Fluke IDA-5 User manual

FBC-0039
April 2013 Rev. 1
© 2013 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
IDA-5
Infusion Device Analyzer
Users Manual

Warranty and Product Support
Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from
the date of original purchase OR two years if at the end of your first year you send the instrument to a Fluke
Biomedical service center for calibration. You will be charged our customary fee for such calibration. During
the warranty period, we will repair or at our option replace, at no charge, a product that proves to be
defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the
original purchaser only and is not transferable. The warranty does not apply if the product has been
damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke
Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL,
INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA,
ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number
tag. Recalibration of instruments is not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different
jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of
incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this
warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such
holding will not affect the validity or enforceability of any other provision.
7/07

Notices
All Rights Reserved
Copyright 2013, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a
retrieval system, or translated into any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials
for use in service training programs and other technical publications. If you would like other reproductions or distributions,
submit a written request to Fluke Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Checkthe shipping carton for damage. If damage is
found, stop unpacking the instrument. Notifythe carrier and ask for an agent to be present while the instrument is
unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it.
Inspect the instrument for physical damage such as bent or broken parts, dents, or scratches.
Technical Support
For application support or answers to technical questions, either email techservices@flukebiomedical.com or call 1-800-
850-4608 or 1-440-248-9300. In Europe, email techsupport.emea@flukebiomedical.com or call +31-40-2965314.
Claims
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all
packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered
in good physical condition but does not operate within specifications, or if there are anyother problems not caused by
shipping damage, please contact Fluke Biomedical or your local sales representative.
Returns and Repairs
Return Procedure
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factorylocation. When
you return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel
Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be
responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or
handling.
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide
for repackaging:
Use a double–walled carton of sufficient strength for the weight being shipped.
Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all
projecting parts.
Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the
instrument.
Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number,
obtained from our Order Entry Group at 1-440-498-2560.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service or
In the U.S.A.:
Cleveland Calibration Lab
Tel: 1-800-850-4608 x2564
Everett Calibration Lab
Tel: 1-888-99 FLUKE (1-888-993-5853)
In Europe, Middle East, and Africa:
Eindhoven Calibration Lab
Tel: +31-40-2675300
In Asia:
Everett Calibration Lab
Tel: +425-446-6945

To ensure the accuracy of the Product is maintained at a high level, Fluke Biomedical recommends the product
be calibrated at least once every 12 months. Calibration must be done by qualified personnel. Contact your local
Fluke Biomedical representative for calibration.
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing
specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of
Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-
house performance standards using accepted test procedures.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards
or improper operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment
modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical.
Changes made to the information in this document will be incorporated in new editions of the publication. No
responsibility is assumed by Fluke Biomedical for the use or reliability of software or equipment that is not
supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location
The IDA-5 Infusion Device Analyzer is manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA,
U.S.A.

i
Table of Contents
Title Page
Introduction........................................................................................................ 1
Intended Use...................................................................................................... 1
Unpack the Product............................................................................................ 1
Safety Information............................................................................................. 2
Symbols ............................................................................................................. 3
Instrument Familiarization................................................................................. 4
Product Connections.......................................................................................... 6
Connect Infusion Devices.............................................................................. 6
Connect Drains to the Product....................................................................... 7
Connect Accessories.......................................................................................... 7
Keyboard ....................................................................................................... 8
Bar Code Reader............................................................................................ 8
Printer............................................................................................................ 8
Product Operation.............................................................................................. 8
Preferences .................................................................................................... 9
How to Test Infusion Devices....................................................................... 10
Flow Tests.......................................................................................................... 13
Occlusion Tests.................................................................................................. 17
How to Test PCA Pumps................................................................................... 20
Dual Flow Pump Test........................................................................................ 22
Templates........................................................................................................... 22
Define a Template ......................................................................................... 23
Template Operation....................................................................................... 24
Utilities .............................................................................................................. 25
Recall Tests ................................................................................................... 25
Set Clock................................................................................................... 26
User Preferences........................................................................................ 26
Printer Setup/Test...................................................................................... 26
Cancel Print............................................................................................... 26
Instructions.................................................................................................... 26
Report Header ........................................................................................... 26
Default Test Preferences ........................................................................... 26
Calibration..................................................................................................... 27
Edit Templates............................................................................................... 27
Troubleshooting................................................................................................. 28

IDA-5
Users Manual
ii
Product Maintenance ......................................................................................... 29
Clean the Product .......................................................................................... 29
Outside ...................................................................................................... 29
Inside......................................................................................................... 29
Test Fluid....................................................................................................... 29
Storage........................................................................................................... 30
Shipping......................................................................................................... 30
Specifications..................................................................................................... 30
General Specifications................................................................................... 30
Performance Specifications........................................................................... 30
Other Specification........................................................................................ 31
Appendices
Remote Operation..............................................................................................A-1

Contents (continued)
iii


v
List of Tables
Table Title Page
1. Symbols.................................................................................................................. 3
2. Front-Panel Controls and Connections .................................................................. 4
3. Rear-Panel Controls and Connections.................................................................... 5
4. Channel Setup Screen Controls.............................................................................. 10
5. Device Information Fields...................................................................................... 11
6. Device Information Controls.................................................................................. 12
7. Device Detail Information Fields........................................................................... 12
8. Flow Screen Menu Controls................................................................................... 14
9. Channel Flow Measurement Parameters................................................................ 14
10. Flow Graph Screen Controls.................................................................................. 15
11. Flow Screen in End Mode...................................................................................... 16
12. Occlusion Screen in Start Mode............................................................................. 17
13. Occlusion Screen in Active Mode.......................................................................... 18
14. Occlusion Screen in End Mode.............................................................................. 19
15. PCA/Dual Information Screen............................................................................... 20
16. PCA Screen in Active Mode.................................................................................. 21
17. Template Detail Screen.......................................................................................... 23
18. Recall Tests Screen ................................................................................................ 25
19. Bubble Errors......................................................................................................... 28
20. Air Lock Errors...................................................................................................... 28

IDA-5
Users Manual
vi

vii
List of Figures
Figure Title Page
1. Infusion Device Connections to the Product.......................................................... 6
2. Drain Connections to the Product .......................................................................... 7
3. Status All Channels Screen.................................................................................... 8
4. Utilities Screen....................................................................................................... 9
5. Channel Flow Screen ............................................................................................. 13
6. Channel Flow Graph .............................................................................................. 15
7. Occlusion Graph Screen......................................................................................... 18
8. PCA Screen - Prime Mode..................................................................................... 21
9. PCA Graph Screen................................................................................................. 22
10. Utilities Menu Screen............................................................................................. 25
11. Default Test Preferences Screen ............................................................................ 26

IDA-5
Users Manual
viii

1
Introduction
The Fluke Biomedical IDA-5 Infusion Device Analyzer (the Product) is a precision
instrument that examines the performance of medical infusion devices. The Product
measures the flow rate and volume supplied, and the pressure generated in occlusion or
blockages of the fluid line. A maximum of 4 infusion devices can be independently
examined with the four-channel version of the Product.
Intended Use
The Product is to be used by infusion device manufacturers, hospital biomedical
engineering departments, and third-party service organizations. Use the Product to verify
accurate performance of infusion devices through measurement of flow, volume, and
pressure. The performance of a wide range of infusion devices can be analyzed including
syringe, drop counting, peristaltic, and volumetric types. Non-steady flow rate pumps can
also be analyzed. The Product uses distilled or deionized water with an optional wetting
agent only.
Unpack the Product
Carefully unpack all items from the box and check that these items are included:
•The Product
•Power Cord
•Accessory Set (syringe, stopcocks, drain tubing, and Micro 90)
•CD (contains Users Manual and Hydrograph software)
•USB Cable

IDA-5
Users Manual
2
Safety Information
A Warning identifies hazardous conditions and actions that could cause bodily harm or
death. A Caution identifies conditions and actions that could harm the Product, the
equipment under test, or cause permanent loss of data.
Warning
To prevent possible electrical shock, fire, or personal injury:
•Read all safety Information before you use the Product.
•Read the User Manual before you operate the Product.
•Use the product only as specified, or the protection supplied
by the product can be compromised.
•Do not use the Product if it operates incorrectly.
•Do not use an incompatible power receptacle. The Product
power cord must be connected to a power receptacle that
provides voltage and current within the specified rating for
the system.
•Connect an approved three-conductor mains power cord to
a grounded power outlet.
•Never use a two-prong plug adapter to connect primary
power to the Product.
•Turn the Product off and remove the mains power cord
before cleaning the outer surface of the Product.
•Do not open the Product unless you are qualified.
•Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
•Do not use the Product on infusion devices that are
attached to patients.
•Do not reuse test tubing or syringes for patient infusion
•Avoid possible contamination of reusable components due
to backflow conditions. Some older style infusion devices
may have reusable components that could come in direct
contact with the fluids being pumped. When testing these
types of devices take care to avoid possible contamination
of reusable components.
•Do not use delivery set or components that have been used
for testing for patient infusion.

Infusion Device Analyzer
Symbols
3
Caution
To prevent possible damage to the product or to equipment
under test:
•Only qualified service personnel should service the Product.
•Only qualified technical personnel should perform
troubleshooting and service procedures on internal
components.
•Only use degassed de-ionized water with the Product.
Wetting agent may be added.
•Do not use high-viscosity fluids. Oils (solvents, or strong
chemicals) may also damage or contaminate the Product.
•Do not use "Bleach" sterilizing agents or alcohols.
•Do not rapidly switch the Product On or Off, nor remove the
line cord while energized.
•Remove internal water before shipping or storing. Do not
use compressed air to clean out the Product.
•Do not expose the Product to temperature extremes. For
proper operation, ambient temperatures should be from
15 °C to 30 °C (59 °F to 86 °F). Performance may be
adversely affected if temperatures fluctuate above or below
this range. For Storage Temperature limits, see the
Specifications section.
•Do not use the Product in close proximity to sources of
strong electromagnetic radiation (e.g., unshielded
intentional RF sources). These sources may interfere with
proper operation.
Symbols
Table 1. Symbols
Symbol Description Symbol Description
Risk of Danger. Important
information. See Manual. Hazardous voltage. Risk of electric shock.
Conforms to European Union
directives. CAT II
Measurement category II is applicable to
test and measuring circuits connected
directly to utilization points of low voltage
mains installation.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The
affixed label indicates that you must not discard this electrical/electronic product in domestic
household waste. Product Category: With reference to the equipment types in the WEEE
Directive Annex I, this product is classed as category 9 "Monitoring and Control
Instrumentation” product. Do not dispose of this product as unsorted municipal waste. Go to
Fluke’s website for recycling information.

IDA-5
Users Manual
4
Instrument Familiarization
Tables 2 and tell you about the controls and connections on the front and rear panels of
the Product.
Table 2. Front-Panel Controls and Connections
1
2
4
3
5
6
gir01.eps
Item Description
Display (LCD)
Power on indicator
ESC (escape) button – Moves back one step or does the operation given on the display.
ENTER button – Operates the highlighted function or moves to the subsequent data-entry field.
Arrow buttons – Moves the highlight on menus in the direction of the arrow or operates the
function shown on the display.
Flow inlet ports – One port for each measurement channel.

Infusion Device Analyzer
Instrument Familiarization
5
Table 3. Rear-Panel Controls and Connections
2
1
3
4
5
6
7
gir02.eps
Item Description
Handle
Power switch
Power inlet
Equipotential post
USB “B” connector – Computer connection.
USB “A” connectors – Connect a maximum of four accessories such as:
Keyboard
Printer
Bar-Code reader
Fluid outlets – One per measurement channel. Channel 1 at right and channel 4 at left.

IDA-5
Users Manual
6
Product Connections
The Product connects to infusion devices through the front-panel inlet ports. Fluid drain
hoses and accessories connections are made through the rear-panel connections.
Connect Infusion Devices
As shown in Figure 1, it is recommended that all infusion device connections be made to
the Product front-panel inlet connectors through 3-way stopcocks.
20 ml Syringe
to aid priming
3-way Stopcock
From Infusion
Pumps
gir03.eps
Figure 1. Infusion Device Connections to the Product
The channel 4 inlet shown in Figure 1, shows a 20 ml syringe attached to one 3-way
stopcock inlet. The syringe can be used to help priming. It can be used as shown or can
be connected further away from the inlet to help for flow tests. The syringe can be shared
among the channels and removed after the channel is primed.
Follow these recommendations when you connect to the inlet tubing circuits:
•Use adequate prime volumes (for example, 10 ml) to push through any bubbles.
•Use the stopcocks at the inlet to prevent fluid backflow out of inlets between tests.
•When you connect to the inlet circuits (for example, when you attach the priming
syringes to the stopcocks) make sure no new bubbles are introduced.

Infusion Device Analyzer
Connect Accessories
7
Caution
Do not use delivery set or components that have been used for
prior testing for patient infusion.
Note
Before you use the delivery set (tubing, syringe, etc.), make sure it is within
the specified use period of the manufacturer. Many sets are made to be used
only once.
Connect Drains to the Product
Figure 2 shows tubing connected to the rear-panel outlets of the Product.
gir04.eps
Figure 2. Drain Connections to the Product
When you connect drain tubing to the Product outlets:
•Connect different drain tubes to each channel.
•Do not connect the drain tubes together.
•The drain tubes should not be allowed to rise more than 10 cm (4 in) at any point
above the height of the inlet ports of the Product.
•The discharge end of the drain tubes must not be more than 10 cm (4 in) below the
bottom of the Product.
Connect Accessories
Accessories connect to any of the four USB “A” ports on the rear-panel of the Product.
Use a USB cable that is less than 3 meters long.
Note
When an accessory is connected to a Product that is ON, some seconds are
necessary before the accessory is recognized.

IDA-5
Users Manual
8
Keyboard
It is recommended that a small footprint USB keyboard be used with the Product. The
keyboard is necessary to record data about the infusion device under test.
Note
The keyboard must not have an internal USB hub (for example, no extra
USB ports).
Bar Code Reader
A bar code reader can also be used to scan infusion device data into the Product.
Printer
A printer that supports PCL-5 (or higher) printer-command language can be used with the
Product to print test results and reports.
Product Operation
Control the Product with the four arrow buttons and the ENTER and ESC (Escape)
buttons on the front panel:
•Four Arrows move the highlight between menu options.
•ENTER operates the highlighted option.
•ESC functionality is described on the bottom of each screen.
Note
The Arrow, Enter, and Esc buttons of a connected keyboard replicate the
front-panel buttons.
The Product power switch is on the rear panel. The Status All Channels screen in Figure
3 is shown when the Product is prepared to operate and is the “home” screen.
Status All Channels
Channel 1 Channel 2 Channel 3 Channel 4
00: 00:00 00: 00:00 00: 00:00 00: 00:00
SETUP SETUP SETUP SETUP
FLOW FLOW FLOW FLOW
OCCL OCCL OCCL OCCL
Press ESC for Utilities
gir05.eps
Figure 3. Status All Channels Screen
Other manuals for IDA-5
1
Table of contents
Other Fluke Measuring Instrument manuals

Fluke
Fluke VIBXPERT II User manual

Fluke
Fluke OptiFiber PRO OTDR User manual

Fluke
Fluke 789 Use and care manual

Fluke
Fluke 05-450 User manual

Fluke
Fluke 1736 User manual

Fluke
Fluke 377 User manual

Fluke
Fluke 902 User manual

Fluke
Fluke Hart Scientific 914 Series User manual

Fluke
Fluke 922 User manual

Fluke
Fluke 1653B User manual