GE Proteus XR/a User manual

GE Medical Systems
Technical
Publications
Direction 2259724-100
Revision 11
Proteus XR/a
Operator Manual
0459
Copyright © 2000, 2001, 2003, 2004, 2005, 2006
By General Electric Company
Operating Documentation
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
ii
This page intentionally left blank
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
i
IMPORTANT!...X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the
instructions herein contained should be thoroughly read and understood by
everyone who will use the equipment before you attempt to place this equipment
in operation. The General Electric Company, Medical Systems Group, will be glad
to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation
other than the useful beam, no practical design of equipment can provide complete
protection. Nor can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves or
others to radiation.
It is important that everyone having anything to do with x-radiation be properly trained
and fully acquainted with the recommendations of the National Council on Radiation
Protection and Measurements as published in NCRP Reports available from NCRP
Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of
the International Commission on Radiation Protection, and take adequate steps to
protect against injury.
The equipment is sold with the understanding that the General Electric Medical Systems,
its agents, and representatives have no responsibility for injury or damage which may
result from improper use of the equipment.
Various protective material and devices are available. It is urged that such materials or
devices be used.
United States Federal law restricts this device to be used by or on the order of a
physician.
CAUTION
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
ii
All electrical installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. In addition, electrical feeds into the Power
Distribution Unit shall be performed by licensed electrical contractors. Other connections between pieces of
electrical equipment, calibrations, and testing shall be performed by qualified GE Medical personnel. The
products involved (and the accompanying electrical installations) are highly sophisticated, and special
engineering competence is required. In performing all electrical work on these products, GE will use its own
specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements
of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of
third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on
the equipment.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT
If you have any comments, suggestions or corrections to the information in this document,
please write them down, include the document title and document number, and send them to:
GENERAL ELECTRIC MEDICAL SYSTEMS
MANAGER - INFORMATION INTEGRATION
A
MERICAS, X-RAY W-622
P.O. BOX 414
MILWAUKEE, WI 53201-0414
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
iii
REGULATORY REQUIREMENTS
This product complies with the regulatory requirements of the
following:
• Council Directive 93/42/EEC concerning medical devices: the CE label affixed to the
product testifies compliance to the Directive.
The location of the CE label on the product is described page 2-4.
European registered place of business:
GE Medical System Europe
Quality Assurance and Safety Regulatory Manager
BP 34
F 78533 BUC CEDEX FRANCED
Tel: +33 (0) 1 30 70 40 40
• Green QSD 1990 Standard issued by MDD (Medical Devices Directorate, Department
of Health, UK).
• Quality System Regulation issued by the FDA (Food and Drug Administration,
Department of Health, USA).
• Underwriter’s Laboratories, Inc. (UL), an independent testing laboratory.
• Canadian Standards Association (CSA).
• International Electrotechnical Commission (IEC).
The following equipment classifications are applicable to the
product:
• Equipment classification with respect to protection from electric shock: Class 1
• Degree of protection from electric shock: Type B
• Degree of protection against ingress of liquids: not classified
• Equipment not suitable for use in the presence of a flammable anaesthetic mixture with
air or with nitrous oxide; mode of operation: continuous
• Mode of operation: continuous with intermittant loading
• The Proteus XRa has only level 1 EMC susceptability immunity responses.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
iv
This page intentionally left blank
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
v
ELECTROMAGNETIC COMPATIBILITY (EMC)
This product conforms with IEC 60601-1-2:2001 EMC standard for medical devices.
Note:This equipment generates, uses, and can radiate radio frequency energy. The
equipment may cause or subject to radio frequency interference with other
medical and non–medical devices and radio communications. To provide
reasonable protection against such interference, the Proteus XR/a System (32,
50, 65, 80kW) complies with emissions limits for a Group 1, Class A Medical
Devices and has applicable immunity level as stated in EN 60601-1-2:2001.
However, there is no guarantee that interference will not occur in a particular
installation. Special precautions and other information regarding EMC provided
in the accompanying documents of this equipment shall be observed during
installation and operation of this equipment.
Note: If this equipment is found to cause interference (which may be determined by
switching the equipment on and off), the user (or qualified service personnel)
should attempt to correct the problem by one or more of the following
measure(s):
• Reorient or relocate the affected device(s).
• Increase the separating space between the equipment and the affected
device.
• Power the equipment from a source different from that of the affected
device.
• Consult the point of purchase or service representative for further
suggestions.
Use of accessories, transducers, cables and other parts other than those
specified by the manufacturer of this equipment may result in increased
emissions or decreased immunity of the equipment. The manufacturer is not
responsible for any interference caused either by the use of interconnect
cables other than those recommended, or by unauthorized changes or
modifications to this equipment. Unauthorized changes or modifications could
void the user’s authority to operate the equipment.
WARNING
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
vi
ELECTROMAGNETIC COMPATIBILITY (EMC) (CONT.)
Note: To comply with the regulations applicable to an electromagnetic interface for a
Group 1, Class A Medical Device, and to minimize interference risks, the
following requirements shall apply:
• All interconnect cables to peripheral devices must be shielded and
properly grounded. Use of cables not properly shielded and grounded
may result in the equipment causing radio frequency interference in
violation of the European Union Medical Device directive and FCC
regulations.
• All of those recommended guidance regarding electromagnetic
environment should be followed.
Note: Do not use devices that intentionally transmit RF signals (Cellular Phones,
Transceivers, or Radio Controlled Products) in the vicinity of this equipment as it may
cause performance outside the published specifications. Keep the power to these
type devices turned off when near the equipment.
The medical staff in charge of this equipment is required to instruct technicians,
patients, and others.
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The Proteus XR/a system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Proteus XR/a system should assure that it is used in an
electromagnetic environment as described below:
Emissions Test Compliance Electromagnetic Environment
RF Emissions
CISPR11
Group1
The Proteus XR/a system uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF Emissions
CISPR11
Class A
Harmonic emissions
IEC 61000-3-2 Not
applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not
applicable
The Proteus XR/a system is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
vii
ELECTROMAGNETIC COMPATIBILITY (EMC) (CONT.)
Guidance and manufacturer’s declaration - Electromagnetic Immunity (1)
The Proteus XR/a system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Proteus XR/a system should assure that it is used in an electromagnetic
environment as described below:
Immunity Test IEC 60601-1-2
Test Level Compliance
Level Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air ±6 kV contact
±8 kV air Floors are wood, concrete, or ceramic
tile, or floors are covered with synthetic
material and the relative humidity is at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output
lines
±2 kV for
power supply
lines
±1 kV for
input/output
lines
Mains power quality is that of a typical
commercial and/or hospital environment
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV
differential
mode
±2 kV common
mode
Mains power quality is that of a typical
commercial and/or hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5 % UT
(> 95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
< 5 % UT
(> 95 % dip in UT)
for 5 s
0 % UTfor 5
sec
Mains power quality is that of a typical
commercial and/or hospital environment.
If the user of the Proteus XR/a system
requires continued operation during
power mains interruptions, it is
recommended that the Proteus XR/a
system be powered from an
uninterruptible power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m Power frequency magnetic fields are at
levels characteristic of a typical location
in a typical commercial and/or hospital
environment.
Note: These are guidelines. Actual conditions may vary.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
viii
ELECTROMAGNETIC COMPATIBILITY (EMC) (CONT.)
Guidance and manufacturer’s declaration - Electromagnetic Immunity (2)
The Proteus XR/a system is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Proteus XR/a system should assure that it is used in an electromagnetic
environment as described below:
Immunity
Test IEC 60601-1-2
Test Level Compliance
Level Electromagnetic Environment
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V
150 kHz to
80 MHz
3 V/m
80 kHz to
800 MHz
[V1=] 3 V
[E1=] 3 V/m
Portable and mobile RF communications
equipment are used no closer to any part of the
[EQUIPMENT and/or SYSTEM], including cables,
than the recommended separation distance
calculated from the equation appropriate for the
frequency of the transmitter.
Recommended separation distance
d= 1.2
d= 1.2 80 MHz to 800 MHz
d= 2.3 800 MHz to 2,5 GHz
Note: P is the power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,*
are less than the compliance level in each
frequency range.**
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
*Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios,
amateur radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If
the measured field strength exceeds the RF compliance level above, observe the Proteus XR/a system to
verify normal operation in each use location. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the [EQUIPMENT and/or SYSTEM].
**Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
The Recommended Separation Distances are listed in the next table.
Note: These are guidelines. Actual conditions may vary.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
ix
ELECTROMAGNETIC COMPATIBILITY (EMC) (CONT.)
Recommended Separation Distances for Portable and Mobile RF Communications
Equipment and the Proteus XR/a system
Frequency of
Transmitter
150KHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
Equation
d= 1.2
d= 1.2
d= 2.3
Rated Power of
Transmitter
(W)
DISTANCE
(meters)
DISTANCE
(meters)
DISTANCE
(meters)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the
equation in the corresponding column, where P is the power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Note: These are guidelines. Actual conditions may vary.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
x
This page intentionally left blank
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
xi
WARNING
SAFETY
ELECTRIC SHOCK HAZARD! DO NOT REMOVE COVERS OR PANELS.
GENERATOR CABINET CONTAINS HIGH VOLTAGE CIRCUITS FOR GENERATING
AND CONTROLLING X-RAYS. PREVENT POSSIBLE ELECTRIC SHOCK BY
LEAVING COVERS AND PANELS ON THE EQUIPMENT. THERE ARE NO
OPERATOR SERVICEABLE PARTS OR ADJUSTMENTS INSIDE THE CABINETS
UNDER THE TABLE. ONLY TRAINED AND QUALIFIED PERSONNEL SHOULD BE
PERMITTED ACCESS TO THE INTERNAL PARTS OF THIS EQUIPMENT.
FOR CONTINUED SAFE USE OF THIS EQUIPMENT, FOLLOW THE INSTRUCTIONS
CONTAINED IN THIS OPERATING MANUAL. STUDY THIS MANUAL CAREFULLY
BEFORE USING THE EQUIPMENT AND KEEP IT AT HAND FOR QUICK
REFERENCE.
RADIOGRAPHIC EQUIPMENT MUST BE OPERATED BY QUALIFIED PERSONNEL
AND ONLY AFTER SUFFICIENT TRAINING.
UNITED STATES FEDERAL LAW RESTRICTS THIS DEVICE TO USE BY OR ON
THE ORDER OF A PHYSICIAN.
IT IS THE RESPONSIBILITY OF THE OPERATOR TO ENSURE THE SAFETY OF
THE PATIENT WHILE THE MACHINE IS IN OPERATION BY CHECKING PROPER
PATIENT POSITIONING AND USING THE EQUIPMENT PROTECTIVE DEVICES.
TO AVOID INJURY TO FINGERS AND HANDS OF PATIENT AND OPERATOR
CAUSED BY TABLE TOP MOVEMENT, HANDS MUST BE KEPT AWAY FROM
TABLE TOP EDGES AT ALL TIMES.
USE A SID AS LARGE AS POSSIBLE IN ORDER TO KEEP THE ABSORBED DOSE
TO THE PATIENT AS LOW AS REASONABLY ACHIEVABLE.
IT IS THE RESPONSIBILITY OF THE OPERATOR TO PROVIDE MEANS FOR AUDIO
AND VISUAL COMMUNICATION WITH THE PATIENT FROM THE CONTROL ROOM.
PERFORM PERIODIC MAINTENANCE TO ENSURE CONTINUED SAFE USE OF THE
EQUIPMENT. (See chapter 11 Planned Maintenance).
IF ANY SAFETY PROBLEM OCCURS, PLEASE STOP USING THIS DEVICE AND
CONTACT OUR SERVICE AT ONCE.
Always be alert to safety when you operate this equipment. You must be familiar
enough with the equipment to recognize any malfunctions that can be a hazard. If
a malfunction occurs or a safety problem is known to exist, do not use this
equipment until qualified personnel correct the problem.
Apply necessary sterilization with 75% medical Alcohol to components which are
possible to be contacted with the patients, such as Table top, Wall Stand
(including SG120 Wall Stand) front panel, etc.
WARNING
WARNING
WARNING
WARNING
WARNING
CAUTION
WARNING
WARNING
WARNING
CAUTION
WARNING
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
xii
ESTABLISH EMERGENCY PROCEDURES
ESTABLISH PROCEDURES FOR HANDLING THE PATIENT IN CASE OF THE LOSS
OF RADIOGRAPHIC IMAGING OR OTHER SYSTEM FUNCTIONS DURING AN EXAM.
ESTABLISH PROCEDURES FOR HANDLING THE PATIENT IN
CASE OF THE LOSS OF RADIOGRAPHIC IMAGING OR OTHER SYSTEM
FUNCTIONS DURING AN EXAM.
POSSIBLE PATIENT INJURY!
TO AVOID POSSIBLE PATIENT INJURY, BE SURE THAT SYSTEM POWER IS
APPLIED BEFORE THE PATIENT ENTERS THE ROOM. THE OVER HEAD TUBE
SUSPENSION MOVEMENT EM LOCKS AND TABLE LONGITUDINAL TRAVEL
LOCKS FUNCTION ONLY WHEN SYSTEM AC POWER IS APPLIED.
IF POWER IS DISCONNECTED, THE OTS AND THE TABLE TOP (LONGITUDINAL)
WILL MOVE FREELY, POSSIBLE CAUSING THE PATIENT TO FALL.
• DO NOT ALLOW THE PATIENT TO MOUNT OR DEMOUNT THE SYSTEM.
• DO NOT ALLOW THE PATIENT TO USE THE OTS AS A SUPPORT.
OPERATIONAL CHECKS
Be sure the equipment is functioning properly and safely before each examination:
Verify that the following controls are operating correctly:
• Motion controls, and Lock Releases
• Audible and visual alarms
Visually inspect the equipment and make sure that:
• Equipment is not damaged or missing parts
• All cover panels are in place prior to turning on electrical power (hazardous electrical
or mechanical parts could be exposed).
APPROVED OPERATING PROCEDURES AND ACCESSORIES
Be sure to use the equipment and the approved accessories according to
approved operating procedures:
• Perform X-ray tube warm up procedure prior to the exam. Failure to perform this
procedure could damage the X-ray Tube assembly.
• Do not exceed tabletop rating of a 220 kg (484 lbs.) patient. Excessive loading could
damage the tabletop and/or cause the patient to fall.
• Accessories should be properly attached to the table and positioned so as not to
interfere with system motions.
• Avoid unnecessary exposure to radiation. Stay behind the lead glass radiation shield
or lead screen. When in unshielded areas, wear protective apparel such as goggles,
lead aprons, and gloves.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
xiii
PLANNED MAINTENANCE
POSSIBLE PATIENT OR OPERATOR INJURY!
TO AVOID POSSIBLE PATIENT OR OPERATOR INJURY, BE SURE TO PERFORM
THE PERIODIC INSPECTIONS AND MAINTENANCE PROVIDED IN THIS
DOCUMENT. FAILURE TO PERFORM THESE INSPECTIONS COULD ALLOW
DETERIORATING CONDITIONS TO DEVELOP WITHOUT BEING DETECTED. THIS
DETERIORATION COULD RESULT IN EQUIPMENT FAILURES WHICH COULD
CAUSE SERIOUS INJURY EQUIPMENT DAMAGE.
RADIATION SAFETY
Always use proper technique factors for each procedure to minimize x-ray exposure and
to produce the best diagnostic results. In particular, you must be thoroughly familiar with
safety precautions before operating this System.
It is not always possible to determine when some components, such as x-ray tubes, are
nearing the end of their operating lives. These components could stop operating during a
patient examination.
KNOW THE EQUIPMENT
Read and understand all the instructions in the operating manuals before attempting to
use the product and request training assistance from GE Medical System if needed.
Keep the operating manuals with the equipment at all times and periodically review the
procedures and safety precautions.
This system contains operating safeguards to provide maximum safety. Before calling for
service, be certain proper operating procedures are being used.
Satisfactory equipment performance requires the use of service personnel specially
trained on x-ray apparatus. GE Medical Systems is responsible for the effects on safety,
reliability, and performance only if the following conditions are met:
•The electrical wiring of the relevant rooms complies with all national and local
codes as well as the Regulations for the electrical equipment of buildings
published by the Institution of Electrical Engineers.
•All assembly operations, extensions, re-adjustments, and modifications or repairs are
carried out by GE Medical Systems’ authorized service representatives.
•The equipment is used in accordance with the instructions for use.
WARNING
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
xiv
This page intentionally left blank
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
xv
TABLE OF CONTENTS
CHAPTER TITLE PAGE NUMBER
1 QUICK START 1-1
1-1 Turn System On 1-1
1-2 Tube Warm-Up 1-1
1-3 Set Technique APR 1-2
1-4 Set Manual Technique 1-3
1-5 Set AEC Technique 1-4
2 SYMBOLS 2-1
2-1 Special Notices 2-1
2-2 X-ray Tube 2-1
2-3 Power ON and OFF 2-2
2-4 Electrical Type 2-2
2-5 Electrical Current 2-2
2-6 Ground 2-2
2-7 Proteus XR/a Collimator / Eclipse Proteus Collimator 2-2
2-8 Emergency Button 2-2
2-9 Warning Signs and Labels 2-3
2-10 System Labeling 2-4
3 SYSTEM DESCRIPTION 3-1
3-1 System Components/Features 3-1
3-2 HHS Compliance Compatibilities 3-3
4 PROTEUS XR/A SYSTEM START UP AND SHUT DOWN 4-1
4-1 Turn the power on 4-1
4-2 Turn Power off 4-1
4-3 Daily Warm Up Procedures 4-2
4-4 System Status Display 4-2
4-5 Radiography Control Key 4-3
5 PROTEUS XR/A SYSTEM CONSOLE 5-1
5-1 Introduction 5-1
5-2 Procedure Edit 5-9
5-3 Application 5-13
6 PROTEUS XR/A TABLE COMPONENTS 6-1
6-1 Safe Operation Precautions 6-1
6-2 Introduction 6-3
6-3 Table Operation 6-5
6-4 Cassette Tray Operation 6-6
7 PROTEUS XR/A OVERHEAD TUBE SUSPENSION (OTS) 7-1
7-1 Introduction 7-1
7-2 Overhead Rail System 7-1
7-3 Telescopic Column and Carriage 7-2
7-4 X-ray Tube Support 7-4
7-5 OverHead Tube Suspension User Interface 7-6
7-6 Proteus XR/a Automatic Collimator 7-8
7-7 Proteus XR/a Manual Collimator (Optional) 7-14
7-8 Eclipse Proteus Collimator 7-15
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
xvi
TABLE OF CONTENTS (CONT.)
CHAPTER TITLE PAGE NUMBER
8 PROTEUS XR/A WALL STAND (GPCP No.: 2260354) COMPONENT 8-1
8-1 Introduction 8-1
8-2 Operation
9 PROTEUS XR/A SG120 WALL STAND (GPCP No.: 2402562) COMPONENT 9-1
9-1 Safe Operation Precautions 9-1
9-2 Introduction 9-2
9-3 Applications 9-4
9-4 Operation 9-4
10 ACCESSORIES 10-1
10-1 Introduction 10-1
10-2 Accessories 10-1
11 PLANNED MAINTENANCE 11-1
11-1 General 11-1
11-2 HHS Testing 11-1
11-3 Qualified Service 11-2
11-4 Periodic Maintenance 11-2
11-5 Recycling 11-5
12 SYSTEM FAULTS 12-1
12-1 Introduction 12-1
12-2 General Trouble Shooting 12-1
12-3 Other Operator Fault Analysis 12-4
12-4 Resetting Faults 12-4
13 PHYSICAL REQUIREMENTS OF ROOM 13-1
13-1 Environmental Requirements/Limitations 13-1
13-2 Equipment Heat output 13-2
13-3 Radiation Protection 13-3
14 SPECIFICATION 14-1
14-1 General System Specifications 14-1
14-2 Table Specifications 14-2
14-3 Generator Specifications 14-3
14-4 System Console Specifications 14-10
14-5 OTS Specifications 14-10
14-6 Collimator Specifications 14-11
14-7 Wall Stand (GPCP No.: 2260354) Specifications 14-14
14-8 SG120 Wall Stand (GPCP No.: 2402562) Specifications 14-15
14-9 X-ray Tube Specifications 14-16
14-10 Printer Specifications 14-18
APPENDIX
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
xvii
REVISION HISTORY
REV DATE TYPE OF MODIFICATION
0 10/01/2000 Initial production Release
1 20/07/2000 Add system function description and system specification
2 05/12/2000 Update OTS user’s interface description
3 01/02/2001 Update the regulatory requirements.
3 01/02/2001 Add a warning to wall stand operators.
3 01/02/2001 Update system labeling.
3 01/02/2001 Add printer information.
3 01/02/2001 Update wall stand illustration.
3 15/02/2001 Add notes.
3 19/02/2001 Add new wall stand.
3 19/02/2001 Add a maintenance item.
3 01/03/2001 Add manufacturer’s name
3 07/03/2001 Add a warning.
4 27/09/2001 Add notes, change specs.
5 14/05/2003 Add description of MX100 X-ray tube and SG100 Wall Stand
6 12/04/2004 Add a caution about the shroud of the elevating table.
7 25/06/2005 Add EMC and WEEE Rules.
7 25/06/2005 Add description of SG120 Wall Stand.
8 05/12/2005 Add description of Eclipse Proteus collimator
8 05/12/2005 Add a warning
9 15/02/2006 Add description of Reciprocating Bucky and AID Ion Chamber.
10 20/06/2006 Update warning label to meet HHS requirements in Chapter 2.
11 08/10/2006 Add mA and mAs
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
xviii
LIST OF EFFECTIVE PAGES
PAGE REVISION PAGE REVISION
NUMBER NUMBER NUMBER NUMBER
i thru xii 1 Address page 5
1-1 thru 1-4 1 6-5 6
2-1 thru 2-2 1 iii 7
3-1 thru 3-4 1 v thru xi 7
4-1 thru 4-4 1 xv thru xvi 7
5-1 thru 5-12 1 2-1and 2-4 7
6-1 thru 6-10 1 3-1 thru 3-3 7
7-1 thru 7-14 1 5-13 7
8-1 thru 8-4 1 7-5 7
9-1 thru 9-2 1 chapter 9 7
10-1 thru 10-6 1 11-4 thru 11-5 7
11-1 thru 11-4 1 13-1 thru 13-2 7
12-1 thru 12-2 1 14-9 7
13-1 thru 13-14 1 14-14 7
APPENDIX 1 xi 8
7-7 and 7-8 2 xv thru xviii 8
ⅲ3 2-2 thru 2-4 8
2-4 3 3-1 thru 3-3 8
3-1 thru 3-3 3 6-6 8
5-1 thru 5-12 3 7-1 thru 7-2 8
6-1 thru 6-5 3 7-15 thru 7-19 8
7-6 thru 7-10 3 14-13 8
8-2 and 8-3 3 2-4 9
10-4 3 2-3 10
11-3 3 14-6 11
13-11 and 13-16 3
APPENDIX 3
2-3 4
4-3 4
5-4 4
6-3 4
7-7 thru 7-8 4
8-1 thru 8-2 4
9-2 4
chapter12 thru chapter13 4
2-4 5
3-1 thru 3-3 5
5-13 5
6-5 5
6-8 5
7-5 and 7-7 5
8-1 5
8-2 5
chapter 9 5
10-3 5
11-4 5
13-1 thru 13-2 5
14-9 5
APPENDIX 5
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).
Table of contents
Other GE Medical Equipment manuals