manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. GE
  6. •
  7. Measuring Instrument
  8. •
  9. GE 3300 User manual

GE 3300 User manual

Specifications and Ordering Information
Part Number 141500-01
Rev. J 06/07)
Page 1 of 6
3300/20 Dual Thrust Position Monitor
BUFFER ED TR ANSD UCER S
DANGER
A
GAP
ALERT
B
BYPASS
ALERT
DANG ER
A
B
OK
3300/20
DUAL THRUST
D
D
12
4
R
N
T
0
25
20
S
P
A
C
E
M
E
L
15
10
U
N
T
E
C
O
8
5
I
0
R
N
25
20
T
A
10
15
U
E
N
T
O
C
M
E
E
C
0
5
S
L
P
I
MONITOR
vdc
16
N
20
M
S
L
I
/
5
10
15
M
O
R
20
25
24
20
N
10
5
L
/
S
I
M
15
M
R
O
25
Description
The 3300/20 Dual Thrust Position Monitor provides early warning of thrust
bearing failure. It continuously measures and monitors two independent
channels of shaft axial position relative to the axial clearances within the
machine. Ideally, the axial probes are installed to observe the thrust collar
directly, so the measurement represents the position of the collar relative to the
thrust bearing clearance.
Caution
Because thrust measurements are made by observing the gap voltage of the
proximity probe used as an input, a transducer failure gap out of range) can be
interpreted by the monitor as thrust position movement and result in a false
thrust alarm. For this reason, Bently Nevada LLC. does not recommend the use
of a single probe for thrust position applications. Instead, these applications
should use two proximity probes observing the same collar or shaft and
configure the monitor as AND voting whereby both transducers must
simultaneously reach or exceed their alarm setpoints for the monitor’s alarm
relays to actuate. This 2-out-of-2 voting scheme also known as AND voting)
provides optimum protection against both false trips and missed trips. While the
3300/20 monitor can be programmed for either single voting OR) or dual voting
AND), dual voting is strongly recommended for all thrust position applications.
Caution
Probe adjustment and range is critical in this monitor for machinery protection.
Improper adjustment of the transducer may prevent the monitor from alarming
no machinery protection). For proper adjustment, follow instructions in the
manual.
Specifications and Ordering Information
Part Number 141500-01
Rev. J 06/07)
Page 2 of 6
Specifications
Inputs
Signal:
Accepts one or two proximity
probe signal inputs.
Input
Impedance:
10 k Ω.
Sensitivity:
User-programmable for 200
mV/mil 8 V/mm) or 100 mV/mil 4
V/mm).
Power:
Nominal consumption of 1.5
watts.
Signal Conditioning
Accuracy:
Within ±0.33% of full-scale
typical, ±1% maximum. All
specified at temperature of +25°C
+77°F).
Outputs
Recorder:
User-programmable for +4 to +20
mA, 0 to -10 Vdc, or +1 to +5 Vdc.
Voltage or current outputs are
proportional to programmed
monitor full-scale. Independent
recorder outputs are provided for
each channel. Monitor operation
is unaffected by short-circuits on
recorder outputs.
Recorder
Accuracy (in
addition to
signal
conditioning
accuracy) at
+2 °C (+77°F):
• +4 to +20 mA: ±0.7% of
signal, ±0.09 mA offset.
• +1 to + Vdc: ±1.1% of
signal, ±10 mV offset.
• 0 to -10 Vdc: ±1.1% of
signal, ±15 mV offset.
Output
Impedance
(voltage
outputs):
100 Ω. Minimum load resistance
is 10 k Ω.
Voltage
Compliance
(current
outputs):
0 to +12 Vdc range across load.
Load resistance is 0 to 600 Ω
when using +4 to +20 mA option.
Buffered
Transducer
Outputs:
There is one coaxial connector
per channel on the front panel
and one terminal connection per
channel on the rear panel. All are
short-circuit protected.
Output
Impedance:
100 Ω.
Transducer
Supply Voltage:
User-programmable in power
supply for -24 Vdc or -18 Vdc.
Current limited on individual
monitor circuit board.
Note: Contact your nearest sales professional if 3000 series
transducers are to be used in the monitoring system which also
uses 3300 and/or 7200 series transducers.
Alarms
Alarm
Setpoints:
Both alarms Alert and Danger)
are bi-directional normal and
counter), digitally adjustable from
0 to 100% of full-scale and can be
set within LCD resolution ±1.6%)
to desired level. Once set, alarms
are repeatable within 0.39% of
full-scale.
Specifications and Ordering Information
Part Number 141500-01
Rev. J 06/07)
Page 3 of 6
Displays
Meter:
Nonmultiplexing vertical bargraph
type Liquid Crystal Display LCD).
Individual 63 segment LCD per
channel. Probe Gap indicated on
a third, center scale. LCD also
displays error codes and monitor
ADJUST mode.
Resolution:
Within ±1.6% of monitor full-
scale.
Size:
83 mm 3.25 inches), vertical
dimension.
LED Indicators
OK:
One constant ON green LED per
channel indicates OK condition of
monitor, transducers, and field
wiring. Constant OFF indicates
NOT OK condition or Channel
Bypassed red Bypass LED will be
ON). OK LED flashing at 5 Hz
indicates error code s) stored in
memory.
Alarm:
Two red LEDs per channel
indicate alarm status individually
for Alert and Danger). Flashing
alarm LED indicates First Out
independent for Alert and
Danger).
Bypass:
Two red LEDs indicate status of
Danger Bypass and Rack/
Channel Bypass functions
individually per channel).
Environmental Limits
Operating
Temperature:
0°C to +65°C +32°F to +150°F).
Storage
Temperature:
-40°C to +85°C -40°F to +185°F).
Relative
Humidity:
To 95%, noncondensing.
CE Mark Directives
EMC Directive
Certificate of Conformity: 158710
Low Voltage
Directive
Certificate of Conformity: 135300
Hazardous Area Approvals
CSA/NRTL/C
Class I, Div 2
Groups A, B, C, D
T4 @ Ta = +65 °C
Certification
Number
150368 – 1002151 LR 26744)
ATEX
II 3 G
EEx nC[L] IIC
T4 @ Ta = -20°C to +60°C
When installed per document
number 132577-01.
Certification
Number
BN26744C-55A
P ysical
Space
Requirements:
One rack position any position
except 1 and 2, which are
reserved for Power Supply and
System Monitor, respectively).
Weight:
1 kg 2.2 lbs.).
Specifications and Ordering Information
Part Number 141500-01
Rev. J 06/07)
Page 4 of 6
Ordering Information
For spares, order the complete catalog number as
described below. This includes a front panel
assembly, monitor PWAs with sheet metal, and
appropriate relay module. This unit is optioned,
tested and ready to install in your system. Spare
relay modules can be ordered separately.
Dual T rust Position Monitor
3300/20-AXX-BXX-CXX-DXX-EXX
Option Descriptions
A: Full-scale Range Option
0 1 25-0-25 mils
0 2 30-0-30 mils
0 3 40-0-40 mils
0 5 50-0-50 mils
0 6 75-0-75 mils
1 1 0.5-0-0.5 mm
1 2 1.0-0-1.0 mm
1 3 2.0-0-2.0 mm
B: Transducer Input Option
0 1 3300 or 7200 Pro imitor®
systems, 200 mV/mil (Ranges
01, 02, 03, 11, and 12 only.)
0 2 7200 11 mm (not 3300XL)
Pro imitor system, 100 mV/mil
0 3 7200 14 mm or 3300 HTPS
Pro imitor systems, 100
mV/mil
0 4 3000 Pro imitor® 200 mV/mil
(Transducer Output Voltage in
power supply must be set for –
18 Vdc or use power conver-
ter. Ranges 01 and 11 only.)
0 5 3300XL NSv and 3300 RAM
Pro imitor Sensor, 200 mV/mil
(Ranges 01 and 11 only).
Note:
1. For 3300 XL 11mm transducers, use mod 146300-01.
2. Contact your nearest sales professional if 3000 series
transducers are to be used in a monitoring system
which also uses 3300 and/or 7200 Series transducers.
C: Alarm Relay Option
0 0 No Relays
0 1 Epo y-sealed
0 2 Hermetically-sealed
0 3 Quad Relay (Epo y-sealed
only)
0 4 Spare Monitor-No SIM/SIRM
Notes:
1. AND voting logic is not available with Quad Relays.
2. At least one relay module must be ordered with each
3300 System. If one common relay module per system
has been ordered, all monitors of this type must be
jumper programmed at the factory to activate a relay
bus. Order SCK Special Configuration Kit) 157520-119
& -120 for bus one or 157520-121 & -122 for bus two.
3. Agency approval places limitations on the relay
module. Refer to the Relay Module data sheet for
information.
4. Quad Relays are not available with the Internal Safety
Barriers option.
D: Agency Approval Option
0 0 Not required
0 1 CSA/NRTL/C
0 2 ATEX self certification
Note: ATEX approval requires the monitor rack be installed in a
weatherproof housing.
E: Safety Barrier Option
0 0 None
0 1 E ternal
0 2 Internal
Spare Relay Module Assemblies
Order the option in parenthesis for ATEX approved
spares)
81 44-01(02)
No Relays
81 4 -01(02)
Dual Epoxy Relays
81 46-01(02)
Dual Hermetic Relays
841 2-01(02)
Quad Relays
88984-01(04)
Dual Hermetic, Internal Barriers
88984-02(0 )
Dual Epoxy, Internal Barriers
88984-03(06)
No Relay, Internal Barriers
Notes:
1. External Safety Barriers must be ordered separately.
Specifications and Ordering Information
Part Number 141500-01
Rev. J 06/07)
Page 5 of 6
2. Quad Relays are not available with Internal Safety
Barriers option.
Field-programmable Options
These options are field-programmable via plug-in
jumpers.
Bold text indicates options as shipped from the
factory.
First Out Option
Enabled
Disabled
OK Mode Option
Nonlatching
Latching
Alarm Time
Delay Option
0.1 second
1 second
3 seconds
6 seconds
Alert Reset
Mode Option
Latching
Nonlatching
Danger Reset
Mode Option
Latching
Nonlatching
Alert Relay
Mode Option
Normally de-energized
Normally energized
Danger Relay
Mode Option
Normally de-energized
Normally energized
Recorder
Outputs Option
+4 to +20 mA
+1 to + 5 Vdc
0 to -10 Vdc
Danger Relay
Voting Option
AND voting for relay drive
OR voting for relay drive
Note: For Quad Relays, AND voting logic must be done externally
by wiring the contacts in series.
Normal Thrust
Direction
Option:
(programmable
per channel)
Toward probe
Away from probe
Accessories
89634-01
-24V to -18V Proximitor Power
Converter
128112
Galvanic Isolator Kit
0224 002
External Barrier
02200214
Surge Protector
Specifications and Ordering Information
Part Number 141500-01
Rev. J 06/07)
Page 6 of 6
Field wiring diagram
3300/20 Dual Thrust Position Monitor
Bently Nevada and Proximitor are trademarks of General Electric Company.
Copyright 1999. Bently Nevada LLC.
1631 Bently Parkway South, Minden, Nevada USA 89423
Phone: 775.782.3611 Fax: 775.215.2873
www.ge-energy.com/bently
All rights reserved.
OUTPUT
COM
T
-V
COM
CH A
CH B
ALERT
DANGE R
PWR
COM
IN
BUF
REC
NC
ARM
NO
NC
NO
ARM
Channel A
Channel B
To Channel A
Recorder
To Channel B
Recorder Output
Alert
Relay
Danger
Relay
Contacts
Dual Relays Shown
Quad Relays also available
3 Conductor
Proximitor
Channel B
Channel A
Shaft
Thrust Collar
BUFFERED TRANSDUCERS
DANGER
A
A
B
B
GAP
ALERT
BYPASS
ALERT
5
5
L
DANGER
OK
25
0
25
20
T
N
15
E
R
4
10
E
M
E
C
A
O
U
N
T
C
8
20
15
10
/
I
P
S
D
0
12
L
S
5
M
10
16
0
5
10
15
20
20
M
R
N
O
25
vdc
24
M
15
O
N
R
20
25
3300/20
DUAL THRUST
MONITOR
Contacts
Shielded Cable
R
P
A
C
E
M
E
N
T
L
C
T
N
U
O
R
E
M
S
L
I
D
S
/
(To single earth
point ground) 4 pl.
Shield (Green)
Probe
Probe
Output

Other manuals for 3300

2

This manual suits for next models

2

Other GE Measuring Instrument manuals

GE MAC 5 A4 User manual

GE

GE MAC 5 A4 User manual

GE CTF878 Owner's manual

GE

GE CTF878 Owner's manual

GE moisture.IQ User manual

GE

GE moisture.IQ User manual

GE MAC 800 User manual

GE

GE MAC 800 User manual

GE AutoSigma 3000 DL Use and care manual

GE

GE AutoSigma 3000 DL Use and care manual

GE Centricity MP3510 User manual

GE

GE Centricity MP3510 User manual

GE Transport PT878 User manual

GE

GE Transport PT878 User manual

GE MT144 User manual

GE

GE MT144 User manual

GE Druck ADTS 405 User manual

GE

GE Druck ADTS 405 User manual

GE RHM 100 User manual

GE

GE RHM 100 User manual

GE RHM 40 User manual

GE

GE RHM 40 User manual

GE Transport PT878 User manual

GE

GE Transport PT878 User manual

GE Paywise User manual

GE

GE Paywise User manual

GE Ultrasonic USM Go User manual

GE

GE Ultrasonic USM Go User manual

GE MFE DMS 2 User manual

GE

GE MFE DMS 2 User manual

GE DMS Go User manual

GE

GE DMS Go User manual

GE StressTel T-MIKE E User manual

GE

GE StressTel T-MIKE E User manual

GE Sievers 900 Series User manual

GE

GE Sievers 900 Series User manual

GE Baker Hughes PanaFlow LC XMT1000 User manual

GE

GE Baker Hughes PanaFlow LC XMT1000 User manual

GE Druck DPI 620 Genii User manual

GE

GE Druck DPI 620 Genii User manual

GE DigitalFlow DF868 Owner's manual

GE

GE DigitalFlow DF868 Owner's manual

GE RHM 80 User manual

GE

GE RHM 80 User manual

GE Multilin EPM 2200 User manual

GE

GE Multilin EPM 2200 User manual

GE TOP-FREEZER 16 User manual

GE

GE TOP-FREEZER 16 User manual

Popular Measuring Instrument manuals by other brands

NDT FERRITE-CHECK 110 manual

NDT

NDT FERRITE-CHECK 110 manual

Zenner WSD installation instructions

Zenner

Zenner WSD installation instructions

ELGO Electronic MTM-A-000 Series operating manual

ELGO Electronic

ELGO Electronic MTM-A-000 Series operating manual

Hanna Instruments AquaDip instruction manual

Hanna Instruments

Hanna Instruments AquaDip instruction manual

Fluke ProcessMeter 787 Safety information

Fluke

Fluke ProcessMeter 787 Safety information

TSI Incorporated Alnor TH710 Operation and service manual

TSI Incorporated

TSI Incorporated Alnor TH710 Operation and service manual

Rigol DS6000 Series Application note

Rigol

Rigol DS6000 Series Application note

SMART EX500B Installation and operation manual

SMART

SMART EX500B Installation and operation manual

Copper Mountain Technologies R60 Operating and programming manual

Copper Mountain Technologies

Copper Mountain Technologies R60 Operating and programming manual

Monitor Monidrop W operating manual

Monitor

Monitor Monidrop W operating manual

EKF Lactate Scout+ user manual

EKF

EKF Lactate Scout+ user manual

Pyle PTHM20 instruction manual

Pyle

Pyle PTHM20 instruction manual

KWS Electronic VAROS 109 manual

KWS Electronic

KWS Electronic VAROS 109 manual

Blue-White FLOWREAD R-300 quick start guide

Blue-White

Blue-White FLOWREAD R-300 quick start guide

OHAUS STARTER 2100 Service manual

OHAUS

OHAUS STARTER 2100 Service manual

Tonghui U822 Operation manual

Tonghui

Tonghui U822 Operation manual

ASA Electronics ChemScan mini Installation, operation and maintenance manual

ASA Electronics

ASA Electronics ChemScan mini Installation, operation and maintenance manual

Leica Leica UTILIFINDER+ user manual

Leica

Leica Leica UTILIFINDER+ user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.