Emerson dixell XM664K User manual

1598013141 XM664K GB CLABO rev1.2 2012.01.26 XM664K 1/6
XM664K
1. GENERAL WARNING____________________________________________________________ 1
2. GENERAL DESCRIPTION________________________________________________________ 1
3. CONTROLLING LOADS__________________________________________________________ 1
4. FRONT PANEL COMMANDS _____________________________________________________ 1
5. SECTIONS MENU ______________________________________________________________ 2
6. HOW TO CHANGE THE ACTIVE MAP______________________________________________ 2
7. INSTALLATION AND ASSEMBLY__________________________________________________ 2
8. ELECTRICAL CONNECTIONS ____________________________________________________ 2
9. TTL SERIAL LINE_______________________________________________________________ 2
10. HOW TO USE THE HOT KEY _____________________________________________________ 2
11. ALARMS ______________________________________________________________________ 3
12. TECHNICAL DATA ______________________________________________________________ 3
13. CONNECTIONS ________________________________________________________________ 3
14. DEFAULT SETTING VALUES _____________________________________________________ 4
1. GENERAL WARNING
1.1 PLEASE READ BEFORE USING THIS MANUAL
This manual is partof the product and should be kept neartheinstrumentfor easy and quick reference.
The instrument shall not be used for purposes different from those described hereunder. It cannot beused
as a safety device.
Checkthe application limits before proceeding.
1.2 SAFETY PRECAUTIONS
Check the supply voltage is correct before connecting the instrument.
Do not expose to water or moisture: use the controller only within the operating limits avoiding sudden
temperature changes with high atmospheric humidity to prevent formation of condensation
Warning: disconnect all electrical connections before any kind of maintenance.
Fit the probe where it is not accessible by the End User. The instrument must not be opened.
In case of failure or faulty operation send the instrument back to the distributor (see address) with a
detailed description of the fault.
Consider the maximum current which can be applied to each relay (see Technical Data).
Ensure that the wires for probes, loads and the power supply are separated and far enough from each
other, without crossing or intertwining.
In case of applications in industrial environments, the use of mains filters (our mod. FT1) in parallel with
inductive loads could be useful.
2. GENERAL DESCRIPTION
The XM664K is microprocessor controller suitable for medium and low temperature multiplexed cabinet
applications composed of as many as 8 sections. This makes it possible to create a LAN (local area network)
made up of the instruments which, depending on the programming, can operate as single controllers or
managing the controls which arrive from one or more sections. The XM664K is equipped with 6 relay outputs
which control the compressor, the defrost 1 and 2 (which can be either electric or with hot gas, the evaporator
fans, the lights and an ON/OFF output. It is supplied with three NTC probe inputs, one for thermostating, one for
temperature control at the end of defrost on evaporator 1 and one for end of defrost 2 control. There are also
two digital inputs which can be configured by parameters.
The two LAN outputs provide an easy connection between the various controllers while the TTL connector
programsthecomplete parameter list by means ofa programming keycalled “Hot Key”.
3. CONTROLLING LOADS
3.1 COMPRESSOR
The regulation is performed according to the temperature measured by the thermostat probe with a positive
differential from the set point: if the temperature increases and reaches set point plus differential the compressor
is started and then turned off when the temperature reaches the set point value again.
In case of fault in the thermostat probe the start and stop of the compressor are timed through parameters Con
andCoF.
3.2 DEFROST
There are three types of defrost available that can be selected with the parameter tdF: defrost with electric
heater (tdF=EL), with hot gas (tdF=in) or disabled (tdF=oFF). When defrost is over, the dripping time starts,
which is managed by the parameter Fdt.
Management of two defrosts is configured by parameters dMo, dd1, dEM and dEn. In particular:
dMo Defrost 1-2 management mode: (ind; 2-1; dEL) if dMo=ind defrost 1 e 2 managements are independent;
if dMo=2-1 activation of defrost 2 entails activation of defrost 1; if dMo=dEL activation of defrost 2 entails
activation of defrost 1 with delay set by parameter dd1;
dd1 Defrost 1 activation delay compared to defrost 2: (0 to 255 min) as previously mentioned, it describes
the activation delay of defrost 1 after activation of defrost 2 if dMo=dEL;
dEM Defrost synchronisation in LAN: (ind; Sin) if dEM=ind the end of defrosts is independent. If dEM=Sin,
the end of defrosts is synchronised;
dEn Synchronisation of local defrost end (def1 and def2): (ind; Sin) if dEM=ind the end of defrosts is
independent. If dEM=Sin, the end of defrosts is synchronised.
The start of a defrost cycle can be controlled locally (manual activation by the keyboard, digital input or at the
end of the time interval). Otherwise the command can arrive from any controller inserted in the LAN. In this last
case, the controller will start a defrost cycle and will bring it to the end following the indications present in its
programming parameters. But at the end of the cycle, it will wait for all the controllers present in the LAN to finish
before returning to normal temperature regulation.
3.3 CONTROL OF EVAPORATOR FANS
The fan control mode is selected by means of the FnC parameter:
FnC = C_n: fans will switch ON and OFF with the compressor and not run during defrost;
FnC = o_n: fanswill run even if the compressor isoff, and not run during defrost.
After defrost, there is a timed fan delay allowing for drip time, set by means of the Fnd parameter:
FnC = C_Y: fans will switch ON and OFF with the compressor and run during defrost;
FnC = o_Y: fans will run continuously also during defrost.
An additional parameter FSt provides the setting of temperature, detected by the evaporator probe, above which
the fans are always OFF. This is used to make sure circulation of air only if his temperature is lower than set in
FSt.
4. FRONT PANEL COMMANDS
To display and modify target set point; in programming mode it selects a parameter or confirm
an operation.
By holding it pressed for 3 sec when max or min temperature is displayed it will be erased.
To see the max stored temperature value.
In programming mode it browses the parameter codes or increases the displayed value.
By holding it pressed for 3s it gives access to the “Section” menu.
To see the min stored temperature value.
In programming mode it browses the parameter codes or decreases the displayed value.
By holding it pressed for 3 sec the defrost is started.
Switch ON and OFF the room light.
Keep it pressed more than 3 sec to switch ON and OFF the instrument.
KEY COMBINATIONS
+To enter the programming mode.
+To exit the programming mode.
4.1 USE OF LEDS
Each LED function is described in the following table.
LED MODE Function
ON Compressor enabled
Flashing Programming phase (flashing with the LED )
Anti-short cycle delay enabled
ON Fans enabled
Flashing Programming phase (flashing with the LED )
ON Defrost enabled
Flashing Defrost or drip time enabled
ON The controller is working in “ALL” mode
Flashing The controller is working in remote virtualdisplay mode
ON An alarm is active
4.2 HOW TO SEE AND MODIFY THE SET POINT
1. Push for about 3 sec the SET key: the display will show the Set point value;
2. The measurement unit starts blinking;
3. To change the SET value push the o ornarrows within 10 sec.
4. To store the new set point value push the SET key again or wait 10 sec.
4.3 HOW TO START A MANUAL DEFROST
1. Push the DEF key for more than 3 seconds and a manual defrost will start.
2. Use the UP and DOWN keys to select the dEF1 or the dEF2.
4.4 TO ENTER IN PARAMETERS LIST “PR1”
To enter the parameter list “Pr1” (user accessible parameters) operate as follows:
1. Enter the Programming mode by pressing the SET+DOWN key for few seconds
( and start blinking).
2. The instrument will show the first parameter present in “Pr1”.
4.5 TO ENTER IN PARAMETERS LIST “PR2”
To access parameters in “Pr2”:
1. Enter the “Pr1” level.
2. Select “Pr2” parameter and press the SET key.
3. The “PAS” flashing message will be displayed shortly, followed by “0 - -” with a flashing zero.
4. Use oor nto input the security code in the flashing digit; confirm the figure by pressing SET. The
security code is “321“.
5. If the security code is correct the access to “Pr2” is enabled by pressing SET on the last digit.
Another possibility is the following: after switching ON the instrument the user can push SET+DOWN keys within
30 sec.
NOTE: each parameter in “Pr2” can be removed or put into “Pr1” (user level) by pressing SET+DOWN. When a
parameter is present in “Pr1” LED is on.
4.6 HOW TO CHANGE THE PARAMETER VALUE
1. Enter the Programming mode.
2. Select the required parameter with o orn.
3. Press the SET key to display its value (measurement unit starts blinking).

1598013141 XM664K GB CLABO rev1.2 2012.01.26 XM664K 2/6
4. Useo ornto change its value.
5. Press SET to store the new value and move to the following parameter.
To exit: Press SET+UP or wait for 15 sec without pressing any key.
NOTE: the new programming is stored even when the procedure is exited by waiting the time-out.
4.7 ON/OFF FUNCTION
By pushing the ON/OFF key, the instrument shows “OFF”. During the OFF status, all the relays are
switched OFF and the regulations are stopped; if a monitoring system is connected, it does not
record the instrument data and alarms.
N.B. During the OFF status the Light and AUX buttons are active.
4.8 TO VIEW THE VALUE OF THE PROBES
1. Enterthe protected menu “Pr2”.
2. Using the keys oorn, select the parameter “dP1” for probe 1, “dP2” fortheevaporator probe and “dP3”
for the display probe.
3. Press the SET key to view the relative value.
4. Press the SET key to pass to the next parameter.
5. SECTIONS MENU
This menu allows to access the special function (characteristic of the XM series) relative to the LAN (Local Area
Network) of the controllers. Depending on how a keyboard is programmed, it can assume control of the section
it is connected to, control any of the other sections of the LAN or control all the sections.
To enter the sections menu, do as follows:
1. Press omore than 3 sec
2. The message corresponding to the current programming of the keyboard is displayed.
3. With keys oornselect the section you wish to control.
4. Press the SET key to confirm and exit programming.
The following messages appear on the display:
LOC: The keyboard shows the measured temperature values, the states of the outputs and the alarms of the
section to which it is connected. All the commands given by the keyboard will be carried out only by the local
section.
SE1…SE8: the keyboard controls the section of the LAN corresponding to the number (remote control) and
shows the values of the measured temperature, the states of the outputs and the alarms corresponding to that
section. All the commands given by the keyboard will be carried out by that section alone.
ALL: The keyboard shows the measured temperature values, the states of the outputs and the alarms of the
section to which it is connected, but the commands given by the keyboard are transferred to all the sections
making up the LAN.
In the event of an alarm, the label “Asn” will appear on the display, with “n” standing for the number of the
section in alarm. To view the type of alarm in detail, program the keyboard so that it takes control of the
corresponding section
6. HOW TO CHANGE THE ACTIVE MAP
This device is provided with 5 parameter maps. Passage from one parameter map to another takes place by
entering a password and selecting the index of the desired map. When the map is selected, the instrument
performs a reset to load the new functioning settings.
The procedure for changing maps is as follows:
1. enterin parameter programming and search fortheparameterMAP;
2. press SET;
3. enter the password “019”;
4. selectthe index of the map desired and pressSET once again;
5. The label “- - -” will appear on the display to indicate a reset of the instrument after changing map.
7. INSTALLATION AND ASSEMBLY
The XM664K must be installed inside the machine with two or more through screws and connected to the
keyboard by means of two 1mm cables. The allowed temperature range for correct functioning is between 0
and 60°C. Avoid places subject to heavy vibrations, corrosive gas, excessive filth or humidity. The same
instructions hold for the probes as well. Leave the area near the cooling louvers ventilated.
7.1 CUT OUT
The T640 keyboard shall be mounted on vertical panel, in a 150x31 mm hole, and fixed using two screws 3 x
2mm. To obtain an IP65 protection grade use the front panel rubber gasket (mod. RG-L).
165
150
31
+0.5
-0
+
0
.
5
-
0
+1
-1
Ø3 x2
7.2 MOUNTING WITH KEYBOARD COVER OPENING UPWARD
CLICK!
1
2
2
3
3
1
1
7.3 MOUNTING WITH KEYBOARD COVER OPENING DOWNWARD
CLICK!
1
2
2
2
3
3
1
1
8. ELECTRICAL CONNECTIONS
The XM664K power modules are provided with screw terminal block to connect cables with a cross section up
to 2.5 mm2. Relays and power supply have Fast-on type connection (6.3 mm). Heat-resistant cables have to be
used. Before connecting cables make sure the power supply complies with the instrument’s requirements.
Separate the probe cables from the power supply cables, from the outputs and the power connections. Do not
exceed the maximum current allowed on each relay, in case of heavier loads use a suitable external relay.
N.B. Maximum current allowed for all the loads is 20A.
8.1 PROBE CONNECTION
The probes shall be mounted with the bulb upwards to prevent damages due to casual liquid infiltration. It is
recommended to place the thermostat probe away from air streams to correctly measure the average room
temperature.Place the defrost termination probe among the evaporator fins in the coldest place, where most ice
is formed, far from heaters or fromthewarmest place during defrost, to prevent premature defrost termination.
9. TTL SERIAL LINE
The TTL connector allows, by means of the external module XJ485, to connect the unit to a network line
ModBUS-RTU compatible as the dIXEL monitoring system XWEB.
The same TTL connector is used to upload and download the parameter list of the “Hot Key “.
10. HOW TO USE THE HOT KEY
10.1 DOWNLOAD (FROM THE “HOT KEY” TO THE INSTRUMENT)
1. Turn OFF the instrument by means of the ON/OFF key, remove the TTL serial cable if present, insert the
“Hot Key” and then turn the Wing ON.
2. Automatically the parameter list of the “Hot Key” is downloaded into the Wing memory, the “DoL”
message is blinking. After 10 seconds the instrument will restart working with the new parameters.
3. Turn OFF the instrument and remove the “Hot Key”;plug the TTL serial cable (if was previously
present) and then turn it ON again.
At the end ofthe data transfer phase the instrument displays thefollowing messages:
“End”: for right programming. The instrument starts regularly with the new programming.
“Err”: forfailed programming.In this case turntheunit offand then on ifyouwant to restart the download
again or remove the “Hot key” to abort the operation.
10.2 UPLOAD (FROM THE INSTRUMENT TO THE “HOT KEY”)
1. Turn OFF the instrument by means of the ON/OFF key and remove the TTL serial cable if present; then
turn it ON again.
2. When the unit is ON, insert the “Hot key” and push okey; the “uPL” message will appear.
3. PushSET key to start the UPLOAD; the “uPL” message will start blinking.
4. Turn OFF the instrument and remove the “Hot Key”; plug the TTL serial cable (if was previously present)
and then turn it ON again.
At the end ofthe data transfer phase the instrument displays the following messages:
“End”: for right programming.
“Err”: for failed programming. In this case push SET key if you want to restart the programming again or
removethe not programmed “Hot key”.

1598013141 XM664K GB CLABO rev1.2 2012.01.26 XM664K 3/6
11. ALARMS
The alarm signals displayed on the keyboard can be divided into 2 groups:
1. Local alarms, including all the alarms relating to the section directly controlled by the keyboard.
2. Remote alarms, including all those alarms relating to the LAN.
The signal on the display remains until the alarm condition has ceased. All of the alarm signals flash, alternating
with the probe temperature, except for "P1" which is always flashing.
The "EE" alarm can be cancelled by pressing any key during the alarm signal. Then the message “rSt” is
displayed for about 3s before going back to normal functioning.
11.1 LOCAL ALARMS
Messagge Cause Outputs
PoF Keyboard locked
Pon Keyboard unlocked
rst Alarm muting
noP Probe not present
noL Absence of communication keyboard-main unit
P1 Room probe failure Compressor output acc. to par. Con
andCoF
P2 Evaporator probe failure Outputs unchanged
P3 Auxiliary probe failure Outputs unchanged
HA Maximum temperature alarm Outputs unchanged
LA Minimum temperature alarm Outputs unchanged
MSn Serious alarm because of reaching the max
number of pauses
rtF Real time clock board failure Alarm output ON; other outputs
unchanged
rtc Real time clock alarm Outputs unchanged
EA External alarm Outputs unchanged
CA Compressoralarm Regulation outputs deactivated
PAL Pressure sensor alarm
dA Door open alarm
StP Stop regulation
EE Eeprom failure
11.2 REMOTE ALARMS
Mess. Cause
“ASn” Section nis in alarm, with n= serialLAN address ("LAN"parameter). this alarmis displayed onlyif
thekeyboard isin “ALL” mode. Thisis a genericindication ofthestate of alarm. To have more
detailed information, set the keyboard so that it controls that specific section.
“nLn” No link alarm (no communication) with section n (“Lan”parameter)
“nLn” No link alarm (no communication) while a remote probe is viewed. This alarm appears only if the
LdS parameter of the section is set at "y".
“rPE” Remote viewer error. More than one keyboard has been set for remote control
11.3 BUZZER SILENCING AND ALARM EXIT
Once the alarm signal has been detected, the buzzer and alarm exit can be deactivated by pressing any key.
However the signal on the display remains until the alarm condition has ceased. It is possible to inhibit the alarm
relay deactivation by setting the parameter tbA=n. In this way the alarm relay remains active as long as the
alarm condition lasts.
The signal buzzer is found in the keyboardT640.
11.4 “EE” ALARM
The instruments of the Dixell series are equipped with an internal control which checks the integrity of the data.
The flashing “EE” alarm which alternates with the temperature signals the presence of a data anomaly. .
11.5 MODE FOR CEASING ALARMS
The probe alarms “P1”, “P2” and “P3” are triggered about 10 sec after the probe failure. They automatically
cease 10 sec after the probe goes back to regular functioning. Before replacing a probe, it is recommended to
check the connections.
The temperature probes “HA” and “LA” automatically cease as soon as the thermostat temperature goes back
to normal, when defrost starts or when the door opens.
The No link alarms “nLn”and “nLn”cease automatically as soon as correct communication is re-established
between the sections
The digital input alarms “EAL” and “BAL” cease automatically when the input is deactivated.
12. TECHNICAL DATA
Keyboard
Housing: self extinguishing ABS.
Case: facia 38x185 mm; depth 23mm.
Mounting: panel mounting in a 150x31 mm panel cut-out with two screws. 3 x 2mm.
Distance between the holes 165mm.
Frontal protection: IP65 with frontal gasket mod RG-L (optional).
Connections: Screw terminal block 2.5 mm2heat-resistant wiring.
Power supply: from XM664K power module.
Display: 3 digits, red LED, 14.2 mm high.
Optional output: buzzer.
Mounting Din
Housing self extinguishing ABS
Dimensions 8 Din
Connections Screw 1,6mm2and FASTON 6,3mm
PowerSupply 230Vac 10%
Powerabsorption 9VA max.
Display NO
Temperature inputs 3 sensors: NTC (10KΩ at 25°C); PTC (806Ω at 0°C)
-40 to +150°C; Pt1000 sensor
Digital inputs Two free voltage digital inputs
Relay outputs SPST 20 (8) A
Relay Defrost 1 SPST 8 (3) A
Relay Defrost 2 SPST 8 (3) A
Relay Fans SPST 8 (3) A
Relay Light SPDT 16 (5) A
Relay ON/OFF SPDT 16 (5) A
Data storing On the non-volatile memory (EEPROM)
Kind of action 1B
Pollution degree Normal
Softwareclass A
Operating temperature 0 to 60°C
Storage temperature -25to 60°C
Relative humidity 20 to 85% (not condensing)
Measuring and regulation range NTC probe: -40 to 110°C
PTC probe: -50 to 150°C
Resolution 0.1°C
Accuracy ±0.5°C ±1 digit
13. CONNECTIONS

1598013141 XM664K GB CLABO rev1.2 2012.01.26 XM664K 4/6
14. DEFAULT SETTING VALUES
FACTORY DEFAULT PARAMETER TABLE
The indication for just one of the 5 available parameter maps is carried.
(*) Parameters visible on Pr1 level.
LABEL VALUE DESCRIPTION RANGE
SET -20 Regulation set point LS to US
MAP -- - Parameter map switch 1 to 5
AtA - - - Alternative parameter map access - - -
REGULATION
Hy 2,0 HY is the ON/OFF heat regulation differential [0.1to 25.5°C]
LS -30 Minimum Set-Point that can be set by keyboard [-55.0°C to SET]
US 10 Maximum Set-Point that can be set by keyboard [SET to 150.0°C]
odS 0 Delayin minutes until switch-on before allregulation outputs are activated. 0 to 255 min
AC 1 Anti-short cycle delay of relay cold regulation output 0 to 60 min
CCt 0 Duration of continuous cycle 0.0 to 24h00min, res. 10 min
CCS 0 Continuous cycle set point [-55.0 to 150°C]
Con 15 Time compressor ON with probe failure. 0 to 255 min
CoF 30 Time compressor OFF with probe failure. 0 to 255 min
CH cL Type of regulation action CL; Ht
CF °C Temperature measurement unit: °C or °F °C; °F
rES In Display resolution dE; in
Lod P1 Display probe: from P1 to P3. tEr= not used dEF=not used. If Err appears on the display, the probe is in error, or nP=not
present in parameter P1C..P3C. nP; P1; P2; P3; tEr; dEF
dLy 0.0 Delay in minutes/ seconds of the display of temperature rise to limit effects of too quick variations of a temperature sensor. 0.0 to 24h00min, res. 10 min
DEFROST
dPA P2 Selection of 1st defrost probe input: from P1 to P3, nP=not present. nP; P1; P2; P3
dPb nP Selection of 2nd defrost probeinput:from P1to P3, nP=not present. nP; P1;P2; P3
tdF In Defrosting mode EL –in - OFF
tbO 0 Time duration of black-out/stand-by reset of intervals between defrosts (0=instant reset). If tb0 is different than0, use 0 to 255 min
dtE 8 Probe dPA defrost 1 end temperature [-55.0°C ÷ 50.0°C]
dtS 8 Probe dPb defrost 2 end temperature [-55.0°C ÷ 50.0°C]
idF 8 Defrost 1 interval if EdF=in (intervals) If idF=0 defrost is forced manually only, from serial, from outside contact, from
coordinated defrost LAN. 0 to 120 hours
IdS 0 Defrost 2 interval if EdF=in (intervals) If idS=0 defrost is forced manually only, from serial, from outside contact, from
coordinated defrost LAN. 0 to 120 hours
MdF 8 Maximum duration forcing defrost 1 endif it is notcut-off before the dTE temperature. 0 to 255 min
MdS 0 Maximum duration forcing defrost 2 endif it is not cut-off before the dTS temperature. 0to 255 min
dMo Ind Defrost management mode (independent, 2 force 1,1 delayed compared toactivationof 2) Ind; 2-1; dEL
dd1 0 Defrost 1 activation delay compared tostart of 2 0 to 255 min
dEM Ind Defrost end synchronisation in LAN (independent – synchronised) ind; Sin
dEn Ind Local 1-2 defrost endsynchronisation (independent– synchronised) ind;Sin
dSd 0 Defrost delay from call 0 to 255 min
dFd rt Display during defrost rt; it; Set; dEF
dAd 60 Temperature display delay after defrost blocking display. 0 to 255 min
Fdt 0 Post-defrost dripping time keeping the ventilation outputs and heat regulation stopped. 0 to 255 min
dPo n Defrost at switch-on (yes /no) n; Y
dAF 0 Defrost delay after continuous cycle 0.0 to 24h00min, res. 10 min
FAN REGULATION
FPA P2 Fan probe control input selection: from P1 to P3, nP=not present. nP; P1; P2; P3
FnC O-n Fans functioning mode. C= active only with cold output ON, O= always active, n= stopped for defrost, Y= active for defrost. C-n; O-n; C-Y; O-Y
Fnd 1 Fan switch-on delay after defrost 0 to 255 min
FSt 10 Fan block temperature on FPA fan probe [-55.0 to 50.0°C]
FHy 1 Return differential from fans stop FSt set [0.1 to 25.5°C]
Foo 10 Timefans on afterinstrumentOFF 0 to 255 min
Fon 15 Time fans ON with compressor off 0 to 15 min
FoF 15 Time fans OFF with compressor off 0 to 15 min
TEMPERATURE ALARMS
rAL P1 Probe selection for HA and LA temperature alarm nP; P1; P2; P3

1598013141 XM664K GB CLABO rev1.2 2012.01.26 XM664K 5/6
ALC Ab Alarms: rE=relating to temperature set point [SET+ALU] and [SET-ALL];
Ab= absolute value ALUorALL within scale limits ofselectedsensor rE; Ab
ALU 60 High temperature alarm threshold generating HA message after Ald minutes. [0.0 to 50.0°C or ALL to 150.0°]
ALL -50 Low temperature alarm threshold generating LA message after Ald minutes. [0.0 to 50.0°C or -55.0°C to ALU]
AHy 2 Temperature alarms differential [0.1 to 25.5°C]
ALd 15 HA and LAtemperature alarm signal delay 0 to 255 min
dAo 1.3 Alarm signal delay at hour switch-on.10 minutes. 0.0 to 24h00min, res. 10 min
EdA 30 Exclusion of temperature alarm after defrost in minutes. 0 to 255 min
dot 30 Duration of exclusion oftemperature alarmafter door closing. 0 to 255 min
tbA n Silencing of alarm relay by pressing any key on the keyboard. n; Y
CONFIGURATION OF RELAY OUTPUTS: Cpr=cold output, dEF=defrost, dF2=defrost2 FAn=fans, ALr=alarm, LiG=light,
AUS=auxiliary, OnF=ON/OFF
oA2 dEF Configuration of 2nd relay function CPr; dEF; dF2; FAn; ALr; LiG; AUS; OnF
oA5 dF2 Configuration of 5th relay function CPr; dEF; dF2; FAn; ALr; LiG; AUS; OnF
AOP oP Relay output polarity configured as alarm. OP= alarm with open contact. OP; CL
iAU n Auxiliary relay output independent from ON/OFF status of board. n; Y
DIGITAL INPUTS
Key: EAL=external alarm, Bal=regulation block alarm, PAL=do not use, dor=door switch, dEF=external defrost contact, AUS=
auxiliary relay, LiG=light relay, OnF= On/OFF board, ES=energy saving, AtA=alternative map activation CbL= compressor
block
i1P CL Input polarity DI1 CL=active for closed contact OP; CL
i1F dor Digital input function configuration DI1 EAL; bAL; PAL; dor; dEF; AUS; LiG; OnF; ES;
ATA; CbL
d1d 0 Alarm delay for configurable digital inputDI1 0 to 255 min
i2P CL Input polarity DI2 CL=active for closed contact OP; CL
i2F AtA Digital input function configuration DI2 EAL; bAL; PAL; dor; dEF; AUS; LiG; OnF; ES;
ATA; CbL
d2d 0 Alarm delay for configurable digital inputDI2 0 to 255 min
nPS 10 Number of interventions at status before block 0 to 15
OdC FAn Loads control with door open: Fan=fans, Cpr=compressor, F-C fans and compressor. Valid if one of the parameters i1,2F= dor no; FAn; CPr; F-C
rrd 30 Forcing of fan compressor restart after door opening Valid if one of the parameters i1,2F= dor 0 to 255 min
KEYBOARD CONFIGURATION
bbC 6bb 6-key or 8-key keyboard configuration 6bb; 8bb
LAN MANAGEMENT
rdA y Remote function management activation (RVD) n; Y
SEb y Sections that can be selected from keyboard n;Y
LSn - - - Number of devices detected in LAN 1 to 8
LAn - - - Serial addresses of devices detected in LAN 1 to 247
L1d y Defrost 1 synchronisation in LAN n; Y
L2d y Defrost 2 synchronisation in LAN n; Y
LSP n Y= forcing of same set point for all boards, n= independent set point. n; Y
LdS n Y= Synchronise board display n=independent display. n;Y
LOF n Y= Synchronise ON/OFF n= ON/OFF independent n; Y
LLi y Y= Synchronise Light n= Light independent n; Y
LAU n Y= Synchronise AUX n= AUX independent n; Y
LES n Y= Synchronise Energy Saving n= Energy Saving independent n; Y
LSd n Y= Remote probe display n= no remote probe display n; Y
INPUT CONFIGURATION. NTC (10KΩ at 25°C), Ptc (806Ω at 0°C)
P1C Ntc Probe P1 Configuration nP= not present nP; PtC; ntC
Ot 0 Probe P1 calibration [-12.0 to 12.0°C]
P2C Ntc Probe P2 Configuration nP= not present nP; PtC; ntC
oE 0 Probe P2 calibration [-12.0 to 12.0°C]
P3C nP Probe P3 Configuration nP= not present nP - Ptc - ntc
O3 0 Probe P3 calibration [-12.0 to 12.0°C]
SERVICE
Adr 1 Cold output opening ON/OFF percentage 1 to 247
OnF OFF ON/OFF key function configuration (not used – ON/OFF – Energy saving) nu; OFF; ES
LPC LiG Light key function configuration (not used– Light – auxiliary – FHU not used) nu; LiG; AUS; FHU
dPC dEF Defrost key function configuration (defrost – auxiliary) dEF; AUS
SERVICE DISPLAY
rEL 1.1 Firmware release code (read only) Read only
Ptb EEPROM map identification Read only
Pr2 - -- PR2 MENU input (protected parameters) (Protected parameters input)
dP1 -- - Probe 1display Read only
dP2 -- - Probe 2display Read only

1598013141 XM664K GB CLABO rev1.2 2012.01.26 XM664K 6/6
dP3 -- - Probe 3display Read only
L°t - - - Minimum temperature display Read only
H°t - - - Maximum temperature display Read only
ALTERNATIVE PARAMETER MAP
SET 4 Set point LS to US
Hy* 3 HY is the ON/OFF heat regulation differential [0.1to25.5°C]
LS* 0 Minimum Set-Point that can be set by keyboard [-55.0 to SET]
US* 20 Maximum Set-Point that can be set by keyboard [SET to 150.0°C]
tdF* OFF Defrosting mode EL; in; OFF
dtE* 8 Probe dPA defrost 1 end temperature [-55.0 to 50.0°C]
dtS* 8 Probe dPb defrost 2 end temperature [-55.0 to 50.0°C]
idF* 8 Defrost 1 interval if EdF=in (intervals) If idF=0 defrost is forced manually only, from serial, from outside contact, from
coordinated defrost LAN. 0 to 120 hours
IdS* 0 Defrost 2 interval if EdF=in (intervals) If idS=0 defrost is forced manually only, from serial, from outside contact, from
coordinated defrost LAN. 0 to 120 hours
MdF* 30 Maximum duration forcing defrost 1 end if itis not cut-off before the dTE temperature. 0to 255 min
MdS* 0 Maximum duration forcing defrost 2 endif itis not cut-off before the dTS temperature. 0to 255 min
dFd* rt Display during defrost rt; it; Set, dEF
FnC* O-y Fans functioning mode. C= active only with cold output ON, O= always active, n= stopped for defrost, Y= active for defrost. C-n; O-n; C-Y; O-Y
Fnd* 0 Fan switch-on delay after defrost 0 to 255 min
FSt* 10 Fan block temperature on fan probe [-55.0 to 50.0°C]
Ot* 0 Probe P1 calibration [-12.0 to 12.0°C]
P1C* Ntc Probe P1 Configuration nP= not present nP; PtC; ntC
P2C* nP Probe P2 Configuration nP= not present nP; PtC; ntC
P3C* nP Probe P3 Configuration nP= not present nP; PtC; ntC
Table of contents
Other Emerson Control Panel manuals

Emerson
Emerson Fisher LCP100 User manual

Emerson
Emerson Comverge IntelliTEMP 900 Assembly instructions

Emerson
Emerson Appleton ALPN User manual

Emerson
Emerson Vission Use and maintenance manual

Emerson
Emerson SW102 User manual

Emerson
Emerson NELSON HEAT TRACE SMMC-3 User manual

Emerson
Emerson Fisher LCP100 User manual

Emerson
Emerson E2 User manual