CORNING OV-1000 Service manual

1
Quick Reference Manual
OV-1000
Issue 1.1
March 2007
Note!! Please read before proceeding.
Please print “Release Notes” from the
installation CD before installing software
and follow instructions closely. Failure to
do so will cause communication issues
between OTDR and computer.

2
Corning Cable Systems
PO Box 489
Hickory, NC 28601
1-828-901-5000 or 1-800-743-2675
Corning Cable Systems reserves the right to improve, enhance and modify the
features and specifications of Corning Cable Systems’ products without prior
notification. Corning Cable Systems is ISO 9001 certified.
© 2006 Corning Cable Systems. All Rights Reserved. Printed in USA.
Table of Contents
1. OV-1000 OVERVIEW .................................................................................................................................................................3
2. GETTING STARTED..................................................................................................................................................................4
2A.INSERTING AND REMOVING TEST MODULES ............................................................................................................................4
2B.TURNING THE OV-1000 ON OR OFF OR GOING INTO “SLEEP”MODE. ........................................................................................4
2C.TO START/EXIT A MODULE OR THE VFL/POWER METER OR PROBE APPLICATION ...................................................................5
2D.INSTALLING THE UNIVERSAL INTERFACE (UI) ADAPTER ..........................................................................................................6
2E.ENVIRONMENTAL SETTINGS FOR THE OV-1000........................................................................................................................6
3. INSTALLING COMPUTER SOFTWARE ...............................................................................................................................7
4. TESTING WITH THE OTDR ....................................................................................................................................................8
4A.GENERAL TESTING PROCEDURE ................................................................................................................................................8
4B.SETTING UP YOUR OTDR..........................................................................................................................................................8
4C.AUTO-MODE TESTING.............................................................................................................................................................11
4D.ADVANCED MODE TESTING...................................................................................................................................................16
4E.TEMPLATE MODE TESTING .....................................................................................................................................................18
5. USING THE VFL......................................................................................................................................................................18
6. USING THE POWER METER................................................................................................................................................19
7. CREATING AND PRINTING REPORTS...............................................................................................................................20
7A.SETTING UP THE HEADER INFORMATION ON TRACES.............................................................................................................20
7B.PRINTING A REPORT................................................................................................................................................................21
8.0 TRANSFERRING DATA.......................................................................................................................................................21
9.0 USING THE PROBE TO INSPECT A CONNECTOR.......................................................................................................23
10.0 MISCELLANEOUS INFORMATION................................................................................................................................24
11.0 ADVANCED OTDR USE – MANUAL MEASUREMENTS.......................................................................................25
IMPORTANT
•This manual is intended as a quick reference guide only. For detailed
instruction on how to use your OV-1000 OTDR, please consult the Reference
Manual included on the OV-1000 CD-ROM.
•Electronic test equipment is exempt from Part 15 compliance (FCC) in the
United States.
•Electronic test equipment is subject to the EMC Directive in the European
Union. The EN61326 standard prescribes both emission and immunity
requirements for laboratory, measurement, and control equipment. This unit
has undergone extensive testing according to the European Union Directive
and Standards.

3
1. OV-1000 Overview
The OV-1000 is an optical time domain reflectometer (OTDR) with integrated color
touchscreen and interchangeable modules for both single- and multi-mode fiber systems. It
also can include a VFL for additional continuity testing and a power meter for end-to-end
system attenuation testing. Also, The OV-1000 allows for automatic (auto mode), manual
(advanced mode), and reference (template) testing. The OV-1000 is battery operated or can
be plugged into an AC/DC converter (24V, 4A output).
Figure 1. OV-1000 Features
VFL port
Battery slot
Module slots
Power meter
Compact
Flash card
reader
RJ-45 port
(lab only)
Probe port
USB client port
USB host port
DC connector
“dead”
microphone
p
ort
LEDs
touchscreen
shortcut keys
On/off key
function keys
dial w/ enter key
brightness key

4
2. Getting Started
2a. Inserting and Removing Test Modules
To insert a module into or remove a module from your unit:
1. Power off your unit.
2. To take a module out of the OV-1000, loosen the retaining screw on the
bottom of the unit using a coin (in a counter-clockwise direction) until the
module pops out and pull the module out. It is not necessary to remove the
screw to remove the module.
3. To put a module into the OV-1000, make sure the connector pins are at the
back and on the same side as the retaining screw. Look into the OV-1000, and
make sure you can see the screw to align it with the connector pins. On some
of the modules, there is a message that states: “This end towards back”.
Ignore the message because it is meant for module installation in an
OPTIVISOR 400 not the OV-1000. Once the pins are aligned, slide the module
in gently. Once the module slides as far as it can go, put pressure on the
module and tighten the retaining screw (turn in a clockwise direction) until the
module is flush with the unit. Do not use a screwdriver because you may over
tighten the screw and break the module (the pins may break).
4. When you turn on the unit, the startup sequence will automatically detect the
newly installed module. Also, if the module port is empty, please make sure it
has a blank port cover for protection.
2b. Turning the OV-1000 On or Off or going into “sleep” mode.
IMPORTANT
•The OV-1000 has both a sleep and off mode. In sleep mode (1 beep)
the unit is still consuming battery power.
Module
retaining screws

5
1. To turn the unit on hold down the green button on the front panel for one or
two seconds. The startup screen briefly appears and the unit starts.
2. To turn the unit off press and hold the green button until the unit beeps twice,
and the unit will then shut itself down.
3. To put the unit into sleep mode press and hold the green button until the unit
beeps once, and the unit will go into sleep mode. To “wake up” the unit, touch
the green power button, and it should take about 4 seconds for the unit to load
up. This power mode can be used throughout the day, but should be fully
powered down at the end of the day.
2c. To start/exit a Module or the VFL/Power Meter or Probe Application
Once the startup screen is loaded, you should see a list of your modules on the left
and a main menu on the right. If you have more than one module, both modules will
be listed. To start a module select the one you would like to use by touching the
screen with your finger/stylus or by touching the button/hot key on the front panel of
the unit (corresponds to the slot). Under the module list, a list of applications
associated with the module should appear. Select the application you would like to
use then touch the start button on screen.
To start a VFL/Power Meter or Connector Probe application, look under the main
menu list and select the application you would like to use. Data storage is located
under the File Manager tab. The Setup tab is used to calibrate the stylus and
customize the system settings. See screenshot below.
Please note that each application is started in a new window. You must exit an
application to get back to the main menu. To exit any application, look for a red X at
the bottom right hand corner of the screen and touch it. Touching the X will close the
application and return you to the startup menu.
Figure 2. Main Menu

6
2d. Installing the Universal Interface (UI) adapter
1. To install a UI connector adapter onto the UI baseplate begin by holding the UI
connector adapter by the base and not the cap. The cap will break off if you
twist the adapter by the cap.
2. Close the dust cap in order to hold the connector adapter more firmly.
3. Align keys and insert the connector adapter into the base plate.
4. While pushing firmly, turn the connector adapter clockwise on the base plate to
lock it in place. To remove turn counterclockwise.
2e. Environmental Settings for the OV-1000
To setup your OV-1000 press the Setup button in the main menu on the start-up
screen. There are several things you can setup, but we are only going to point out a
few important things in the screenshot below.
Figure 3. Auto Mode – OTDR tab
1. Date/Time icon. Select icon to set date and time for unit.
2. Application startup icon. Select icon to setup modules to load on startup.
3. Volume & Sounds icon. Select icon to set volume and sound types.
4. Stylus icon. Select icon to calibrate stylus.
3.
2.
1.
4.

7
3. Installing Computer Software
You can install OTSView on a computer for data post-processing. Use the
CD-ROM that came with your OV-1000 and follow the instructions.
To install OTSView on a computer:
Insert the installation CD in the CD-ROM drive. The CD should autoload to a menu
screen, if not load the CD by clicking on my computer and selecting the CD-ROM/DVD
drive. At the menu screen select the link to the RELEASE NOTE. Once the RELEASE
NOTE has launched please print the instructions and follow each step. Failure to
follow these instructions will result in lack of communication between the
device and computer.
Figure 4. Auto Mode – OTDR tab
1. Release note link. Click on the link to open the release notes. Again please print
the release note and follow the directions to avoid any communication issues.
2. Start Installation link. Clicking on this link will start the process of installing the
required software on your computer. Please note that it will not install the trace
viewing software. That software is in the Viewers and Utilities links.
3. Viewers and Utilities link. Click on this link to install the trace viewing software
and other programs such as Adobe Acrobat.
IMPORTANT
•Plugging the unit into the computer without installing the software first
will result in driver errors.
1.
2.
3.

8
4. Testing with the OTDR
4a. General testing procedure
To test in any mode, simply attach the fiber to the correct port. Note that some
modules can test both MM and SM fiber. Make sure that you connect the fiber to the
correct port which is labeled on the module. To start the test you need to select the
green Start button. The parameters you set vary from mode to mode and will be
discussed in the following sections.
4b. Setting up your OTDR
The following screenshots were taken from the Advanced Mode Testing module. Also,
the create ref/template setup tab in Template Mode is not shown. Please refer to the
main reference manual to see that tab. To get to the OTDR setup, press the OTDR
Setup button in the main menu on the right side of the screen(see Fig.9,button 9).
The setup tabs will load up and should look like the following.
Figure 5. OTDR setup – General tab
1. General Setup tab- Select this tab to access the general setup functions.
2-4. Display settings- Select whether or not you want a grid, the file name, or the
zoom window to appear on the trace.
5. Distance units settings- Select what units you would like to work in.
6. Automatic zoom button- Select this option if you want the OTDR to automatically
zoom in on a defined span after each trace is shot.
7. Touchscreen keyboard setting- Press this button if you want to turn on/off the
touchscreen keyboard.
8. Exit OTDR Setup button- press this button to exit setup and save settings
9. Close module button- press this button to close module, not the setup window.
1.
2.
3.
4.
5.
6.
7.
8.
9.

9
Figure 6. OTDR setup – Acquisition tab
1. Acquisition Setup tab- Press this tab to access the acquisition settings.
2. Acquisition Time setting- Set the time period for auto testing here.
3-4. Same pulse + time per wavelength setting- You can choose to set the pulse
and time acquisition settings to be the same for all wavelengths.
5. Fiber selection- Choose which fiber type to enter IOR, backscatter, and helix
factor for.
6-8. Fiber specifications- Enter the IOR, backscatter, and helix factor for the selected
fiber type or choose the default setting by hitting the default button.
9. Custom acquisition settings – Set the time and pulse width for each wavelength
1.
2.
3.
4.
5.
6.
7.
8.
9.

10
Figure 7. OTDR setup – Analysis tab
1. Analysis Setup tab- Press this tab to access analysis settings.
2. Auto-analyze option- Press this tab to automatically analyze traces after
shooting. If not selected the event table will not be populated until analyzed by the
trace viewing software on your computer.
3-5. Threshold settings- Set the minimum values for splice loss,
reflectance, and end of fiber threshold, or choose the default settings by
hitting the default button. For example, if the splice threshold were set to
0.02dB, then a step of 0.01dB would not be detected, but a step of 0.02dB or
greater would be.
6. Span start setting- Set the system start by event number from the beginning or
by distance from the beginning. This is useful when you want to eliminate the
launch cable in your system calculations.
7. Span end setting- Set the system end by number from the end or by distance
from the end.
1.
2.
3.
4.
5.
7.
6.

11
Figure 8. OTDR setup – Event Table tab
1. Event table tab- Press this tab to access the Event table settings
2. Mark faults option- Select this option to mark event table faults in red(fail) or
green(pass) after analyzing the trace. For example, if a loss threshold were set to
0.5dB, then an event with a 0.3dB loss would show up green on the table.
3. Pass/Fail Message- Press this button to turn on/off the pass/fail message.
Typically this button is left unchecked.
4-6. Threshold values- Enter the detection threshold values for event loss, event
reflectance, and fiber section attenuation.
7-8. Span start and end loss- Select to show span start and end loss on event table.
9. Fiber sections- Select to show fiber spans on event table. If left unchecked then
the event table will show events only.
10. Injection level- Select to show injection level of the source laser.
4c. Auto-mode Testing
Auto Mode testing automatically determines fiber length, sets acquisition parameters,
acquires and displays traces, and displays an event table at the touch of a button.
Also, the only inputs required are your test wavelengths and fiber type (buttons 2 & 3
in the screenshot on the next page).
2.
7.
3.
6.
5.
4.
1.
8.
9.
10

12
Figure 9. Auto Mode – OTDR tab
1. OTDR trace display. The trace will be displayed here regardless of testing mode.
2. Test wavelength selection box. This box will typically show 1310nm and 1550nm
for single-mode (9µm) and 850nm and 1300nm for multimode fiber (50µm and
62.5µm)
3. Select mode button. If you press this button you can change between single
mode (9µm) and the two standard multimode (50µm and 62.5µm) core sizes. This
tells the unit the correct IOR and backscatter coefficient.
4. Status window. This window shows which fiber type you selected.
5. Help button. This loads the help index for troubleshooting.
6. The close module button. Hit the red X to close the OTDR application.
7. About OV-1000 button. Hit this button for copyright information.
8. Print button. Press this button to print a trace and/or trace report to a compatible
Bluetooth wireless printer.
9. OTDR Setup. Press this button to access the setup menu.
10. Report button. Press this button to setup the format of the trace report.
11. Storage button. Press this button to get to auto file naming.
12. Quick save. Press this button to save your trace to a preset destination and with
a preset root file save name.
13. Start button. Press the green start button to run the trace.
14. Battery status and plugged/unplugged status
15. Testing Mode status window. Says which application you are using.
1.
2.
3.
4.
13.
12.
11.
10.
9.
8.
7.
6.
5.
14.
15.

13
Figure 10. Auto Mode – Events tab
1. “A” marker. This marker appears over the event you select from the event
table to locate the event quickly.
2. Event table. This table displays all detected events and event types in the
order they are found along the trace. Note: Events that “fail” are in red.
3. Magnifying glass. Press this button to zoom in on the selected event from the
event table. For example, in the screenshot above pressing the button would
zoom in on event #3.
Definitions:
Location(Loc.)- Distance from span start to the event
Loss- Event or cable span loss in dB
Cumulative loss(Cumul.)- Loss from span start to selected event
Reflectance(Refl.)- Reflectance value for reflective events only(ex. connector pairs)
Attenuation(Att.)- dB/km or dB/mile for cable span lengths
Event types
- Span start - Span end - Reflective event - Non-reflective event
- “gainer” - fiber section/span - “ghost”
1.
2.
3.

14
Figure 11. Auto Mode – Trace Info. tab
This tab is displays information about the trace and testing parameters.

15
Figure 12. Auto Mode – Source tab
This tab allows you to use your OTDR port as a light source with a
standalone power meter.
1. Source wavelength selection. You can select from any of the available
wavelengths in the current module.
2. On/Off button for source wavelength. Please be aware of which wavelength type
(multimode or single-mode) you are using. The light will come out of the
corresponding OTDR port. For example, 850nm would come out the MM port and
1310 will come out the SM port.
3. On/Off light.
4. Frequency selection/modulation. You can choose what frequency the source laser
will emit light. Typically, continuous is selected.
Note: This light source and meter do not have automatic wavelength detection
capabilities like some attenuation test sets. The wavelength must be selected
manually for both the source and meter.
1.
2.
3.
4.

16
4d. Advanced Mode Testing
Advanced Mode Testing allows for manual control over the testing parameters (range,
pulse width, average time, etc) that define the OTDR trace. The following section will
only highlight differences in the tabs from the Auto Mode testing.
Figure 13. Advanced Mode – OTDR tab
1. Wavelength selection. In order to shoot “Real time”, you can only choose
one wavelength at a time.
2. Distance “knob”. This knob modifies the range of the trace. Set the range to the
next highest step. For example, if the length of system is 2km, set it to 2.5km.
3. Pulse width “knob”. This knob sets the pulse width of the laser. Note: The unit
will not allow you to set a too large pulse width based on the distance of the system.
The higher pulse width settings will be grayed out.
4. High-resolution acquisition button. If you require a higher resolution trace select
this button, and the computer will collect more data points.
5. Time acquisition “knob”. This knob sets the time for trace acquisition
The only difference in the events tab in advanced mode is an additional analyze
button. This allows you to reanalyze trace if you add an event, change pass/fail
threshold, etc..
5.
4.
3.
2.
1.

17
Figure 14. Advanced Mode – Measure tab
1. Marker select. Press this to select the marker you wish to move. You move
either A, a, B, b, all of the markers, or the trace itself.
2. Marker Information table. This table shows the location of each marker, the
signal strength at that marker, the distance between the markers, and the loss
between the markers.
3+4. Moving arrows. The buttons move the selected marker left or right.
5+6. Zoom in or out. Pressing these buttons zooms in or out one step. When you
zoom in a box appears that you can place and set to the area you want to magnify.
7. “Full Trace” button. This button allows you to zoom all the way out.
8. Event loss button (4-pt Ev. loss value). Pressing this button allows you to
measure an event loss using the markers.
9. Attenuation button (A-B LSA Loss). Pressing this button allows you to measure
the loss/distance (dB/km or dB/Kft) for a cable section.
10. Reflectance button. Pressing this button allows you to measure a 3-point
evaluation of an event’s reflectance.
11. ORL button. Pressing this button allows you to measure a point to point ORL
between the start and end of span limits. Please note you only have this option
with a single mode system.
12. Marker measurement results window.
13. Full trace preview window. Shows you which part of the trace that you are
zoomed in on.
1.
2.
3.
4.
5.
6.
7.
8.
12.
9.
10.
11.
13.

18
The Trace Info tab in advanced mode is the same as in auto mode, except for the
fiber settings button. Here you can manual change the settings of the fiber (ex. IOR,
back scatter, helix factor). If you do this then you will need to re-analyze the trace.
The source tab in Advanced Mode testing is the same as in auto-mode testing.
4e. Template Mode Testing
Template Mode Testing allows the user to shoot trace files using a reference or
“standard” trace for comparison. The user creates a sample trace as a reference and
then compares that to all the following traces. This insures that all traces have the
same number of events regardless in the event table. This avoids the issue where a
known splice or connector pair is skipped due to it falling below the detection
thresholds. Please refer to the full user manual for additional information.
5. Using the VFL
Figure 16. VFL tab
1. VFL tab. Press this tab if you want to use the VFL.
2. VFL “figure”. This figure will turn red if the VFL is on and will turn black when the
VFL is off.
3. On/off button. This button turns the VFL on or off.
4. Modulation setting. These arrows change the setting of the VFL from
continuous(CW) to modulating(2kHz). Typically, the VFL is set to CW.
5. Close window “red x” button. Closes the VFL/Power Meter application.
1.
2.
3.
4.
5.

19
6. Using the Power Meter
Please note that not all OTDR are configured with power meter ports.
Figure 15. Power meter tab
1. The VFL tab. Select this tab if you want to use the VFL.
2. The Power Meter tab. Select this tab if you want to use the power meter.
3. Power Meter readout screen. This screen displays the loss measured and the
corresponding wavelength.
4. Wavelength selection button. Choose the wavelength you want to use.
5. Unit selection button. Cycle between dBm(baseline), dB(reference), and W.
6. Reference button. Press this button to reference out the existing readout. This
will zero out the value and default the units to dBr.
7. Nulling button. Press this button to null the meter. Make sure the meter is
capped with the plastic cover that comes with unit when nulling.
8. Close window (“red x”) button.
9. Power result add button. Press this button to add the result to your data table.
10. Setup button. Press this button to go into power meter setup. Here you can
setup your system thresholds for pass/fail results.
11. Storage button. Press this button to open up the data storage window.
Note: This meter is not capable of automatic wavelength detection.
9.
3.
1.
2.
4.
5.
6.
7.
10.
11.
8.

20
7. Creating and Printing Reports
7a. Setting up the Header Information on Traces
To setup header information on your trace report, first press the report button on the
right when in an OTDR module. Then enter the appropriate information for the
current trace. If you want to apply the same info to future fibers, press the “Save as
Template” button. Pressing this button will apply the current header information to
all future traces. If you want the Fiber ID to auto-increment, select the autonaming
function, and the Fiber ID will increase automatically as in the Autosave function. To
save the information you just entered, press the “Exit Report” button.
Figure 17. Report Template screen
1-4. Location A, Cable ID, Fiber ID, and Location B. Fill in the blank with the
appropriate information. Note: Fiber ID is unavailable if using autonaming.
5. Autonaming button. Press this button if you want to use autonaming.
6. Job ID. Fill in the box with the job name.
7. Date, Time, Unit, and Serial no. These boxes are automatically filled for you.
8. Comments box.
9. Save as Template button. Press button to save current report info as a template.
10. Customer ID. Fill in the box with the customer name.
11. Exit report button. Press button to exit report screen and save any information
that was entered to the current trace.
12. Clear all button. Press button to clear all information entered.
1-4.
5.
6.
7.
10.
8.
12.
9.
11.
Table of contents
Other CORNING Measuring Instrument manuals

CORNING
CORNING ONE HEU User manual

CORNING
CORNING 440 pH Meter User manual

CORNING
CORNING ONE IHU User manual

CORNING
CORNING ONE RAU5x User manual

CORNING
CORNING ONE OIU User manual

CORNING
CORNING Lambda Plus User manual

CORNING
CORNING 315 User manual

CORNING
CORNING ONE GX User manual

CORNING
CORNING MRU User manual

CORNING
CORNING ONE GX Configuration guide