Fluke Biomedical IMPULSE 6000D User manual

Impulse 6000D
Defibrillator Analyzer
Impulse 7000DP
Defibrillator/Transcutaneous Pacer Analyzer
Users Manual
August 2007, Rev. 1, 1/08
© 2007 Fluke Corporation, All rights reserved.
All product names are trademarks of their respective companies.

Warranty and Product Support
Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of origi-
nal purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for
calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our
option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to
Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if
the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized
Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE
EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSE-
QUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration
of instruments is not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since
some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages,
this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or
other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
7/07

Notices
All Rights Reserved
©Copyright 2007, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into
any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs
and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument.
Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to dam-
age the instrument when unpacking it. Inspect the instrument for physical damage such as bent orbroken parts, dents, or scratches.
Technical Support
For application support or answers to technical questions, either email [email protected] or call 1-800- 648-7942 or 1-425-446-6945.
Claims
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original
condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifica-
tions, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative.

Standard Terms and Conditions
Refunds and Credits
Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for
partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for re-
turn or refund. Onlyproducts returned within 90 days from the date of original purchase are eligible for refund/credit. In order to receive a partial re-
fund/credit of a product purchase price on a serialized product, the product must not have been damaged by the customer or by the carrier chosen by the cus-
tomer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in “as new” and resalable con-
dition. Products not returned within 90 days of purchase, or products which are not in “as new” and resalable condition, are not eligible for credit return and
will be returned to the customer. The Return Procedure (see below) must be followed to assure prompt refund/credit.
Restocking Charges
Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days af-
ter purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additionalchargesfor damageand/ormissing parts and accesso-
ries will be applied to all returns.
Return Procedure
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to
Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its
actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper
packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:
Use a double–walled carton of sufficient strength for the weight being shipped.
Use heavy paper or cardboard to protectall instrument surfaces. Use nonabrasive material around all projecting parts.
Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.

Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our Order Entry Group at
1-800-648-7952 or 1-425-446-6945.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service or
In the U.S.A.:
Cleveland Calibration Lab
Tel: 1-800-850-4606
Everett Calibration Lab
Tel: 1-888-99 FLUKE (1-888-993-5853)
In Europe, Middle East, and Africa:
Eindhoven Calibration Lab
Tel: +31-402-675300
In Asia:
Everett Calibration Lab
Tel: +425-446-6945
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shippedfrom the
factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibra-
tion standards are measured against in-house performance standards using accepted test procedures.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Bio-
medical will not be responsible for any injuries sustained due to unauthorized equipment modifications.

Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in
this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability
of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location
The Impulse 6000D and 7000DP Defibrillator/Transcutaneous Analyzers are manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA,
U.S.A.

i
Table of Contents
Title Page
Defibrillator Analyzer...............................................................................................................1
Introduction ....................................................................................................................1
Intended Use..................................................................................................................1
Unpacking the Analyzer .................................................................................................1
Safety Information..........................................................................................................2
Instrument Familiarization..............................................................................................4
Turning the Analyzer On and Off....................................................................................7
Accessing the Analyzer Tests ........................................................................................8
Analyzing Defibrillators...................................................................................................8
Testing Energy Levels...............................................................................................8
Testing Defibrillator Synchronization.........................................................................10
Testing Defibrillator Charge Time..............................................................................11
Analyzing Pacemakers (7000DP only)...........................................................................12
Setting Up the Analyzer for Pacer Testing.................................................................12
Performing a Pacer Asynchronous Test....................................................................14

7000DP, 6000D
Users Manual
ii
Performing a Pacer Demand Test.............................................................................15
Performing a Pacer Sensitivity Test..........................................................................16
Performing a Pacer Refractory Period Test ..............................................................17
Simulating ECG Signals.................................................................................................18
Connecting to the ECG Terminals.............................................................................19
Setting a Normal Sinus Rhythm ECG Signal ............................................................19
Setting a Performance ECG Signal...........................................................................19
Setting Pacer Interactive ECG Waves (7000DP only)...............................................21
Selecting ECG Arrhythmias ......................................................................................22
Selecting TV Paced...................................................................................................23
Testing R Wave Detection ........................................................................................23
Performing a Noise Immunity Test............................................................................24
Setting Analyzer Setup Functions..................................................................................24
Setting Up the Battery...............................................................................................24
Setting Up the Display...............................................................................................25
Setting Up Sound......................................................................................................26
Displaying Instrument Information.............................................................................26
Controlling the Analyzer Remotely.................................................................................26
Maintenance..................................................................................................................27
Cleaning the Analyzer...............................................................................................27
Maintaining Peak Battery Condition..........................................................................28
Accessories ...................................................................................................................29
Specifications.................................................................................................................30
General Specifications..............................................................................................30
Defibrillator Analyzer Specifications..........................................................................31
Transcutaneous Pacemaker Analyzer Specifications (Impulse 7000DP only)..........38

Contents (continued)
iii
Appendix A - Impulse 6000D/7000DP Remote Operation .....................................................41
Ansur Test Guide ...........................................................................................................41
Defibrillator Tests ...........................................................................................................43
Energy Measurement Test ........................................................................................43
Charge Time Test......................................................................................................45
Sync Time Test..........................................................................................................45
Pacemaker Tests ...........................................................................................................46
Pacer Parameter Test ...............................................................................................46
Pacer Refractory Test................................................................................................46
Pacer Sensitivity Test................................................................................................47
Pacer Demand Mode Test.........................................................................................47
Asynchronous Mode Test..........................................................................................48
ECG Pacer Interactive Test.......................................................................................48
ECG Waveform Simulation Tests...................................................................................48
Normal Sinus Wave Simulation Test .........................................................................48
Arrhythmia Wave Test...............................................................................................49
Performance Wave Simulation..................................................................................49
ECG R-Wave Test.....................................................................................................49
ECG Noise Immunity Test.........................................................................................50
Battery Performance Tests.............................................................................................50
Battery Capacity Test................................................................................................50
Defib Pulse Repetition Test.......................................................................................51
Appendix B - Impulse 6000D/7000DP Test Templates..........................................................53
Introduction ....................................................................................................................53
Creating Test Templates................................................................................................53
Using Defibrillator Test Elements...................................................................................60

7000DP, 6000D
Users Manual
iv
Energy Measurement Test........................................................................................60
Charge Time Test .....................................................................................................62
Synchronization Time Test........................................................................................63
Using Pacemaker Test Elements (Impulse 7000DP only) .............................................64
Pacer Parameter Test...............................................................................................64
Pacer Refractory Test...............................................................................................66
Pacer Sensitivity Test................................................................................................67
ECG Pacer Interactive Test ......................................................................................68
Pacer Demand Mode Test........................................................................................70
Asynchronous Mode Test .........................................................................................71
Using ECG Waveform Simulation Test Elements..........................................................71
Normal Sinus Wave Simulation.................................................................................71
Arrhythmia Wave Test...............................................................................................72
Performance Wave Simulation..................................................................................73
ECG R-Wave Test ....................................................................................................74
ECG Noise Immunity Test.........................................................................................75
Using Battery Performance Test Elements....................................................................75
Battery Capacity Test................................................................................................75
Defib Pulse Repetition Test.......................................................................................77

v
List of Tables
Table Title Page
1. Symbols................................................................................................................................. 2
2. Top-Panel Controls and Connections.................................................................................... 5
3. Rear-Panel Connections ....................................................................................................... 7
4. Accessories........................................................................................................................... 29
5. Energy Measurement Test Measurements............................................................................ 61
6. Energy Measurement Test Custom Parameters ................................................................... 61
7. Charge Time Test Measurements......................................................................................... 62
8. Charge Time Test Custom Parameters................................................................................. 63
9. Synchronization Time Test Measurements ........................................................................... 63
10. Synchronization Time Test Custom Parameters................................................................... 64
11. Pacer Parameter Test Measurements................................................................................... 65
12. Pacer Parameter Test Custom Parameters........................................................................... 65
13. Pacer Refractory Test Measurements................................................................................... 66
14. Pacer Refractory Test Custom Parameters........................................................................... 67
15. Pacer Sensitivity Test Measurements ................................................................................... 67
16. Pacer Sensitivity Test Custom Parameters........................................................................... 68
17. ECG Pacer Interactive Test Custom Parameters.................................................................. 69

7000DP, 6000D
Users Manual
vi
18. Pacer Demand Mode Test Custom Parameters ................................................................... 70
19. Normal Sinus Wave Simulation Test Custom Parameters.................................................... 71
20. Arrhythmia Wave Advisory Test Custom Parameters........................................................... 72
21. Performance Wave Simulation Test Custom Parameters..................................................... 73
22. ECG R-Wave Test Custom Parameters ............................................................................... 74
23. ECG Noise Immunity Test Custom Parameters.................................................................... 75
24. Battery Capacity Test Measurements................................................................................... 76
25. Battery Capacity Test Custom Parameters........................................................................... 76
26. Defib Pulse Repetition Test Measurements.......................................................................... 77
27. Defib Pulse Repetition Test Custom Parameters.................................................................. 77

vii
List of Figures
Figure Title Page
1. Top-Panel Controls and Connections.................................................................................... 4
2. Rear-Panel Connections ....................................................................................................... 6
3. Analyzer Ready Display ........................................................................................................ 7
4. Defibrillator Menu .................................................................................................................. 8
5. Cursor Navigation Example................................................................................................... 8
6. Defibrillator Test Connections ............................................................................................... 9
7. Defibrillator Energy Test........................................................................................................ 10
8. Defibrillator Synchronization Test.......................................................................................... 10
9. Defibrillator Charge Time Test............................................................................................... 11
10. Pacemaker Brand Selection.................................................................................................. 12
11. Connecting a Pacemaker to the Analyzer ............................................................................. 13
12. Displayed Pacer Parameters................................................................................................. 14
13. Pacer Async Overdrive Mode................................................................................................ 15
14. Pacer Demand Overdrive Test.............................................................................................. 16
15. Pacer Sensitivity Test Display............................................................................................... 16
16. Paced Refractory Period (PRP)............................................................................................. 17
17. Sensed Refractory Period (SRP)........................................................................................... 18

7000DP, 6000D
Users Manual
viii
18. ECG Main Menu ................................................................................................................... 19
19. Normal Sinus Rhythm Rate Selection................................................................................... 19
20. ECG Connections................................................................................................................. 20
21. Performance Wave Selection................................................................................................ 21
22. Pacer Simulation Interactive Setup Screen........................................................................... 21
23. Ventricular Parameter Selection ........................................................................................... 22
24. TV Paced Selection .............................................................................................................. 23
25. AV Sequential Screen........................................................................................................... 23
26. R Wave Detection Screen..................................................................................................... 23
27. Pacer Noise Immunity Test................................................................................................... 24
28. Battery Setup Screen............................................................................................................ 25
29. Analyzer Information Screen................................................................................................. 26
30. Ansur Test Guide Window .................................................................................................... 42
31. Graph of Discharge Curve .................................................................................................... 44
32. Test Template with Selected Test Element........................................................................... 54
33. User-Definable Parts of the General Setup Tab.................................................................... 55
34. Expected Results Options for User Input.............................................................................. 56
35. Changing the Operand in Expected Results......................................................................... 57
36. Add or Delete Limits Pop-up Menu....................................................................................... 58
37. Custom Setup Page for Pacer Parameter Test Element....................................................... 59

1
Defibrillator Analyzer
Introduction
The Impulse 6000D and 7000DP (hereafter the Analyzer)
are portable, battery-powered precision instruments for
testing external defibrillators. The 7000DP has the added
capability of testing trancutaneous pacemakers. Where
the additional pacemaker testing capability is applicable,
this manual qualifies the description with “7000DP only.”
The model 7000DP appears in all product illustrations.
Intended Use
The Analyzer is used to determine that defibrillators and
transcutaneous pacemakers are performing within their
performance specifications through measurement of
energy output.
Unpacking the Analyzer
Carefully unpack all items from the box and check that
you have the following items:
•Impulse 6000D or 7000DP
•Battery charger
•Getting Started Manual
•Users Manual CD
•Defib paddle contact plates
•Impulse 6000D 7000DP Ansur Software CD (demo)

Impulse 6000D, 7000DP
Users Manual
2
Table 1. Symbols
Symbol Description
WImportant information; refer to manual.
~Do not dispose of this product as
unsorted municipal waste. Go to Fluke’s
website for recycling information.
;Conforms to relevant Australian EMC
requirements
)Conforms to relevant Canadian and US
standards
XHazardous voltage
PConforms to European Union directives
CAT I
IEC Measurement Category I – CAT I
equipment designed to protect against
transients in equipment on circuits not
directly connected to MAINS. Under no
circumstances should the terminals of
the Analyzer be connected to any
MAINS voltage.
Safety Information
In this manual, a Warning identifies hazardous conditions
and actions that could cause bodily harm or death. A
Caution identifies conditions and actions that could
damage the Analyzer, the equipment under test, or cause
permanent loss of data.
XW Warning
To avoid possible electrical shock or personal
injury, follow these guidelines:
•Use this Analyzer only in the manner
specified by the manufacturer or the
protection provided may be impaired.
•Read the Users Manual before operating the
Analyzer.
•Do not use the product if it operates
abnormally.
•Do not use the product in wet locations,
around explosive gases or dust.

Defibrillator/Transcutaneous Pacemaker Analyzer
Safety Information
3
•Do not operate the Analyzer with the battery
eliminator connected, unless connected
directly to mains power. During battery
operation, completely remove the battery
eliminator/charger from both the Analyzer
and wall socket.
•Do not connect the Analyzer to a patient or
equipment connected to a patient. The
Analyzer is intended for equipment
evaluation only and should never be used
in diagnostics, treatment or in any other
capacity where the Analyzer would come in
contact with a patient.
•Observe all precautions noted by the
Device Under Test (DUT) equipment
manufacturer when analyzing the DUT.
•Use extreme caution when working with
voltages above 30 volts.
•Use the proper terminals, functions and
ranges for the test being performed.

Impulse 6000D, 7000DP
Users Manual
4
Instrument Familiarization
Figure 1 and Table 2 describes the top-panel controls and connections of the Analyzer.
DEFIBRILLATOR
5000
Vp
MAX
275
V MAX
1
6
2
3
4
7
55
8
10
11
9
fak07.eps
Figure 1. Top-Panel Controls and Connections

Defibrillator/Transcutaneous Pacemaker Analyzer
Instrument Familiarization
5
Table 2. Top-Panel Controls and Connections
Item Name Description
1 ECG lead connectors Outputs of low-level ECG signals (RA/R, LL/F, LA/L, RL/N, V1/C1, V2/C2, V3/C3, V4/C4,
V5/C5, and V6/C6).
2 Backlight button Turns the LCD display backlight on and off.
3 Power button Turns the Analyzer on and off.
4 Navigation buttons Cursor control buttons for navigating menus and lists.
5 Defib connectors Defibrillator connections (Shown with removable defib paddle contact plates installed).
6 Function softkeys Keys F1 through F5 are used to select from a number of selections that appear in the LCD
display above each function softkey.
7 Setup button Opens the setup menu.
8 ECG button Opens the main menu for ECG test functions.
9 Pacemaker inputs Input for low-level Pacer signal (7000DP only).
10 Pacer button Opens the main menu for pacer test functions (7000DP only).
11 Defibrillator button Opens the main menu for defibrillator test functions.

Impulse 6000D, 7000DP
Users Manual
6
Figure 2 and Table 3 describes the rear-panel connections of the Analyzer.
FLUKE BIOMEDICAL
6920 SEAWAY BLVD
EVERETT,WA 98203
www.flukebiomedical.com
MADE INUSA
SNSERIAL NUMBER COMPUTER
PORT
SCOPE
OUTPUT
CHARGE STATUS
15VDC / 1.5A
HIGH LEVEL
ECG OUTPUT
21 3 4 5
fak08.eps
Figure 2. Rear-Panel Connections
Other manuals for IMPULSE 6000D
1
This manual suits for next models
1
Table of contents
Other Fluke Biomedical Measuring Instrument manuals

Fluke Biomedical
Fluke Biomedical ESA615 User manual

Fluke Biomedical
Fluke Biomedical RaySafe Xi 8201011 User manual

Fluke Biomedical
Fluke Biomedical RAYSAFE 452 User manual

Fluke Biomedical
Fluke Biomedical 180 User manual

Fluke Biomedical
Fluke Biomedical ESA601 Parts list manual

Fluke Biomedical
Fluke Biomedical IDA-1S User manual

Fluke Biomedical
Fluke Biomedical Victoreen 07-494 User manual

Fluke Biomedical
Fluke Biomedical ASM-990 User manual

Fluke Biomedical
Fluke Biomedical VT305 Instruction Manual

Fluke Biomedical
Fluke Biomedical ESA620 User manual