Fluke 8808A User manual

Fluke 8808A
Digital Multimeter
Extended Specifications
Making measurements is
as simple as pushing a
button
The Fluke 8808A 5.5 digit
multimeter has a broad range
of functions, measuring volts,
ohms and amps with a basic
V dc accuracy of 0.01 %. It is
remarkably easy to use, even by
unskilled operators, because it
makes the measurements you
perform most often extremely
easy and fast to do.
Six setup buttons on the
8808A front panel operate like
a car radio’s station presets.
Simply set up the meter for a
common measurement, then
press shift followed by a setup
button (S1 to S6) to save the
setup. Now each time you
perform that measurement, you
simply press the appropriate
setup key. It’s that easy!
The setup buttons eliminate
the need to follow complex work
instruction sheets. Operators no
longer need to press multiple
buttons to set up a measurement
function and range, test limits, or
enter other parameters to make
a measurement.
Eliminate production
mistakes
The Fluke 8808A 5.5 digit
multimeter dependably performs
the most common measurements
required by today’s applications.
Features at a glance
•5.5 digit resolution
•Basic V dc accuracy of 0.01 %
•Dual display
•Dedicated dc leakage current measurement
•2x4 ohms 4-wire measurement technique
•Six dedicated buttons for fast access to instrument setups
•Hi.Lo limit compare for Pass/Fail testing
•Fluke 45 remote command emulation
Whether you are performing
functional tests or making
critical measurements on test
points, using the limit compare
mode with pass/fail indicators
eliminates production mistakes,
especially those where results
are “on the edge.”
The 8808A display has
built-in enunciators that clearly
show the operator whether a
test passes or fails. The pass/fail
indicators take the guesswork
out of testing: the result is either
within limits or it’s out.

2 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
8808A Fluke Corporation 1
8808A Specifications
Voltage
100V Setting .............................................................90 V to 110 V
120V Setting ............................................................108 V to 132 V
220V Setting ............................................................198 V to 242 V
240V Setting ............................................................216 V to 264 V
Frequency..................................................................47 Hz to 440 Hz.
Power Consumption ................................................15 VA peak (10 W average)
Dimensions
Height .........................................................................88 mm (3.46 in)
Width ..........................................................................217 mm (8.56 in)
Depth .........................................................................297 mm (11.7 in)
Weight ........................................................................2.1 kg (4.6 Ibs)
Display
Vacuum Fluorescent Display, segment
Environment
Temperature
Operating................................................................0 rC to 50 rC
Storage....................................................................-40 rC to 70 rC
Warm Up.................................................................½ hour to full uncertainty specifications
Relative Humidity (non-condensing)
Operating................................................................Uncontrolled (< 10°C)
<90 % (10 °C to 30 °C)
<75 % (30 °C to 40 °C)
<45 % (40 °C to 50 °C)
Storage ...................................................................-40 rC to 70 rC <95 %
Altitude
Operating ...............................................................2,000 Meters
Storage .......................................................................12,000 Meters
Vibration....................................................................Complies with MIL-PRF-28800F Class 3
Safety
Complies with IEC 61010-1:2001, ANSI/ISA 61010-1 (S82.02.01):2004, UL 61010-1:2004, CAN/CSA C22.2 No.
61010.1:2004, CAT I 1000V/CAT II 600 V.
EMC
Designed to comply with IEC 61326-1:1997+A1:1998+A2:2000
Triggering
Trigger Delay ...........................................................400 ms
External Trigger Delay ..........................................<2 ms
External Trigger Jitter ...........................................<1 ms
Trigger Input ...........................................................TTL Levels
Trigger Output..........................................................5 V max
Math Functions
Min/max, relative, hold, compare and dB functions.
Electrical
Input Protection ......................................................1000 V all ranges
Overrange..................................................................10 % on the largest ranges of all functions except continuity and diode
test
Remote Interfaces
RS-232C
Warranty
One year

Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 3
2 Fluke Corporation 8808A
Electrical Specifications
Accuracy specifications are valid for 5-½ digit mode and after at least a half-hour warm-up.
DC Voltage Specifications
Maximum Input ........................................................1000 V on any range.
Common Mode Rejection .......................................120 dB at 50 or 60 Hz ധ0.1㧑(1 k˖unbalance)
Normal Mode Rejection..........................................80 dB at Slow Rate
A/D Nonlinearity ......................................................15 ppm of range
Input Bias Current ...................................................<30 pA at 25 rC
Settling Considerations..........................................Measurement settling times are affected by source impedance, cable
dielectric characteristics, and input signal changes
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast Input Impedance
200 mV 199.999 mV 1 µV 10 µV 10 µV >10 G [1]
2 V 1.99999 V 10 µV 100 µV 100 µV >10 G [1]
20 V 19.9999 V 100 µV 1000 µV 1000 µV 10 M ±1 %
200 V 199.999 V 1 mV 10 mV 10 mV 10 M ±1 %
1000 V 1000.00 V 10 mV 100 mV 100 mV 10 M ±1 %
Notes:
[1] At some dual display measurements, the input impedance of 200 mV and 2 V ranges may be changed to 10 M .
Accuracy
Accuracy [1]
90 days 1 year
Range
23 °C ± 5 °C 23 °C ± 5°C
Temperature Coefficient/°C
Outside 18 – 28 °C
200 mV 0.01 + 0.003 0.015 + 0.004 0.0015 + 0.0005
2 V 0.01 + 0.002 0.015 + 0.003 0.001 + 0.0005
20 V 0.01 + 0.003 0.015 + 0.004 0.0020 + 0.0005
200 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
1000 V 0.01 + 0.002 0.015 + 0.003 0.0015 + 0.0005
Notes:
[1] Accuracy given as ± (% of reading + % of range)

4 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
8808A Fluke Corporation 3
AC Voltage Specifications
AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range, and for 50kHz to 100 kHz, add 0.13 % of range.
Maximum Input ........................................................750 V rms or 1000 V peak or 8 x 107Volts-Hertz product
Measurement Method .............................................AC-coupled true-rms. Measures the ac component of input with up to
1000 V dc bias on any range.
AC Filter Bandwidth:
Slow .......................................................................20 Hz – 100 kHz
Common Mode Rejection .......................................60 dB at 50 Hz or 60 Hz (1 k unbalance)
Maximum Crest Factor ..........................................3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) .........Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Input Impedance
200 mV 199.999 mV 1 uV 10 uV 10 uV
2 V 1.99999 V 10 uV 100 uV 100 uV
20 V 19.9999 V 100 uV 1000 uV 1000 uV
200 V 199.999 V 1 mV 10 mV 10 mV
750 V 750.00 V 10 mV 100 mV 100 mV
1 M ±2 % shunted by
<100 pf
Accuracy
Accuracy [1]
90 days 1 year
Range Frequency
23 °C ± 5 °C 23 °C ± 5 °C
Temperature
Coefficient/°C
Outside 18 – 28 °C
200 mV 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
2 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
20 V 20 Hz – 45 Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
200 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
750 V 20 Hz – 45Hz 0.8 + 0.05 0.9 + 0.05 0.01 + 0.005
45 Hz – 20 kHz 0.15 + 0.05 0.2 + 0.05 0.01 + 0.005
20 kHz – 50 kHz 0.3 + 0.05 0.35 + 0.05 0.01 + 0.005
50 kHz – 100 kHz 0.8 + 0.05 0.9 + 0.05 0.05 + 0.01
Notes:
[1] Accuracy given as ± (% of reading + % of range)

Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 5
4 Fluke Corporation 8808A
Resistance
Specifications are for 4-wire resistance function, or 2-wire resistance with REL. If REL is not used, add 0.2 for 2-wire
resistance plus lead resistance.
Measurement Method .............................................Current source referenced to LO input
Max Lead Resistance (4-wire ohms)...................10 % of range per lead for 200 , 2 k ranges. 1 k per lead on all other
ranges.
Input Protection .......................................................1000 V on all ranges
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Current Source
200 199.999 0.001 0.01 0.01 0.8 mA
2 k 1.99999 k 0.01 0.1 0.1 0.8 mA
20 k 19.9999 k 0.1 1 1 0.08 mA
200 k 199.999 k 1 10 10 0.008 mA
2 M 1.99999 M 10 100 100 0.9 µA
20 M 19.9999 M 100 1 k 1 k 0.16 µA
100 M 100.000 M 1 k 10 k 10 k 0.16 µA || 10 M
Accuracy
Accuracy [1]
90 days 1 year
Range
23 °C ± 5 °C 23 °C ± 5 °C
Temperature Coefficient/°C
Outside 18 – 28 °C
200 0.02 + 0.004 0.03 + 0.004 0.003 + 0.0006
2 k 0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
20 k 0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
200 k 0.015 + 0.002 0.02 + 0.003 0.003 + 0.0005
2 M 0.03 + 0.003 0.04 + 0.004 0.004 + 0.0005
20 M 0.2 + 0.003 0.25 + 0.003 0.01 + 0.0005
100 M 1.5 + 0.004 1.75 + 0.004 0.2 + 0.0005
Notes:
[1] Accuracy given as ± (% of reading + % of range)
DC Current
Input Protection ......................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses.
Shunt Resistance .....................................................0.01 for 2 A and 10 A ranges
1 for 20 mA and 200 mA
Burden voltage < 1 mV for 200 uA and 2 mA range.
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Burden Voltage
200 uA 199.999 µA 0.001 µA 0.01 µA 0.01 µA <1 mV
2 mA 1999.99 µA 0.01 µA 0.1 µA 0.1 µA <1 mV
20 mA 19.9999 mA 0.1 µA 1 µA 1 µA <0.05 V
200 mA 199.999 mA 1 µA 10 µA 10 µA <0.5 V
2 A 1.99999 A 10 µA 100 µA 100 µA <0.1 V
10 A 10.0000 A 100 µA 1 mA 1 mA <0.5 V

6 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
8808A Fluke Corporation 5
Accuracy
Accuracy [1]
90 days 1 year
Range
23 °C ± 5 °C 23 °C ± 5 °C
Temperature Coefficient/°C
Outside 18 – 28 °C
200 µA 0.02 + 0.005 0.03 + 0.005 0.003 + 0.001
2 mA 0.015 + 0.005 0.02 + 0.005 0.002 + 0.001
20 mA 0.03 + 0.02 0.04 + 0.02 0.005 + 0.001
200 mA 0.02 + 0.005 0.03 + 0.008 0.005 + 0.001
2 A 0.05 + 0.02 0.08 + 0.02 0.008 + 0.001
10 A 0.18 + 0.01 0.2 + 0.01 0.008 + 0.001
Notes:
[1] Accuracy given as ± (% of reading + % of range)
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs
from 1 % to 5 % of range, add an additional error of 0.1 % of range.
Input Protection .......................................................Tool accessible 11 A / 1000 V and 440 mA / 1000 V fuses
Measurement Method .............................................AC-coupled True RMS
Shunt Resistance .....................................................0.01 for 2 A and 10 A ranges
1 for 20 mA and 200 mA
AC Filter Bandwidth:
Slow ........................................................................20 Hz – 100 kHz
Maximum Crest Factor............................................3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz) .........Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale
Input Characteristics
Resolution
Range Full-Scale
(5-1/2 Digits) Slow Medium Fast
Burden Voltage
20 mA 19.9999 mA 0.1 µA 1 µA 1 µA <0.05 V
200 mA 199.999 mA 1 µA 10 µA 10 µA <0.5 V
2 A 1.99999 A 10 µA 100 µA 100 µA <0.1 V
10 A 10.0000 A 100 µA 1 mA 1 mA <0.5 V
Accuracy
Accuracy [1]
90 days 1 year
Range Frequency
23 °C ± 5 °C 23 °C ± 5 °C
Temperature
Coefficient/°C
Outside 18 – 28 °C
20 mA 20 Hz - 45Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
200 mA 20 Hz - 45Hz 0.8 + 0.05 1 + 0. 06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
2 A 20 Hz - 45Hz 1 + 0.05 1.25 + 0.06 0.015 + 0.005
45 Hz - 2 kHz 0.25 + 0.05 0.3 + 0.06 0.015 + 0.005
10 A 20 Hz - 45Hz 1 + 0.1 1.25 + 0.12 0.015 + 0.005
45 Hz - 2 kHz 0.35 + 0.1 0.5 + 0.12 0.015 + 0.005
Notes:
[1] Accuracy given as ± (% of reading + % of range)

Fluke 8808A Digital Multimeter Extended Specifications Fluke Corporation 7
6 Fluke Corporation 8808A
Frequency
Gate Time...................................................................131 ms
Measurement Method .............................................AC-coupled input using the ac voltage measurement function.
Settling Considerations..........................................When measuring frequency after a dc offset voltage change, errors
may occur. For the most accurate measurement, wait up to 1 second to
allow input blocking RC time constant to settle.
Measurement Considerations ...............................To minimize measurement errors, shield inputs from external noise
when measuring low voltage, low frequency signals.
Accuracy
Accuracy
90 days 1 year
Range Frequency
23 °C ± 5 °C 23 °C ± 5 °C
Temperature
Coefficient/°C
Outside 18 – 28 °C
20 Hz – 2 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
2 kHz – 20 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
20 kHz – 200 kHz 0.01 + 0.002 0.01 + 0.003 0.002 + 0.001
100 mV to
750 V[1,2]
200 kHz – 1 MHz 0.01 + 0.004 0.01 + 0.006 0.002 + 0.002
Notes:
[1] Input > 100 mV
[2] Limited to 8* 107 V Hz
Continuity
Continuity Threshold ..............................................20
Test Currents ............................................................1 mA
Response Time ........................................................100 samples/sec with audible tone
Rate…….…………………………………………… Fast
Maxiumum Reading………………………………199.99
Resolution……….….. ……………….……………0.01
Diode Test
Response Time.........................................................100 samples/sec with audible tone
Rate………………………………………………….Fast
Maxiumum Reading………………………..….…1.9999 V
Resolution….………..….. ………………………..0.1 mV

Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2007 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 9/2007 3093965 D-EN-N Rev B
Pub_ID: 11288-eng Rev 01
Fluke. Keeping your world
up and running.®
Ordering information
Models Description
8808A 120V 5.5 Digit Multimeter
8808A 220V 5.5 Digit Multimeter
8808A 100V 5.5 Digit Multimeter
8808A 240V 5.5 Digit Multimeter
8808A/SU includes
8808A package plus, FlukeView Forms basic software, USB to RS-232
interface adapter cable.
8808A/SU 120V 5.5 Digit Multimeter, SW USB Cable Kit
8808A/SU 220V 5.5 Digit Multimeter, SW USB Cable Kit
8808A/SU 100V 5.5 Digit Multimeter, SW USB Cable Kit
8808A/SU 240V 5.5 Digit Multimeter, SW USB Cable Kit
8808A includes
Meter, TL71 test leads, line cord, spare line fuse, statement of cal prac-
tices, WEEE information sheet, Warranty statement, Getting Started
guide (English, French, German, Spanish, Italian, Simplified Chinese,
Japanese), CD Rom with user manual (English).
8 Fluke Corporation Fluke 8808A Digital Multimeter Extended Specifications
Other manuals for 8808A
6
Table of contents
Other Fluke Multimeter manuals

Fluke
Fluke 77 User manual

Fluke
Fluke 1550B User manual

Fluke
Fluke 8050A User manual

Fluke
Fluke 287 Installation and operating instructions

Fluke
Fluke 8022A User manual

Fluke
Fluke 80 Series III User manual

Fluke
Fluke 83 User manual

Fluke
Fluke 12B/18 User manual

Fluke
Fluke 8010A User manual

Fluke
Fluke 8508A User manual

Fluke
Fluke 15B MAX Use and care manual

Fluke
Fluke 365 Use and care manual

Fluke
Fluke 1508 User manual

Fluke
Fluke 80 Series III User manual

Fluke
Fluke 77 Series III User manual

Fluke
Fluke 27-II User manual

Fluke
Fluke 87 III Series Operating and installation instructions

Fluke
Fluke LinkRunner Pro User manual

Fluke
Fluke 8508A User manual

Fluke
Fluke 175 Owner's manual