Crowcon SprintPro 3 User manual

User Manual
M07016
July 2019
Issue 3
Multi-function Flue Gas Analyzer
Sprint Pro 3
5430910000J18
Your company name
Your Company Phone

Safety and product related information:
•Read and understand all instructions in the operation sec-
tion of this manual before use.
• Observe all warnings and instructions marked on Sprint
Pro, or as prompted by the analyzer and within this manual.
• Sprint Pro must only be operated with the Sprint Pro Flue
Probe assembly and always with the water trap and filters
in place.
• Connect the Flue Probe Assembly to Sprint Pro before
switching on and do not insert probe into flue until after
the zeroing process is complete.
• Sprint Pro water trap must be dried after use and before
Sprint Pro Flue Probe Assembly is returned to carry case.
• Before use ensure Sprint Pro is in good repair and do not
use if damaged or if calibration has expired.
• Sprint Pro passes the requirements of EN50379-1 and
EN50379-3 and has been independently tested by BSI.
• Sprint Pro is designed for use in ambient temperatures in
the range 14°F to 122°F and should not be used outside
this range.
• Sprint Pro contains a re-chargeable Lithium ion battery.
• Sprint Pro’s battery must not be charged at temperatures
below 32°F or above 104°F.
• If Sprint Po is damaged do not use and return to a quali-
fied service center for repair/replacement.
• Service, calibration & repair of Sprint Pro must only be
undertaken by a qualified service center.
• Do not substitute components as this may impair safety
and invalidate warranty.
• Repair of Sprint Pro and gas sensor replacement shall be
carried out by the manufacturer or qualified service center
in accordance with the applicable code of practice.
• If this product is not working properly, read the trouble-
shooting guide or contact the service center.
• Batteries are non-replaceable by the user and must only
be replaced by authorised service centers.
• Sprint Pro battery is rechargeable via a USB-C input con-
nector.
• Sprint Pro must be charged via a USB charger that meets
the specification as detailed in the user manual.
• Sprint Pro is not to be used as a permanent fixed detector.
• Sprint Pro is not designed or tested for continuous use.
• Sprint Pro is not a safety alarm.
• Only use probes, leads and accessories supplied by the
manufacture.
• Do not subject the product to cleaning fluids, such as
those containing high concentrations of acetone and
silicone compounds (such as silicone grease).
Summary of Screen Icons
Select, pick or action
Up item or entry
Down item or entry
Left field
Right field
Zero pressure
Start timed test
Stop timed test
Restart timed test
Yes, pass or done
No or fail or cancel
Purge or pump running icon
Select temp 1 (flow)
(Differential Temperature Test)
Select temp 2 (return)
(Differential Temperature Test)
Save log
Export report/log to IR Printer
Export report/log to Mobile App
Delete
Edit value
Increase value
Decrease value
Descend through character list
(String edit)
Ascend through character list
(String edit)
CO Alarm
IR Printer Alignment

• Do not immerse in water.
• Do not use silicon grease on the O-rings.
• Fuel options available are: natural gas, LPG, heavy
oil, light oil, kerosene, coal, wood, wood pellet
dry, coke, Biomass and Bagasse.
Additional information:
Sprint Pro is designed to meet the requirements of
EN50379-1 and EN50379-3 and to support the work-
ing practices defined in Standard BS7967.
It is highly recommended that users are fully con-
versant with local regulations when using a flue gas
analyzer for servicing or installing a boiler system.
When performing any CO measurements ensure
the unit is zeroed in clean air. A suitable location
for sampling clean air will be outside of the build-
ing where the boiler system is installed, away from
exhaust vents.
Sprint Pro offers a timed let-by/tightness test.
If the Sprint Pro has not been used for more than
6 months it should be put on charge for at least 4
hours without use. This will allow the oxygen chan-
nel and if present the NO channel, to read correctly.
Failure to do this may cause the oxygen measure-
ment, the NO measurement and any calculations
depending upon them to be incorrect.
Magnets on the reverse of Sprint Pro can be used to
place the Sprint Pro in location on the boiler system
for easy hands-free operation.
Diagnostics and usage data from Sprint Pro is acces-
sible at service and can be viewed at service centers.
We reserve the right to update and enhance the soft-
ware at times of service without prior knowledge.
WARNING: Take care not to place items which may
be sensitive to strong magnetic fields near this mag-
net, e.g. credit cards or magnetic storage devices like
computer hard drives.
FCC compliance
Sprint Pro FCC number: A8TBM78ABCDEFGH
© Copyright Crowcon Detection Instruments Ltd 2019.
All rights are reserved. No part of the document may be
photocopied, reproduced, or translated to another language
without the prior written consent of Crowcon Detection
Instruments Ltd.
Publication number: M07016
Second edition: May 2019
NOTE:
This product uses the FreeRTOS.org real time kernel available from http://www.FreeRTOS.org
The FreeRTOS source code is licensed by the GNU General Public License (GPL) with an exception.
The full text of the GPL is available here: http://www.freertos.org/license.txt.
The text of the exception is available on FreeRTOS official website: http://www.FreeRTOS.org - License and Warranty Page


Contents
Sprint Pro Multi-function Flue Gas Analyzer.................................. 2
Unpacking ................................................................................................... 2
I. Sprint Pro Orientation ................................................................ 3
Overview ..................................................................................................... 3
Probe connections ....................................................................................... 4
Switch on and using the Menu .................................................................... 4
Switch off .................................................................................................... 4
Charging battery.......................................................................................... 5
Pump........................................................................................................... 8
Keypad overview.......................................................................................... 8
Menu structure ............................................................................................ 9
II. Set Up & Menu Configuration Options...................................... 10
Fuel options ............................................................................................... 10
Units of measurement................................................................................ 10
Analyzer settings........................................................................................ 10
III. Performing tests..................................................................... 14
Zeroing process.......................................................................................... 14
1. Flue gas analysis..................................................................................... 15
2.
Differential temperature test........................................................................... 16
3.
Room Safety Test........................................................................................... 19
4.
Appliance sweep test................................................................................ 20
IV. Logging.................................................................................. 24
V. Test Report/Log exporting ....................................................... 25
VI. Maintenance and Calibration................................................. 26
VII. Sprint Pro Specification ......................................................... 28
VIII. Troubleshooting guide.......................................................... 31
Appendix I: Room Safety Test...................................................... 33
Warranty .................................................................................... 37
Sprint Pro Warranty Limitations........................................................ 38

2
Unpacking Sprint Pro
Sprint Pro
Multi-function Flue Gas Analyzer
Thank you for purchasing the Sprint Pro Multi-function Flue Gas Analyzer.
Sprint Pro will give you years of unparalleled service and reliability if looked
after correctly.
There are three versions covered in this manual as follows:
Please read the instructions carefully before use. Keep the manual for
future reference.
Unpacking
Important: Please fully charger Sprint Pro before use.
Remove the Sprint Pro from the packaging. The Sprint Pro accessories will
be located in the carry case. Check the contents are complete, you should
have:
• Sprint Pro Analyzer.
• USB charger and USB lead.
• Flue probe assembly, including water trap and filters.

3
Sprint Pro Sprint Pro Orientation
I. Sprint Pro Orientation
Overview
Display
Showing start-up
splash screen
ON/OFF/ENTER key
Press and hold button for two
seconds to switch on and switch
off Sprint Pro. Use this key to
finish edits, save changes and
view next page of multiple
paged screens.
ESCAPE key
Use this button to quit
tests and step back
through menus.
Purge icon
Rotating icon shows
pump is purging
USB connector
For charging and
communications to PC
IR window
Printer output
Exhaust exit
Warning: do not
inhale exit gas exhaust
or block exit
Thermocouple
sockets
K-type connectors
Flue sample inlet
Sprint Pro 3
5430910000J18
Your company name
Your Company Phone
Soft keys
Use these unmarked buttons
to make selections from the
menu, start and stop tests,
print and store results.
See section Operation for more
details
Battery icon
Three bars show
complete charge

4
Sprint Pro Orientation Sprint Pro
Probe connections
Flue gas analysis
Connect the flue probe to the flue sample inlet and the thermocouple to either
K-connector.
To measure flue draught pressure connect pressure tube to (-) pressure inlet.
Differential temperature test
Connect one (or two) thermocouple probes to the K-type connectors.
When using one probe, Sprint Pro will display a soft key option to switch
between T1 and T2 snapshot measuring points.
Note: When using (a) two thermocouples on metal pipes, or (b) any thermocouples
whilst the Sprint is on charge - electrically isolated thermocouples should be used.
Alternatively, cover the thermocouple contacts with plumbers tape to electrically isolate
the thermocouple contacts from the pipes.
Switch on and using the Menu
Press and hold the ON/OFF/ENTER key for two seconds.
Sprint Pro will emit a few rising beeps, display the
splash screen and then enter the Main Menu.
Use the soft keys to scroll, select and edit menu items
(see icon list – inside front cover). Press accept to
make a change or the ESC key to cancel. Press the ESC
key to return to the Main Menu.
Switch off
Note: Ensure flue probe is attached to Sprint Pro (including water trap and filters) prior
to following the switch off procedure and allow the sensor readings to return to their
ambient levels.
Tip: see soft key
icon list on the
inside front cover
of this manual
Test Menu
Fuel Options
Units of Measurement
Zero Gas Sensors
Analyser Settings
Stored Logs
Information
Main Menu
14:44:54 02-Nov-2018
Cal due 02 Nov 2019

5
Sprint Pro Sprint Pro Orientation
Turn off in clean air and allow Sprint Pro to complete purge cycle. Press and
hold the ON/OFF/ENTER button for approximately two seconds. The power off
screen will be displayed and the pump will run to purge the sensors. The pump
will run for up to 30 or 40 seconds to purge if gas is present. Sprint Pro will
normally switch off in 10 seconds. Press the ESC key to abort the switch off
sequence.
Charging battery
It is recommended that Sprint Pro is fully charged before first use.
Note: Sprint Pro’s battery must not be charged at temperatures below 32°F or above 140 °F.
1. Plug the USB charger into a mains socket.
2. Connect the charger to the Sprint Pro using the USB-C connector on
the unit.
3. Sprint Pro would normally be left switched off for charging. The
display will show the battery charging icon during charging. The
battery will recharge in three hours from flat and when fully charged
the battery icon will turn green in color.
Please see specification in section VII for Sprint Pro run times. A shorter charge
time can be applied, such as 1/2 hour, though it will give limited length of
operation from that charge.
Low battery
When the battery is low, Sprint Pro will display a low battery icon. If the battery
gets too low, then Sprint Pro will give further warning before switching off.
On a low battery alert, plug in the charger. If this cannot be done immediately
finish the test in progress and then place on charge as soon as possible.
Use of the charger as a power adaptor
The Sprint Pro charger can be used to power the unit and will continue to
charge whilst operating Sprint Pro.
USB C

6
Sprint Pro Orientation Sprint Pro
If the battery has been allowed to become exhausted or nearly exhausted:
When turning on the Sprint Pro, if the Time & Date
have been lost, a Time & Date entry screen will be
presented to allow the correct current Time & Date to
be entered.
If an attempt is made to turn the instrument on when
the battery level is too low for the instrument to oper-
ate, a warning screen will be displayed.
Automatic battery saver
If Sprint Pro is not undertaking a test it will automatically power down if left
unused. Sprint Pro will warn user when power down is imminent.
Set Time & Date
16:32:59 05-Feb-2019
Use to save date
15:12
2019 - Feb - 05
Set Time & Date
16:32:59 05-Feb-2019
Is time & date correct?
15:12
2019 - Feb - 05
Battery Exhausted
The battery is too low
Fully charge the battery
for correct instrument
performance.
to continue
Please plug the
charger in.
Battery Exhausted
The battery is too low
Fully charge the battery
for correct instrument
performance.
to continue
The instrument is
switching off

7
Sprint Pro Sprint Pro Orientation
If the instrument enters the auto turn-off state due to
an exhausted battery, this screen will be displayed, with
the red, empty, battery icon flashing:
If the instrument is turned off in a low-battery alert
state (battery icon red, double beep alert tune sound-
ing) then this screen will be displayed. The title will be
alternating ‘Low Battery Warning’ and ‘Turning Off’.
When the progress bar has completed (after about 8
seconds) or the Continue (middle) button is pressed,
the normal purge screen will be shown with the turn
off proceeding normally.
Sprint Pro has a lithium-ion rechargeable battery.
Batteries are non-replaceable by the user and must only be replaced by autho-
rised service centres.
Note: Fully charging the battery before putting Sprint into storage for an extended
period will help to maintain the battery’s life expectancy.
O2 Sensor Warning
If Sprint Pro battery becomes extremely exhausted,
when Sprint Pro is subsequently charged and turned
on the O2 sensor will require an extended stabilisation
time. The instrument will indicate this on the display,
it is important that the charger remain connected for
the duration of the stabilisation time until the display
reports ‘Sensor ready’.
Please plug the
charger in.
Fully charge the battery
for correct instrument
performance.
Low Battery Warning
16:32:59 05-Feb-2019
Continue?
Warning: Only use USB chargers provided with Sprint Pro and standard USB-C
charger leads.
Ensure the charger is
connected
The O2 sensor needs to
stabilise. This can take
up to 60 minutes.
O2 Sensor Warning
16:32:59 05-Feb-2019
Continue?
Sensor not ready
If the battery has been allowed to become exhausted or nearly exhausted:
When turning on the Sprint Pro, if the Time & Date
have been lost, a Time & Date entry screen will be
presented to allow the correct current Time & Date to
be entered.
If an attempt is made to turn the instrument on when
the battery level is too low for the instrument to oper-
ate, a warning screen will be displayed.
Automatic battery saver
If Sprint Pro is not undertaking a test it will automatically power down if left
unused. Sprint Pro will warn user when power down is imminent.
Set Time & Date
16:32:59 05-Feb-2019
Use to save date
15:12
2019 - Feb - 05
Set Time & Date
16:32:59 05-Feb-2019
Is time & date correct?
15:12
2019 - Feb - 05
Battery Exhausted
The battery is too low
Fully charge the battery
for correct instrument
performance.
to continue
Please plug the
charger in.
Battery Exhausted
The battery is too low
Fully charge the battery
for correct instrument
performance.
to continue
The instrument is
switching off

8
Sprint Pro Orientation Sprint Pro
Once the sensor is ready for use, pressing the
‘Continue’ (middle) button will acknowledge this
warning and allow normal instrument function.
Pump
Sprint Pro runs the internal pump during purge and when certain tests are
selected, and during and after some tests. A rotating purge icon will appear on
the screen when purging. To save on the battery life, Sprint Pro will turn off
the pump when it is not required.
When the pump is running ensure the exit gas exhaust is not blocked and do
not breathe in the exhaust gases. The purge function is often used by Sprint Pro
to put clean air over the sensors prior to turning off. This is ideal for realizing the
full potential lifespan of the sensors, especially the CO sensor filters.
Keypad overview
Sprint Pro provides a large color LCD with backlight. Navigation and functions
are provided by three soft key buttons which change according to what you are
doing.
Ensure the charger is
connected
The O2 stabilisation
has completed.
O2 Sensor Warning
16:32:59 05-Feb-2019
Continue?
Sensor ready
Use the soft keys to
navigate menus,
select, start and end
tests, change options
and select actions;
such as print, log or
zeroing.
Use the ESC key
to exit menus, exit
tests and step back
through screens.
ON/OFF/ENTER switch.
Use this key to finish
edits, save changes
and view next page of
multiple paged screens.

9
Sprint Pro Sprint Pro Orientation
Menu structure
After your Sprint Pro has been switched on the display will show the Main Menu
screen ready for use.
To select a menu item, use the soft keys below the and screen icons to
scroll the menu list and press the soft key underneath the screen icon to
select. Some menus also have their own submenu.
Press the ESC key to exit a menu, press ESC twice to return to the Main Menu
from a submenu.
The soft keys control the function displayed above them on the screen. These
will change depending on the test or menu, or if the ESC key is pressed.
To display Sprint Pro serial number, identity and software version press the ESC
key from the Main Menu.
The menu structure is shown below:
Test Menu
Main Menu
Flue Gas
Analysis
Room
Safety
Diff
Temperature
Analyser Settings
Pressure
Units of Measurement
Temperature
Stored Logs
Browse
Today’s logs
Browse
all logs
Delete
all logs
Press ESC to go
to Main menu
Key clickAuto off
timeout
Supervisor
settings
Efficiency
Set time
Disable/Enable
CO/NO/CO2
Alarms (where fitted)
& date
Edit report
header
Change
password
Find log by
test & date
Back Light Report
Appliance
Sweep
Find log
by number
Fuel Options
Zero Gas Sensors
Information

10
Set Up & Menu Configuration Options Sprint Pro
II. Set Up & Menu Configuration
Options
This section of the manual describes how to change the default configured
options if required, if not required please refer to Section III for details on
Performing Tests.
Fuel options
Sprint Pro displays the current fuel option during flue gas testing on the Flue
Gas Analysis screen.
To change the fuel option required, from the Main Menu select Fuel Options.
Use the soft keys and to navigate the available fuels and select the
required fuel using the soft key.
The new fuel option will be displayed on the Flue Gas Analysis screen during
testing.
Fuel options available are: Natural gas, LPG, Heavy oil, Light oil, Kerosene, Coal,
Wood, Wood pellet dry, Coke, BioMass and Bagasse.
Units of measurement
To change the units of pressure, temperature or efficiency from the Main Menu
select Units of Measurement.
Use the soft keys and to navigate the available submenus and units and
select the required option using the soft key.
Select one of the following options.
Temperature: degrees Celsius (ºC) or degrees Fahrenheit (ºF).
Pressure: mBar, Pa, hPa, kPa, PSI, inWG, mmWG, inHG and mmHG.
Analyzer settings
The Analyzer settings menu allows you to alter the settings for the display,
auto off timeout, back light, key pad and Supervisor settings. To change any of
these settings from the Main Menu select Analyzer Settings. Use the soft keys
and to navigate the available submenus select the required option using
the soft key.

11
Sprint Pro Set Up & Menu Configuration Options
Auto off timeout
The Auto off timeout screen allows the number of minutes at which the Sprint
Pro will automatically switch off to be selected. Use the soft keys – and + to
reduce or increase the number of minutes. Press the key to accept the change
or ESC to cancel. To disable Auto off timeout press – key until 'disabled' is dis-
played.
Backlight
Use the soft keys and to select one of the following options: Dim, Mid
or Bright.
Key click
The Key click settings allows the audible 'clicks' when the keypad is pressed to
be enabled or disabled.
Use the soft keys to select Enabled or Disabled from options. Press the key to
accept the change or ESC to cancel.
Report
The report option allows the selection of the means by which the logs and test
reports will be exported.

12
Set Up & Menu Configuration Options Sprint Pro
Supervisor settings
CO/NO/CO2 Alarm
Gas alarms for CO / NO /CO2 can be enabled or disabled (options will be pres-
ent depending on model). Use the soft keys and to change the setting.
Set time & date
On the Time & Date screen the current time and date are shown. Use the soft
keys – and + to alter the values of hours, minutes, day, month and year. Use the
key to select each unit. Press the ESC key to accept the change.
Edit report header
Use the soft keys and to select the report header text line one or two.
Press key to edit text. The screen displays the character lists and highlights the
current list in use. Use the and keys to scroll through character values in
each list and to move on to the next letter in the header text. The character
lists are shown below.
Press ON/OFF/ENTER to delete characters to the right. Press to move the cur-
sor to the end of the text and to accept change and return to Edit report
header screen.
Change password
Press the soft key to edit password. Use the and keys to edit or create
a password as described above in Edit report header. Press to accept change
and return to Password screen. Press ON/OFF/ENTER to store the change.
When a password has been created, Sprint Pro will display the Supervisor pass-
word screen on entering Supervisor settings.
See table of settings.
! .. / !"# $% & ' ( ) * + , - _ / A .. Z Uppercase alphabet
0 .. 9 0123456789 a .. z Lowercase alphabet
: .. @ : ; < = > ? @ Space
Tip: It is a good idea to enter your name/company name and phone number to identify
Sprint Pro and then set a password to prevent others changing these settings.

13
Sprint Pro Set Up & Menu Configuration Options
CO2Zero
In normal operation Sprint Pro will perform its zeroing function whilst in clean
air. CO2in ambient air contains 0.04% CO2.
However if it is not possible to undertake the zeroing function in clean air then
a CO2scrubber can be utilized.
In clean air: Sprint Pro will undertake the zeroing function in clean air and a CO2
scrubber must not be used whilst zeroing.
With scrubber: Sprint Pro will undertake the zeroing function in clean air and a
CO2scrubber must be used whilst zeroing.

14
Performing tests Sprint Pro
III. Performing tests
Important:
Sprint Pro must only be used with the Sprint Pro flue probe assembly
and always with the water trap and filters in place.
Please ensure the flue probe is connected before switching on your
Sprint Pro unit. Ensure the water trap is empty and is fitted in the right
direction. Always start in clean air and ensure that the gas exhaust is
not blocked. Do not insert the probe into the flue until after the zeroing
process is complete.
Note: Sprint Pro will switch off the pump when not performing tests or purging.
Zeroing process
Following Switch on Sprint Pro will require a ‘zero’ before selection of any of
the following tests: Flue Gas Analysis, Room Safety test, Appliance Sweep test
or Ambient Air Monitor test.
Ensure you are in clean air before proceeding with the zeroing process.
The zeroing process will then not be required when selecting these menu
options unless Sprint Pro is subsequently switched off and on.
Press the key to confirm you are in clean air. The Sprint Pro will then perform
a ‘zero’. Provided the zeroing process is successful Sprint Pro will display the
menu option chosen.
Tip: To ensure the zero is performed in clean air undertake the zeroing process outside
of the building or well away from the heating appliance to avoid any potential gases in
the vicinity affecting the zero process.
This is especially important for Sprint Pro 6 because of the additional carbon dioxide
sensor.
If a CO2sensor is fitted and it is not possible to undertake the zeroing process in clean
air a CO2scrubber can be utilized, please see Analyzer Settings section.
Warning: During testing, ensure the combined filter and water trap is not blocked or
full. Failure to do so may result in an error message.

15
Sprint Pro Performing tests
1. Flue gas analysis
Before performing the flue gas analysis test, check the water trap is clean and
is upright (arrow should point in direction of gas flow). To begin test, select Flue
Gas Analysis from the Test Menu. If the zeroing process has not previously been
performed Sprint Pro will carry out a 'zero' and then enter the test screen. Check
the fuel type displayed on the screen is correct. Use the soft keys to log or print
results.
Press the ON/OFF/ENTER button
to display the screens available:
Screen 1: O2, CO, CO2, CO/
CO2ratio, Pressure.
Screen 2: O2, excess (XS) air,
Temperature Flue
and Efficiency.
Screen 3: combines flue
elements from previous
screens.
Screen 4: (if NO sensor not
fitted): Temperature Flue,
Temperature Inlet, Net
Temperature
Screen 4: (if NO sensor fitted): NO, NOx, CO, CO2, Ratio, O2
Screen 5: (if NO sensor fitted): Temperature Flue, Temperature Inlet, Net
Temperature
Fuel
CO
ppm
CO2
%
Ratio
Pressure
mBar
Natural gas
0
###.#
#.####
!<
Flue Gas Analysis
07:33:03 14-Oct-2017
1
Tip: Press the
ON/OFF/ENTER
button to cycle
through the
screens.
Tip: The screen
number is
displayed here.
Tip: If # symbol
is displayed -
flue probe is
not in flue or
not connected.
P=0
O2
CO
CO2
Ratio
Pressure
5.0
10
9.1
0.0001
3.00
Natural Gas
07:32:55 23-Jul-2018
%
mBar
ppm
%
1
P=0
O2
CO2
XS air
Effcy
5.0
9.1
31.4
94.0
Flue Gas Analysis.
07:32:55 23-Jul-2018
%Net
%
%
3
%
CO
Ratio
Temp. flue
Press.
10
0.0001
162
3.00
mBar
ppm
C
η
O2
XS air
Temp flue
Efficiency
Flue Gas Analysis
07:32:55 23-Jul-2018
C
%Net
%
%
2
5.0
31.4
158
94.2
Flue Gas Analysis...
07:32:55 23-Jul-2018
4Flue Gas Analysis..
07:32:55 23-Jul-2018
5
Temp. flue
Temp. inlet
Net Temp
C
C
C55
24
31
NO
NOx
CO
Ratio
20
21
10
0.0001
ppm
ppm
ppm %
%
NO
NOx
CO2
O2
21
22
9.1
5.0
mg/m3
C
mg/m3
Screen 1 Screen 2
P=0
O2
CO
CO2
Ratio
Pressure
5.0
10
9.1
0.0001
3.00
Natural Gas
07:32:55 23-Jul-2018
%
mBar
ppm
%
1
P=0
O2
CO2
XS air
Effcy
5.0
9.1
31.4
94.0
Flue Gas Analysis.
07:32:55 23-Jul-2018
%Net
%
%
3
%
CO
Ratio
Temp. flue
Press.
10
0.0001
162
3.00
mBar
ppm
C
η
O2
XS air
Temp flue
Efficiency
Flue Gas Analysis
07:32:55 23-Jul-2018
C
%Net
%
%
2
5.0
31.4
158
94.2
Flue Gas Analysis...
07:32:55 23-Jul-2018
4Flue Gas Analysis..
07:32:55 23-Jul-2018
5
Temp. flue
Temp. inlet
Net Temp
C
C
C55
24
31
NO
NOx
CO
Ratio
20
21
10
0.0001
ppm
ppm
ppm %
%
NO
NOx
CO2
O2
21
22
9.1
5.0
mg/m3
C
mg/m3
Screen 3

16
Performing tests Sprint Pro
To perform a pressure zero, use the soft key marked P=0 on the relevant screen.
Note: for personal safety a carbon monoxide (CO) alarm will activate at 350ppm. This will
deactivate when CO levels drop below 200ppm. This is to inform the user of potentially
hazardous exhaust levels.
To end test press the ESC key.
Note: The minimum test duration should exceed both the time taken for burners
to heat up, during which the CO and NO produced may not be representative of
subsequent operation, and also the time needed for the boiler output to stabilise. Both
these periods are less than 5 minutes for most boilers hence flue test duration should
be a minimum of 5 minutes and a maximum of 30 minutes.
Values of Oxygen (O2), carbon monoxide (CO), nitric oxide (NO), pressure, inlet
temperature and flue temperature are measured and displayed by Sprint Pro.
Values of carbon dioxide (CO2), carbon monoxide to carbon dioxide ratio, excess air (XS
Air), efficiency, NOx and net temperature are calculated and displayed by Sprint Pro.
Note: When using Sprint pro with oil or solid fuel boilers, to ensure correct operation
and prolong the life of the filters, testing guidance in CD/11 (Oil Firing, Servicing and
Commissioning standard) should be followed.
CD/11 recommends undertaking initial tests to check cleanliness of combustion
prior to flue gas analysis testing. Thus please light the burner and allow it to run
for at least ten minutes to establish settled conditions. At this time take a smoke
spot reading at the test point provided. Adjust until a satisfactory figure is obtained.
Now it is possible to carry out a flue gas analysis test.
Screen 4 Screen 5
P=0
O2
CO
CO2
Ratio
Pressure
5.0
10
9.1
0.0001
3.00
Natural Gas
07:32:55 23-Jul-2018
%
mBar
ppm
%
1
P=0
O2
CO2
XS air
Effcy
5.0
9.1
31.4
94.0
Flue Gas Analysis.
07:32:55 23-Jul-2018
%Net
%
%
3
%
CO
Ratio
Temp. flue
Press.
10
0.0001
162
3.00
mBar
ppm
C
η
O2
XS air
Temp flue
Efficiency
Flue Gas Analysis
07:32:55 23-Jul-2018
C
%Net
%
%
2
5.0
31.4
158
94.2
Flue Gas Analysis...
07:32:55 23-Jul-2018
4Flue Gas Analysis..
07:32:55 23-Jul-2018
5
Temp. flue
Temp. inlet
Net Temp
C
C
C55
24
31
NO
NOx
CO
Ratio
20
21
10
0.0001
ppm
ppm
ppm %
%
NO
NOx
CO2
O2
21
22
9.1
5.0
mg/m3
C
mg/m3
This manual suits for next models
1
Table of contents
Other Crowcon Measuring Instrument manuals

Crowcon
Crowcon Tetra 3 User manual

Crowcon
Crowcon APM 155 User manual

Crowcon
Crowcon Gas-Pro User manual

Crowcon
Crowcon Gas-Pro User manual

Crowcon
Crowcon Xgard ESU C01987 Manual

Crowcon
Crowcon Gas-Tec User manual

Crowcon
Crowcon APM 145 User manual

Crowcon
Crowcon Gasman User manual

Crowcon
Crowcon M07630 User manual

Crowcon
Crowcon Gas-Pro User manual