
7560100010L02 1805V002 3
Important information
2.5 General safety notes
iAlways comply with the specifications of all
guidelines, laws, and other rules and regula-
tions applicable at the site of operation for the
operation of this unit.
iCheck the function and condition of the unit
prior to every use.
iDo not convert or modify the unit.
iObserve the installation instructions.
iMake the installation instructions available to
the operator of the unit at all times.
2.6 Specialist personnel
Installation and repairs
iInstallation, readjustments, alterations, up-
grades and repairs must be carried out by
Dürr Dental or by qualified personnel specifi-
cally approved and authorized by Dürr Dental.
2.7 Protection from electric shock
iComply with all the relevant electrical safety
regulations when working on the unit.
iNever touch the patient and unshielded plug
connections on the device at the same time.
iReplace any damaged cables or plugs imme-
diately.
Observe the EMC rules concerning medical
devices
iThe appliance is designed for the use in health
care establishments (in accordance with IEC
60601-1-2). If the appliance is operated in an-
other environment, potential effects on elec-
tromagnetic compatibility must be taken into
account.
iDo not operate the unit in the vicinity of HF
surgical instruments or MRT equipment.
iMaintain a minimum distance of at least 30
cm between the unit and other electronic de-
vices.
iKeep a minimum distance of 30cm between
the unit and mobile radio devices.
iNote that cable lengths and cable extensions
have effects on electromagnetic compatibility.
2 Safety
Dürr Dental has designed and constructed this
unit so that when used properly and for the in-
tended purpose it does not pose any danger to
people or property. Nevertheless, residual risks
can remain. You should therefore observe the
following notes.
2.1 Intended purpose
The purpose of the place selection valve is to
specifically open the volume flow between the
suction unit and the hose manifold.
2.2 Intended use
The installation position and technical data must
be observed. Installation as a suction valve in a
dental treatment unit or in other areas of a den-
tal suction unit for one treatment chair. Only the
media associated with dental treatment
(e.g.water, saliva, dentine and amalgam) must
be suctioned.
2.3 Improper usage
Any other usage or usage beyond this scope is
deemed to be improper. The manufacturer ac-
cepts no liability for damages resulting from im-
proper use. In these cases the user/operator will
bear the sole risk.
2.4 Systems, connection with other
devices
Additional devices connected with medical elec-
trical devices must be proven to conform with
their corresponding IEC or ISO standards. All
configurations must continue to comply with the
standard requirements for medical systems (see
IEC 60601-1).
Whoever connects additional devices to medical
electrical devices automatically becomes the
system configurator and is responsible for en-
suring that the system corresponds with the
standard requirements for systems. Local laws
take priority over the requirements outlined
above.
EN