EMIT 50557 Installation guide

TB-6520 March 2010 Page 1 of 7
Surface Resistance Test Kit
Operation and Maintenance
Figure 1. Item 50557 Surface Resistance Test Kit
Description
The Surface Resistance Test Kit is a portable, accurate,
and versatile battery operated instrument designed
to measure resistance between two points (RTT) and
resistance to ground (RTG) in accordance with ANSI/ESD
S4.1 including:
• Resistancemeasuringaccuracy±10%
(E11&greater±20%)
• Resistancerange<1.0x10E3ohmsto>10E12ohms
• Opencircuitvoltagesof10and100volts±5%
• Electrificationperiodof15seconds
• Electrodes(two)5pounds±2ozwith50-70durometer
conductive pads
In addition, the Meter measures ambient temperature (in
degrees Fahrenheit and Celsius) and relative humidity.
Voltage set to “Auto”, will measure at 10 volt automatically
switchingto100voltsifmaterialbeingtestedis>10E5
ohms. No switching test voltage is necessary.
The Surface Resistance Test Kit is referenced and
designed to be used to make measurements in accordance
with the test methods in:
• Worksurfaces-ANSI/ESDS4.1Worksurfaces
• ComplianceVerification-ESDTR53-Resistance
Measurements
• Floors-ANSI/ESDS7.1-Resistive
CharacterizationofMaterialsFloor
Materials
•Footwear-ANSI/ESDS9.1-Footwear-Resistiv
Characterization
• Garments-ANSI/ESDSTM2.1Garments
• Seating-ANSI/ESDSTM12.1-Seating-Resistive
Measurement
• Floor/Footwear-ANSI/ESDSTM97.1-FloorMaterials
andFootwear-ResistanceMeasurementinCombination
with a Person
• Workstations-ESD-ADV53.1-ESDProtective
Workstations
Versatility:
Measures RTT, RTG or Resistivity (requires optional
Resistivity Attachment, item 50560)
•Quick Checks -Surfaceresistanceexponentnumber
illuminates immediately i.e. 8 = 10E8 ohms or
100,000,000 ohms
• Compliance Verifications of Installed Products
-Surfaceresistanceexponentnumberilluminates
immediately and displays Temperature and Relative
Humidityduring15secondelectrificationperiodmaking
numerous measurements and calculations to then display
mantissa i.e. if LED displays 8 and LCD displaying 7.14
asmantissa=7.14x10E8ohmsor714,000,000ohms
• Test Lab Evaluation of Product-Displaysameas
Periodic Check above
The Surface Resistance Test Kit is NIST calibrated and
items available are:
Made in America
©2010DESCOINDUSTRIESINC.
EmployeeOwned
Model Description Power
50557 SurfaceResistanceTestKit 2xAABatteries
50560 Resistivity Attachment N/A
Inspection
Remove the Meter from the carton and inspect for damage.
Each Surface Resistance Test Kit should include the
following:
1 Protective carrying case
1 Meter
2 Test leads
2 5-poundelectrodes
2 AA alkaline batteries
1 Certificate of Calibration
Properly store the Meter and component assemblies when
not in use.
TECHNICAL BULLETIN TB-6520
EMIT-3651WalnutAvenue,Chino,CA91710•(909)664-9980•Fax(909)627-7449•Website:DescoEMIT.com

TB-6520 Page 2 of 7
ESD Control items should be tested:
A. Prior to installation to qualify for listing in user’s ESD
control plan approved ESD control items
B. Duringinitialinstallation
C. For periodic checks of installed products as part of ESD
control plan compliance verification plan
“A Compliance Verification Plan shall be established
toensuretheOrganization’sfulfillmentofthetechnical
requirements of the ESD Control Program Plan. Process
monitoring (measurements) shall be conducted in
accordance with a Compliance Verification Plan that
identifies the technical requirements to be verified, the
measurement limits and the frequency at which those
verifications shall occur. The Compliance Verification Plan
shall document the test methods and equipment used
for process monitoring and measurements. If the test
methodsusedbytheOrganizationdifferfromanyofthe
standards referenced in this document, then there must
be a tailoring statement that is documented as part of
the ESD Control Program Plan. Compliance verification
records shall be established and maintained to provide
evidence of conformity to the technical requirements. The
test equipment selected shall be capable of making the
measurements defined in the Compliance Verification Plan.”
(ANSI/ESDS20.20-2007section7.3)
Electrification Period
The Surface Resistance Test Kit provides the proper
electrificationperiodof15secondsperANSI/ESDS4.1,
afternumerousreadingsandcalculationsareexecuted,
then displays surface resistance mantissa measurement
(Note: most analog type meters display measurements
instantaneously).
WhentheTestButtonisdepressed,theliquidcrystal
display (LCD) will indicate:
• TemperatureindegreesFahrenheit(tolerance±5°F,
typical)
• TemperatureindegreesCelsius(tolerance±3°C,typical)
• Humidityaspercentage(from5%-95%tolerance,±10
typical)
• Surfaceresistancemantissa(withexponentdisplayedvia
illuminated LED, measurement in ohms)
Measurement being displayed is identified by the illuminated
functionLED.Thesurfaceresistanceexponent(orpowerof
number) is immediately illuminated and remains illuminated
measuring the range the surface resistance in ohms.
Reference Literature
In addition to those noted above:
ANSI/ESDS20.20-DevelopmentofESDControlProgram
ESDADV1.0-GlossaryofTerms
ANSI/ESDS6.1Grounding
These documents can be obtained directly from the
ESDAssociation,7902TurinRd.,Suite4,Rome,NY
13440-2069,(315)339-6937orwww.esda.org.
Otherstandardsareavailablefromtheagencieswho
producethem.MIL-HDBK263A,EIA-IS-5-A,ASTM-F-150,
EN100015,andJESD625-A.
Features
©2010DESCOINDUSTRIESINC.
EmployeeOwned
A. LED Displays:Surfaceresistanceexponentisdisplayed
via light emitting diodes illuminating 1/8” high number.
Thereare12LEDsdisplayingsurfaceresistanceexponent
measurement. They are color coded for quick checks:
Exponent Color
<3,3 Red
4,5 Green
6,7,8 Blue
9,10,11 Green
12,>12 Yellow
Five Function LED’s identify the measurement being
displayed(seeBbelow).
• Whenbatteryvoltagedropstoapproximately2volts,one
of the Function LED’s to the right of the red pushbutton
switch will begin to flash, indicating the need to replace
batteries.
B. Liquid Crystal Display (LCD): Mantissa is displayed via
9/16"highliquidcrystaldisplayandprovideseasytoread
resistance (or resistivity) measurements.
WhentheTestButtonisdepressed,theLCDwillindicate:
• TemperatureindegreesFahrenheit
• TemperatureindegreesCelsius
• Humidityaspercentage
• Surfaceresistancemantissa(withexponentdisplayedvia
LED, measurement in ohms)
Figure 2. Features of the Surface Resistance Meter
EMIT-3651WalnutAvenue,Chino,CA91710•(909)664-9980•Fax(909)627-7449•Website:DescoEMIT.com
A
B
C
D
E
F
G

TB-6520Page3of7
Surfaceresistanceohmvaluesareexpressedwitha
mantissaandexponentorpowerofthenumber.For
example,if“8”isilluminatedbyitsLEDandtheLCD
displaysamantissaof“7.14”,themeasurementis7.14x
10E8 ohms or 714,000,000 ohms
If surface is over 10E12 ohms, the following will appear:
• “1____”ontheLCDdisplayindicateOverrangeorthatthe
readingexceedsthedisplaycapabilities
C. Test Button: This red button activates electrical power
totheMeter.Theexponentorpowerofthenumberis
displayed immediately with LED illuminating number. If “8”
is illuminated, the measurement is in the 108or 10E8 ohm
range. To make a measurement in accordance with
ANSI/ESDS4.1,thebuttonistobedepressedfor10-20
seconds for the unit to make numerous readings and
calculations. (Testing in accordance with ANSI/ESD S4.1
requires15secondsofelectrification;incontrast,most
analog type meters display measurements instantaneously.)
The LCD will display temperature (F), then temperature
(C), then relative humidity, and then the surface resistance
mantissa.Forexample,iftheLCDdisplaysamantissaof
“7.14”, and “8” is illuminates by its LED, the measurement is
7.14x10E8 ohms or 714,000,000 ohms.
Duringtheentireperiod,surfaceresistanceexponentwillbe
displayed by LED illuminating number.
D. Override Test Voltage Button:Wheninthe“up”
position, during resistance portion of test, Meter will
automatically switch to the correct voltage for the resistance
range. LED will illuminate noting selected voltage.
Material10E5 ohms or less should be measured at 10
volts.Material10E6ohmsorgreatershouldbetested
at 100 volts. The button is a switch, which if depressed
will override automatic voltage selection and test will be
performed at 10 volts regardless of resistance level.
E. Jacks:OneendofTestLeadshas3.5mmplug(fits
Meterleftjack).The3.5mmplugisshielded(identified
byinsulatedtipandbarrelblacklead)-seeRTGtest
procedure: Per ANSI/ESD S4.1 “The sensing lead of the
resistance meter shall be connected to groundable point.”
OneendofTestLeadhasstandardbananaplug(fits5#
Electrode jack).
F. AA Battery Compartment.
G. Resistivity Parallel Electrodes Attachment.
(optional Item number 50560)
Cleaning
Per ANSI/ESD S4.1 “Clean the electrodes with a minimum
70%isopropanol-watersolution.”Makesureconductive
pads are dry prior to use.
See specific product test standard for test lab specimen
cleaninginstructions.PerANSI/ESDS4.1Worksurfaces
“The test specimens and electrodes shall be cleaned twice
withaminimum70%isopropanol-watersolutionusinga
clean,low-lintingclotheachtime.”(Note:thenconditioned
for 72 hours, minimum).
For installed product periodic testing, do not clean surfaces.
However, if any measurements lie outside acceptable
range,thencleanthesurfaceandre-test(Note:for
worksurfaces, use an ESD cleaner not containing silicone).
Besurethesurfaceisdrybeforetesting.
Periodic maintenance-Theareasurroundingthecable
jacks at the top end of the meter should be wiped with a
clean cloth moistened with alcohol to remove skin oils that
will accumulate and affect the accuracy at high resistances.
Thefrequencyofcleaningwilldependonusage;oncea
monthwouldbeagoodstartingpoint.Otheritemsthat
should also be cleaned in this fashion are the cable jackets
and the resistivity attachment, if included.
Power Requirements
The Meter is powered by two replaceable alkaline AA
batteries.
Battery Replacement
Depress buttons on both sides of bottom end cap of meter
housing. Remove cap, press on right edge of protruding
battery door, unlatching it, and swing door open. Carefully
replace batteries with alkaline Type AA. Polarity must be
correct or damage may occur. Close battery door and
re-installbottomendcaponmeterhousing.
Test Procedure
General Guidelines:
• Useboth5PoundElectrodesforRTT
• Useone5PoundElectrodeandblacksensingleadto
equipment ground to groundable point for RTG
• UseoptionalResistivityAttachment(removingleads&
electrodes) for Resistivity measurements
• Ensurethatitembeingmeasurediselectricallyisolated
(i.e. placed on an insulative surface) as Meter will
measure lowest resistance path
• EnsurethattestleadsareseparatedorMetermay
measure lower resistance path
• Whenusing5PoundElectrodes:
• Placenocloserthan2"fromedgeofsurfacebeing
measured.
• Placenocloserthan3"toanygroundablepoint
• Place5PoundElectrodesabout10"apartforRTTof
worksurfacesandabout3'apartforRTTofflooring.
• Preferredplacementsinclude:mostcommonlyused
surface portion, most worn, center, and furthest from
groundable point.
• Ifsurfacehassections(likefloortilesorgarmentpanels),
forRTTplacea5poundelectrodesondifferentsections
• CleansurfaceforTestLab,butdonotinitiallyclean
surface for installed products (if fails, clean and retest).
• OnlyuseanapprovedESDcleanerandallowtodryprior
to testing.
• LayMeteronsurface;holdinginhandcancause
measurent distortions.
©2010DESCOINDUSTRIESINC.
EmployeeOwned
EMIT-3651WalnutAvenue,Chino,CA91710•(909)664-9980•Fax(909)627-7449•Website:DescoEMIT.com

TB-6520 Page 4 of 7
Test Lab Test Procedure Guideline
FortestlabuseofESDWorksurfaces,FloorMaterials,
Footwear, Garments, or Seating, best advice is to follow
procedures in applicable ESD Association standards (see
above) which include details regarding:
• Cleaning(ForexampleperS4.1“Thetestspecimensand
electrodesshallbecleanedtwicewithaminimum70%
isopropanol-watersolutionusingaclean,low-lintingcloth
each time). Allow to dry.
• Environmentalchamber(ForexampleperS4.1control
relativehumidityto12±3%RHand50±5%RHand
temperatureto23±1degreesC)
• Specimensupportsurface(ForexampleperS4.1
greaterthan1.0x10E12ohmssuchasPMMA,PTFEor
polycarbonate)
• SpecimenPre-Conditioning(ForexampleperS4.1at23
±1degreesC;3specimensat12±3%relativehumidity
for72hoursminimum,and3specimensat50±5%
relative humidity for 72 hours minimum)
• ReportingTestResults,forexampleperS4.1Reporting
Test Results, Report:
• Minimum,medianandmaximumreadingsforboth
resistance-to-groundablepointandpoint-to-point
resistance in ohms at low relative humidity
• Minimum,medianandmaximumreadingsforboth
resistance-to-groundablepointandpoint-to-point
resistance in ohms at moderate relative humidity
• Temperature
• Relativehumidity
• Actualdurationofconditioning
• Testequipmentused
Periodic Checks of Installed Product Test
Procedure Guideline
The ESD Association lists test procedures and
troubleshootingtipsinComplianceVerificationESDTR53.
Werecommendthisdocumentbepurchasedfrom
www.ESDA.org
Measure RTG Resistance to Ground
Test Procedure in reference with ANSI/ESD S4.1 section
6.4PeriodicWorksurfaceTesting:
• Donotcleanthesurface
• Removefromthesurfaceonlythoseitemsthatmight
interfere with the test.
• ESDsensitivedevicesshallalsoberemoved
• Theblacksensingtestleadwithshieldedplugshouldgo
to equipment ground or a groundable point
• UseoneElectrodeonothertestleadandplaceElectrode
the farthest convenient point on the surface
• PressbuttonandholdTestButtonuntilmeasurementis
displayed
• PerformadditionalmeasurementsplacingElectrodeon
the most commonly used or most worn area
If measurement is outside acceptable limits, clean surface
and retest to determine if cause of failure is insulative dirt
layerortheESDcontolitem.Note:UseanESDcleaner
containing no insulative silicone.
Measure RTT Resistance Between Two Points
on the Surface
• Donotcleanthesurface
• Removefromthesurfaceonlythoseitemsthatmight
interfere with the test.
• ESDsensitivedevicesshallalsoberemoved
• Usetwo5PoundElectrodes,placeinthemostcommonly
usedportionofthesurface10"apart(2"fromanyedge,
3"fromanygroundablepoint)forworksurfaceand3'for
flooring
• PressTestbuttonandholdButtonuntilmeasurementis
displayed
• Ifthemostusedportionisnotobvious,usetwopoints
near the center of the surface
If measurement is outside acceptable limits, clean surface
and retest to determine if cause of failure is insulative dirt
layerortheESDcontolitem.Note:UseanESDcleaner
containing no insulative silicone.
©2010DESCOINDUSTRIESINC.
EmployeeOwned
Figure 3. Setting up for RTG testing.
10”
Figure 4. Setting up for RTT testing. (Electrodes are to be
about 10" apart for worksurfaces and about 3' for flooring)
Reporting and Using Test Results
Differentstandardshavedifferentrequirements;use
requirements as specified in user’s ESD control plan.
Examplesare:
Per ANSI/ESD 4.1 (worksurfaces), Report:
• RTGmaximumandminimumvaluesmeasuredfor
resistance-to-ground
• RTTmaximumandminimumvaluesmeasuredfor
point-to-pointresistanceinohms
EMIT-3651WalnutAvenue,Chino,CA91710•(909)664-9980•Fax(909)627-7449•Website:DescoEMIT.com

TB-6520Page5of7
Per ANSI/ESD 7.1 (Floor) Report:
• RTGallvaluesinohmsforresistancetoground
• RTTallvaluesinohmsforpoint-to-pointresistance
• Voltagelevel
• Testdate
• Temperature
• Relativehumidity
• Testequipmentused
Summarizetestdatabyreportingminimum,maximum,
median, and mean values obtained. Include a diagram
showingapproximateelectrodepositionsandground
connections used.
Recommended Frequency of Compliance
Verification Periodic Checks of ESD Control
Items
PerESDTR53ComplianceVerificationofESDProtective
Equipment and Materials ANNEX A “Each user will need
to develop their own set of test frequencies based on the
critical nature of those ESD sensitive items handled and
the risk of failure for the ESD protective equipment and
materials.
Note: “The frequency of periodic testing is normally
specified in corporate operating procedures. …The
frequency of testing is driven by the amount of risk
exposurethatcanoccurbetweentests.For,example,what
is the quantity of product handled between test periods?”
(See ESD Handbook ESD TR20.20) However, a common
guide would be:
• Worksurface-atleastquarterly(seeESDTR20.20
section5.3.1.13Periodicverificationtesting)
• Footwear-“Incominginspectiononalotsamplingbasis
should be performed for all static control footwear.” (see
TR20.20section5.3.3.4Testing)
• Floor -(SeeESDTR20.20section5.3.4.5Performance
Monitoring) Test methods, standards, and specifications.
“Typically, once qualified, floors may use simple
resistance checks to ensure the flooring does not change.
The frequency of testing is also a consideration. Some
materials, such as floor finishes, may require more
frequent testing because of their lack of permanency.”
[ESDTR20.20section5.3.4.15TestMethods,
Standardsm and Specifications
• Seating-“Therecommendedelectricalresistancerange
forseatingislessthan1x10E9ohmsastestedin
accordance with ANSI/ESD STM2.1. This value should
be during acceptance testing, installation and periodically
thereafter.”(seeESDTR20.20section5.3.5.3Testing)
• Garments-“Tomaintainprocesscontrol,itisimperative
that the garment be tested per ANSI/ESD STM 2.1. The
point to point and sleeve to sleeve resistance test should
bemade.”(seeESDTR20.20section5.3.13.3.1.8
Periodic Testing)
Surface Resistivity
Theoretically Resistivity is 10 times greater than Resistance,
i.e. a material that measures 10E7 ohms RTT, Surface
Resistivity should measure 10E8 ohms/square.*
Ref: ANSI/ESD S11.11 section 12.0 Conversion to
Resistivitystates,“Whenitisappropriatetoconverta
resistance obtained by this test method to an equivalent
resistivity in ohms per square, multiply the resistance
measurements obtained by this method by 10. The
conversion factor of 10 is derived from the geometry of the
electrode assembly.”
No conversion is required with the Surface Resistance Test
Kit. The Meter incorporates provision to install optional
Resistivity Attachment which has parallel electrodes
on bottom of the unit to make direct surface resistivity
measurements. These parallel electrodes allow for
measurement of resistivity or when quick testing without the
useof5poundelectrodesisdesired.
*Note: Per ESD Association the proper unit of measurement
of resistivity is ohms or even though. ANSI/ESD S11.11
uses ohms/square.
Battery Life:Approximately1,500measurements
Resistivity Electrodes: Two parallel conductive silicone
rubber electrodes (optional item)
External Electrodes:Two5-pound(±2oz)electrodes,2.5"
in diameter, complies with ANSI/ESD S4.1
Display:One3.5digit.500”LiquidCrystalDisplay(LCD)
Display and seventeen Light Emitting Diodes (LED’s)
illuminateexponentnumbersandthefunctionbeing
performed.
*Accuracy:Resistancemeasurementswithin±10%(E11
&greater±20%),complieswithANSI/ESDS4.1.Open
circuitvoltagesof10volts±5%and100volts±5%halfthe
accuracy required of ANSI/ESD S4.1
Meter Weight: 15.5oz.
Dimensions:8"Lx4.3"Wx1.6"H
Test Switch:Presstheredtestbuttonfor15-20second
electrification period. Meter will immediately display surface
resistanceexponentviaLED.LCDwilldisplaytemperature
(both Fahrenheit & Celsius), then relative humidity, and
after taking numerous readings and making numerous
calculations display surface resistance mantissa.
Note: Reporting test results, you may want to record the
resistance, humidity, temperature, and test voltage.
Maintenance
YourSurfaceResistanceTestKitwillrequirelittle
maintenance, and there are no user serviceable parts. If
your Meter requires service beyond cleaning the Electrodes
or replacing the batteries, please contact the factory.
Thisproductutilizesahighfrequencyswitchingcircuit
tostepupthe3voltsfromthebatteriestothe100volt
test level. Some users are able to discern a slight hum
orbuzzing.Thisisperfectlynormalandshouldnotbe
considered a flaw or defect.
©2010DESCOINDUSTRIESINC.
EmployeeOwned
EMIT-3651WalnutAvenue,Chino,CA91710•(909)664-9980•Fax(909)627-7449•Website:DescoEMIT.com

TB-6520Page6of7 ©2010DESCOINDUSTRIESINC.
EmployeeOwned
Calibration
The Surface Resistance Test Kit is calibrated to NIST
traceable standards. Most users require calibration
annually. Please call our Customer Service Department at
909-664-9980(Chino,CA)or781-821-8370(Canton,MA)
fordetails.In-housecalibrationcanbeperformedbyusing
1%resistorsineachofthemeterranges.Simplyattachthe
resistors to the cords using grounding clips and recording
the meter display. Keep the cords separated. Should
adjustment be necessary, it is recommended that the unit
be returned to the factory as access to internal adjustments
requires special equipment and test cables.
CALIBRATION EQUIPMENT:
• DigitalMultimeter-accurateto1.25%@10VDCand
100VDC
• FixedDecadeBoxvalue10E3-10E12-accurateto
±2.5%,exceptat10E11and10E12(±5%)
• Thermometer-accurateto±1°F
• Humiditymeter-accurateto2%
• Testleads
• 99%Isopropylalcoholandcleaningwipes
SET UP:
A. Test Area-Areaneedstobefreeofanyhighvoltage
transformer or power supply. Not to be performed under
any type of fluorescent lighting or high power lighting.
B. Worksurface-needstobecoveredwithconductivemat
at1.0x10E3orless,connectedtoequipmentground.
C. Technician-needstobegroundedwithzeroohm
resistor to equipment ground.
D. Decade Box -needstobegroundedtoequipment
ground.
Accuracyismeasuredafternormalizingtheinstrumentat
70-80°Fand30-50%RHforaminimumof4hours.
NORMALIZATION OF SURFACE RESISTANCE TEST KIT
Temperatureinsidetestingareaneedstobe75°F@
±6.6%40%to60%RH.Testersneedtostayatconstant
temperature75°F@±6.6%forabout2hoursforproper
reading.Metercannotbeinsideobjects,enclosedboxes,
containers or cases unit is supplied with (temperature inside
case will differ from outside temperature, cases will act like
an insulator to the Meter), Meter will have to be stationary
in testing area for about 2 hours with no significant
temperature changes.
CALIBRATION TEST PROCEDURE
• Whentestinguseonlytheleadssuppliedwiththemeter.
DONOTTESTWITHRESISITIVITYATTACHMENT,
Resistivity Attachment will give a much higher reading
and is only designed for spot checking.
• With10V/AUTOswitchdown,pressTESTbutton-
voltagebetweenthetwoleadsshouldbe10V±5%
• With10V/AUTOswitchup,pressTESTbutton-voltage
betweenthetwoleadsshouldbe100V±5%
• Usingthecleaningwipesand99%isopropyl,clean
around banana jack and mono jack where leads connects
to, oil from human fingers can alter accuracy.
• Makesure10V/AUTOswitchissettoAUTO(switchup).
Testingeachdecadestartingfrom1.0x10E12anddown,
donotstartfrom1.0x10E3.
Temperature Fahrenheit =75°F±6.6%
Celsius=23.8°C±10.2%
RelativityHumidity±10Digits
A. 1.0 x 10E12
+20% LED=12YellowMantissa1.20
0% LED=12YellowMantissa1.00
-20% LED=11YellowMantissa8.00
B. 1.0 x 10E11
+20% LED=11YellowMantissa1.20
0% LED=11YellowMantissa1.00
-20% LED=10Green Mantissa8.00
C. 1.0 x 10E10
+10% LED=10Green Mantissa1.10
0% LED=10Green Mantissa1.00
-10% LED=9Green Mantissa9.00
D. 1.0 x 10E9
+10% LED=9Green Mantissa1.10
0% LED=9Green Mantissa1.00
-10% LED=8Blue Mantissa9.00
E. 1.0 x 10E8
+10% LED=8Blue Mantissa1.10
0% LED=8Blue Mantissa1.00
-10% LED=7Blue Mantissa9.00
F. 1.0 x 10E7
+10% LED=7Blue Mantissa1.10
0% LED=7Blue Mantissa1.00
-10% LED=6Blue Mantissa9.00
G. 1.0 x 10E6
+10% LED=6Blue Mantissa1.10
0% LED=6Blue Mantissa1.00
-10% LED=5Green Mantissa9.00
H. 1.0 x 10E5
+10% LED=5Green Mantissa1.10
0% LED=5Green Mantissa1.00
-10% LED=4Green Mantissa9.00
I. 1.0 x 10E4
+10% LED=4Green Mantissa1.10
0% LED=4Green Mantissa1.00
-10% LED=3Red Mantissa9.00
J. 1.0 x 10E3
+10% LED=3Red Mantissa1.10
0% LED=3Red Mantissa1.00
-10% LED=<3Red Mantissa<0.90
EMIT-3651WalnutAvenue,Chino,CA91710•(909)664-9980•Fax(909)627-7449•Website:DescoEMIT.com

TB-6520 Page 7 of 7 ©2010DESCOINDUSTRIESINC.
EmployeeOwned
Specifications
Ranges: 1x10E3-<1x10E6ohms@10Volts
complies with ANSI/ESD S4.1.
1x10E6-1x10E12ohms@100Volts
complies with ANSI/ESD S4.1.
Power Supply: Two AA alkaline batteries
EMIT-3651WalnutAvenue,Chino,CA91710•(909)664-9980•Fax(909)627-7449•Website:DescoEMIT.com
Limited Warranty
EMITexpresslywarrantsthatforaperiodoffive(5)yearsfromthedateof
purchase, EMIT Surface Resistance Test Kits will be free of defects in material
(parts)andworkmanship(labor).Withinthewarrantyperiod,acreditforpurchase
of replacement EMIT products, or, at EMIT’s option, the product will be repaired or
replaced free of charge. If product credit is issued, the amount will be calculated by
multiplyingtheunusedportionoftheexpectedfiveyearlifetimestheoriginalunit
purchaseprice.CallourCustomerServiceDepartmentat909-664-9980(Chino,CA)
or781-821-8370(Canton,MA)foraReturnMaterialAuthorization(RMA)andproper
shipping instructions and address. Please include a copy of your original packing
slip, invoice, or other proof of date of purchase. Any unit under warranty should be
shippedprepaidtotheEMITfactory.Warrantyreplacementswilltakeapproximately
two weeks.
If your unit is out of warranty, call our Customer Service Department at
909-627-8178(Chino,CA)or01892665313(Crowborough,U.K.)foraReturn
MaterialAuthorization(RMA)andpropershippinginstructionsandaddress.EMIT
will quote repair charges necessary to bring your unit up to factory standards.
Warranty Exclusions
THEFOREGOINGEXPRESSWARRANTYISMADEINLIEUOFALL
OTHERPRODUCTWARRANTIES,EXPRESSEDANDIMPLIED,INCLUDING
MERCHANTABILITYANDFITNESSFORAPARTICULARPURPOSEWHICHARE
SPECIFICALLYDISCLAIMED.Theexpresswarrantywillnotapplytodefectsor
damage due to accidents, neglect, misuse, alterations, operator error, or failure to
properly maintain, clean or repair products.
Limit of Liability
In no event will EMIT or any seller be responsible or liable for any injury, loss or
damage, direct or consequential, arising out of the use of or the inability to use
theproduct.Beforeusing,usersshalldeterminethesuitabilityoftheproductfor
their intended use, and users assume all risk and liability whatsoever in connection
therewith.
This manual suits for next models
1
Table of contents
Other EMIT Test Equipment manuals

EMIT
EMIT SmartLog V5 Installation and operating instructions

EMIT
EMIT TB-6568 Installation guide

EMIT
EMIT eScan2 User manual

EMIT
EMIT 50512 Service manual

EMIT
EMIT SmartLog V5 Service manual

EMIT
EMIT ESD Glove Test Fixture Installation and operating instructions

EMIT
EMIT Limit Comparato User manual

EMIT
EMIT SmartLog V4 Installation and operating instructions