DLab Thermo Control H100-Pro User manual

Thermo Control & Thermo Mix
User Manual
H100-Pro
Thermo Control
HC110-Pro
HM100-Pro
Thermo Mix
HC
M
100-
Pr
o
Please read the User Manual carefully before use, and follow all operating
and safety instructions!
We cannot be responsible to inform at real-time if the outline and
specifications are subject to change for improvement.
12301086 Ver.20191025

Contents
Preface.................................................................................1
Service.................................................................................1
Warranty..............................................................................1
1. Safety Instructions.......................................................3
2. Proper Use...................................................................4
3. Inspection....................................................................4
3.1 Receiving Inspection.............................................4
3.2 Listing of Items.....................................................4
4 Control & display........................................................5
4.1 Control ..................................................................5
4.2 Display..................................................................6
5 Operation.....................................................................7
5.1 Parameter setting methods ....................................8
5.2 Start and Stop........................................................8
5.3 Running.................................................................9
6 Trial Run ...................................................................13
7 Faults.........................................................................13
8 Maintenance and Cleaning........................................14
9 Associated Standards and Regulations......................15
10 Specifications............................................................16

1
Preface
Welcome
to
the
T
h
ermo
Control & Thermo
Mix User Manual. Users
shou
l
d
r
e
a
d th
is
Manu
al
carefully, follow the instructions and
procedures, and be aware of all the cautions when
using this instrument.
Service
When help needed, you can always contact the
Service Department of manufacturer for technical
support in the following ways:
Please provide the customer care representative
with the
following information:
·Serial number(on the rear panel)
·Certification
·Description problem (i.e., hardware or software)
·Methods and procedures adopted to resolve the
problems
·Your contact information
Warranty
Changes or modifications not expressly approved
by the party responsible for compliance could void
the user's authority to operate the equipment.
NOTE: This equipment has been tested and found
to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide
reasonable protection against harmful
interference when the equipment is operated in a
commercial environment. This equipment
generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance
with the instruction manual, may cause harmful

interference to radio communications. Operation
of this equipment in a residential area is likely to
cause harmful interference in which case the user
will be required to correct the interference at his
own expense.
This instrument is warranted to be free from
defects in materials and workmanship under
normal use and service, for a period of 24 months
from the date of invoice. The warranty is
extended only to the original purchaser. It shall
not apply to any product or parts which have been
damaged on account of improper installation,
improper connections, misuse, accident or
abnormal conditions of operation.
For claims under the warranty please contact your
local supplier. You may also send the instrument
directly to manufacturer, enclosing the invoice
copy and by giving reasons for the claim
1 Safety Instructions
Warning!
•Read the operating instructions carefully
before use.
•Ensure that only trained staff works with the
instrument.
Risk of burn!
•Caution when touch the housing parts and
the dry bath which can reach temperature of
110°C(Please refer to technical
parameters)
•Pay attention to the residual heat after
switching off.
Protective ground contact!
•Make sure that socket must be grounded
(protective ground contact) before use.
•When working wear personal safety guards to
avoid the risk from:
- Splashing and evaporation of liquids
- Release of toxic or combustible gases
•Set up the instrument in a spacious are on a stable,

3
clean, non-slip, dry and fireproof surface. Do not
operate the instrument in explosive atmospheres,
with hazardous substances or under water.
•Temperature must always be set at least 50℃
lower than the fire point of the media used.
•Be aware of hazards due to:
- Flammable materials or media with a low
boiling temperature
- Overfilling of media
- Unsafe container
•Process pathogenic materials only in closed
vessels.
•Check the instrument and accessories prior to
each use.
•Do not use damaged components. Safe operation
is only guaranteed with the accessories provided
by the manufacturer. Accessories must be securely
attached to the device and can’t come off by
themselves. Always disconnect the plug before
fitting accessories.
•The instrument can only be disconnected from the
main power supply by pulling out the main or the
connector plug.
•The voltage stated on the label must correspond
to the main power supply.
• Ensure that the main power supply cable does
not touch the dry bath. Do not cover the device.
• The instrument may only be opened by experts.

2Proper Use
The instrument is designed for heating liquids in
schools, laboratories or factories.
•Observe the minimum distances between the
devices, between the device and the wall and
above the assembly (min. 100 mm)
•This device is not suitable for using in
residential areas or other constraints mentioned
in Chapter 1.
>100mm
Fig. 1
3Inspection
3.1 Receiving Inspection
Unpack the equipment carefully and check for any
damages which may have arisen during transport.
Please contact manufacturer/supplier for technical
support.
Note:
If there is any apparent damage to the system,
please do not plug it into the power line.
3.2 Listing of Items
Items
Qty
Main unit
1
Power cable
1
Block
1
User Manual
1
Tab. 1

4Control & Display
4.1 Control
Fig.2 Thermo Mix
Fig.3 Thermo Control
Fig.4 Interface & Power Switch

Item
Description
Block
Changeable unit, suit for
different tubes
LCD Screen
Display the setting and actual
parameters
Function
Button
Switch the input parameters or
programming
Adjustable
Knob
At parameter setting state, rotate
this knob can input values; After
setting, push it to start work. At
working state, short push to
pause; Long push to stop
working
RS232 Interface
Used for temperature calibration
Power Interface
Connect the power cable
Power Switch
Power on/off
Tab. 2
4.2 Display
Items
Description
Set
Display the setting parameters
Actual
Display the actual parameters
Prog
Program button, used for
programming
Temp
Temperature setting button,
when press this button, the
temperature characters of the
setting area are flashing.
Time
Time setting button, when press

this button, the time characters of
the setting area are flashing . In
timed working mode, the time
shows the countdown time. In
continuous working mode, the
time show running time. Refer
to below figure.
Speed
Mixing speed setting button, when
press this button, the speed
characters of the setting area are
flashing (Thermo mix models)
Adjustable knob
At parameter setting state, rotate
this knob can input values; When
finish the parameters setting, push
it to start work. At working state,
short push to pause; Long push to
stop working
5Operation
5.1 Parameter setting methods
A. Respectively, press the function button,
the corresponding characters of the setting area are
flashing
B. Rotate parameter knob, input the target parameters.
C. Wait 3s, the corresponding character will stop
blinking, complete parameter setting.
Note: when power on, the setting area of the screen will
display parameters of the last run.
When the time is adjusted to 00:00, means the product
will run the continuous mode.
5.2 Start and Stop
1.Start: When finish the parameter input, push the
adjustable knob to start work. And in the last line in
the actual value area will display the characters
“Run”. Refer to below figure.

Temp.
Speed
Timer
Prog.
Set
30.5
1500
07:00
Run
30.5
----
07:00
Actual
2. Pause: In the working state, short push the
adjustable knob to pause. Refer to below figure.
Temp.
Speed
Timer
Prog.
Set
30.5
1500
07:00
Pause
30.5
----
07:00
Actual
And in the last line in the actual value area will
display the characters “Pause”. When in this state,
only the timer and mixing functions are paused,
the temperature control function is still working.
(Thermal Control doesn’t support mixing function).
Resume: In the pause state, push the adjustable
knob, will resume to working state. And in the last
line in the actual value area will display the
characters “Run” again.
3. Stop: In the working state, long push the
adjustable knob to stop the current work.
Temp.
Speed
Timer
Prog.
Set
30.5
1500
07:00
Actual
(Thermal Control doesn’t support mixing
function). Refer to below figure.
5.3 Running
5.3.1 Single Step Work Without Saving
When power on, the screen shows the parameters of
the last run, or factory settings. Push the Temp,

Speed, Time button respectively, the corresponding
characters will flash. Rotate the Adjustable knob to
set the parameters. If time is set to 00:00, it means
continuous working mode. After the setting
completed, push the adjustable knob to start
work.Refer to below figure.
Temp.
Speed
Timer
Set Acutal
30.5
1500
07:00
Prog.
Temp.
Speed
Timer
Set Actual
30.5
1500
07:00
30.5
800
06:59
RUN
Prog.
5.3.2 Multiple Step work without Saving
A.The program set in the single step is the
first step by default. Refer to below figure.
Temp.
Speed
Timer
Set Actual
30.5
1500
07:00
1-1
Prog.
B.Push the Prog. button, start to set the second
step. The last line of in the Set column shows 1-
2 now. Refer to below figure.

Temp.
Speed
Timer
Set Actual
---.-
-----
--:--
1-2
Prog.
Note: Before the setting of the second
step complete, system consider there is only one
step by default. But after the setting complete,
the actual steps will be shown correctly. Refer
to below figure.
Temp.
Speed
Timer
Set Actual
30.5
1500
07:00
2-2
Prog.
CIn the same way, complete the later steps. Then
push the adjustable knob to start work. Refer to
below figure.
Temp.
Speed
Timer
Set Actual
30.5
1500
07:00
30.5
800
06:59
2-2 RUN
Prog.
Prog. shows the program and step number. Refer to
the figure, there are 2 steps totally, now it’s the
second step.
Note: Before the time is set, the step will not take
effect. If set time to 00:00, it means continuous
working mode. Before complete the setting for
the current step, can’t enter next step.

5.3.3 Program with Saving mode
Programming method is similar with Multiple Step
without Saving mode. After the programming
complete, long push the adjustable knob, program
name characters flash (refer to the red part in
below figure). Rotate the Adjustable knob to
input the number of the name, then push the
adjustable knob to complete the input. The program
saving is complete now.
Temp.
Speed
Timer
Set Actual
30.5
1500
07:00
P1 5-2
Prog.
Prog. shows P1, 5-2,means the program and step
number. Refer to the figure, there are 5 steps totally,
now it’s the second step.
5.3.4 Load Program
In the power on state, long push the prog.
button , enters the load program mode. Rotate the
adjustable knob to show the saved program, from
P1 to the last, at most 9 programs. Switch to the
program you want, and short push the prog.
Button to check the program in detail. Then
push the adjustable knob to start the program.
5.3.5 Single function running
If Temp. is set to ---.-, then it means working
without temperature control function. If Timer is set
to 00:00, it means work continuous. If speed is set
to -----, it means work without mixing function.

5.4 Temperature calibration
1. The temperature of the instrument had been
calibrated before delivery, but due to some other
factors the temperature may vary from the actual
temperature and the deviation can be corrected by
temperature calibration.
2.In order to ensure the temperature control
accuracy, based on the temperature calibration
function, it is recommended to calibrate the
temperature whenever the heating block is changed.
The device adopts the three-point temperature
calibration method, HC110-Pro, HCM100-Pro
calibration temperature is 10℃, 50℃ and 90℃,
H100-Pro,HM100-Pro calibration temperature is
37℃,60℃ and 90℃.
The linear calibration of the three temperature
points ensures that the accuracy of the three
temperature points of the system within ±0.5℃.
At the time of temperature calibration, the ambient
temperature must be 15 to 25℃.
Caution:
The temperature of device has been calibrated
before delivery, so please don’t use this
function at will.
Calibration procedure:
Inject silicon oil into the hole in the center of the
heating block. Place the thermometer probe in the
oil filled hole.
1. Switch power ON the instrument,Press both the
Prog and Temp buttons at the same time to enter
into calibration mode while self-checking only.
(Please note that the instrument cannot enter into
calibration mode if the self-check is complete.)
The picture followed is Calibration interface:

2. Rotate the parameter adjustment knob to set
target temperature for calibration. Press the
parameter adjustment knob and confirm the set
values.
3. If the actual temp. reaches the set temperature
and remains 20min-30min,and other blocks will be
stable after 5min), The calibration area will display
the corresponding temperature on display.
Record the measurement on a Thermometer,
rotate the parameter knob to adjust the calibration
Temp. to actual thermometer temp. value
(Calibration temp. adjustment range is Set
temp±5℃) and press the same knob to confirm
settings.
4. To ensure the calibration temperature adjustment
set correctly, press the push knob and the
instrument will enter the second calibration
temperature, repeat step 3, after the third
temperature calibration completed, the screen
displays “Calibration Success" on interface and the
instrument will automatically enter into the self-
check mode. Then calibration is complete.。
If the display screen shows the “Cali Failed”
interface, the input value of calibration may be
wrong. Hence restart the instrument and repeat the
calibration steps again

Note: If the deviation between the actual
temperature and the calibration temperature is
within ±2℃ during calibration, a single point
calibration can be performed. If the deviation
between the actual temperature and the calibration
temperature exceeds ±2 ℃, a three-point calibration
is required.
5.5 Restore the factory settings
1. Switch power ON the instrument, Press Temp
and Prog buttons at the same time to enter into
calibration mode while the instrument is in self-
check mode. (Please note that the instrument cannot
enter into calibration mode if the self-check is
complete.)
2. After entering the calibration mode, press both
the Prog and Time keys to reset, the interface will
display “RESET Success," and the instrument will
restart automatically.
6Trial Run
•Make sure the required operating voltage
and power supply voltage match.
•Ensure the socket must be properly grounded.
•Add the medium into the vessel
•Place vessel on the block.
•Plug in the power cable,ensure the power is
on and begin initializing.
•Set the target parameters or pragraming
•Start working
•Observe the LCD display
•Stop working, and power off.
If these operations above are normal, the device is
ready to operate. If not, the device may be damaged
during transportation, please contact
manufacturer/supplier for technical support

7Faults
•Instruments can’t be power ON
-Check whether the power line is
unplugged
-Check whether the fuse is broken or loose
•Fault in power on self test
- Switch OFF the unit, then switch ON and reset
the instruments to factory default setting.
If these faults are not resolved, please
contact manufacturer/supplier.
8Maintenance and Cleaning
•Proper maintenance can keep
instruments working properly and lengthen
its lifetime.
•Do not spray cleanser into the instrument when
cleaning.
•Unplug the power line when cleaning.
•Only use recommended cleansers:
Dyes
Isopropyl alcohol
Construction
materials
Water containing tenside/
Isopropyl alcohol
Cosmetics
Water containing tenside/
Isopropyl alcohol
Foodstuffs
Water containing tenside
Fuels
Water containing tenside
•Before using other method for cleaning
or decontamination, the user must ascertain
with the manufacturer that this method will not
harm the instrument. Wear the proper
protective gloves during cleaning of the
instrument.

9Associated Standards and Regulations
Construction in accordance with the following
safety standards:
EN 61010-1
UL 3101-1
CAN/CSA C22.2(1010-1)
EN 61010-2-10
Construction in accordance with the following EMC
standards:
EN 61326-1
Associated EU guidelines:
EMC-guidelines: 89/336/EWG
Instrument guidelines: 73/023/EWG
Note:
•Electronic device can not clean with
cleanser.
•If you require maintenance service, must
be cleaned the instrument in advance to
avoid pollution of hazardous substances,
and to send back into original packing.
•If the instrument will not use for a long time,
please switch off and place in a dry, clean,
room temperature and stable location.

10 Specifications
Model
HCM100-
Pro
HM100-Pro
HC110-Pro
H100-Pro
Functions
Heating
&cooling
& mixing
Heating &
mixing
Heating &
cooling
Heating
Temp.
range
15℃below
room
temp.
~100℃
room
temp.+5℃~
100℃
25℃ below
room temp.
~ 110℃
25℃ below
room temp.
~ 100℃
Temp.
setting
range[℃]
0.1℃/100
℃
15℃/100℃
-5℃/110℃
15℃/110℃
Control
accuracy
[℃ @20-
45℃]
±
0.5
±
0.5
±
0.5
±
0.5
Uniformity
[℃
@20-45℃]
Max
±
0.5
Max
±
0.5
Max
±
0.5
Max
±
0.5
Max
heating
rate[℃/s]
5.5
5.5
5.5
5.5
Max
cooling
5
(100℃-
-
5(110℃-
Room
rate[℃/s]
Room
temp.)
0.5(Below
room
temp.)
temp.)
0.5
(Below
room
temp.)
Speed
range
[rpm]
200-1500
200-1500
-
-
Mixing
diameter
[mm]
3
3
-
-
LCD
display
LCD
LCD
LCD
LCD
Program
6 stages,
9programs
Voltage
100-240V
100-240V
100-240V
100-240V
Frequency
50/60Hz
50/60Hz
50/60Hz
50/60Hz
Power
200W
200W
200W
200W
This manual suits for next models
3
Table of contents
Other DLab Laboratory Equipment manuals

DLab
DLab Liquid Handling HiPette User manual

DLab
DLab RE100-Pro User manual

DLab
DLab MX-T6-Pro User manual

DLab
DLab OS40-Pro User manual

DLab
DLab SK-O330-Pro User manual

DLab
DLab MS7-H550-S User manual

DLab
DLab MS-H380-Pro User manual

DLab
DLab D3024 User manual

DLab
DLab OS20-Pro User manual

DLab
DLab D2012 User manual