Table 2-1: Troubleshooting
(continued)
Error Recommended action(s)
Communication issues due to
blocked network ports
•Blocked network ports on the computer running
MeterLink or on a company LAN can prevent
connections to the meter or prevent certain features
from working. These issues may occur over Ethernet,
Modem and Direct serial connections. Reference the
list of network ports used by MeterLink in the Help file
and the symptoms of having blocked ports. Contact
your IT department for assistance in resolving these
issues.
•Error condition of a blocked network:
—Cannot connect to a meter
—Cannot collect Archive log files
—Cannot view or stream waveforms in Waveform
Viewer or Signal Analyzer
—Cannot upgrade firmware
—Communications lost over serial or modem
connections while MeterLink is idle on a screen
•Symptoms of blocked network:
—If a PING is blocked on this network port,
serial or modem connections could be lost after
approximately 15 seconds of inactivity. This issue
can be confirmed by checking the log_meter log
file in the Temp data folder. The path of the Temp
data folder is shown in the MeterLink About dialog.
—A blocked FTP port will generally not prevent
a connection to the meter, but will prevent log
collections and program downloads. A blocked
FTP port could prevent a connection in the
event the meter is running a newer version of
firmware for which MeterLink does not currently
have a database configuration file. If this is a
case, a message stating “Error reading database
config file dbconfig<databaseversion>.xml from
the meter.” will be displayed.
—A blocked DB API port will report “Error 10001
opening database connection to <IP address>”.
—A blocked Streaming port will report an error
message “Unable to open a control socket”. This
will occur when opening the Signal Analyzer
window or clicking Read or Stream to File in the
Waveform Viewer.
Communicating with meter
but all chords display failures
•Verify that the resistance of transducers is within
specification (2 Ω).
•Check the Acquisition Board.
•Check the interconnect cables between the Base
Enclosure and the Transmitter Electronics Enclosure.