Fluke 80T-150 User manual

LL
(6)
Instruction
Manual
Model
80T-150
Temperature
Probe


INTRODUCTION
The
Model
80T-150
Temperature
Probe
is
a
self-contain-
ed
temperature-to-voltage
converter.
И
is
designed
to
provide
a
direct
temperature
reading
when
connected
to
any
high
impedance
voltmeter
that
is
capable
of
1
mV
resolution,
and
at
least
a
300
count
full-scale
readout
capability.
Two
temperature
ranges
are
provided,
and
either
can
be
jumper
selected;
-58
to
+302°F
or
-50
to
+150°C.
The
unit
is
housed
in
two
separate
assemblies;
atempera-
ture
probe
and
a
temperature-to-voltage
converter.
The
probe
contains
the
temperature
sensing
element
and
is
electrically
connected
to
the
temperature-to-voltage
converter
through
a
46-inch
shielded
cable.
Two
banana
plugs
with
standard
0.75
inch
spacing
are
provided
for
connecting
the
80T-150
to
the
voltmeter.
Operating
power
for
the
80T-150
15
derived
from
two
internal
batteries.
Typically,
they
will
provide
up
to
1000
hours
of
continuous
operation
before
replacement
is
necessary.
An
ON/OFF
switch
is
provided
on
the
temperature-to-voltage
converter
to
allow
battery
conservation
when
the
unit
is
not
in
use.
Temperature
is
measured
by
exposing
the
probe
tip
directly
to
the
material
to
be
measured
(liquid,
gas
or
solid).
A
direct
temperature
reading
is
displayed
on
the
voltmeter.
SPECIFICATIONS
The
80T-150
will
achieve
rated
accuracy
when
used
with
any
0.25%
DVM
which
has
an
input
impedance
of
>
1
MO.
Specifications
for
the
80T-150
are
given
in
Table
1.
80T-150
OPERATING
NOTES
The
following
paragraphs
are
intended
to
familiarize
the
operator
with
the
80T-150,
as
shown
in
Figure
1.
They
should
be
read
before
attempting
to
operate
the
probe.
Probe
Limitations
The
80T-150
probe
is
constructed
of
a
highly
durable
plastic
and
is
suitable
for
measuring
the
temperature
of
liquids,
gases
and
solid
surfaces
up
to
150°C.
When
measuring
temperature,
observe
the
following
pre-
cautions
to
prevent
damage
to
the
probe:
а
Do
not
expose
the
probe
end
(probe
tip
plus
~2
inches
of
probe
body)
to
temperatures
in
excess
of
+150
С.
The remainder
of
the
probe
body
should
not
be
exposed
to
temperatures
above
+70
С.
b.
For
liquid
measurements,
recommended
applica-
tions
range
from
water,
lubricants
and
fuels
to
most
solvents.
Liquids
as
shallow
as
4”
can
be
measured
since
the
temperature
sensor
is
in
the
probe
tip.
CAUTION
Long
term
exposure
of
the
probe
to
corrosive
environments
will
result
in
pitting
and
deteri-
oration
of
the
aluminum
probe
tip.
Temperature
Scale
The
80T-150
is
capable
of
providing
temperature
mea-
surements
in
°C
or
°F.
See
decal
opposite
power
switch
for
assigned
temperature
scale.
Instructions
for
changing
temperature
scale
are
given
later
under
General
Mainten-
ance.
Table
1.
Specifications
ELECTRICAL
Range
(field
selectable.
. .
50°С
to
+150°C
or
—58
F
by
internal
jumpers).
to
+302°F
Accuracy
1159106
+35°C
7
Best)
їн
М
ambient
+1
C
MS
FIO
C
ton
OLG,
decreasing
linearly
to
+
oe
(5.4°F)
at
-50°C
and
+150°C
0°C
to
+15°C
and
+35°C
to
+50°
ambient
Add
1°C
(1.8°F)
to
above
Sensitivity
(807—150
output)
1
mV
de/
°C
or
1
mV
de/
°F
350V
dc
or
peak
ac
5.5
seconds
to
settle
within
2°
Voltage
Standoff
Settling
Time
for
a
50°
change.
Overall
Length
Power.
. .
Two
internal
disposable
batteries
(1.5V,
Size
№”)
Output
Termination....
GENERAL
100
gm
(3.5
oz.)
1.33
meters
(52.5
in.)
1000
hours
continuous
use
Standard
0.75“
spaced
double
banana
plug
ENVIRONMENTAL
Entire
Unit
Storage
Operating
Altitude
<50000
ft.
<10000
ft.
—40
to
+70°C
О
to
+50°C
80%
non-
condensing
Temperature
Humidity
Application
Force
20
pounds
max.
(Probe
tip-to-measurements
surface)

80T-150
TEMPERATURE-
TO-VOLTAGE
CONVERTER
POWER
SWITCH
Va
ON/OFF
BANANA
PLUGS
(0.75
”
SPACING)
PROBE
TIP
ЕЕ]
PROBE
CABLE
PROBE
BODY
Figure
1.
80T-150
Features
Error
Sources
When
the
probe
пр
15
applied
to
a
solid
surface
it
draws
or
sinks
heat
from
the
surface.
Therefore,
if
the
measured
surface
has
a
low
mass
(е.р.,
a
transistor
case),
the
in-
dicated
temperature
may
be
lower
than
the
actual
temperature.
Similarly,
a
steady-state
error
or
gradient
exists
between
the
measured
surface
and
the
sensing
device
in
the
probe
tip.
This
15
due
to
the
flow
of
heat
from
the
measurement
surface
to
the
probe
body.
The
effect
of
the
steady-state
error
increases
as
the
differential
between
ambient
and
surface
temperature
increases.
To
determine
the
actual
surface
temperature
of
a
device,
both
the
heat-sinking
and
steady-state
errors
must
be
considered.
The
correction
curve
given
in
Figure
2
approximates
the
effect
of
both
error
sources
on
TO-3,
TO-5
and
TO-18
transistor
cases.
son
HH
naa
зи
Й
|
ГТА
ви
ЕД
|
4
Ӯ
|]
um
i
=
`
DEGREES
ABOVE
AMBIENT
(°C)
в
в
u
в
в
y
4”.
r
4
в
[1
в
1
в
в
в
в
Ц
INITIAL
SURFACE
TEMPERATURE
IN
INDICATED
SURFACE
TEMPERATURE
DEGREES
ABOVE
AMBIENT
Figure
2.
Initial
Case
Temperature
Above
Ambient
vs.
Meter
Reading
Above
Ambient
RF
signals
applied
to
the
80T-150
probe
tip
can
also
cause
errors
in
temperature
measurement.
Figure
3
defines
the
rf
signal
limits
that
can
be
tolerated
without
degrading
measurement
accuracy.
AN
m
'
da
ma
ай
ва
ма
|
|
|
aa
Figure
3.
Maximum
Signal
RF
Limits
(Vrms)
at
Probe
Tip
OPERATION
Use
the
following
procedure
to
operate
the
80T-150
probe:
a.
Connect
the
banana
plugs
on
the
80T-150
to
the
input
terminals
of
a
high
impedance
voltmeter.
Observe
polarity.
b.
Select
a
dc
voltage
range
that
will
provide
at
least
1
mV
resolution
(1
mV/degree),
and
a
full
scale
readout
that
will
encompass
the
expected
temperature.
The
2V
range
of
a
3%
digit
DVM
is
adequate.
Ignore
readings
of
less
than
1°
when
a
more
sensitive
DVM
is
used.
с:
Set
the
80T-150
power
switch
to
ON
and
energize
the
voltmeter.
d.
Firmly
touch
the
probe
tip
to
the
surface
to
Бе
measured,
or
expose
it
to
a
liquid
or
gas.
The
volt-
meter
will
display
the
temperature
in
degrees.
Vary
probe
angle
and
pressure
when
measuring
solid
surface
temperatures;
the
highest
stabilized
reading
will
be
the
most
accurate.
CAUTION
The
force
exerted
on
the
probe
tip
should
not
exceed
20
pounds.
THEORY
OF
OPERATION
The
Model
80T-150
utilizes
the
negative
temperature
coefficient
of
a
semiconductor
(P-N)
junction
to
measure
temperature.
The
PN
junction
is
thermally
integrated

into
the
probe
tip
and
comprises
one
leg
of
a
bridge
circuit
as
shown
in
simplified
circuit
diagram
section
of
Figure
4.
Two
1.5V
batteries
are
used
to
power
both
the
bridge
circuit
and
operational
amplifier
U2.
Since
the
bridge
must
be
balanced
to
provide
0°C
and
0°
Е
indica-
tions,
separate
range
or
temperature
scale
resistors
RA
and
RB
are
included
in
the
bridge
circuit.
When
RB
is
shorted,
the
°C
scale
is
selected
and
the
bridge
is
calibrat-
ed
by
R2
to
null
at
0°C.
Conversely,
when
КВ
is
open,
the
02
Е
scale
is
selected
and
the
bridge
is
calibrated
by
R2
to
null
at
0"
Е.
Deviations
above
and
below
0°
provide
a
bridge
output
of
approximately
2.45
mV/*C.
Figure
4.
Simplified
Circuit
Diagram
Operational
amplifier
U2
is
used
to
measure
the
bridge
output
and
scale
itto
a
|
mV/degree
signal.
Since
the
°C
and
the
°F
scale
are
sloped
differently,
the
scale
for
U2
must
be
matched
with
the
scale
selected
for
the
bridge
circuit.
Shorting
resistor
RD
selects
the
°C
scale.
Con-
versely,
when
КО
is
open
the
°
F
scale
is
selected.
Resistor
К4
calibrates
both
scales.
The
output
voltage
used
to
drive
the
external
voltmeter
is
taken
from
the
output
of
U2(P2)
and
the
reference
side
of
the
bridge
(P1).
Since
U2
is
operating
as
an
inverting
amplifier,
its
output
is
used
as
the
low
input
to
the
volt-
meter.
This
enables
the
voltmeter
to
display
an
increase
in
temperature
as
an
increase
in
voltage.
GENERAL
MAINTENANCE
Access
Information
The
battery,
the
temperature
scale
jumpers
and
the
calibration
pots
are
located
on
the
interior
of
the
temp-
80T-150
erature-to-voltage
converter
assembly.
Access
to
these
locations
is
accomplished
by
removing
the
three
screws
from
the
bottom
side
of
the
assembly
and
removing
the
plastic
box.
Battery
Replacement
The
807-150
uses
two
1.5
volt,
size
N,
alkaline
batteries.
They
provide
a
useful
operating
life
of
=1000
hours.
To
replace
the
batteries
pull
the
old
ones
out
and
press
the
new
ones
into
the
battery
holders.
Match
polarity
mark-
ed
on
the
pcb.
Mercury
or
equivalent
N-size
batteries
may
be
installed
in
the
80T-150.
Temperature
Scale
Jumpers,
°C
or
°F
The
jumper
locations
for
selecting
either
the
°C
or
°F
temperature
scale
are
shown
in
Figure
5.
The
absence
of
jumpers
at
both
points
indicates
selection
of
the
°F
scale.
The
presence
of
solder
bridges
at
both
points
indicates
selection
of
the
°C
scale.
Under
no
circumstances
should
one
point
be
open
while
the
other
is
bridged.
TEMPERATURE
SCALE
JUMPERS
80T-3001
Figure
5.
Temperature
Scale
Jumpers
PERFORMANCE
TEST
Complete
the
calibration
procedure
without
opening,
the
temperature-to-voltage
converter
assembly
and
with-
out
making
any
calibration
adjustments.
Observe
the
readings
given
in
[brackets].
Other
readings
are
for
cal-
ibration
only.
CALIBRATION
A
calibration
cycle
of
|
year
is
recommended
to
maintain
the
unit
within
the
specifications
given
earlier.
The
equip-
ment
required
for
calibration
is
listed
in
Table
2.
NOTE
Values
given
in
brackets
apply
to
the
Per-
formance
Test.

80T-150
a.
Access
the
interior
of
the
temperature-to-voltage
NOTE
converter.
If
the
unit
is
calibrated
on
the?
Е
scale,
inter-
action
will
occur
between
adjustments
R2
and
R4.
Repeat
steps
d,
e,
f
and
g
until
read-
ings
are
obtainable
without
adjustment.
b.
Determine
the
selected
temperature
(°C
or
°F).
See
marking
on
bottom
cover
decal.
c.
Connect
the
80T-150
to
a
DVM
and
select
the
Table
2.
Test
Equipment
Requirements
100
mV
ас
range
(10
uV
resolution).
Instrument
Minimum
Use
Recommended
Type
Specifications
Model
а.
Turn
the
80T-150
power
switch
to
ON
and
immerse
Mercury
0.1°C
i
Princo
Model
the
probe
tip
=2
inches
into
a
thermometer
moni-
Thermometer
Resolution
SAMA-CP45
tored
0°C
bath.
Allow
the
DVM
reading
to
stabil-
Dewar
Flask
1-pint
capacity
Thermos
ize
(2230
seconds).
and
Cap
(for
¡ce
bath)
Bottle
Metal
or
1-pint
capacity
Suitable
for
y
e
4
я
Glass
Container
boiling
water
e.
Adjust
R2
(see
Figure
6.)
to
obtain
the
appropriate
S
DVM
reading:
Digital
Volt
100
mV
range
Fluke
Model
У
Meter
with
10
uV
8600A
°C
0.00#0.05
mV
de
[0+2
mV
dc]
2E
320501
мас
[32:04
то
М
de]
If
necessary,
additional
cw
range
can
be
added
to
R2
by
adding
a
solder
bridge
across
E8
and
E9
as
shown
in
Figure
5.
Resolution
1000
mV
range
with
100
uV
Resolution
PROBE
REPLACEMENT
f.
Immerse
the
probe
tip
approximately
2
inches
into
a
70
to
100°C
bath
and
allow
the
DVM
reading
to
A
probe
kit
(80T-150-7001k)
is
available
for
replacing
stabilize.
Change
DVM
range,
if
necessary,
for
damaged
or
defective
probes.
The
kit
includes
a
probe
°F
calibration.
and
cable
assembly,
four
metal
film
resistors
(R3),
and
an
installation
and
calibration instruction
sheet.
р.
Adjust
R4
to
obtain
а
DVM
reading
which
agrees
LIST
OF
REPLACEABLE
PARTS
with
the
bath
temperature
(BT)
as
monitored
by
a
mercury
thermometer.
Component
location
drawings
and
a
list
of
replaceable
ЗОЎ
BA<E0.05
mV
дет,
[BTÆ2
mMmVide]
parts
are
shown
in
Figure
6.
When
ordering
parts,
рго-
Cy
BTEO
mV
de®
ВУСА
ОУ
de]
vide
description,
Fluke
ра
and
quantity
required.

NOTES:
Sensing
Probe
and
Cable
Assembly,
80T-150-7001
K
(includes
selected
resistor
R3)
Temperature-to-Voltage
Converter
PCB
Assembly
(order
80T-150)
Battery,
Alkaline,
1.5V,
size
N
Socket,
1-pin
Plugs,
banana
Resistor,
selected
(supplied
with
A1
Assy)
Button,
switch,
green
(not
shown)
Housing,
bottom
(not
shown)
Housing,
top
(not
shown)
Shock
absorbers
(not
shown)
Decal,
bottom
(not
shown)
Decal,
top
(not
shown)
Screw
(not
shown)
[7
Part
of
resistor
network
U1
2:
Unless
otherwise
noted,
all
capacitance
is
in
ur
and
all
resistance
is
in
ohms.
All
voltages
measured
with
respect
to
TP1.
431023
478735
343285
352716
432575
527747
527739
428441
534792
534784
183574
80T-150
80T-1601
Temperature-to-Voltage
Converter
PCB
Assembly
R3
SELECT
J2
Cap,
disc,
0.01
uF
+20%,
100V
ас
Cap,
Ta,
2.2
uF
+20%,
20V
de
Res,
metal
film,
1.62k
+1%,
1/8W
Res,
variable,
2k,
1/2W
See
Final
Assembly
Res,
variable,
50k
1/2W
Switch,
slide
Resistor
Network
Resistor
Network
IC,
linear
op
amp
Diode,
zener,
1.22V
de
Contact,
Battery
Figure
7.
80T-150
Temperature
Probe
Schematic
5
149153
161927
282697
226076
225417
393629
616615
616656
418913
452771
477463
80T-1201

www.manualsplus.com
P/N
433565
|
June
1976
1020806
Кем
4
8/82
Litho
in
U.S.A.
01982,
John
Fluke
Mfg.
Co.,
Inc.,
all
rights
reserved.
Litho
in
U.S.A.
John
Fluke
Mfg.
Co.,
Inc.
/
P.O.
Box
С9090
/
Everett,
WA
98206
Table of contents
Other Fluke Measuring Instrument manuals

Fluke
Fluke 830 User manual

Fluke
Fluke 381 Use and care manual

Fluke
Fluke 525B User manual

Fluke
Fluke i3000s Flex Series User manual

Fluke
Fluke PLS 6R User manual

Fluke
Fluke 971 User manual

Fluke
Fluke Hart Scientific 914 Series User manual

Fluke
Fluke ScopeMeter 192 Configuration guide

Fluke
Fluke 724 Use and care manual

Fluke
Fluke C9970 User manual