Operating Instructions DAC UNIVERSAL 1.8Intended use
båÖäáëÜ
1.8 Intended use
The DAC UNIVERSAL is intended to clean, lubricate and steam sterilize
dental handpieces. Dental handpieces is defined as rotating instruments
split up into the categories straight handpieces, contra angles and
turbines.
The DAC UNIVERSAL is intended to clean and steam sterilize solid
dental instruments.
Intended use_FLEX lid
The DAC UNIVERSAL is intended to clean and thermally disinfect Sirona
ultrasonic scaler handpieces of type SIROSONIC, SIROSONIC L,
SIROSONIC TL, PerioSonic and similar dental ultrasonic scaler
handpieces, which are comparable concerning weight, materials,
surface, construction principles, gaps and geometry.
The DAC UNIVERSAL is intended to clean and thermally disinfect Sirona
Scaler tips of the types Instrument L, PE, PS, SIRORETRO, SIROPREP
and similar dental scaler tips, which are comparable concerning weight,
materials, surface, construction principles, gaps and geometry including
the associated torque wrench.
The DAC UNIVERSAL is intended to clean and thermally disinfect Sirona
Sprayvit nozzles of the types SPRAYVIT, SPRAYVIT 4000, SPRAYVIT
4000E and similar dental sprayvit nozzles, which are comparable
concerning weight, materials, surface, construction principles, gaps and
geometry.
The DAC UNIVERSAL is intended to clean and steam sterilize EMS
Piezon ultrasonic scaler handpieces of type EN-041, EN-060 and similar
dental ultrasonic scaler handpieces, which are comparable concerning
weight, materials, surface, construction principles, gaps and geometry.
The DAC UNIVERSAL is intended to sterilize wrapped solid instruments,
straight handpieces, contra angles and turbines.
1.9 Contraindications
The DAC UNIVERSAL is not intended to clean, lubricate or sterilize/
disinfect liquids, cloth loads, biomedical waste or other devices and
materials not specified above.
Furthermore DAC UNIVERSAL is not intended to sterilize/ disinfect
medical devices which are not compatible with steam sterilization.
The processing of articles other than specified above may result in
incomplete cleaning or sterilization/ disinfection, damage to the
incompatible articles, and/or damage to the DAC UNIVERSAL.
The DAC UNIVERSAL is not intended to reprocess Implant, Endo and
CEM tips.
The DAC UNIVERSAL is not intended to reprocess ultrasonic scaler
handpieces of type SIROSON and SIROSON S.