Maintenance
Cleaning and disinfection
Clean on a regular basis. Clean surfaces with
a damp cloth using an appropriate pH-neutral
detergent. Do not use solvents or strong liquids,
this may damage surfaces on the hoist. For
disinfection when needed; use isopropyl alcohol.
Avoid abrasive cleaning products. Remove hair
and pile from the castors and verify that the
castors rotates freely. Clean contact and hand
control with isopropyl alcohol to remove grease
and dirt. Check emergency stop and emergency
lowering after cleaning. The lift should not be
exposed to running water.
Make sure not to damage or
remove labels when cleaning.
Reconditioning
Follow cleaning and assembly procedure, complete
periodic inspection and use checklist after assem-
bly to recondition the hoist.
Recycling
Refer to “Recycling instructions” for how to
properly dispose of product. This can be found on
www.etac.com.
Spare parts
A list of spare parts is available on request.
Periodic inspection
Periodic inspection scope
Periodic inspection is a visual examination (par-
ticularly of the hoist’s load bearing structure and
lifting mechanism with attachments, brakes, con-
trols, safety devices and person-support devices)
according to Periodic inspection report for Molift
Smart 150. This can be found on www.etac.com.
If there is any problems with the
hoist that could jeopardize some-
ones safety, the hoist shall imme-
diately be taken out of service and
marked “out of order”. Do not use
the hoist untill it is repaired.
Periodic inspection shall be performed annually or
more frequently if required by local requirements.
The inspection must be performed by service
personnel authorized by Etac.
Contact Etac at molift@etac.com for training and
authorization or recommendation of an approved
service partner.
When performing a periodic inspection, the
inspector shall fill out the inspection report for
Molift Smart 150. The reports should be retained
by the person(s) responsible for servicing the hoist.
If the inspection reveals defects and damages, the
owner shall be notified.
When periodic inspection is completed the
inspector shall mark the hoist with a sticker on
the control label showing the date when periodic
inspection is performed and a label (with 6-digit
ID from Etac or distributor) showing who has
performed the periodic control. The control label
can be found on the right side of the battery
holder, and this will then indicate the date when
next Inspection is due.
Mark label
with month
and year of
inspection
ECP ID:Rev. Description: Released Date:Revised Date: Revised By: Released By:
Description / Title:
Designed By:
Date Created:
Category: Project ID:
Scale: Format: State:
Volume (mm3): Weight (g):
Material:
Drawing / Document number:
M_3004
APPROVED
2022-04-19
ELJ Label, safety check 03-270
2022-05-03
CO772 ELJ ELJ
2022-04-255
Ø30 mm
Material: Vinyl sticker
2
0
2
7
2
0
2
6
2
0
2
5
2
0
2
4
2
0
2
3
ECP ID:Rev. Description: Released Date:Revised Date: Revised By: Released By:
Description / Title:
Designed By:
Date Created:
Category: Project ID:
Scale: Format: State:
Volume (mm3): Weight (g):
Material:
Drawing / Document number:
APPROVED
2022-04-20
ELJ 13-224
2022-05-04
CO773 ELJ ELJ
2022-05-048
Material: Transparent vinyl sticker
60 mm
35 mm
Next periodic
inspection
Inspection performed by: