Cybertek CPL8100A User manual

Shenzhen Zhiyong Electronics Co., Ltd.
www.cybertek.cn
AC/DC Current Probe
CPL8100A (100A/600kHz)
CPL8100B (100A / 2MHz)
Shenzhen Zhiyong Electronics Co., Ltd
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
1
Preface
First of all, thank you for purchasing our products, the product specification is about the
function, usage, operation attention points, etc. Before use, please read the instructions
carefully and used correctly.
Manual annotation will use the following symbols to distinguish.
To avoid short out and deadly accident, the circuit under test should not surpass 600VAC
This probe is not designed for naked conductor.
Do not touch the conductor under test and the transducer probe during measuring.
When the oscilloscope is connected to other test terminal, please be careful for these
points:
Please use the fundamental insulation equipment with fitful voltage range and
pollution rate while the probe’s test terminal is connected to the other terminals.
Please do not surpass the safe voltage range if the fundamental insulation of the test
terminal cannot be met.
Please operating under correct guidance about electric safety.
There could be deadly electric shock if the device is damp or the users’ hand is wet.
Warning
Notice
Warning
=n
In the case of wrong operation, the user has the threat of injury, it records the
relevant matters needing attention to avoid such dangers.
The user may have suffered minor injuries and material damage while with the
wrong operation, to avoid such situation, note the matters needing attention.
This symbol means it has a harm to the machine and human body, you must
refer to manual operation.
Note
Recording of important notes while using the machine.
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
2
The probe is consist with precision components such as Magnetic core and Hall sensor,
which could be possibly damaged because of the sudden temperature change or external
force impact. Please be careful for vibration or impact.
The product is neither waterproof nor dustroof, please do not operate in the environment
filled with dust and water.
The probe’s upper and lower contact surfaces is made by precise polishing process.
Please protect this part or the probe could be malfunction.
When the battery voltage is lower than 6.5V, the probe will indicate a low voltage alert,
please change the battery . You can use a 9V alkaline battery for longer time.
CPL8100A/B Brief Summary
Type
Max current
(DC+Pk)
Max RMS
value
Band width
(-3dB)
Range
Current
transfer
ratio
CPL8100A
100A
70.7 Arms
600kHz
100A
0.01V/A
10A
0.1V/A
CPL8100B
100A
70.7 Arms
2MHz
100A
0.01V/A
10A
0.1V/A
Attention
Note
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
3
Catalog
Preface…………………………………………… ……………………………………………1
CPL8100A/B Brief Summary……………………………………………………………….…2
Summary ……………………………………………………………………………….……...4
Introduction about probe and accessories……………………………………………………...4
Electronic Characteristics……………………………………………………………………...6
Operating Method…………………………………………………………...…………………8
Mechanic Characteristics………………………………………………………………….…...9
Environment Characteristics…………………………………………………………………...9
Maintenance……………………………………………………………………………………9
Treatment if anything wrong……………………………………………………………….…10
Packing List……………………………………………………………...………………….. 10
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
4
1. Summary
CPL8100A/B is a kind of current probe that can measure both DC and AC current up
to 100A (70.7Arms). The band width of CPL8100A is 600 kHz (-3dB), and the band width of
CPL8100B is 2MHz (-3dB).CPL8100 series provide two range option of 10A and 100A, the
users may choose the correct range according to the current value. The auto zero function
make it easy for the users to operate, and it also has power and low voltage indicator as well
as overload sound alert. The probe can be powered by battery or external power supply,
which makes the measurement a lot more convenient. With standard BNC output port, the
probe can connect to any oscilloscope, and it can also connect to multimeter with
BNC-banana plug. The CPL8100 series is usually used in power supply , motor drive and
power supply.
2. Introduction about probes and accessories
Probe
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
5
Power Switch:
Power indicator: When the power switch is on, the indicator is lighted green.
Battery Low Voltage Indicator: when the voltage of the battery is lower than 6.5V, the
power indicator will lighted red, reminding the users to change battery.
Range Switch: press the button switch to change between H and L Range.
Range Indicator: indicating the current Range choice. H level: 100A; L level: 10A.
Auto Zero button: to make accurate measurement, users need to do zero degaussing
before measuring to avoid the influence of environment factors like earth magnetic field
and temperature shift.
Auto Zero Indicator: when Auto Zero, the indicator is lighted on and in green and will
be turned off when the Auto Zero is finished.
Clamp: the clamp used for current measurement; the max diameter for the cable is
12mm.
External Power Supply Plug: powered by external DC supply.
Battery Box: standard 9V alkaline battery is used. Please turn off the switch and
disconnect from the conductor under test & oscilloscope when changing battery.
Accessories
Output Cable (CK-310):1 meter
Power Supply Adaptor (CK-612):DC12V/1.2A
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
6
3. Electronic Characteristics
Test Condition: 23℃, 60%RH, cable under test get through the test center, load
resistance 1MΩ
Range level
L
H
Current range
50mA~10A Peak
1A~100A Peak
Attenuation accuracy
0.1V/A
0.01V/A
Typical DC precision
3%±50mA
500mA~40A Peak :
4%±50mA;40A~100A Peak :
±15% Maximum
Band-width
(-3dB)
CPL8100A
DC-600kHz Refer to the typical Amplitude frequency
characteristic (Figure 1)
CPL8100B
DC-2MHz Refer to the typical Amplitude frequency
characteristic (Figure 2)
Max current VS frequency
characteristics curve
Figure 3
Phase shift
DC~65Hz:<1.5°
DC~65Hz:<1°
Typical DC linearity
The typical DC linearity at H level(0.01V/A), Figure 4
Rise time
CPL8100A
≤583ns
CPL8100B
≤175ns
Max operation current
Please refer to voltage and current rating table
Max operation voltage
Please refer to voltage and current rating table
Max floating voltage
Please refer to voltage and current rating table
Operating voltage RMS
CATI 600V CATII 600V CATIII 300V
Common mode voltage RMS
CATI 600V CATII 600V CATIII 300V
Typical battery type and life.
9V alkaline layer-built battery/ 15H
Low power indication
When battery voltage is lower than 6.5V, battery indicator
will turned red and alert.
Overload indication
When the current under test surpasses the range, the buzzer
will buzz
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
7
Voltage and current rating table
Parameter
Max operating current (A)
Max
operating
voltage (V)
Max
floating
voltage(V)
H(0.01V/A)
L(0.1V/A)
DC
100
10
600
600
DC+AC peak value
100
10
600
600
AC peak value
100
10
600
600
AC peak to peak value
200
20
1200
- -
RMS CAT III
70.7
7.07
300
300
RMS CAT II
70.7
7.07
600
600
RMS CAT I
70.7
7.07
600
600
Figure 1 CPL8100A Typical Gain VS Frequency curve
Figure 2 CPL8100B Typical Gain VS Frequency curve
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
8
Figure 3 Max current VS Frequency curve
Figure 4 Typical DC linearity (0.01V/A level)
4. Operating instruction
Set the coupling mode of the oscilloscope to DC and the input impedance to 1MΩ; To
make it easy to read, users can set the display unit of the oscilloscope from voltage to
current display. If user’s oscilloscope doesn’t have such function, user have to manually
set the corresponding attenuation ratio, for instance, if the probe is H level (0.01V/A),
the oscilloscope should be 100X; if the probe is L level (0.1V/A), oscilloscope is 10X.
User can connect the BNC output port with the input port of the oscilloscope with
standard double terminal BNC cable.
Turn the switch to ON, and the power indicator will light up green.
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
9
According to the value of the current under test, choose the correct range by pressing
button.
Attention: Different range correspond to different attenuation ratio of the oscilloscope.
Press the Auto Zero button . The buzzer will beep for twice, or it will have a long and
continuous beep meaning the failure of Auto Zero .The external magnetic field can have a
little influence to the DC zero, so do not move the device after Auto Zero.
Open the clamp of the probe and clamp the conductor under test.
Attention: There’s direction tip on the current probe clamp. If the direction of the current
fits the tip, output will be positive; On the contrary, the result is negative.
Adjust the vertical sensitivity of the oscilloscope to get a stable waveform.
5. Mechanic Characteristics
Size of the front current clamp
About 100*20*60mm
Size of the back output box
About 137*33*35mm
Operation altitude
0~2000 meter
Maximum diameter of the conductor
under test
12mm ø
Length of the cable connecting
current clamp and output box
1 meter
Length of double terminal BNC cable
1 meter
Weight
About 223g(without battery)
6. Environment characteristics
Operating temperature
0℃~+50℃
Storing temperature
-20℃~+80℃
Operating relative humidity
0℃to +40℃,95%RH;
+40℃~+50℃,45%RH
Pollution Degree
Level 2
7. Maintenance
If the product is in the warranty period and used correctly, and at the same time
the malfunction is caused by the flaws of product itself and have not been
disassemble, the company will give free repair to the product.
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
10
Clamp: Please keep the clamp clean. After long time use, please use soft cloth together
with alcohol to erase the dirt. Please keep the clamp away from damp environment, not
to mention direction contact with water.
Handle: Please use clean cloth or sponge to clean the handle. Please use little alcohol
instead of water to erase the dirt and parch it.
To maintain the performance of the product, there can be one check or calibration every
year.
8. Troubleshoot
Problem
Reason
Treatment
No output signal or
amplitude is low
Power is off
Turn on the power.
Oscilloscope is AC mode
Please set it to DC
mode
Clamp isn’t closed
Check the clamp, close
it completely.
Powerindicator is off
when the probe is on.
Battery voltage is lower
than 6.5V
Change the battery
Amplitude is half
The input impedance is 50Ω
Set it to 1MΩ
9. Packing List
Packaging
Name
Quantity
Current Probe
1
9V Battery
1
DC12V/1.2A Adaptor (CK-612)
1
BNC Output Cable(CK-310)
1
Tool box
1
Instruction book
1
Warranty Card
1
CYBERTEK

Shenzhen Zhiyong Electronics Co., Ltd.
11
CYBERTEK
SHENZHEN ZHIYONG ELECTRONICS CO., LTD.
Addr:RoomA1702, Building 4, TianAn Cyber Park, HuangGe North Road, LongGang
District, ShenZhen City, China
Tel: +86-400 852 0005
+86-755-86628000
Q Q: 400 852 0005
Fax: +86-755-86620008
Email: cy[email protected] © Zhiyong Electronics, 2016
URL: http://www.cybertek.cn Published in China, June 06, 2016
CYBERTEK
This manual suits for next models
1
Table of contents
Other Cybertek Measuring Instrument manuals

Cybertek
Cybertek DP6065 User manual

Cybertek
Cybertek HCP8000 Series User manual

Cybertek
Cybertek EM5030 User manual

Cybertek
Cybertek DP6000 Series User manual

Cybertek
Cybertek CPL2000 User manual

Cybertek
Cybertek EM5018 User manual

Cybertek
Cybertek DP6000 Series User manual

Cybertek
Cybertek CP1000A User manual

Cybertek
Cybertek CTB20B User manual

Cybertek
Cybertek OP6031 User manual