Cybertek DP6020 User manual

Shenzhen Zhiyong Electronics Co., Ltd
www.cybertek.cn
Low Voltage Differential Probe
DP6020
(20V/ 200MHz)
Shenzhen Zhiyong Electronics Co., Ltd
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd
1
Preface
First of all, thank you for purchasing our products, this instruction manual is the description about the
function, usage, operation attention points, etc. Before use, please read the instructions carefully and use
correctly.
Manual annotation will use the following symbols to distinguish.
To the safely use the machine, you must abide by the following safety precautions strictly. The violation
against the manual is likely to damage the protective function of the machine. In addition, the company is
not responsible for any safety problem caused by the violation of matters needing attention in operation.
Please watch out for the maximum input voltage to avoid shock accident
Do not use the device under humid or explosive environment
Make sure the circuit under test is off before the probe is plugged in
Turn off the circuit and then take off the probe after the testing
Make sure the BNC port is well grounded when the probe BNC output cable is connected to the
oscilloscope or other devices.
Check the external surface of the probe before testing. If there’s any damage upon the probe,
please stop the usage.
Power the probe with the standard adapter.
DP6020 Brief Description
Type
Maximum input
differential voltage
Bandwidth
Attenuation ratio
DP6020
20V
200MHz
10X
Notice
Warning
In the case of wrong operation, the user risk injury. The content under this mark records the
relevant matters needing attention to avoid such dangers.
The user may have suffered minor injuries and material damage while with the wrong
operation, to avoid such situation, note the matters needing attention.
This symbol means it is harmful to the machine and human body; you must strictly follow the
instruction manual to operate.
Note
This symbolizes important note about how to use the machine.
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd
2
Contents
Preface………………………………… ……………………………………………………………..1
DP6020 Series Brief Description…… … …… ……………………… ……… ……….……………1
Summary……… …………………………………………………………… ………..………..…….3
Application……………… …… ………………………………… ……………………….………….3
Products and Accessories………………………… ……… …………… …….……………….…… 3
Main Part of Probe………………………………………………………………………..….…..3
Accessories………………………………………………………………………………...……..4
Electrical Specifications…………………………………………………………………..……….…..5
Machinery Specification…………………………………………………………………………..…. 6
Environment Specifications……………………………………………………………………...……6
Operating Steps……………………………… …………………………………………….…..….….6
Maintenance……………………………………………………………………………….………… 7
Warranty………………………………………………………………………………………..…… 7
Packing List………………………………………………………………………………….....….….7
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd
3
1. Summary
DP6020 probe is the low voltage differential probe with float ground testing function. Test voltage
±20V(DC+Pk), maximum bandwidth up to 200MHz, with 10:1 attenuation setting, can largely decrease the
circuit load. With the overload alarm function, the DP6020 can adapt with any oscilloscope with 50Ω BNC
input, and can be powered by the USB port on oscilloscope or other computers. Widely used in many
applications, DP6020 can provide excellent general differential signal testing for high speed power
measurement, vehicle bus testing and digital system design.
2. Application
Floating voltage measurement
High speed power testing
Digital differential bus
Vehicle serial bus(CAN、LIN、FlexRay)
3.Product and Accessories
■Product
Detailed Instruction
Input Cable: length around 15cm, used to measure the voltage signal
OffsetAdjust: adjust the resistor to realize the adjusting of the output
Connecting Cable: the connector of the front and back side of the probe, length around 70cm
Power Supply Port: standard USB Type B port, power up by standard USB adaptor. Can also
be powered by oscilloscope, easy to use
Overload alarm indicator: when the measuring range surpass the limit, the overload indicator
will be lighted, and buzzer will be activated.
Output Port: standard BNC output port can be connected to the oscilloscope of any brand. The
input impedance of the oscilloscope is required to be 50Ω, or connect to thorough 50Ω load,
with input impedance set to 1MΩ
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd
4
■Accessories
Alligator Clips (CK-261 red one pair) Thorough type 50Ω load(CK-50)
Piston Clips (CK-281 red one pair) Test Hook (CK-284 red one pair)
Coaxial Output Cable (CK-310)
USB Cable(CK-315 AM-BM,1.5m)
Power SupplyAdaptor (CK-605) USB 5V/1A
PowerAccessories Instruction
Alligator clips(CK-261)
CATIII 1000V CATIV 600V
Piston clips(CK-281)
CATIII 1000V
Test hook(CK-284)
CATII 1000V
Thorough type 50Ω load(CK-50)
50Ω 1W
Coaxial output cable (CK-310)
Double-ended BNC port coaxial line, 1m
USB cable(CK-315)
AM-BM,1.5m
Power supply adaptor(CK-605)
USB 5V/1A
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd
5
4.Electronics Specification
Bandwidth(-3dB)
200MHz(see Fig 1)
Rise time
≤1.75ns
Accuracy
±2%
Attenuation ratio
10:1
(DC + Peak AC)
Maximum differential voltage
±20V
Common mode voltage
±60V
Maximum rated input voltage (to earth)
±60V
Input impedance
Single to earth
500kΩ
Double input
1MΩ
Input capacitance
Single to earth
<7pF
Double input
<3.5pF
Output voltage fluctuation
±2V(50Ω Oscilloscope Input)
Offset (Typical value)
±2mV
Offset adjust range (Typical value)
±95mV
CMRR
50Hz/60Hz
>80dB
10MHz
>50dB
Noise(Vrms)
6mV
Overload indicator voltage threshold
≥20V
Delay time
Probe main
11ns
BNC(1m)
5ns
Overload indicator
Indicator will be lighted red when
overload occurs
Terminal load requirement
50Ω
Power supply
USB 5V/1AAdaptor
Fig 1: Frequency Response Curve
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd
6
5. Machinery Specification
Type
Parameter
Differential input cable
15cm
BNC output cable (CK-310)
1m
Alligator clips (CK-261)
85*40*17mm
Piston clips (CK-281)
152*50*13mm
Test hook (CK-284)
121*23*23mm
Probe dimension
Front
116*22*15mm
Back
124*49*27mm
Probe weight
172g
6. Environmental Characteristics
Type
Parameter
Pollution level
2
Operating temperature
0℃~50℃
Storage temperature
-30℃~70℃
Operating humidity
≤85%RH
Storage humidity
≤90%RH
Operating altitude
3000m
Storage altitude
12000m
7. Operating Steps
The range of the voltage under test should be estimated before the test. Voltage surpassing the
measuring range could possibly damage the probe and the product.
Connect the input and output cable to the probe and connect the probe with oscilloscope or other test
devices.
The green power indicator will be lighted when the power supply adaptor connects with the voltage
probe. When the voltage under test surpass the range, the indicator will be lighted along with the
buzzer.
Set the attenuation ratio of the oscilloscope or other devices to 10:1, input impedance to 50Ω (If the
probe output port is connected with 50Ω thorough type load, the input impedance should be 1MΩ)
Adjust the sensitivity of the oscilloscope according to the voltage under test
Connect the clips according to needs and start the test. The main body of the probe should be put
away from the high voltage pulse circuit to minimize the interference.
Turn off the power of the circuit under test after the test is over, then turn off the power of the probe
and disconnect the two input ports from the point under test, and unplug the BNC connector from
oscilloscope.
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd
7
8. Maintenance
Keep the probe dry and clean
Clean the probe with dry cloth instead of chemical potion
Put the probe back to its pack and store in shady, clean and dry places
The pack of our company can provide quakeproof protection to the probe, please make sure the probe
is packed during transportation
Do not pull or draw the cable to avoid over twisting or knot
9. Warranty
Please refer to the instruction on warranty card
10. Packing List
Packing List
Voltage probe
1
USB 5V/1A adaptor (CK-605)
1
Alligator clips (CK-261)
1
Insulated piston clips (CK-281)
1
Test hook (CK-284)
1
BNC output cable(CK-310)
1
50Ω thorough type load(CK-50)
1
USB connecting cable (AM-BM, 1.5m)
1
High class tool box
1
Instruction manual
1
Warranty card
1
Test report
1
CYBERTEK
SHENZHEN ZHIYONG ELECTRONICS CO., LTD.
Addr:RoomA1702, Building 4, TianAn Cyber Park, HuangGe North Road, LongGang
District, ShenZhen City, China
Tel: +86-400 852 0005
+86-755-86628000
Q Q: 400 852 0005
Fax: +86-755-86620008
Email: cybertek@cybertek.cn © Zhiyong Electronics, 2018
URL: http://www.cybertek.cn Published in China,Aug. 1, 2018
CYBERTEK
Table of contents
Other Cybertek Measuring Instrument manuals

Cybertek
Cybertek CPL8100A User manual

Cybertek
Cybertek CPL2000 User manual

Cybertek
Cybertek EM5030 User manual

Cybertek
Cybertek CP1000A User manual

Cybertek
Cybertek DP6000 Series User manual

Cybertek
Cybertek EM5018 User manual

Cybertek
Cybertek CTB20B User manual

Cybertek
Cybertek HCP8000 Series User manual

Cybertek
Cybertek DP6000 Series User manual

Cybertek
Cybertek OP6031 User manual