Drive DeVilbiss Healthcare Theia User manual

Dynamic System
Instructions for use

2
MUTE STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
• Mattress only - For profiling beds, it is
essential that straps are secured around the
movable sections of the bed frame - damage
will be incurred when profiled if secured to
fixed parts of the frame (FIG 1).
• Cushion only: It is the responsibility of the
carer to ensure the chair is suitable for
product compatibility and the safety of the
patient. Position the cushion with the pipes at
the rear of the chair
• To avoid any risk of damage to the mattress/
cushion ensure there are no sharp objects
which may come into contact with it.
• Position the control unit by hanging the
hooks over the foot board. If there is no foot
board or the cushion is in use place the unit
on the floor with the front facing upwards.
Ensure the rear of the unit is not obstructed
by carpet, rugs etc. It is advisable to place the
unit on a firm surface (FIG 2).
• The mattress /cushion will start to inflate.
Inflation can take up to 40mins. Once inflated,
ensure the straps attaching the mattress
to the bed frame are secure and hold the
mattress in place. Secure sheets loosely
enough to ensure they do not interfere with
cell alternation.
CPR - mattress only
• Rapid deflation of the mattress may be required
for emergency treatment or system deflation.
The CPR pull tag is located at the foot end of
the mattress.
• Pull the CPR tag to disengage from the
mattress, once done the entire system will
rapidly deflate.
• To re-inflate, re attach the CPR tag (FIG 4).
• Wait for the mattress system to reach optimal
pressure prior to a return to normal use.
QUICK REFERENCE GUIDE P1 of 2
FIG 1. FIG 2.
FIG 3. FIG 4.
• Connect the mattress/cushion to the control
unit (FIG 5) ensuring the feed pipes do not
kink or become trapped between bed/chair
parts.
• Switch pump on by the ON/OFF switch (FIG 3).
• Mattresses - ensure that the CPR tag is
fully engaged within the mattress and all
connectors are lined up correctly (FIG 4).
TO BE USED IN CONJUNCTION WITH THE FULL INSTRUCTIONS FOR USE FOR THIS
PRODUCT. THIS QUICK GUIDE DOES NOT REPLACE THE FULL DOCUMENT.
THEIA Pump with mattress options and cushion
Overlay Mattress Air on Foam Mattress Dynamic Cushion
The mattress is intended to support a
single patient who is up to 180kg in weight
and 185cm in height. For those patients
of a very low weight, typically less than
40kg and a physical size less than 146cm,
clinical judgment is to be used to determine
suitability. A lower (or upper) age limit is not
defined as it depends on the physical size of
the patient in relation to the proportions of
the mattress and bed frame.
The mattress is intended to support a
single patient who is up to 180kg in weight
and 185cm in height. For those patients
of a very low weight, typically less than
40kg and a physical size less than 146cm,
clinical judgment is to be used to determine
suitability. A lower (or upper) age limit is not
defined as it depends on the physical size of
the patient in relation to the proportions of
the mattress and bed frame.
The cushion is intended to support a single
patient who is up to 120kg in weight. A
lower (or upper) age limit is not defined
as it depend on the physical size of the
patient in relation to the proportions of
the cushion and the platform. Clinical
judgement is to be used to determine
patient suitability.

3
QUICK REFERENCE GUIDE P2 of 2
TO BE USED IN CONJUNCTION WITH THE FULL INSTRUCTIONS FOR USE FOR THIS
PRODUCT. THIS QUICK GUIDE DOES NOT REPLACE THE FULL DOCUMENT.
No. Symbol Description
1
MUTE STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
Power Button: Press to turn
system ON/OFF, green light will
indicate pump is ON.
2
MUTE STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
Comfort Control: Press the UP
arrow to increase pressure and
the DOWN arrow to decrease
pressure, comfort setting
shown in the digital display.
It’s recommended to adjust
comfort control to the max when
inflating, before adjusting the
pressure to the patient’s needs.
Before changing the comfort
control a clinical judgement is
required.
THEIA Pump with mattress options and cushion
MUTE STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
No. Symbol Description
3
MUTE
STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
Static Mode: Press to set the
air within the support surface
to stay static, green light will
indicate static mode is selected.
The system will revert back to
alternating mode after 20 mins.
MUTE
STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
Mute : The audible alert will
sound when the pressure within
the system is low or a power
failure is detected. To mute,
press the mute button and to
reactivate the alert re-press the
mute button.
Transport Mode
You can achieve 8 hours transport time by
carrying out the following procedure:
• Disconnect the umbilical cord from the power
unit by squeezing the two tabs and pulling
away (FIG 5).
• Seal using the cap marked ‘Transport’ (FIG 6).
• Switch o the control unit. 1
1
2
FIG 5. FIG 6.
1342

4
CONTENTS
1. INTRODUCTION ...................................................................................
2. CONTACT INFORMATION ..................................................................
3. PRODUCT DESCRIPTION ...................................................................
3.1 Environment .............................................................................................
3.2 Intended User Group.............................................................................
3.3 Intended Use...........................................................................................
3.4 Product Overview ..................................................................................
3.5 Indications ...............................................................................................
3.6 Features ..................................................................................................
4. SAFETY ................................................................................................
4.1 Warnings and Cautions ........................................................................ .
4.2 Risk Assessment ...................................................................................
4.3 Contraindications ..................................................................................
4.4 Training ....................................................................................................
4.5 Biocides ...................................................................................................
4.6 General Warnings ..................................................................................
5. TRANSPORT AND STORAGE .............................................................
5.1 Storage .....................................................................................................
5.2 Transportation .......................................................................................
6. SYMBOL DEFINITION .........................................................................
7. INSTALLATION ....................................................................................
8. OPERATION .........................................................................................
8.1 Preparing for Use ...................................................................................
8.2 Control Interface ...................................................................................
8.3 CPR Function ..........................................................................................
8.4 Low Pressure Warning ..........................................................................
8.5 Power Failure Warning .........................................................................
9. DECONTAMINATION/CLEANING ......................................................
9.1 Control Unit ......................................................................................... ....
9.2 Mattress and Cushion ...........................................................................
9.3 Alternative Cover Cleaning Instructions ...........................................
10. MAINTENANCE ..................................................................................
11. DISPOSAL ...........................................................................................
12. SPECIFICATION .................................................................................
12.1 18” Cushion Specification ....................................................................
12.2 Air on Foam Mattress Specification ..................................................
12.3 Overlay Mattress Specification .........................................................
12.4 Air on Air Mattress Specification .......................................................
12.5 Hybrid Mattress Specification ...........................................................
13. ELECTROMAGNETIC COMPATIBILITY (EMC) .................................
14. WARRANTY ........................................................................................
5
5
6
6
6
6
6
6
7
7
7
7
8
8
9
9
10
10
10
11
13
15
15
16
17
17
17
18
18
19
19
20
21
22
22
23
24
25
26
27
30

5
Forassistanceinsettingup,using,maintainingyoursystem,toreportunexpected
operation or for any service, warranty, sales or customer service information
regarding this product please contact your provider or if in doubt contact Drive
DeVilbiss Healthcare Ltd. at the following address:
Thank you for purchasing this product. These instructions for use should be read
carefully before operating the system. Please ensure that you understand all
instructions, if you have any questions concerning the operation or maintenance
of the system please contact your provider/supplier who will provide you with
expert professional advice. Drive DeVilbiss Healthcare Ltd. recommend the
system is maintained by Drive DeVilbiss Healthcare Ltd. service engineers or
qualified personnel.
1. INTRODUCTION
2. CONTACT INFORMATION
Drive DeVilbiss Healthcare Ltd.
Sidhil Business Park,
Holmfield,
Halifax,
HX2 9TN
United Kingdom
For Service & Support outside the United Kingdom & Northern Ireland please
contact the local distribution company from where this equipment was
purchased. Failure to do so may result in the manufacturer’s warranty becoming
void.
Service & Maintenance
Tel: +44 (0)1422 233136
Fax: +44 (0)1422 233010
Spares
Tel: +44 (0)1422 233138
Fax: +44 (0)1422 233010
Sales
Tel: +44 (0) 845 0600 333
Fax: +44 (0) 845 0600 334
www.drivedevilbiss.co.uk
Airflo (xiamen) Medical Co., Ltd
1F, 3F, 4F, No. 6, East Haijing Road, Haicang Xiamen, Fujian,
China
Y. Sung Handelsvertretung
Duesselihaler Str. 24, 40211 Duesseldorf Germany
Drive DeVilbiss Healthcare Ltd.
Sidhil Business Park, Holmfield, Halifax, HX2 9TN, UK
IMPORTER:

6
3. PRODUCT DESCRIPTION
3.1 Environment
Your dynamic system is intended for use in the following environments:
• A hospital where intensive/acute care is provided and medical supervision
is required and monitoring provided.
• A long term care area where medical supervision is required and
monitoring is provided if necessary (e.g. nursing homes, rehabilitation
facilities, geriatric facilities).
• A domestic area.
3.2 Intended User Groups
3.2.1 Cushion
The cushion is intended to support a single patient who is up to 120kg in weight.
A lower (or upper) age limit is not defined as it depend on the physical size of
the patient in relation to the proportions of the cushion and the platform. Clinical
judgement is to be used to determine patient suitability.
3.2.2 Mattresses
The Air on Foam and Overlay mattresses are intended to support a single
patient who is up to 180kg in weight and 185cm in height. For those patients of
a very low weight, typically less than 40kg and a physical size less than 146cm,
clinical judgment is to be used to determine suitability. A lower (or upper) age
limit is not defined as it depends on the physical size of the patient in relation to
the proportions of the mattress and bed frame.
3.3 Intended Use
The intended use of the support surface is for sleeping/resting/sitting and it is
intended to assist in alleviation or prevention of a pressure related injury.
The support surface is intended to support a single adult. A risk assessment
must always be performed on the suitability of the patient to the support surface
to ensure they are both mentally and physically suited to using the support
surface with minimal risk of personal injury. The patient is only defined as such
when situated on the mattress/cushion.
The system is designed for use on both standard and profiling bed frames.
Ideally, patients allocated to this system will have some degree of independent
mobility or can be repositioned according to individual needs.
3.4 Product Overview
The Thiea control unit controls air cell pressure within the support surface,
which is attached by an umbilical tube. The support surface is made up of a
single layer of air cells or two layers, a top layer of air cells, and a bottom layer of
foam to provide additional support where needed. The control unit is intended
for continuous use.
The control unit is intended to be positioned on compatible bed frames only.

7
4. SAFETY
4.1 Warnings and Cautions
4.2 Risk Assessment
Cautions in these instructions for use highlight potential hazards that if
disregarded could lead to equipment damage or failure.
Warnings in these instructions for use highlight potential hazards that if
disregarded could lead to injury or death.
Caution
Warning
Control Unit
• Provides an air supply to the
mattress or cushion
• Rear bed hooks
•• Accessible rear filterAccessible rear filter
• 1 in 2 alternating cycle
•• 12 minute cycle time12 minute cycle time
•• Static inflate modeStatic inflate mode
• Adjustable comfort control
• Fault indicators with visual and
audible alerts
3.6 Features
3.5 Indications
To assist as part of an overall programme of care when active load distribution
through mechanical means is required.
Bed frames used with the system can vary greatly depending on the specific
healthcare setting (i.e. hospitals, nursing homes, home care etc). It is the
responsibility of the carer to carry out the necessary risk assessment to ensure
the safety of the patient.
Before a patient uses the dynamic mattress system a risk assessment must be
performed on a patient by patient basis. The risk assessment should include,
but is not limited to:
Cushion
• 2 way stretch, vapour permeable
and waterproof cover
• High frequency welded top
cover
• 360° zip
• Covered feed pipes
Mattress
• Bed platform securing straps
• 2 way stretch, vapour permeable
and waterproof cover
• High frequency welded top
cover
• 360° zip
• Covered feed pipes

8
• Product combinations (bed frame, mattress, side rails etc.).
• Extent of tissue damage (if any).
• Entrapment.
• Patient falls.
• Small adults (and children).
• Patients who have reduced capacity and are agitated and/or restless.
• Patients with burns.
• Unauthorised people with access to the controls.
4.3 Contraindications
Patient conditions for which the application of pressure relief on an alternating
mattress system is a contraindication are as follows:
• Cervical or skeletal traction.
• Unstable skeletal fractures.
• Unstable spinal injury.
• Exceeds maximum patient weight of the mattress.
• Gross Oedema (when using alternating mode only).
Other contraindications may be relevant which are specific to the patient or care
environment.
4.4 Training
If these instructions for use are not deemed sucient and the need for
training is required please contact your distributer who will be able to define
the intention and outcomes of any necessary training, who should attend, its
duration and any potential costs involved.
4.5 Biocides
Support surface covers contain a anti-fungal agent to control microbial
deterioration. The active ingredient is 3-iodo 2-propynyl butylcarbamate. The
active ingredient is fully encapsulated within the polymer coating. There are
no special handling requirements.
This product does not contain any Nano-materials and all components are
latex free.

9
4.6 General Warnings
• The system is to be installed and put into service in accordance with the
information provided in these instructions for use.
• The system is typically not suitable for child use, if it is to be used for
child occupancy ensure a risk assessment has been undertaken taking
into account the proportions of the child and dimensions of the system.
• Misused electrical equipment can be hazardous.
• Exposure of the control unit to any liquid while it is plugged in could
result in a severe electrical hazard.
• The control unit is a precision electronic product. Use care when
handling or transporting it. Dropping or other sudden impacts may result
in damage to the unit.
• Do not open the control unit – Risk of electrical shock.
• Repairs and service are to be conducted by suitably trained personnel.
If the control unit is not functioning properly, or has been damaged,
unplug the unit and take it out of service immediately.
• Modification of the mattress or control box is not allowed without the
permission of Drive DeVilbiss Healthcare Ltd. – A hazard could be
introduced.
• Occupants and users of this equipment must never smoke in close
proximity to the control unit, mattress, cushion or bedding being used
with it - Risk of fire.
• Accessories that have not been approved or designed for use with the
system are not be used.
• Control unit shall not be used in the presence of flammable gasses or
used in oxygen rich environments – Risk of explosion/fire.
• Control unit functions must be locked out when a patient is left
unattended.
• If children, adults who lack capacity or even pets pose a potential
risk of intentional or unintentional tampering with the control unit its
suitability for use is to be considered during the initial patient/product
risk assessment.
• The support surface is for single occupancy use. Additional weight
could damage the mattress or aect the performance of the mattress
system.
• Minimise articles (e.g. bedding) between the support surface and
patient, and secure bed sheets loosely so as not to aect product
functionality.
• Perform regular patient skin checks – Any tissue deterioration may
require equipment reallocation and/or a re-assessment of the care
being provided.
• Incompatible support platforms (eg. bed frame or mattress) can create
safety/stability hazards.
Warning

10
5. TRANSPORT AND STORAGE
5.1 Storage
• Detach the control unit from the support surface.
• Release CPR or rotate the CPR dial until it is open.
• Lay the mattress or cushion out flat and position upside down.
• Ensure there is no air trapped in the cells.
• Position the control unit on the mattress or cushion.
• The product should be rolled from the head end towards the foot end
(ensuring the control unit is fully covered).
• Place into storage bag/holdall to protect from dirt, debris, fluids etc.
• To prevent the risk of cross infection, when removing the system from
an end user’s residence ensure that all activities in relation to the
system are carried out using disposable gloves and that they are then
discarded appropriately, unless it can be verified that the cushion,
mattress and control unit have been suitably cleaned and disinfected
prior to collection.
• On the return of the system from an end users residence, prior to
putting into storage ensure it has been cleaned and disinfected in line
with the local infection control policy and/or as defined in section 10
of these instructions for use.
• Do not crease or stack mattresses, cushion and/or control units -
damage could be incurred.
• Do not store whilst inflated - damage could be incurred.
• Do not store objects such as side rails on top of the mattress or
cushion - damage could be incurred.
Warning
Caution
5.2 Transportation
Where possible, it is recommended the transport of mattresses should be
carried out on a flat based trolley or mattress trolley. Do not drag or pull the
mattress by its cover or foam core. Please follow local moving and handling
policies and guidelines when handling a mattress. It is recommended that two
people manoeuvre the mattress.
You can achieve 8 hours transport time by
carrying out the following procedure:
• Disconnect the umbilical cord from the
control unit by squeezing the two tabs and
pulling away (FIG 5).
• Seal using the cap marked ‘Transport’
(FIG 6).
• Switch o the control unit.
QUICK REFERENCE GUIDE P2 of 2
TO BE USED IN CONJUNCTION WITH THE FULL INSTRUCTIONS FOR USE FOR THIS
PRODUCT. THIS QUICK GUIDE DOES NOT REPLACE THE FULL DOCUMENT.
No. Symbol Description
1
LOW PRESSURE
NORMAL PRESSURE
MIN MAX
W
E
I
G
H
T
R
A
N
G
E
MID
MUTE STAT IC
Weight Range Control: Turn dial
clockwise to increase mattress
pressure and anti-clockwise to
decrease mattress pressure. It
is recommended to adjust the
control to the max when first
inflated. Before changing the
control a clinical judgement is
required.
2
LOW PRESSURE
NORMAL PRESSURE
MIN MAX
W
E
I
G
H
T
R
A
N
G
E
MID
MUTE
STATIC
Static Mode: Press to set the
air within the mattress to stay
static, green light will indicate
static mode is selected. The
system will not revert back to
alternating mode. The user must
press alternate button to return
to alternating mode.
Drive DeVilbiss Sidhil Ltd | APOLLO II | Quick Reference guide | REV 5
EROS Pump with mattress options and cushion
LOW PRESSURE
NORMAL PRESSURE
MIN MAX
W
E
I
G
H
T
R
A
N
G
E
MID
MUTE STATIC
3 21
No. Symbol Description
3
LOW PRESSURE
NORMAL PRESSURE
MIN MAX
W
E
I
G
H
T
R
A
N
G
E
MID
MUTE
STATIC
Alarm Mute: The audible alarm
will sound when the pressure
within the system is low or a
power failure is detected. To
mute, press the mute button and
to reactivate the alarm, re-press
the mute button.
LOW PRESSURE
NORMAL PRESSURE
MIN MAX
W
E
I
G
H
T
R
A
N
G
E
MID
MUTE STAT IC
Low Pressure : When abnormal
low pressure occurs, the low
pressure indicator will illuminate
and the alarm will be activated
to alert of a low pressure
condition.
Transport Mode
You can achieve 8 hours transport time by
carrying out the following procedure:
• Disconnect the umbilical cord from the power
unit by squeezing the two tabs and pulling
away
(FIG 5).
• Seal using the cap marked ‘Transport’
(FIG 6).
• Switch o the control unit.
1
1
2
FIG 5. FIG 6.

11
6. SYMBOL DEFINITION
The following symbols are found on this bed:
Refer to instructions for use - Recommended
Failure to read the instructions for use could introduce a hazard
Refer to instructions for use - Mandatory
Failure to read the instructions for use could introduce a hazard
Warning
Beware of potential hazard
Caution
Beware of potential product damage
Manufacturer
Date of manufacture
Lot number
Class II electrical device
The user/occupant is protected by at least two layers of insulation between the current
carrying parts (e.g. control box and mains cable) and the metal accessible parts – If
damage is noticed to any electrical component, turn off at the mains and contact your
provider or Drive DeVilbiss Healthcare Ltd. immediately.
Catalogue number
Medical device
MD
Authorized representative in the European Community
Serial number
SN
No Smoking
CE certification

12
Type BF applied part
Applied Part: The parts of the bed that come into physical contact with the user/occupant in
order for the bed to carry out its intended function (refer to section 16.2 for a list of applied
parts).
Type BF: Applied parts which are electrically isolated from earth and other parts of the
medical equipment - Complying with specific requirements for protection against electric
shock to EN 60601-1.
95°C Machine wash at 95OC
Do not iron
Do not dry clean
Tumble dry on low heat
Keep out of direct sunlight
Drip Dry
Maximum patient weight
(mattress)
Maximum patient weight
(cushion)
CPR Release
Safe working load
Mattress

13
7. INSTALLATION
• Ensure the support surface is used with a compatible side rail
and bed frame combination – Incorrect combinations can lead to
entrapment and/or falls hazards.
• Ensure the support surface is of the correct type for the patient –
Incorrect specification could lead to an injury.
• The mains plug is the disconnect device for the means of isolating
the control unit from the mains supply, the plug must be accessible
at all times.
• Ensure the mains cable is plugged into an appropriate power source
at all times.
• To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with a protective earth.
• Ensure the mains cable is not in tension, paying particular attention
to when the bed travels up/down.
• Precautions are to be taken when routing the mains cable around
the bed to ensure that it does not become squeezed, trapped or
damaged by the bedframe or other ancillary equipment - Risk of
electrocution.
• Any electrical cable that is part of the system or associated
ancillary equipment that is found to be damaged must be replaced
immediately - Damaged electrical cables can create a risk of
electrocution and/or fire.
• A CE marked extension cable must only be used when it is not
possible to reach a wall socket with the equipment mains cable –
Contact Drive DeVilbiss Healthcare Ltd. for detail in regards to safe
use of extension cables.
• If an extension cable is used never overload it by plugging in
appliances that together will exceed the maximum current rating
stated for the extension cable – Risk of fire.
• Block adaptors are not to be used.
• Ensure multiple socket outlets are not positioned under the bed
frame - Liquids that leak onto such a socket could pose an electrical/
fire risk.
• Consideration is to be taken in the positioning of the mains cable
and air hose to minimise the risk of accidental strangulation resulting
from patient, baby or child entanglement.
• Keep away from sources of heat and naked flames (e.g. cigarettes,
fireplaces, electric fires, fan heaters, electric blankets etc.) – Close
proximity could damage the electrical system and/or cover, bedding
could catch fire etc.
• Do not place any objects or items, such as blankets, on or over the
control unit - Risk of fire.
• Avoid placing the system in a moisture rich environment - Prolonged
exposure to moisture could damage the electrical system and pose
an electrical/fire risk.
Warning

14
• For profiling beds, it is essential that the straps are secured around
the movable sections of the bed frame – Damage will be incurred
when profiled if secured to fixed parts of the frame.
• To avoid any risk of damage to the mattress ensure there are no
sharp objects which may come in contact with it.
Caution
• Position the control unit by hanging the hooks over the foot board. If there
is no foot board place the unit on the floor with the front facing upwards.
Ensure the rear of the unit is not obstructed by carpet, rugs etc. It is advisable
to place the unit on a firm surface.
• If the system has come from a storage / transport
temperature environment near to the minimum or maximum
values stated allow the cushion/mattress and control unit
to adjust to room temperature for a minimum of 2 hours
prior to plugging into the mains supply - Risk of electrical
system damage if operated outside of the recommended
temperatures.
Caution
• Open all packaging with care.
• Once removed from the packaging check the product for any signs of
damage. If damaged do not put into use and contact your provider or Drive
DeVilbiss Healthcare Ltd. (See Section 2).
• Remove all covers, sheets and the existing mattress/cushion from the bed/
chair.
• Position the mattress on top of the bed frame, top cover facing upwards
and air hose at the foot of the bed for control unit positioning.
• If using a cushion, position the cushion onto a fixed chair which has a padded
seat, with the top cover facing upwards and air hose at a rear corner of the
seat for control unit positioning.
• If using the mattress attach to the bed frame by securing the adjustable
straps to the moving sections of the bed. If using a cushion, loosely secure
the cushion to the chair frame by using the attached securing straps.

15
8. OPERATION
8.1 Preparing for Use
Prior to patient usage of the system the following must be performed:
• Ensure the bed and mattress system are at room temperature.
• Ensure the bed and mattress system have been cleaned and disinfected.
• ‘Ensure the mattress cover has been checked for tears, punctures,
abrasion marks etc. and that their are no signs of fluid ingress.
• Attach the male air connector to the control unit, ensuring the air hose does
not kink or become trapped between parts of the bed frame.
• Plug the mains cable into a suitable mains supply and switch on the control
unit.
• The system will start to inflate. Inflation will be complete within 40 minutes.
• Once inflated and the low pressure light is out, ensure the straps that
attach the mattress/cushion to the bed frame/chair are secure and hold
the mattress/cushion in place, adjust as necessary.
• Once the mattress is fully inflated, the bedding can be replaced. Secure
sheets loosely enough to ensure they do not interfere with cell alternation.

16
MUTE STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
MUTE STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
MUTE STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
MUTE
STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
MUTE
STATIC
POWER FAILURE
LOW PRESSURE
WEIGHT RANGE
1
MIN MAXMID
4 8
Power Button - Press to turn system ON/OFF, green light will indicate
pump is ON.
Comfort Control - Press to UP arrow to increase pressure and the
DOWN arrow to decrease pressure, comfort setting shown in the
digital display. It’s recommended to adjust comfort control to the max
when inflating, before adjusting the pressure to the patient’s needs.
Before changing the comfort control a clinical judgement is required
from frequent monitoring and repositioning of the patient.
Static Mode - Press to set the air within the support surface to stay
static, green light will indicate static mode is selected.
Mute - The audible alert will sound when the pressure within the
system is low or a power failure is detected. To mute, press the mute
button and to reactivate the alert re-press the mute button.
8.2 Control Interface

17
8.3 CPR Function
When there is an emergency requirement to perform CPR on the patient, pull
the CPR strap at the foot end of the mattress to this will release the air quickly
from the mattress. The CPR strap is located at the left-hand side of the mattress.
8.4 Low Pressure Warning
When abnormal low pressure occurs, the low pressure indicator will illuminate
and the audible alert will be activated to alert of a low pressure condition.
Check if the connections are secure and correctly installed according to the
relevant instructions. If the pressure is consistently low, open the zipper and
confirm that all of the hoses are properly connected. Check for any noticeable
leakage from any of the tubes.
If necessary, contact your local dealer to replace any damaged tubes or hoses.
8.5 Power Failure Warning
If there should be power failure, for example caused by power cord unplugged
or power o when pump is working, the audible alert will be activated and power
failure indicator will flash. When power is back, the pump will automatically start
working, the power button does not require pressing again.

18
9. DECONTAMINATION/CLEANING
Infection control and routine cleaning must be carried out in accordance with
your local infection control policy or regulatory body.
Warning
• Always disconnect the support surface and bed frame from
the main power supply prior to cleaning.
• The control unit is rated to IP21 and provides protection
from condensation only, do not immerse or soak the control
unit – Risk of electric shock.
• Regular cleaning and disinfection of the support surface
will help to prevent the risk of infection to the occupant and/
or carer.
• Prior to transferring the system to another user ensure it has
been cleaned and disinfected using the method as detailed
below to help prevent the risk of cross infection.
• Deviations from the specified cleaning and disinfection
instructions can cause serious hazards, and adversely aect
the life and ecacy of the system.
• If any of the below washing instructions are not followed the
product warranty will be invalidated.
• Do not use solvents, neat bleach, phenolic based cleaning
solutions or abrasive products to clean the casing or
mattress.
Caution
9.1 Control Unit
• Check for external damage – If damaged take the control unit out of use.
• All surfaces to be wiped down with a disposable soft cloth moistened
with a mild detergent and diluted in warm water (40°C).
• The control unit is be cleaned by starting with the cleanest parts of it and
systematically moving to the dirtiest parts. Extra care should be taken
around areas where excess dirt or dust may gather.
• The cloth should be changed during the cleaning process if it becomes
soiled.
• Wipe down with a clean cloth moistened with clean water to remove
detergent residue.
• If there are blood spillages or bodily fluids present wipe surfaces down
with 0.1% Chlorine solution (1,000 ppm).
• Wipe down with a clean cloth moistened with clean water.
• Dry o with a paper towel - Always ensure the cleaned surfaces are
allowed to fully dry before putting back into use.

19
9.2 Mattress and Cushion
Before attempting to clean, the top cover is to be checked for physical signs
of damage that may lead to strike-through (ingress of fluid through cover). This
is achieved by unzipping the top cover and looking for signs of staining to the
white underside. Any evidence of strike-through (and/or cover damage) will
require a new cover to be fitted to the system.
Warning
The cover must not be used if strike-through is evident – Risk of
cross infection.
Caution
Frequent or prolonged exposure to higher concentration
disinfectant solutions may prematurely age the fabric cover of the
mattress.
General Cleaning:
• Wipe down with a clean cloth moistened with a mild detergent and
diluted in warm water (40˚C).
• Rinse with cold clean water and a clean cloth and allow to fully dry
before use.
• Ensure the internal foam and air cells are orientated correctly to the
mattress covers when reassembling.
9.3 Alternative Cover Cleaning Instructions
Alternatively disinfection of the cover may be achieved by laundering as follows:
• Remove mattress cover.
• Machine wash at 71°C for no less than 3 minutes or 65°C for no less than
10 minutes. Heavily soiled items should also have a pre-wash/sluice
cycle.
• Allow covers to fully dry before use.
(Refer to the Department of Health document HTM 01-04 for further details).
Decontamination:
• Mop up any fluid with paper towels.
• Wipe cover down using cold clean water.
• Wipe down with a 0.1% Chlorine solution (1,000ppm) in cold water, where
necessary a 1% Chlorine solution (10,000ppm) is to be used instead.
• Rinse down with cold clean water using a clean cloth.
• Dry o with paper towels - Always ensure the cleaned surfaces are
allowed to fully dry before putting back into use.

20
Warning
Only authorised service personnel or Drive DeVilbiss Healthcare Ltd. service
engineers should carry out repairs or service activities. For Service & Support
outside of the UK & Northern Ireland please contact the local distribution
company from where this equipment was purchased. Failure to do so may result
in the product warranty becoming void. The system must be serviced once
yearly, as a minimum. Drive DeVilbiss Healthcare Ltd. also recommends that
the carer performs frequent visual and operational inspections. If there are any
signs of damage or the system is not performing as it should withdraw it from
service until the system has been repaired and is fit for use again. Drive DeVilbiss
Healthcare Ltd. recommends that the following maintenance procedure is
performed every 12 months:
• Check that the air filter is in good condition and replace or clean as
required.
• Check that all electrical functions operate correctly on the control unit.
• Check that all audible and visual indicators work appropriately (when
plugged in and unplugged from mains supply).
• Check that the battery is still functional and operates in the event of a
power loss.
• Check that the mattress reaches the required pressures.
• Check the CPR connection on the mattress.
• Check the cover for tears, punctures, abrasion marks and split seams.
• Check for signs of fluid ingress to the underside of the cover.
• Check that all piping and cells within the mattress are in good condition
and that there is no kinking evident.
10. MAINTENANCE
• Always disconnect the control unit from the main power supply prior
to performing any maintenance procedures (when viable)
• No modification of this equipment is allowed.
• The mattress system should be vacated by the patient before any
maintenance or inspection takes place. If this is not possible due to
the patient’s mobility, care should be taken for the service engineer
not to make contact with the patient when working on electrical items.
• Only Drive DeVilbiss Healthcare Ltd. approved components specified
for the Theia dynamic system are to be used - if in doubt contact Drive
DeVilbiss Healthcare Ltd. or your local distributor.
Table of contents
Other Drive Personal Care Product manuals

Drive
Drive Rollator User manual

Drive
Drive 530 100 100 User manual

Drive
Drive SR8 STEEL ROLLATOR User guide

Drive
Drive FC 2024 Manual

Drive
Drive 530 100 100 User manual

Drive
Drive AGF-3E User manual

Drive
Drive 15019 User manual

Drive
Drive DIGI-SEAT User manual

Drive
Drive PainAway Pro with Heat User manual

Drive
Drive 15065 User manual