www.geratherm.com
7
• Before each use, check the surface of the warming
blankets for any visible damage! Discontinue use
of any damaged blankets immediately!
•PuttheGerathermUniqueTemp°warmingblanketscom-
pletely on the patient. To prevent overheating, do not un-
deranycircumstancesfoldtheequipment!
•Donotplaceanyobjectsontopofthewarmingblankets
to avoid any localised overheating or heat accumulati-
on.
•It must be checked whether any fixture is required to
keep the patient in place. If so, the same must be positi-
oned underneath the warming blankets.
•The temperature setting is the responsibility of the
anaesthetist.
•Do not use Geratherm UniqueTemp° on patients
with impaired circulation! The manufacturer is not
aware of any other contra-indications.
•Wheneverheattreatmentisapplied,thecorebodytem-
perature of the patient must be monitored at regular in-
tervals. The length of the intervals shall be determined
by the medical staff.
If HF surgical instruments or endocardial catheters are
used at the same time as the patient warming system,
equipotential bonding must be ensured.
•Geratherm UniqueTemp° shall only be used under the
supervision of the anaesthetist.
• Do not apply adhesive tape to the warming blan-
kets. This will lead to permanent soiling of the sur-
face.
•Onlyusedrywarmingsystemsonthepatient.
•When using the body temperature sensor, it is abso-
lutely necessary to observe the prescribed connecting
sequence. (Before the temperature sensor is connected
to the patient, the plug on the sensor must be inserted
into the control unit.)
• Warning!
All transducer units connected to the patient by conduc-
ting wires are subject to a certain risk of faulty readings,
localised overheating and possible damage due to in-
tensive HF radiation. Electrosurgical equipment is such
a source, since capacitively coupled currents may flow
to earthing through the probe cables of unconnected
devices. This can lead to the patient suffering burns.
Whenever possible, remove the probe from the patient
completely before activating the surgical equipment or
other HF source. If the probes have to be used simulta-
neously with electrosurgical equipment, the devices to
which the probes are connected should be checked for
adequate protective earthing with high frequencies. The