that this product
will be free from defects in material and workmanship
for a period of three (3) years from date of purchase.
CORNING DISCLAIMS ALL OTHER WARRANTIES
WHETHER EXPRESSED OR IMPLIED, INCLUDING ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR OF
FITNESS FOR A PARTICULAR PURPOSE. Corning’s sole
obligation shall be to repair or replace, at its option, any
product or part thereof that proves defective in material
or workmanship within the warranty period, provided the
purchaser noties Corning of any such defect. Corning is
not liable for any incidental or consequential damages,
commercial loss, or any other damages from the use of
this product.
This warranty is valid only if the product is used for its
intended purpose and within the guidelines specied
in the supplied instruction manual. This warranty does
not cover damage caused by accident, neglect, misuse,
improper service, natural forces, or other causes not
arising from defects in original material or workmanship.
This warranty does not cover the PTFE seal, O-ring, or
shaft. Claims for transit damage should be led with the
transportation carrier.
In the event this product fails within the specied period
of time because of a defect in material or workmanship,
contact Corning Customer Service at: USA/Canada
1.800.492.1110, outside the U.S. +1.978.442.2200, visit
www.corning.com/lifesciences, or contact your local
support oce.
Corning Customer Service will help arrange local service
where available or coordinate a return authorization
number and shipping instructions. Products received
without proper authorization will be returned. All items
returned for service should be sent postage prepaid in
the original packaging or other suitable carton, padded to
avoid damage. Corning will not be responsible for damage
incurred by improper packaging. Corning may elect for
onsite service for larger equipment.
Some states do not allow limitation on the length of
implied warranties or the exclusion or limitation of
incidental or consequential damages. This warranty gives
you specic legal rights. You may have other rights which
vary from state to state.

18 | CORNING LAMBDA PLUS PIPETTORS
No individual may accept for, or on behalf of Corning, any
other obligation of liability, or extend the period of this
warranty.
For your reference, make a note of the model number,
serial number, date of purchase, and supplier here.
Model No.
Serial No.
Date Purchased
Supplier
Warranty/Disclaimer: Unless otherwise specied, all
products are for research use or general laboratory use
only.* Not intended for use in diagnostic or therapeutic
procedures. Not for use in humans. These products are not
intended to mitigate the presence of microorganisms on
surfaces or in the environment, where such organisms can
be deleterious to humans or the environment. Corning Life
Sciences makes no claims regarding the performance of
these products for clinical or diagnostic applications.
*For a listing of US medical devices, regulatory
classications or specic information on claims, visit
www.corning.com/resources.
Other manuals for Lambda Plus
1
Table of contents
Other CORNING Measuring Instrument manuals

CORNING
CORNING 440 pH Meter User manual

CORNING
CORNING 430 pH Meter User manual

CORNING
CORNING ONE HEU User manual

CORNING
CORNING MRU User manual

CORNING
CORNING 315 User manual

CORNING
CORNING ONE OIU User manual

CORNING
CORNING ONE GX User manual

CORNING
CORNING ONE GX Configuration guide

CORNING
CORNING OV-1000 Service manual

CORNING
CORNING ONE RAU5x User manual