manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. GE
  6. •
  7. Medical Equipment
  8. •
  9. GE Vacugene User manual

GE Vacugene User manual

Other GE Medical Equipment manuals

GE Vivid in User manual

GE

GE Vivid in User manual

GE LOGIQ V5 User manual

GE

GE LOGIQ V5 User manual

GE Venue Go User manual

GE

GE Venue Go User manual

GE Marquette 418793-001 User manual

GE

GE Marquette 418793-001 User manual

GE Senographe SecondLook User manual

GE

GE Senographe SecondLook User manual

GE Vivid i SafeLock Cart User manual

GE

GE Vivid i SafeLock Cart User manual

GE LOGIQ C Series User manual

GE

GE LOGIQ C Series User manual

GE MAC 5500 User manual

GE

GE MAC 5500 User manual

GE LOGIQ V2 User manual

GE

GE LOGIQ V2 User manual

GE MAC 5000 User manual

GE

GE MAC 5000 User manual

GE CARESCAPE User manual

GE

GE CARESCAPE User manual

GE LOGIQ C5 Premium User manual

GE

GE LOGIQ C5 Premium User manual

GE LOGIQ C Series User manual

GE

GE LOGIQ C Series User manual

GE Vivid 7 User manual

GE

GE Vivid 7 User manual

GE LOGIQ E9 User manual

GE

GE LOGIQ E9 User manual

GE MAC 5000 Quick start guide

GE

GE MAC 5000 Quick start guide

GE CareScape V100 User manual

GE

GE CareScape V100 User manual

GE T8 User manual

GE

GE T8 User manual

GE LOGIQ 7 Pro User manual

GE

GE LOGIQ 7 Pro User manual

GE Innova 2000 User manual

GE

GE Innova 2000 User manual

GE Dash 2500 User manual

GE

GE Dash 2500 User manual

GE Voluson 730 User manual

GE

GE Voluson 730 User manual

GE Discovery LS User manual

GE

GE Discovery LS User manual

GE LOGIQ P9 User manual

GE

GE LOGIQ P9 User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Instructions 80-1265-18 AH Vacugene Vacuum Blotting Pump
GE Healthcare
Unpacking
Unpack the units and accessories. Do not discard
any packing materials until each item has been
checked against the packing list supplied with the
instrument.
Inspect all items for transit damage. If damage has
occurred, contact your local GE Healthcare
office and advise the transport company concerned.
Installing and operating the pump
1 Place the VacuGene™ Pump within reach of an electrical outlet.
Note: This instrument is for indoor use only.
2 Check that the mains power switch on sawbuck panel of the
pump is OFF (in the 0 position).
3 The mains power inlet, fuses and mains ON/OFF switch are
located on the back panel of the vacuum pump.
4 Fit the mains cable into the mains power inlet and connect it to
the mains power supply. Ensure that the mains cable is plugged
into a fully grounded mains outlet.
5 The power inlet and power supply shall always be easily
accessible.
6 Ensure that all air inlets and outlets are free from obstruction.
7 Make sure the instrument is placed on a stable laboratory work
bench.
8 Using the tubing supplied with VacuGene XL Blotting Unit,
connect the pump inlet on the front panel to the connection on
the blotting unit. The needle valve controls the vacuum pressure
and the vacuum gauge gives readings in the range 0-100 mbar
with an accuracy of + 5%.
9The low pressure VacuGene Vacuum Blotting Pump requires no
regular maintenance.
To protect the pump, the liquid trap supplied with the VacuGene XL
Blotting Unit should be installed between the pump and a blotting
unit. A “T” connector is provided to attach a second blotting unit to the
vacuum pump, if required.
.
Fig.1 Pump unit front panel.
Fig.2 Pump unit back panel.
WARNING! All repairs should be done by personnel
authorized by GE Healthcare. Do not open any covers
or replace any parts unless specifically stated in the
instructions.
Vacuum
mbar
Instructions 80-1265-18 AH 06-2008
For contact information for your local office, please visit:
www.gelifesciences.com/contact
GE Healthcare Bio-Sciences AB
Björkgatan 30
751 84 Uppsala
Sweden
www.gelifesciences.com
GE, imagination at work and GE monogram are trademarks or registered trademarks of General Electric
Company
Vacugene is a trademark of GE Healthcare companies.
All third party trademarks are the property of their respective companies.
© 1989-2008 General Electric Company – All rights reserved.
First published Oct. 1989
All goods and services are sold subject to the terms and conditions of sale of the company within GE
Healthcare which supplies them. Contact your local GE Healthcare representative for the most current
information.
GE Healthcare Europe GmbH
Munzinger Strasse 5, D-79111 Freiburg, Germany
GE Healthcare UK Ltd
Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK
GE Healthcare Bio-Sciences Corp
800 Centennial Avenue, P.O. Box 1327, Piscataway, NJ 08855-1327, USA
GE Healthcare Bio-Sciences KK
Sanken Bldg. 3-25-1, Hyakunincho, Shinjuku-ku, Tokyo 169-0073, Japan
imagination at work
Fig.3 Typical arrangement of pump, liquid trap and blotting unit.
Technical Specifications
Important User Information
CE Certification
This product complies with the European directives listed
below, by fulfillling corresponding standards. A copy of the
Declaration of Conformity is available on request.
• 73/23/EEC, Low voltage Directive
• 89/336/EEC EMC Directive
Note: The declaration of conformity is valid for the instrument when
it is
-used in laboratory locations
- used in the same state as it was delivered from GE Health
except for alterations described in the user manual
- used as a “stand alone” unit or connected to other CE
labelled GE Healthcare instruments or other products as
recommended.
WARNING!
The Warning sign is used to call attention to the
necessity to follow an instruction in detail to avoid
personal injury. Be sure not to proceed until the
instructions are clearly understood and all stated
conditions are met.
CAUTION!
The caution sign highlights instructions that must be followed to
avoid damage to the product or other equipment. Do not proceed
until all stated conditions are met and clearly understood.
Note:
The Note sign is used to indicate information important for trouble-
free and optimal use of the product.
Recycling:
This symbol indicates that the waste of electrical and
electronic equipment must not be disposed as unsorted
municipal waste and must be collected separately.
Please contact an authorized representative of the
manufacturer for information concerning the
decommissioning of equipment.
Vacuum range: 0-100 mbar
Environment: +4 to +40 ºC
20 - 95% relative humidity
Warm-up time: 2 minutes
Supply voltage: 100-110 V ~ 50 Hz (80-1265-14)
100-120 V ~ 50 Hz (80-1265-14)
230-240 V ~ 50 Hz (80-1265-15)
Fuse specifications: (80-1265-14) T1.25A L 250 V
(80-1265-15) T 800mA L 250 V
Power consumption: 135 VA
Dimensions: 160 x 290 x 150 mm
Weight: 3.1 kg
Safety standards: EN61010-1