ETCR 6000 User manual


-1-
CONTENT
Warning.....................................................................................................................................................................2
I Introduction.............................................................................................................................................................2
II ElectricalSigns .....................................................................................................................................................3
TechnicalSpecifications....................................................................................................................................3
ⅣStructure..............................................................................................................................................................4
ⅤLCD Display........................................................................................................................................................4
ⅥOperatingMethod..............................................................................................................................................4
1. Start-up, Shutdown.........................................................................................................................................4
2. Calibration......................................................................................................................................................4
4. Hold,Storage,Access ReadingandDeleting.................................................................................................5
5. DataUpload....................................................................................................................................................5
ⅦHowtoChangeBattery.....................................................................................................................................6
ⅧAccessories.........................................................................................................................................................6

-2-
Warning
Thankyouforpurchasingourcompany's ETCR6000AC/DC ClampLeaker,forbetteruseofthis
product, please:
Readcarefullytheuser'smanual.
Followstrictlysafetyrules and notes listedinthismanual.
uUnderanycircumstances, pleasepayspecialattention toyoursafetyinthe courseofusingthis
leaker.
uGiveheed tolabeltextsandsymbolson panelandback plateofthisleaker.
uPleasebemorecarefulifthe line voltageisabove60VDC or30VAC.
uAfterpoweron, pleasefirst calibratezerobeforemeasuring.
uPutthetested wirethroughthe geometricalcenterofjawwhen measuringACcurrent, errorwill
increaseifdeviatethe center.
uPleasedon'tplaceandstorethisleakerinhotandhumidcondition,locationswithmoisture
condensation andunderdirect sunlight foralongtime.
uIncasevoltageofbatterywaslow, pleasereplacebatteries.
uIncasethisleakerwouldnot beusedforalongtime, pleasetakeout batteries.
uWhen changingbatteries, pleasepayattention topolarityofbattery.
uUse, disassemblyand maintenanceofthisleakershall beoperated byauthorized personnel.
uIncasedangerswouldhavewithcontinuesuseofthisleaker,pleasestoptouseitandsealitfor
safekeepingimmediately; andthen, send it fordisposalofauthorized agency.
uUsersshall carryout operation basedon dangersigns ”on leakerandmanual.
uUsersshall carryoutsafetyoperation basedon instructionslisted inthismanual,e.g. ”and
dangersignson thismanual.
I Introduction
ETCR6000 AC/DC ClampLeaker isspeciallydesigned formeasurementofAC/DCcurrent; by
adoptingup-to-dateCTtechnologyand digitalintegration technology. The small clampissuitableforline
denselyplaces(electricpowermeasurementsystem,the highspeed rail system,carcircuitoverhaul,
and soon), non-contact measurement, toensuresafeoperation. It isaproduct withrelativelysmallsize,
highaccuracy andperfectfunction compared withsimilarleakersinthe world.The leakercouldbe
widelyapplied inthosefieldsaselectricity,communications,meteorology,railroad,oilfield,construction,
measurement, scientific&researchteachinginstitutes,industrialandminingestablishments.itisan
essentialtoolforelectrician safetytesting.
ETCR6000 AC/DC ClampLeaker alsonamed: DC LeakageCurrentClampMeter, Car DC
LeakageCurrentClampMeter,AC/DC ClampLeakageCurrentTester.Ithasthosefunctionsasdata
holdinganddatastorage.ItobtainsoneRS232interface,communication cableandsoftware,through
whichhistoricaldatainquiryisavailable,forreading, savingand printing.

-3-
II Electrical Signs
Extremelydangerous!Operatorsshall strictlyobservesafetyrules;otherwisethere
wouldbedangersofelectricshock tocausepersonalinjuriesorcasualties.
Dangerous!Operatorsshall strictlyobservesafetyrules;otherwisetherewouldbe
dangersofelectricshock tocausepersonalinjuriesorcasualties.
Warning!Operatorsshall strictlyobservesafetyrules;otherwisepersonalinjuriesor
equipment damagesmight becaused.
Doubleinsulation
AC
DC
Technical Specifications
Function Measurement ofAC/DC current, leakagecurrent
PowerSupply Zn-Mndrybattery,6F22,9V
TestMode Clip-on CT, integralmode
ClampSize φ7mm
MeasurementRange 0mA-60.0AAC/DC
Resolution 1mAAC/DC
Accuracy ±2%rdg±5dgt (23℃±5℃, below75%rh)
Display Mode FourdigitsLCD display
Dimension LWH: 168mm×65mm×34mm
LCD Dimension 35mm×21.5mm; displaydomain: 32mm×15mm
Sampling Rate 2times/s
Frequency AC: 45Hz-400Hz
PolarityIndication DC current autoidentified and display ”
TestPosition Tested wireinthe jawcenter
RangeShift Automatically
LineVoltage AC600V
RS232 Interface Datastoredinthememoryofthe meterviaRS232 upload toPC
Com-Configure Baud rate:9600,databit:8, stopbit:1
DataMemory 99units,FULL blinks when the memoryisfull
Reading Hold DHindicatingthe readingishold
OutofRange OLindicatingthecurrent isout ofrange
AutoPower-off 5Minutesafterpoweron, it will poweroffautomaticallytolowerthe
powerconsumption
BatteryVoltage Indicatingthebatteryvoltageislowerthan 7.2V.Then the
batteryhavetobechanged
Weight 180g(includingthe battery)
Working Current 10mA
Consumption 10mW
Working Temperature
and Humidity -10 -50,below80%rh
LimitTemperature
Error -10℃-0℃and 40℃-50℃, within2%rdg
StorageTemperature
and Humidity -10 -60,below70%rh
Insulating Strength AC2kV/rms(between the alloyofthe clampand the housing)
Applicablesafetyrules
IEC1010-1,IEC1010-2-032,Pollution level2, CAT (600V)

-4-
StructureⅣ
1. Clampmouth(φ7mm)
2. DC current positiveinput indication
3. Clamp(slendershape)
4. Toroidopeninglever
5.HOLD key
6. Power/ AC&DC switchkey
7. LCD display
8. RS232Interface
9. Adjust zerokey
10. Batterycoverscrew(1piece)
11. Batterycover
12. Housingscrews(3pieces)
13. Pendant hole
LCD DisplayⅤ
1. DC indication 2.ACindication
3. Negativepolarityindication
4. Lowbatterysymbol
5. Dataholdindication
6. Storeddatacode
7. Current unit8. Value
Operating Method
1. Start-up, Shutdown
SwitchpowerkeytoDC orACgeartostart upthe tester, LCD will begintodisplay;andswitch
the powerkeytoOFF gear,theleakerwill shutdown.Afterstartingupfor5minutes,LCD will flick
notesthattheleakerwillshutdownautomatically;afterflickering for30s,itwillshutdownformally
toreducebatteryconsumption. Incaseyou havepressed HOLD keywhen LCD wasflickering, the
leakerwillcontinuetoworkfor5minutes.Ifthe testerpoweroffautomatically,the gearshouldfirstly
beswitched toOFF forpoweron next time.
2. Calibration
When measuring,first chooseACorDC gear, adjust ZERO keytoreducethe
residualmagnetismtozero, andthenconduct measurement. Rationalusage
ofthisadjust zerofunction will makethe resultsmoreaccurate.Forexample,
afterboot, beforemeasurement, wecan takethe jawclosetothe DC current
wire(showingasrightfigure), LCD will showan inductivecurrent (becauseof
externalelectricfieldinterference).Adjust ZERO keytocalibrate, whichdeduct
the inductivevalue.Aftermeasuringbigcurrent, calibration isverynecessary
tocleartheresidualmagnetism.

-5-
3.Measurement
1)Poweron
2)SwitchAC/DC keytochoosethekind ofmeasurement.
3)Adjust zero
4)Releasethetoroidlevertoopen clampmouthandclampmeasured conductors.
5)Read LCD displaydata.Incase OLA symbolwasdisplayed,it meansthatcurrentofmeasured
line isbeyondthemaximumlimitofthisleaker;withthiscase,pleasechooseleakerwithmuch
higherrangelimit.
Clamppositivewireand negativewiretogetheristomeasureleakagecurrentofDC
circuit.(Note:2wires)
ClamplivewireandnulllinetogetheristomeasureACleakagecurrent.(Note:2wires)
Clampearthwireistomeasuregroundinglineleakagecurrent.(Note:singlewire)
Clampmainwireistomeasurethecurrent.(Note:singlewire)
4. Hold, Storage,Access Reading and Deleting
1)Pressing HOLD keyforashorttimeinthecourseof measurement(less than3seconds), DH
symbolwill display,the leakerwillholdcurrentmeasuringdataand automaticallystored inthe
memorywithacode;press HOLD keyagaintoreleasetheholdstate,andtheleakercontinues
itsmeasuring;incasestored datareached to60 groups,press HOLD keyagain,the FULL”
symbolwilldisplay,whichmeansstoragememoryisfull;press HOLD keytocancelFULL”
flickeringand returntomeasuringmode.
2)Longpress HOLD keytoenterintodataaccess mode and displayUnit1storagedata
automatically; and thenpress HOLD keyagaintoturnthepageofstored data;nullwill display
when thereisno datainstoredinthememory, poweroffthe metertoexit dataaccess mode.
3)Afterenteringintodataaccess mode,press HOLD keyformorethan 3secondswill clean upall
stored data;when the leakerdisplaying dEL”symbol,itmeansthatithasfinished cleanup
process, and thenreturntomeasuringstateautomatically.
5. DataUpload
Makegood connectionofcomputerandthetesterwithRS232communication wire, switchon the
testerandrun monitoringsoftware.Ifthe connectionissuccessful,then itcanread the stored historical
data, upload tocompany, preserveand print.

-6-
ⅦHow toChangeBattery
Warning!Itisdangeroustocarryouttestwhenthebatterycoverplatewas not
on itsposition.
Pleasepay attention topolarityofbatterytoavoiddamagingtheleaker.
Chang thelowbatteryintime
Ifnotuse themeterforalong time, please getoffthebatterytostorage.
1)”symbolmeansthe batteryisunderchargeand needtobereplaced.
2)Shutdownthe leaker;beforeopeningthe batterycover,pleaseconfirmthe leakerisinoff
position,and then replacewithqualified newbattery;specialattention shall bepaidtopolarityof
battery; at last, coverbatterycoverplate.
ⅧAccessories
Clamptester 1pc
RS232 Comcable 1 pc
Software 1 disk
Battery(6F22 9V)1 pc
UserManual 1copy
Warrantycard/Certification 1 copy
Table of contents