manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Eaton
  6. •
  7. Measuring Instrument
  8. •
  9. Eaton Crouse-hinds series User manual

Eaton Crouse-hinds series User manual

DRAFT 22 June 2016
MTL GIR6000
replacement modules
Individual sensor, pump & filter modules
November 2016
INS MTL GIR6000 Rev 2
Instruction sheet
MTL gas analysers & systems
Eaton Electric Limited,
Great Marlings, Butterfield, Luton
Beds, LU2 8DL, UK.
Tel: + 44 (0)1582 723633 Fax: + 44 (0)1582 422283
E-mail: [email protected]
www.mtl-inst.com
© 2016 Eaton
All Rights Reserved
Publication No. INS MTL GIR6000 Rev 2
November 2016
1
The given data is only intended as a product
description and should not be regarded as a legal
warranty of proper ties or guarantee. In the interest
of further technical developments, we reserve the
right to make design changes.
EUROPE (EMEA):
+44 (0)1582 723633
[email protected]
THE AMERICAS:
+1 800 835 7075
[email protected]
ASIA-PACIFIC:
+65 6 645 9888
[email protected]
DRAFT 22 June 2016
MTL GIR6000
replacement modules
Individual sensor, pump & filter modules
November 2016
INS MTL GIR6000 Rev 2
Instruction sheet
MTL gas analysers & systems
Eaton Electric Limited,
Great Marlings, Butterfield, Luton
Beds, LU2 8DL, UK.
Tel: + 44 (0)1582 723633 Fax: + 44 (0)1582 422283
E-mail: mt[email protected]
www.mtl-inst.com
© 2016 Eaton
All Rights Reserved
Publication No. INS MTL GIR6000 Rev 2
November 2016
1
The given data is only intended as a product
description and should not be regarded as a legal
warranty of proper ties or guarantee. In the interest
of further technical developments, we reserve the
right to make design changes.
EUROPE (EMEA):
+44 (0)1582 723633
mt[email protected]
THE AMERICAS:
+1 800 835 7075
mt[email protected]
ASIA-PACIFIC:
+65 6 645 9888
sales.mt[email protected]
Introduction
Modular Gas Sensors are the key feature of the MTL GIR6000
Biogas Analyser design. They are designed to be replaceable,
on-site, with minimal interruption to the operation of the
instrument, given that the location of the instrument is in a
recognised hazardous area.
In addition to the main Gas Sensor modules there are three other
module types, namely, the Pump module, the Filter module
and the Link module. This document describes the replacement
process for any of these module types.
CAUTION
READ THE SAFETY INFORMATION OVERLEAF!
Do not begin work until you have read and understood the
information provided there.
These instructions should be read in conjunction with the
Instruction Manual – INM MTL GIR6000.
Initial instrument shutdown
To add or change any replaceable module, it is important to
follow the procedure described here.
1. Isolate the sample gas inlet supply- using the local manual
isolation valve.
2. Disconnect the sample gas inlet pipe from the analyser.
WARNING!
Failure to isolate the sample gas supply pipe
before disconnection may result in toxic
gas escaping.
3. Leave the analyser running for 5 minutes (to purge air
through system)
NOTE
Purging the gas analyser will result in gas readings falling to
zero, which may generate alarms.
4. Power down the analyser using the external AC power
isolation switch.
WARNING!
Failure to shut down the MTL GIR6000
before replacing a module may result in an
explosion risk.
5. Open the outer door.
Module removal and replacement
CAUTION
It is advised that any person handling an unplugged
module should wear an ESD grounding wrist strap to
prevent inadvertent static damage to any of the module’s
electrical components. The ESD COMMON GROUND jack
socket provided at the lower left corner of the inner door
may be used if the enclosure is provided with a suitable
ground connection.
Each module is held in place with two threaded bolts at the top and
bottom of its front panel. The modules are keyed and will only fit in
the door panel at the appropriate location. A Phillips or flat-bladed
screwdriver is required to loosen the bolts after which they can be
freed by hand.
The module should be drawn straight out of the panel (see
photograph), which will disconnect both the gas and electrical
connections at the rear of the module automatically.
Additional effort will be required to remove a Filter Module
because of the additional O-rings it uses. Remove the Module
by exerting equal force on the both the handle and top bolt.
Module removal Module insertion
Introduction
Modular Gas Sensors are the key feature of the MTL GIR6000
Biogas Analyser design. They are designed to be replaceable,
on-site, with minimal interruption to the operation of the
instrument, given that the location of the instrument is in a
recognised hazardous area.
In addition to the main Gas Sensor modules there are three other
module types, namely, the Pump module, the Filter module
and the Link module. This document describes the replacement
process for any of these module types.
CAUTION
READ THE SAFETY INFORMATION OVERLEAF!
Do not begin work until you have read and understood the
information provided there.
These instructions should be read in conjunction with the
Instruction Manual – INM MTL GIR6000.
Initial instrument shutdown
To add or change any replaceable module, it is important to
follow the procedure described here.
1. Isolate the sample gas inlet supply- using the local manual
isolation valve.
2. Disconnect the sample gas inlet pipe from the analyser.
WARNING!
Failure to isolate the sample gas supply pipe
before disconnection may result in toxic
gas escaping.
3. Leave the analyser running for 5 minutes (to purge air
through system)
NOTE
Purging the gas analyser will result in gas readings falling to
zero, which may generate alarms.
4. Power down the analyser using the external AC power
isolation switch.
WARNING!
Failure to shut down the MTL GIR6000
before replacing a module may result in an
explosion risk.
5. Open the outer door.
Module removal and replacement
CAUTION
It is advised that any person handling an unplugged
module should wear an ESD grounding wrist strap to
prevent inadvertent static damage to any of the module’s
electrical components. The ESD COMMON GROUND jack
socket provided at the lower left corner of the inner door
may be used if the enclosure is provided with a suitable
ground connection.
Each module is held in place with two threaded bolts at the top and
bottom of its front panel. The modules are keyed and will only fit in
the door panel at the appropriate location. A Phillips or flat-bladed
screwdriver is required to loosen the bolts after which they can be
freed by hand.
The module should be drawn straight out of the panel (see
photograph), which will disconnect both the gas and electrical
connections at the rear of the module automatically.
Additional effort will be required to remove a Filter Module
because of the additional O-rings it uses. Remove the Module
by exerting equal force on the both the handle and top bolt.
Module removal Module insertion
The given data is only intended as a product
description and should not be regarded as a legal
warranty of proper ties or guarantee. In the interest
of further technical developments, we reserve the
right to make design changes.
EUROPE (EMEA):
+44 (0)1582 723633
[email protected]
THE AMERICAS:
+1 800 835 7075
[email protected]
ASIA-PACIFIC:
+65 6 645 9888
[email protected]
The given data is only intended as a product
description and should not be regarded as a legal
warranty of proper ties or guarantee. In the interest
of further technical developments, we reserve the
right to make design changes.
EUROPE (EMEA):
+44 (0)1582 723633
mt[email protected]
THE AMERICAS:
+1 800 835 7075
mt[email protected]
ASIA-PACIFIC:
+65 6 645 9888
sales.mt[email protected]
Eaton Electric Limited,
Great Marlings, Butterfield, Luton
Beds, LU2 8DL, UK.
Tel: + 44 (0)1582 723633 Fax: + 44 (0)1582 422283
E-mail: [email protected]
www.mtl-inst.com
© 2016 Eaton
All Rights Reserved
Publication No. INS MTL GIR6000 Rev 2 141116
November 2016
Eaton Electric Limited,
Great Marlings, Butterfield, Luton
Beds, LU2 8DL, UK.
Tel: + 44 (0)1582 723633 Fax: + 44 (0)1582 422283
E-mail: mt[email protected]
www.mtl-inst.com
© 2016 Eaton
All Rights Reserved
Publication No. INS MTL GIR6000 Rev 2 141116
November 2016
MTL GIR6000 replacement modules
November 2016
MTL GIR6000 replacement modules
November 2016
2 2
NOTE
Before ANY module is inserted into a position on the inner
door, ALWAYS check the O-ring seals for any sign of damage;
replace any that appear to show signs of damage or wear. It
is also good practice to do a quick visual inspection on the
pins of the module’s multi-way connector to check for any
bent or distorted pins.
Remove any packaging from the new module and, before
inserting it into the door panel, take care to align it with the
guides on each side of the opening. When the guides are
engaged, insert the module gently into the door until the
connectors are encountered then press firmly to engage both
gas and electrical connectors.
Tighten the two bolts with the fingers and then use a
screwdriver to tighten the bolts to a recommended torque of 1.5
Nm – i.e. sufficient to prevent them being unscrewed by hand.
Leak Testing
It is now necessary, following any module replacement, to
perform a Module Leak Test. All leak-testing procedures are
described in the MTL GIR6000 Instruction Manual (Section 7.4.
Leak-testing)
ATEX Safety Instructions for
MTL GIR6000 sensor, pump & filter modules
The replacement modules form part of the certification of the
MTL GIR6000 and are not separately certified. Refer to the
certificate and instructions of the MTL GIR6000 for full details.
The following information is in accordance with the Essential
Health and Safety Requirements (Annex II) of the EU
Directive 94/9/EC [the ATEX Directive - safety of apparatus]
and is provided for those locations where such requirements
are applicable.
General
a) This module must only be installed, operated and maintained
by competent personnel. Such personnel shall have
undergone training, which included instruction on the various
types of protection and installation practices, the relevant
rules and regulations, and on the general principles of area
classification. Appropriate refresher training shall be given on
a regular basis. [See clause 4.2 of EN 60079-17].
b) The module is designed for operation in a MTL GIR6000 gas
analyser which is mounted in a Zone 2 hazardous area.
c) This equipment has been designed to meet the requirements
of EN 60079-0 and EN 60079-15.
Installation
a) The installation must comply with the appropriate European,
national and local regulations, which may include reference
to the IEC code of practice IEC 60079-14. In addition,
particular industries or end users may have specific
requirements relating to the safety of their installations and
these requirements should also be met. For the majority of
installations the Directive 1999/92/EC [the ATEX Directive -
safety of installations] is also applicable.
b) The module is designed for operation in a Zone 2 hazardous
area.
c) The removable modules must not be inserted or removed
unless the MTL GIR6000 has been de-energised and isolated
from the sample gas supply.
Inspection and maintenance
a) Inspection and maintenance should be carried out in
accordance with European, national and local regulations
which may refer to the IEC standard IEC 60079-17. In addition
specific industries or end users may have requirements which
should also be met.
b) There are potential electrostatic hazards when cleaning the
equipment. Clean only with a damp cloth
Repair
A module cannot be repaired by the user and must be replaced
with an equivalent certified product – see ‘Disposal’ on this
instruction sheet.
Disposal of modules
Any module removed from this equipment must
not be treated as general waste. By ensuring
that this product is disposed of correctly you
will be helping to prevent potentially negative
consequences for the environment and human
health, which could otherwise be caused by
incorrect waste handling of this product.
For more detailed information about take-back and equipment
recycling visit our website.
NOTE
Before ANY module is inserted into a position on the inner
door, ALWAYS check the O-ring seals for any sign of damage;
replace any that appear to show signs of damage or wear. It
is also good practice to do a quick visual inspection on the
pins of the module’s multi-way connector to check for any
bent or distorted pins.
Remove any packaging from the new module and, before
inserting it into the door panel, take care to align it with the
guides on each side of the opening. When the guides are
engaged, insert the module gently into the door until the
connectors are encountered then press firmly to engage both
gas and electrical connectors.
Tighten the two bolts with the fingers and then use a
screwdriver to tighten the bolts to a recommended torque of 1.5
Nm – i.e. sufficient to prevent them being unscrewed by hand.
Leak Testing
It is now necessary, following any module replacement, to
perform a Module Leak Test. All leak-testing procedures are
described in the MTL GIR6000 Instruction Manual (Section 7.4.
Leak-testing)
ATEX Safety Instructions for
MTL GIR6000 sensor, pump & filter modules
The replacement modules form part of the certification of the
MTL GIR6000 and are not separately certified. Refer to the
certificate and instructions of the MTL GIR6000 for full details.
The following information is in accordance with the Essential
Health and Safety Requirements (Annex II) of the EU
Directive 94/9/EC [the ATEX Directive - safety of apparatus]
and is provided for those locations where such requirements
are applicable.
General
a) This module must only be installed, operated and maintained
by competent personnel. Such personnel shall have
undergone training, which included instruction on the various
types of protection and installation practices, the relevant
rules and regulations, and on the general principles of area
classification. Appropriate refresher training shall be given on
a regular basis. [See clause 4.2 of EN 60079-17].
b) The module is designed for operation in a MTL GIR6000 gas
analyser which is mounted in a Zone 2 hazardous area.
c) This equipment has been designed to meet the requirements
of EN 60079-0 and EN 60079-15.
Installation
a) The installation must comply with the appropriate European,
national and local regulations, which may include reference
to the IEC code of practice IEC 60079-14. In addition,
particular industries or end users may have specific
requirements relating to the safety of their installations and
these requirements should also be met. For the majority of
installations the Directive 1999/92/EC [the ATEX Directive -
safety of installations] is also applicable.
b) The module is designed for operation in a Zone 2 hazardous
area.
c) The removable modules must not be inserted or removed
unless the MTL GIR6000 has been de-energised and isolated
from the sample gas supply.
Inspection and maintenance
a) Inspection and maintenance should be carried out in
accordance with European, national and local regulations
which may refer to the IEC standard IEC 60079-17. In addition
specific industries or end users may have requirements which
should also be met.
b) There are potential electrostatic hazards when cleaning the
equipment. Clean only with a damp cloth
Repair
A module cannot be repaired by the user and must be replaced
with an equivalent certified product – see ‘Disposal’ on this
instruction sheet.
Disposal of modules
Any module removed from this equipment must
not be treated as general waste. By ensuring
that this product is disposed of correctly you
will be helping to prevent potentially negative
consequences for the environment and human
health, which could otherwise be caused by
incorrect waste handling of this product.
For more detailed information about take-back and equipment
recycling visit our website.

Other manuals for Crouse-hinds series

95

This manual suits for next models

1

Other Eaton Measuring Instrument manuals

Eaton CMD User manual

Eaton

Eaton CMD User manual

Eaton Crouse-hinds series User manual

Eaton

Eaton Crouse-hinds series User manual

Eaton PXM4K Series User manual

Eaton

Eaton PXM4K Series User manual

Eaton Crouse-Hinds MTL Z230 User manual

Eaton

Eaton Crouse-Hinds MTL Z230 User manual

Eaton IQ 35M Series User manual

Eaton

Eaton IQ 35M Series User manual

Eaton Durant Eclipse Series User manual

Eaton

Eaton Durant Eclipse Series User manual

Eaton Cutler-Hammer Manual

Eaton

Eaton Cutler-Hammer Manual

Eaton IQ 100 series Guide

Eaton

Eaton IQ 100 series Guide

Eaton Eclipse Series User manual

Eaton

Eaton Eclipse Series User manual

Eaton MEMSHIELD3 Manual

Eaton

Eaton MEMSHIELD3 Manual

Eaton GIM 60 User manual

Eaton

Eaton GIM 60 User manual

Eaton SEFELEC 1500M User manual

Eaton

Eaton SEFELEC 1500M User manual

Eaton Cutler-Hammer IQ 300 Series Programming manual

Eaton

Eaton Cutler-Hammer IQ 300 Series Programming manual

Eaton KK650 User manual

Eaton

Eaton KK650 User manual

Eaton PDI WaveStar Infinity 2 Manual instruction

Eaton

Eaton PDI WaveStar Infinity 2 Manual instruction

Eaton Durant 47000-400 Operating and maintenance instructions

Eaton

Eaton Durant 47000-400 Operating and maintenance instructions

Eaton K1650 User manual

Eaton

Eaton K1650 User manual

Eaton Remote Monitoring Device User manual

Eaton

Eaton Remote Monitoring Device User manual

Eaton xComfort CSEZ-01/06 Manual

Eaton

Eaton xComfort CSEZ-01/06 Manual

Eaton 5775X-42X User manual

Eaton

Eaton 5775X-42X User manual

Eaton EMC3P-P2C1 User manual

Eaton

Eaton EMC3P-P2C1 User manual

Eaton Power Xpert PXM 4000 User manual

Eaton

Eaton Power Xpert PXM 4000 User manual

Eaton PXM 2250 User manual

Eaton

Eaton PXM 2250 User manual

Eaton 1MM User manual

Eaton

Eaton 1MM User manual

Popular Measuring Instrument manuals by other brands

HP Q METER 4342A Operating and service manual

HP

HP Q METER 4342A Operating and service manual

Dinel MLM-35 instruction manual

Dinel

Dinel MLM-35 instruction manual

ELGO Electronic EMAX-HI Series operating manual

ELGO Electronic

ELGO Electronic EMAX-HI Series operating manual

BSD LH-TB1000 operation instruction

BSD

BSD LH-TB1000 operation instruction

Danfoss PR-SC4K Modbus installation guide

Danfoss

Danfoss PR-SC4K Modbus installation guide

Teledyne WIRELESS GROUNDLINK WQAR user guide

Teledyne

Teledyne WIRELESS GROUNDLINK WQAR user guide

LORECa Hearing Instrument PKA 84 user manual

LORECa Hearing Instrument

LORECa Hearing Instrument PKA 84 user manual

Molectron EPM1000 user guide

Molectron

Molectron EPM1000 user guide

Leica Disto Special5 user manual

Leica

Leica Disto Special5 user manual

ATP Instrumentation 8820 instruction manual

ATP Instrumentation

ATP Instrumentation 8820 instruction manual

Carrier KW320 specification

Carrier

Carrier KW320 specification

Endress+Hauser Levelflex FMP53 operating instructions

Endress+Hauser

Endress+Hauser Levelflex FMP53 operating instructions

Newtons4th PPA55 Series Startup guide

Newtons4th

Newtons4th PPA55 Series Startup guide

Nuvair Pro VOC Alarm user manual

Nuvair

Nuvair Pro VOC Alarm user manual

Agrosta 100USB quick start guide

Agrosta

Agrosta 100USB quick start guide

SENDOT FluoMini Pro 100 Installation & user manual

SENDOT

SENDOT FluoMini Pro 100 Installation & user manual

AST A450C+ FO PL user manual

AST

AST A450C+ FO PL user manual

GREISINGER GMH 3691 operating manual

GREISINGER

GREISINGER GMH 3691 operating manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.