Esco Medical MIRI AVT User manual

USER MANUAL
Rev. 4.0
Date of revised 16.06.2022
Rx only
Anti-Vibration Table (MIRI®AVT)

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
2
Tel 215-441-9661 Fax 484-698-7757
Tel +65 6542 0833 Fax +65 6542 6920
The information in this manual and the accompanying product is copyrighted and all rights are
reserved by Esco.
Esco reserves the right to make periodic minor design changes without obligation to notify any
person or entity of such change.
Caution: Federal law restricts this device to sale by or on a licensed healthcare practitioner's order.
Only to be used by a trained and qualified professional. The device is sold under exemption 21 CFR
801 Subpart D.
"Material in this manual is provided for informational purposes only. The contents and the product
described in this manual (including any appendix, addendum, attachment or inclusion) are subject to
change without notice. Esco makes no representations or warranties as to the accuracy of the
information contained in this manual. In no event shall Esco be held liable for any damages, direct or
consequential, arising out of or related to the use of this manual."

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
3
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping
carton for damage. If the damage is found, stop unpacking the instrument. Notify the
freight carrier and ask for an agent to be present while the instrument is unpacked. There
are no special unpacking instructions, but be careful not to damage the instrument when
unpacking it. Inspect the instrument for physical damage such as bent or broken parts,
dents, or scratches.
Claims
Our routine method of shipment is via common carrier. If physical damage is found, upon
delivery, retain all packing materials in their original condition and immediately contact
the carrier to file a claim.
If the instrument is delivered in good physical condition but does not operate within
specifications, or if there are any other problems not caused by shipping damage, please
contact your local sales representative or Esco Medical immediately.
Standard Terms and Conditions
Refunds & Credits
Please note only serialized products (products labelled with a distinct serial number) and
accessories are eligible for a partial refund and/or credit. Non-serialized parts and
accessory items (cables, carrying cases, auxiliary modules, etc.) are not eligible for return
or refund. In order to receive a partial refund/credit, the product must not have been
damaged. It must be returned complete (meaning all manuals, cables, accessories, etc.)
within 30 days of original purchase, in "as new" and resalable condition. The Return
Procedure must be followed.
Return Procedure
Every product returned for refund/credit must be accompanied by a Return Material
Authorization (RMA) number obtained from Esco Medical Customer Service. All items
being returned must be sent prepaid (freight, duty, brokerage and taxes) to our factory
location.
Restocking Charges
Products returned within 30 days of original purchase are subject to a minimum
restocking fee of 20% of the list price. Additional charges for damage and/or missing
parts and accessories will be applied to all returns. Products that are not in "as new" and
resalable condition are not eligible for credit return and will be returned to the customer
at their own expense.
Certification
This instrument has been thoroughly tested/inspected and found to meet Esco Medical's
manufacturing specifications when shipped from the factory. Calibration measurements
and testing are traceable and done according to Esco Medicals ISO certification.
Warranty and Product Support
Esco Medical warrants this instrument to be free from defects in materials and
workmanship under regular use and service for two (2) years from the original purchase
date, provided the instrument is calibrated and maintained following this manual. During
the warranty period, Esco Medical will, at our option, either repair or replace a product

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
4
that proves to be defective at no charge, provided you return the product (shipping, duty,
brokerage and taxes prepaid) to Esco Medical. Any transportation charges incurred are
the purchaser's responsibility and are not included within this warranty. This warranty
extends only to the original purchaser. It does not cover damage from abuse, neglect,
accident or misuse, or as the result of service or modification by parties other than Esco
Medical.
IN NO EVENT SHALL ESCO MEDICAL LTD. BE LIABLE FOR CONSEQUENTIAL DAMAGES.
No warranty shall apply when any of the following causes damage:
•Power failure, surges, or spikes
•Damage in transit or when moving the instrument
•An improper power supply such as low voltage, incorrect voltage, defective wiring
or inadequate fuses
•Accident, alteration, abuse or misuse of the instrument
•Fire, water damage, theft, war, riot, hostility, acts of God such as hurricanes, floods,
etc.
Only serialized products (those items bearing a distinct serial number tag) and their
accessory items are covered under this warranty.
PHYSICAL DAMAGE CAUSED BY MISUSE OR PHYSICAL ABUSE IS NOT COVERED UNDER
THE WARRANTY. Items such as cables and non-serialized modules are not covered under
this warranty.
This warranty gives you specific legal rights and you may have other rights, which vary
from province to province, state to state, or country to country. This warranty is limited
to repairing the instrument per Esco Medical's specifications.
When you return an instrument to Esco Medical for service, repair or calibration, we
recommend shipment using the original shipping foam and container. If the original
packing materials are not available, we recommend the following guide for repackaging:
•Use a double-walled carton of sufficient strength for the weight being shipped
•Use heavy paper or cardboard to protect all instrument surfaces. Use non-abrasive
material around all projecting parts
•Use at least four inches of tightly packed, industrial-approved, shock-absorbent
material all around the instrument
Esco Medical will not be responsible for lost shipments or instruments received in
damaged condition due to improper packaging or handling. All warranty claim shipments
must be made on a prepaid basis (freight, duty, brokerage, and taxes). No returns will be
accepted without a Return Materials Authorization ("RMA") number. Please contact Esco

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
5
Medical to obtain an RMA number and receive help with shipping/customs
documentation.
Re-calibration of instruments, which have a recommended annual calibration frequency,
is not covered under warranty.
Warranty Disclaimer
If your instrument is serviced and/or calibrated by someone other than Esco Medical Ltd.
and their representatives, please be advised that the original warranty covering your
product becomes void when the tamper-resistant Quality Seal is removed or broken
without proper factory authorization.
In all cases, breaking the tamper-resistant Quality Seal should be avoided at all costs, as
this seal is key to your original instrument warranty. In an event where the seal must be
broken to gain internal access to the instrument, you must first contact Esco Medical Ltd.
You will be required to provide us with the serial number for your instrument, as well as
a valid reason for breaking the Quality Seal. You should break this seal only after you have
received factory authorization. Do not break the Quality Seal before you have contacted
us! Following these steps will help ensure that you will retain the original warranty on
your instrument without interruption.
WARNING
Unauthorized user modifications or applications beyond the published specifications
may result in electrical shock hazards or improper operations. Esco Medical will not be
responsible for any injury sustained due to unauthorized equipment modifications.
ESCO MEDICAL LTD. DISCLAIMS ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE OR APPLICATION.
THIS PRODUCT CONTAINS NO USER-SERVICEABLE COMPONENTS.
UNAUTHORIZED REMOVAL OF THE INSTRUMENT COVER SHALL VOID THIS AND ALL
OTHER EXPRESSED OR IMPLIED WARRANTIES.

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
6
Table of contents
1 How to use this manual............................................................................................................................ 7
2 Safety warning............................................................................................................................................. 7
3 About the product ...................................................................................................................................... 7
4 Accessories supplied................................................................................................................................. 8
5 Safety symbols and labels ....................................................................................................................... 8
6 Important safety instructions and warnings................................................................................... 9
6.1 Before installation ............................................................................................................................. 9
6.2 During installation............................................................................................................................. 9
6.3 Post-installation ................................................................................................................................. 9
6.4 Bring the following to the installation site............................................................................... 9
7 Assembling Esco Medical MIRI®AVT................................................................................................. 9
8 Cleaning instructions..............................................................................................................................13
9 Maintenance...............................................................................................................................................14
10 User troubleshooting............................................................................................................................14
11 Specifications...........................................................................................................................................15

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
7
1 How to use this manual
The manual is intended to be read from start to finish.
2 Safety warning
•Anyone working with, on or around this equipment should read this manual.
Failure to read, understand and follow the instructions given in this
documentation may result in damage to the unit, injury to the operating personnel
and/or poor equipment performance.
•Any internal adjustment, modification or maintenance to this equipment must be
undertaken by qualified service personnel.
•If the equipment must be relocated, ensure it is adequately fixed on a support
stand or base and move it to a flat surface. When necessary, move the equipment
and the support stand/base separately.
•The use of any hazardous materials in this equipment must be monitored by an
industrial hygienist, safety officer or another suitably qualified individual.
•Before proceeding, you should thoroughly understand the installation procedures
and take note of the environmental requirements.
•In this manual, important safety-related points will be marked with the following
symbols:
NOTE
Used to direct attention to a specific item.
WARNING
Use caution.
•If the equipment is used in a manner not specified by this manual, the protection
provided by this equipment may be impaired.
3 About the product
Esco Medical MIRI®AVT is an anti-vibration table intended to be used with high
magnification microscopes. It isolates the microscope from vibrations generated in the
room where the table is placed and from vibrations generated above or below any part
of the tabletop that is not part of the AVT platform.
The platform consists of a sturdy frame and a tabletop in stainless steel and glass. The
AVT platform is located in the middle of the tabletop and it is floating on the AVT
mechanism.

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
8
The anti-vibration mechanism works as a passive damping device and only partially
damps the vibration that is generated on the AVT platform itself.
The microscope must be placed on the anti-vibration float.
In all buildings, floors vibrate at low frequencies, which may affect the performance of a
high-magnification microscope. The effects of vibration can be significantly reduced by
installing a system that isolates the microscope area. The degree of improvement varies
between instruments and locations, depending on the amplitude and frequency of floor
vibrations.
This product does not carry a CE mark. Only those product categories that are subject to
specific CE marking directives are required (and allowed) to carry CE marking. There are
no such categories or directives that would include a non-electrical anti-vibration table
as MIRI®AVT.
The product is manufactured under ISO 9001 Quality Management System.
Personal Protective Equipment (89/686/EEC) and Machine Directive (2006/42/EC) is
not applicable for MIRI®AVT.
4 Accessories supplied
The User manual
Spanner key (N36) for leg adjustment
A hex key (6x90 mm)
5 Safety symbols and labels
There is one user label on the surface of MIRI®AVT to guide the user. The user label is
shown below.
Table 5.1 Device label
Description
Image
1. Model
2. CE mark
3. Manufacturers address and country
of origin
4. Logo and serial number
5. Observe WEEE
6. Year of manufacture

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
9
6 Important safety instructions and warnings
6.1 Before installation
1. Do not use the product if the package is damaged. Contact Esco Medical or the local
representative.
2. Read the User manual completely before use.
3. Always keep these instructions easily accessible near the device.
6.2 During installation
1. Place this unit on a flat, hard and stable surface.
2. Never place the unit on a carpet or similar surfaces.
3. This unit is intended for the indoor purpose only.
4. Lift the device carefully. All MIRI®AVT parts are heavy and may require additional
people to move them around.
5. Float elements have to go into the hole on a tabletop be careful not to squeeze
fingers or drop the element through the hole.
6.3 Post-installation
1. Refer all servicing procedures to qualified service personnel.
6.4 Bring the following to the installation site
1. 1 level tool.
2. Gloves for installation.
7 Assembling Esco Medical MIRI®AVT
Follow the guidelines in the safety instructions and warnings section.
Clean and disinfect the device before use. It is not delivered sterile or in a
clinically acceptable clean state. Look through this manual's 8th “Cleaning
instructions” section for the manufacturer’s recommended guidelines!
Do not touch the springs or contact points with bare hands! Skin grease will
compromise the performance of vital friction points.
Place the anti-vibration table in an appropriate place on the floor.
Level the table platform horizontally by adjusting the table legs with a spanner key. To
ensure a good result, place the level on a stable platform.

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
10
Figure 7.1 Leveling table legs
Put on gloves.
The 3 springs are already installed in the MIRI®AVT.
Figure 7.2Three springs in the MIRI®AVT
The unit comes with already safety guard installed. It is screwed with 3 screws on the left,
right and the bottom side. 3 screws on the right side are marked red in the image below.
Figure 7.3Right side screws holding the safety guard

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
11
The floated part labeled 1 should be placed on the springs and equal distance left from
the hole on all sides to have space for centering the floating platform. The holding screws
should be facing up.
Figure 7.4 No contact between the floating plate and the table
It is essential that there would be no contact between the floating platform and the table.
Then, place float element 2 on the top of float element 1.
Figure 7.5 Rubber stops facing up
The rubber stops should be facing up. Leave equal distance from the hole on all sides of
the table platform.
Then place the top of the floating element (stainless steel) on the rubber parts again,
leaving an equal distance from the hole on all sides of the table platform.

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
12
Figure 7.6 Float element is placed too high
In the picture above, the floated element is too high. The adjustment can only be made
after the microscope has been placed on the float, as the weight will push it down.
It is the fully laden float that should be level. Thus, the adjustment is dependent on the
microscope weight.
The floated element should be at the same level as the tabletop after placing
the microscope.
The float height is adjusted by loosening 4 securing hexagonal head bolts and turning the
handle.
Figure 7.7 Adjusting the float height
The safety guard plate does not need to be removed to reach these hexagonal head bolts.

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
13
After adjustment, tighten the four securing bolts again. If this is done improperly, the
table will go out of adjustment while in use.
The spanner key for leg adjustment can be placed under the MIRI®AVT in the right
corner.
Figure 7.8 Spanner key saving place
Any additional equipment may now be installed.
Ensure all cables connected to the microscope hang freely between the attachment point
on the microscope and the other side. It can be achieved by stripping the cable to the
points and ensuring that the cable forms an arch between them. Vibration may be
transferred via the cable if the cable is tight or touches an area between the anchor points.
After successful installation and device testing, complete user training and instructions.
8 Cleaning instructions
Always validate cleaning procedures locally; consult either your manufacturer
or the distributor for more guidance.
Switch off the devices on the table and remove the main leads. The combination of
standard cleaning procedures and disinfection procedures is recommended for event-
related concerns such as media spills, visual accumulation of soil and/or other evidence
of contamination. It is also recommended to clean and disinfect the MIRI®AVT
immediately after any media spills.
Periodic cleaning of the device
Wearing gloves and good handling techniques are essential to successful cleaning.
1. It is recommended to clean the unit with aqueous surface disinfectant or
disinfection wipes that have active alcohol or benzyl-alkyldimethyl chloride
ingredients. Clean all the devi external surfaces by rubbing with lint-free cloth
wipes.

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
14
2. Use purified or sterile water to wipe the surface of the device to remove any
residue.
3. Inspect the device if it is visually clean, consider it ready for use.
If the device is not visually clean, repeat the process from step 1.
9 Maintenance
Esco Medical MIRI®AVT is designed to be easy to use and requires almost no
maintenance at all.
Over time, adjustments may become necessary.
If needed, the rubber stops and the metal strips can be changed.
Warranty is considered to be void if service and maintenance procedures are
done not by trained and authorized personnel.
10 User troubleshooting
Table 10.1 Vibrations
Vibrations
Cause
Action
Vibrations from the floating
part
The float touches the table
platform, and vibrations are
transferred directly to the
microscope.
The floater should be in the
center position of the hole. Air
should be all around the floater.
Vibrations coming from
hexagonal head bolts
Loose tightening bolts.
Tighten four tightening bolts.
Vibrations coming from the
spring system (the spring
system is not painted and
must be entirely grease-free
for function; the surface is
rough intentionally).
Grease or oil on the surface of
contact points.
Clean the area with a disinfection
solution.
Vibrations coming from cables
(tight cable of a microscope or
any other equipment placed
on the tabletop will result in
vibrations being transmitted
directly to the microscope).
Cables routed incorrectly.
Tie cables so that they would not
touch the table and would hang
loosely between the attachment
points.
Vibrations coming from the
microscope
The microscope is placed unevenly
on the floater.
The weight of the microscope
has to be distributed evenly on
the floater. Since the back part of
a microscope is the heaviest, it
should be moved more to the
front of the table platform.

Anti-Vibration Table (MIRI®AVT) User Manual Rev. 4.0
15
11 Specifications
Table 11.1 MIRI®AVT specifications
Technical specifications
MIRI®AVT
Overall dimensions (W x D x H)
1200 × 800 × 785 mm
Weight
117 kg
Material
Powder painted mild steel/Stainless steel
Float
Size 540 × 340
Recommended load weight
26 32 kg
Damping coefficient (6 Hz)
~ 0.1
Amplitude (6 Hz)
< 1 µm
The spectrum of the vertical vibration velocity with the additional mass of 47 kg. The
resonance frequency is 5-5.29 Hz.
Figure 11.1 Resonance frequency
Table of contents
Other Esco Medical Medical Equipment manuals

Esco Medical
Esco Medical CE 0123 User manual

Esco Medical
Esco Medical CultureCoin User manual

Esco Medical
Esco Medical Multi-zone ART Workstation User manual

Esco Medical
Esco Medical Mini MIRI Dry User manual

Esco Medical
Esco Medical MIRI II-12 User manual

Esco Medical
Esco Medical MIRI GA User manual

Esco Medical
Esco Medical AVT-1 User manual