Graco PE14 User manual

312865A
INSTRUCTIONS
ME14™and PE14™Meters
- For metered dispense of oils and 50 / 50 water / antifreeze mix fluids -
Maximum Working Pressure: 1500 psi (10 MPa, 103 bar)
Maximum Flow Rate:14 gpm (53.0 lpm)
List of Models page 2
Important Safety Instructions
Read all warnings and instructions in this
manual. Save these instructions.
NOTICE
This dispense valve is designed to dispense
petroleum-based lubricants and antifreeze only.
Do not dispense windshield washer solvent with
this dispense valve.
ti11821
PE14 meter shown

Models
2312865A
Models
ME14 Manual Dispense Electronic Meters
PE14 Electronic Meters
Model
Number Swivel Extension Description Nozzle Fluid Type
255348 1/2” npt(f) Flexible Extension Automatic, non-drip, quick close Oil
255349 1/2” npt(f) Gear Lube Extension None-drip, quick close Gear Lube
255350 1/2” npt(f) Rigid Extension Automatic, non-drip, quick close Oil
255800 3/4” npt(f) Rigid Extension Automatic, non-drip, quick close Oil
255801 3/4” npt(f) Flexible Extension Automatic, non-drip, quick close Oil
255802 1/2” npt(f) Rigid Extension Automatic, non-drip, quick close Anti-freeze
255803 1/2” npt(f) Flexible Extension Automatic, non-drip, quick close Anti-freeze
Model
Number
Type of
Dispense Swivel Extension Description Nozzle Fluid Type
255200 Preset 1/2” npt(f) Rigid Extension Automatic, non-drip,
quick close Oil
255351 Preset 1/2” npt(f) Flexible Extension Automatic, Non-drip,
quick close Oil
255352 Preset 1/2” npt(f) Gear Lube Extension Non-drip, quick close Gear Lube
255353 Preset 3/4” npt(f) Rigid Extension Automatic, non-drip,
quick close Oil
255354 Preset 3/4” npt(f) Flexible Extension Automatic, non-drip,
quick close Oil
255355 Manual 1/2” npt(f) Rigid Extension Automatic, non-drip,
quick close Anti-freeze
255356 Manual 1/2” npt(f) Flexible Extension Automatic, non-drip,
quick close Anti-freeze

Warnings
312865A 3
Warnings
The following warnings are for the setup, use, grounding, maintenance, and repair of this equipment. The exclama-
tion point symbol alerts you to a general warning and the hazard symbol refers to procedure-specific risk. Refer back
to these warnings. Additional, product-specific warnings may be found throughout the body of this manual where
applicable.
WARNING
SKIN INJECTION HAZARD
High-pressure fluid from dispense valve, hose leaks, or ruptured components will pierce skin. This may
look like just a cut, but it is a serious injury that can result in amputation. Get immediate surgical
treatment.
• Do not point dispense valve at anyone or at any part of the body.
• Do not put your hand over the end of the dispense nozzle.
• Do not stop or deflect leaks with your hand, body, glove, or rag.
• Follow Pressure Relief Procedure in this manual, when you stop spraying and before cleaning,
checking, or servicing equipment.
EQUIPMENT MISUSE HAZARD
Misuse can cause death or serious injury.
• Do not operate the unit when fatigued or under the influence of drugs or alcohol.
• Do not exceed the maximum working pressure or temperature rating of the lowest rated system
component. See Technical Data in all equipment manuals.
• Use fluids and solvents that are compatible with equipment wetted parts. See Technical Data in all
equipment manuals. Read fluid and solvent manufacturer’s warnings. For complete information
about your material, request MSDS forms from distributor or retailer.
• Check equipment daily. Repair or replace worn or damaged parts immediately with genuine manu-
facturer’s replacement parts only.
• Do not alter or modify equipment.
• Use equipment only for its intended purpose. Call your distributor for information.
• Route hoses and cables away from traffic areas, sharp edges, moving parts, and hot surfaces.
• Do not kink or over bend hoses or use hoses to pull equipment.
• Keep children and animals away from work area.
• Comply with all applicable safety regulations.
BATTERY SAFETY
The battery may leak, explode, cause burns, or cause an explosion if mishandled:
• You must use the battery type specified for use with the equipment.
• Sparking can occur when changing batteries. Only replace the battery in a non-hazardous location,
away from flammable fluids or fumes.
• Handle and dispose of battery properly - do not short circuit, charge, force over discharge, disas-
semble, crush, penetrate, incinerate, or heat the battery to a temperature exceeding 185° F (85° C).

Warnings
4312865A
FIRE AND EXPLOSION HAZARD
When flammable fluids are present in the work area, such as gasoline and windshield wiper fluid, be
aware that flammable fumes can ignite or explode. To help prevent fire and explosion:
• Use equipment only in well ventilated area.
• Eliminate all ignition sources, such as cigarettes and portable electric lamps.
• Keep work area free of debris, including rags and spilled or open containers of solvent and gasoline.
• Do not plug or unplug power cords or turn lights on or off when flammable fumes are present.
• Ground all equipment in the work area.
• Use only grounded hoses.
• If there is static sparking or you feel a shock, stop operation immediately. Do not use equipment
until you identify and correct the problem.
• Keep a working fire extinguisher in the work area.
WARNING

Installation
312865A 5
Installation
Typical Installations (FIG. 1)
The typical installation shown in FIG. 1 is only a guide. It
is not a complete system design. Contact your Graco
distributor for assistance in designing a system to suit
your needs.
Mounting Bracket (FIG. 2)
Mounting Bracket Kit 249440 is available for resting the
dispense valve on a console.
Drum Mount Bracket (FIG. 3)
Mounting Bracket 15B750 is available for resting the
dispense valve on a drum.
NOTICE
The dispense valve is not designed for in-line
installation. Do not install with a shutoff valve on the
outlet side of the meter. Such installation could result in
damage to the meter housing cover.
FIG. 1
ITEM DESCRIPTION
A Electronic metered dispense valve
B Fluid shut-off valve
CHose
D Hose reel fluid inlet hose
E Hose reel
A Thermal Relief Kit (not shown) is required. The kit
required will vary by pump selected. See Parts, page 34 for
a list of available kits.
B
D
E
C
A
ti11010a
FIG. 2
FIG. 3
ti11822
ti11823

Installation
6312865A
Pressure Relief Procedure
This equipment will stay pressurized until the pressure
has been manually relieved. To reduce the risk of
serious injury from pressurized fluid, accidental spray
from the dispense valve or splashing fluid, follow this
Pressure Relief Procedure when ever you:
• Are instructed to relieve pressure.
• Check, clean or service any system equipment.
• Install or clean fluid nozzles or filter.
1. Turn off power supply to the pump or close
upstream ball valve.
2. Trigger the dispense valve into a waste container to
relieve pressure.
3. Open any bleed-type master air valves and fluid
drain valves in the system.
4. Leave the drain valve open until you are ready to
pressurize the system.
Grounding
The equipment must be grounded. Grounding reduces
the risk of static and electric shock by providing an
escape wire for the electrical current due to static build
up or in the event of a short circuit.
Pump: Follow manufacturer’s recommendations.
Air and fluid hoses: Only use electrically conductive
hoses. Check electrical resistance of hoses. If total
resistance to ground exceeds 29 megohms, replace
hose immediately.
Air compressor: Follow manufacturer’s
recommendations.
Fluid supply container: Follow local code.
To maintain grounding continuity when flushing or
relieving pressure: hold a metal part of the dispense
valve firmly to the side of a grounded metal pail, then
trigger the valve.
Pre-Installation Procedure
1. Relieve pressure, page 6.
2. Close the shut-off valve (B, FIG. 1).
3. Ground the hose and reel or console, page 6. Leave
at least two threads bare when using PTFE tape.
The bare threads ensure a ground is maintained.
Installation Procedure
Flushing
If this is an existing installation, go to Installing Meter
section, page 7. The following procedure, Steps 1-5 are
the Flushing Procedure.
1. Close the fluid shut-off valve (B, FIG. 1, page 5) at
each dispense position.
2. Make sure:
• main fluid outlet valve at the pump is closed,
• air pressure to the pump motor is adjusted,
and
• air valve is open.
FIRE HAZARD: Conductive metal surfaces on the
meter must not make contact with any positively
charged metal surface, including (but not limited to),
the starter solenoid terminal, alternator terminal or
battery terminal. Such contact could cause electrical
arcing and a fire.
NOTICE
• If this is a new installation or if the fluid lines are
contaminated, flush the lines before you install the
metered valve. Contaminated lines could cause the
valve to leak.
• Never dispense compressed air with meter. Doing
so will damage meter.

Installation
312865A 7
3. Slowly open the main fluid valve.
a. Place the hose end (with no dispense valve
connected) into a container for waste oil.
b. Secure the hose in the container so it will not
come out during flushing.
c. If you have multiple dispense positions, first
flush the dispense position farthest from the
pump and work your way toward the pump.
4. Slowly open the shut-off valve (B) at the dispense
position. Flush out a sufficient amount of oil to
ensure that the entire system is clean; then close
the valve.
5. Repeat Step 4 at all other positions.
Installing Meter (FIG. 4)
1. Relieve pressure, page 6.
2. Slide the swivel boot (32) back, over the hose, small
end first, to access the swivel fitting (a).
3. Apply thread sealant to the male threads of the hose
fitting. Thread the hose fitting into the meter swivel
(31). Use two wrenches to tighten securely (FIG. 4).
Installing Tube Extension (FIG. 5)
1.
a. Loosen nut (a) until it is completely off tube
threads.
b. Thread extension (20) into housing (b) until it
bottoms out.
c. Align extension (20) with meter housing and
handle (c).
d. Firmly tighten nut (a).
FIG. 4
Make sure you let the sealant cure to the manufac-
turer’s recommendations before circulating fluid
through the system.
31
ti10983a
32
a
FIG. 5
NOTICE
Do not use PTFE tape or thread sealant on threads of
extension tube (20). This could cause the fitting to
leak.
20
a
c
b
ti10615A

Installation
8312865A
Installing Nozzle (FIG. 6)
1.
a. Thread new nozzle (33) onto extension (20).
b. With an open-end adjustable wrench on flats of
nozzle bushing, tighten firmly.
2. Open automatic twist lock nozzle and all fluid
shut-off valves. Start pump to pressurize system.
See Operation Screens, Dispensing in either the
Manual or Preset Mode instructions beginning on
page 18 for complete meter operation information.
3. To ensure dispensing accuracy, purge all air from
the fluid lines and dispense valve before you use it.
4. Set the system flow to the desired flow rate.
FIG. 6
•Only tighten nozzle with wrench on flats of the
nozzle bushing.
•Do not disassemble the bushing from
nozzle. Disassembly will affect performance of
the nozzle.
20
33
ti10615A
NOTICE
Do not trigger meter when nozzle is closed. Fluid
will build up behind the nozzle, leak from the nozzle,
and unexpectedly be expelled when the nozzle is
opened. If you do accidentally trigger the meter with
the nozzle closed, point the nozzle into a waste
bucket and open the nozzle to relieve pressure and
expel the built up fluid.

Meter Overview
312865A 9
Meter Overview
Navigation and Modes
5-Way Menu Navigation Button (FIG. 7)
• Includes 4 direction ARROWS (UP, DOWN, LEFT,
RIGHT) and a center, ENTER button.
• Pressing the direction ARROWS allows user to
easily scroll through menus. To select/store your
selection, you must press the center, ENTER
button.
• Pressing and holding a direction ARROW down
allows user to scroll through menus quickly.
Meter Display
Adjusting Screen Contrast using ARROWS
On the Setup Home Screen (page 10), use the LEFT
and RIGHT ARROWS to adjust the screen contrast.
•Darken the Screen: Press the RIGHT ARROW
multiple times.
•Brighten the Screen: Press the LEFT ARROW
multiple times.
Asleep / Awake Mode
•Asleep: Battery-saving mode. The display goes
blank after 2 minutes of inactivity during normal
operation. Unit continues to keep track of amount
dispensed while the display is asleep.
•Awake:Display comes awake from sleep mode
when you press any ARROW or the ENTER button
or when you squeeze the trigger to dispense fluid.
Locking and Unlocking Trigger
The PE14 meter only, includes a locking trigger feature
that allows the user to lock the trigger in the dispense
position as shown in FIG. 8. To release the lock, firmly
squeeze the trigger to the handle.
PE14 Meter Only
FIG. 7
ti11824 5-Way Menu
Navigation Button
FIG. 8
Unlocked
Locked

Setup Mode Screens
10 312865A
Setup Mode Screens
If you are in Operation Mode you must be on the Home Screen shown in FIG. 10, to access the Setup Mode Screens.
(A complete description of the Home Screen is provided on page 17).
To display the Home Screen:
1. Wake up the meter by pressing any button on the key pad.
2. When an operation mode screen displays (such as
the one shown in FIG. 9):
a. Use the RIGHT ARROW to move the curser
over the House icon (a).
b. Press center, ENTER button to display the
Home Screen (FIG. 10).
FIG. 9
MOBIL15W-20
10.56 QTS
a
ti12179a
FIG. 10
MOBIL15W-20
MANUAL
TOTALS
MOBIL15W-20
MANUAL
PRESET
TOTALS
PE14 Meter
ME14 Meter
ti12184a
ti12175a

Setup Mode Screens
312865A 11
Main Setup Screen (FIG. 12)
ME14 and PE14 Meters
The Main Setup Screen is the first screen displayed
when you enter the Setup Mode. This screen displays a
list of the available Setup Screens and also includes a
link back to the Home Screen.
The screens available to ME14 Meter users are:
• UNITS/LIMIT
•CALIBRATE
• BANNER
•LANGUAGE
•HOME
PE14 Meters include all the ME14 Setup Screens and
also a PRESET screen.
• UNITS/LIMIT
•CALIBRATE
• PRESET
• BANNER
•LANGUAGE
•HOME
Displaying Setup Screen from an Operation
Mode Screen
1. Hold down the RIGHT ARROW (a) only, for a few
seconds (FIG. 11).
2. Then at the same time, also press the center,
ENTER button (b) (FIG. 11). Hold both buttons down
until the Main Setup Screen shown in FIG. 12 dis-
plays.
Main Screen Features (FIG. 12)
A. Screen Identifier Icon: Wrench icon displays in the
upper right corner when user is on the Main Screen of
the Setup screens.
B. Setup Screens: List/menu of screens available to
user.
1. Use UP or DOWN ARROWS to select a screen from
the list.
2. Press center, ENTER button to confirm selection.
Selected screen displays.
1. Software Version Number: Reference number. You
may be asked to provide this number when contacting
Graco for technical support.
FIG. 11
(a)
(b)
ti11824a
FIG. 12
UNITS/LIMIT
CALIBRATE
PRESET
BANNER
LANGUAGE
HOME
1.23.45
ME14 Meter PE14 Meter
B
A
UNITS/LIMIT
CALIBRATE
BANNER
LANGUAGE
HOME
1.23.45
B
A
ti12177a ti12186a
11

Setup Mode Screens
12 312865A
Units/Limit Screen (FIG. 13)
For ME14 and PE14 Meters
Sets the units of measurement to pints, quarts, liters, or gallons.
Units/Limit Screen Features (FIG. 13)
ME14 and PE14 include C - E
C. Units Mode: Displays measurement unit that was
selected using the Set Measurement Units
button (E).
D. Wrench Icon: Returns user to Main Setup Screen.
E. Set Measurement Units button/field: Sets unit of
measurement as pint, quart, gallon or liter.
To change/set the Unit of Measurement:
1. Use LEFT or RIGHT ARROW to move curser to the
Set Measurement field (E).
2. Use UP or DOWN ARROWS to display measure-
ment unit choices: PTS, QTS, L, GAL.
3. Press center, ENTER button to confirm selection.
Selected measurement unit is displayed on screen
(C).
4. When you are finished making changes, use the
RIGHT ARROW to move curser over Wrench icon.
5. Press center, ENTER button to return to Main Setup
Screen.
PE14 also includes F - G
F. Set Manual Dispense Limit field: Sets maximum
quantity of fluid that can be dispensed in Manual Mode.
To change/set the Manual Dispense Limit:
1. Use RIGHT ARROW to move curser to Manual Dis-
pense Limit field (F).
2. Use UP or DOWN ARROWS to increase and
decrease displayed amount.
3. When amount you want to use is shown in field,
press center, ENTER button to confirm amount.
Confirmed amount displays on screen (G).
4. When you have finished making changes, use the
RIGHT ARROW button to move curser over Wrench
icon.
5. Press center, ENTER button to return to Main Setup
Screen.
G. Manual Limit Confirmation: Displays maximum
quantity of fluid that can be manually dispensed at one
time. Amount is assigned on the task bar Set Manual
Dispense Limit field (F). Preset dispense amounts are
not affected by the manual dispense limit you set.
FIG. 13
QTS
UNITS = QTS
MANUAL LIMIT = 5.0
5.0
D
C
G
EF
PE14 Meter
ME14 Meter
ti12191a
QTS
UNITS = QTS
C
D
E

Setup Mode Screens
312865A 13
Calibrate Screen (FIG. 14)
For ME14 and PE14 Meters
Recalibrates the meter for different fluids.
Table 1. Calibration Factors
• Calibration factors at 70°F (21°C) at 2.0 gpm (7.6 lpm).
• Calibration number may vary slightly due to temperature
or flow rate.
ME14 and PE14 include D - M
D. Wrench Icon: Returns user to Main Screen.
J. Calibration Factor: Displays amount operator
dispensed during meter calibration.
K. Preset Calibration Factor: Preset calibration
amount displays.
Adjusting the Preset Calibration Factor:
1. Use your LEFT or RIGHT ARROW to move the
curser to the Calibration Factor Field (K) .
2. Use the UP or DOWN ARROW to manually adjust
the Preset Calibration Factor up or down (this can
only be done on Calibrate Screen 1 (FIG. 14).
3. Press center, ENTER button to confirm new setting.
New setting displays in Calibration Factor field (J).
L. Start/End button: Clicked at the start and completion
of calibration. See Recalibrating the Meter (page 14).
M. Dispense: Only appears on Screen 2 (FIG. 14).
Displays amount operator dispensed during calibration
test.
Fluids
Calibration Number
Quarts Liters
Oil (10W - 30) 177 187
Gear Lube 177 187
Automatic
Transmission Fluid
177 187
Antifreeze 157 165
FIG. 14
D
J
M
K
L
L
Calibrate Screen 1
Calibrate Screen 2
ti12173a
ti12174a

Setup Mode Screens
14 312865A
Recalibrating the Meter (FIG. 14)
The meter is shipped from the factory with the default
calibration factor for 10W30 motor oil - at 70°F (21°C).
This calibration factor is sufficiently accurate for most
oils. If other fluids are used or if greater accuracy is
required, the meter can be recalibrated by performing
one of the following recalibration methods.
1. Choosing a Calibration Factor From the
Calibration Table 1
1. Use LEFT or RIGHT ARROW to move the curser to
the Calibration Factor Field (K) .
2. Use the UP or DOWN ARROWS to manually adjust
the Preset Calibration Factor up or down until the
number displayed matches the calibration factor you
have chosen from Table 1.
3. Press center, ENTER button to confirm new setting.
New setting displays in Calibration Factor field (J).
4. Use the RIGHT ARROW button to move curser over
Wrench icon.
5. Press center, ENTER button, to return to Main
Setup Screen.
OR
2. Manual Calibration Procedure
See FIG. 14, page 13 for the Calibrate Screens 1 and 2
referenced in the following instructions.
• When an English unit (gallon, quarts, pints) is
set as the measurement unit on the Units/Limit
Screen, page 12, use a one quart graduated
cylinder to calibrate the meter.
• When a System International Unit (SI) (Liter) is
set as the measurement unit on the Units/Limit
Screen, use a one Liter graduated cylinder to
calibrate the meter.
1. Use the LEFT or RIGHT ARROWS to select START
(L) on Calibrate Screen 1.
2. Press the center, ENTER button to confirm selec-
tion. Calibrate Screen 2 will appear.
3. Dispense exactly 1 quart (or 1 liter) of fluid into a
calibrated 1 quart (or 1 liter) container.
The Calibration Factor will appear on the screen in
the Calibration Factor field (J).
4. Use the LEFT or RIGHT ARROWS to select END
(L) on Calibrate Screen 2.
5. Press the center, ENTER button to confirm selec-
tion. Calibrate Screen 1 will appear.
6. When you have finished making changes, use the
RIGHT ARROW button to move curser over Wrench
icon.
7. Press center, ENTER button, to return to Main
Setup Screen.
To attain the most accurate calibration:
• Use the manual calibration procedure (Method 2).
• Use a certified, graduated cylinder; either 1-Quart or
1-Liter volume in the system of Units to be used
(either English or SI) for real-time dispenses.
• Calibrate using the exact fluid to be dispensed, at
the temperature expected during dispense.
• After calibration check your results by measuring a
dispense.
• Be sure to remove all fluid from graduated cylinder
between calibration attempts.
• Be sure meter is set to proper system of units for
container you are using.
Calibration factors provided in Table 1 (page 13),
are only approximate, but sufficient for most appi-
cations. For the most accuracy, use method 2,
Manual Calibration Procedure.
Once the meter has been calibrated, the unit of
measure can be changed to any other unit, without
the need for recalibration.

Setup Mode Screens
312865A 15
Preset Screen (FIG. 15)
For PE14 Meters Only
Sets the default preset amounts. Typically, you would
enter amounts you most frequently dispense.
PE14 Meters includes D, P and Q
D. Wrench Icon: Returns user to Main Screen.
P. Measurement Unit: Set on Units/Limit Screen,
page 12.
Q. Set Preset Amounts: Sets dispensing options for
the Preset Dispense mode.
Sets up to 3, default preset quantities. The amounts can
be listed in any order and do not have to be in a numeric
order.
To set Preset amounts (FIG. 15):
1. Use the LEFT or RIGHT ARROWS to select one of
the three amount fields (Q).
2. Use the UP or DOWN ARROWS to increase or
decrease the number appearing in the field until the
desired amount displays.
Pressing and holding the ARROW button down,
increases the scrolling speed.
3. Press center, ENTER button to confirm amount.
4. When you have finished making changes, use the
LEFT OR RIGHT ARROW button to move curser
over Wrench icon.
5. Press center, ENTER button, to return to Main
Setup Screen.
FIG. 15
PRESET
4.0 6.0 10
123
QTS
D
QP
ti12188a
The amounts displayed do not automatically recal-
culate quantities when changing between measure-
ment units. For example, if you change from quarts
to liters you must manually change the preset
amounts.

Setup Mode Screens
16 312865A
Banner Screen (FIG. 16)
For ME14 and PE14 Meters
Creates the information banner displayed across top of
the Home, Manual and Preset Screens.
ME14 and PE14 Meters include D and R
D. Wrench Icon: Returns user to Main Screen.
R. Banner field: Information line. Provides space for up
to 11 numeric and alpha characters and/or spaces.
Using the DOWN ARROW button when a field is blank
will also provide the following characters for selection:
(period) “ . “; (forward slash) “ / “; or (dash) “ - ”.
To create a banner:
1. Use the UP or DOWN ARROWS to scroll through
numerals, then alpha characters and then
blank/spaces.
2. Press center, ENTER button to select characters.
When character is selected, curser automatically
moves to the next character field.
Use the RIGHT ARROW button to skip fields or if
you do not need to use all 11 spaces, to return
curser to the Wrench icon. After 11th character has
been confirmed, curser automatically moves to
Wrench icon (D).
3. Use the center, ENTER button, to return to the Main
Setup Screen.
Language Screen (FIG. 17)
For ME14 and PE14 Meters
Sets language preference for text displayed on meter.
Choices include English and Spanish.
Language Screen Features D, S, and T
D. Wrench Icon: Returns user to Main Screen.
S. Language field: Identifies selected language.
T. Language field/button: When curser is over this
button, each time UP or DOWN arrow is pressed, the
name of the Language displayed on the button changes.
To select your language:
1. Use the LEFT or RIGHT ARROW to move the
curser to the Language field/button (T).
2. Use UP or DOWN ARROWS to view language
choices.
3. When the name of your language is displayed on
the field/button (T), press the center ENTER button
to confirm selection.
The selected language is then displayed on screen in
Language field (S).
4. When you have finished making changes, use the
LEFT OR RIGHT ARROW button to move curser
over Wrench icon.
5. Press center, ENTER button, to return to Main
Setup Screen.
FIG. 16
MOBIL15W
-20
D
R
ti12181a
FIG. 17
After you press the center, ENTER button, the name on
the field/button will also change from “SPANISH” to “EPA-
NOL” and will appear this way on both the Language but-
ton (T) and in the Language field (S) on the screen.
ENGLISH
LANGUAGE =
ENGLISH
D
S
T

Dispensing Fluid and Operation Mode Screens
312865A 17
Dispensing Fluid and Operation Mode Screens
If you are in the Setup Mode, to display the Operation Mode Screens use the UP or DOWN ARROW button to select
HOME from the list. Press the Center ENTER button to display the Operation Mode Home Screen shown in FIG. 18.
A complete description of the Setup Mode Screens begins on page 10.
Home Screen (FIG. 18)
ME14 and PE14 Meters
ME14 and PE14 Include A - D
A. Screen Identifier Icon: House displays when user is
on the Home Screen.
B. Operation Screens: List/menu of screen choices
available to user. Operation Screen choices include:
ME14 Meters
•MANUAL
•TOTALS
PE14 Meters
•MANUAL
• PRESET
•TOTALS
1. Use UP or DOWN ARROWS to scroll through
options.
2. Press center, ENTER button to confirm selection
and display the screen.
C. Battery Indicator: Appears only on Home Screen.
Shows current battery charge strength. Appears and
flashes on all screens when battery’s charge is low.
D. Information Banner: Configurable information
banner. Data is input on Setup Mode Banner Screen,
page 16.
FIRE HAZARD: Conductive metal surfaces on the
meter must not make contact with any positively
charged metal surface, including (but not limited to),
the starter solenoid terminal, alternator terminal or
battery terminal. Such contact could cause electrical
arcing and a fire.
FIG. 18
MOBIL15W-20
MANUAL
PRESET
TOTALS
C
A
D
B
PE14 MetersME14 Meters
MOBIL15W-20
MANUAL
TOTALS
B
DA
C
ti12175a ti12184a

Dispensing Fluid and Operation Mode Screens
18 312865A
Manual Dispense (FIG. 19)
ME14 and PE14 Meters
See FIG. 19 for terms D - H
D. Information Banner: Configurable information
banner. Data is input on Setup Mode Banner Screen,
page 16.
E. Unit of Measurement: Identifies unit of
measurement, pint, quart, gallon or liter. Measurement
Unit is set on Setup Mode Units/Limits Screen, page
12.
F. Home Icon: Returns user to Home Screen.
G. Reset: Resets counter (H) to zero.
H. Counter: Before dispense has begun, display reads
0.00. As fluid is dispensed, counts up from zero.
To reset counter to zero after a dispense:
1. Use the LEFT or RIGHT ARROW button to move
curser to RESET (G).
2. Press center ENTER button. The displayed amount
in field H returns to zero.
To dispense in Manual Mode:
If display was asleep, press any button to wake it up.
1. From HOME screen, use the UP or DOWN
ARROWS to select MANUAL (FIG. 20).
2. Press center, ENTER button to confirm selec-
tion.The Manual Dispense screen (FIG. 21) appears.
3. Squeeze trigger to dispense fluid. If using a PE
meter trigger may be locked during the dispense.
Fluid flows. The amount displayed counts up from zero or
from the previously dispensed amount
(FIG. 21) and displays in field H.
4. Release, or for PE meters only, unlock trigger when
you have dispensed the desired amount of fluid.
Fluid flow stops. Amount dispensed displays in field H.
5. Use LEFT or RIGHT ARROW to move curser to
RESET button (G). Press center ENTER button to
reset the displayed amount in field H to zero.
FIG. 19
MOBIL15W-20
10.56 QTS E
F
G
H
D
ti12178a
FIG. 20
FIG. 21
MOBIL15W-20
MANUAL
PRESET
TOTALS
ti12180a
PE14 shown
MOBIL15W-20
10.56 QTS
G
H
ti12178a

Dispensing Fluid and Operation Mode Screens
312865A 19
Preset Dispense (FIG. 22)
PE14 Meters Only
See FIG. 22 for terms D - L
D. Information Banner: Configurable information
banner. Data is input in Setup Mode, Banner Screen,
page 16.
E. Unit of Measurement: Identifies unit of
measurement, pint, quart, gallon or liter. Unit is selected
in Setup Mode, Units/Limit Screen, page 12.
F. Home Icon: Returns user to Home Screen.
G. Reset: Resets counter (H) to zero.
H. Counter: Before dispense has begun, display reads
0.00. As fluid is dispensed, counts up from zero. At the
same time the Progression Bar (J) displays a real-time,
visual representation of the progress made toward
completing the dispense.
J. Progression Bar: Displays a real-time, visual
representation of the progress made toward completing
the dispense. Runs in conjunction with Counter (H).
For example, in FIG. 22, a partial dispense of 2.5 quarts
of the total Preset Dispense of 5.0 quarts is shown. The
Progression Bar shows that approximately half the
dispense is complete.
L. Preset Quantities: Use the UP or DOWN ARROW to
select field L on the task bar.
a. Use the UP or DOWN ARROWS to scroll
through the 3 available Preset Quantities (L).
Each time you press the UP or DOWN
ARROWS, one of the 3 preset amounts dis-
plays on the task bar in field L.
b. When the Preset Quantity you want to dispense
is shown in field L, press the center, ENTER
button to select it.
FIG. 22
MOBIL15W-20
2.50 QTS
4.0
XXXXXXXXXXXXXXXX
E
F
G
H
D
J
L
ti122410a
If a suitable preset quantity is not available, select
an amount that is closest to the amount you want to
dispense. Amounts can be modified following the
procedure described on page 20.

Dispensing Fluid and Operation Mode Screens
20 312865A
To modify the amount selected:
a. Use the UP or DOWN ARROW to scroll up or
down to increase or decrease the amount
shown in field L. Each time you press the UP or
DOWN ARROW the amount will increase or
decrease in increments of 0.1 units.
To speed up the scrolling progression, press
and hold the UP or DOWN ARROW.
b. IMPORTANT!!! When the new dispense
amount is shown in field L, the number
flashes, indicating a change has been made
that requires confirmation.
To confirm new amount you MUST PRESS
the center, ENTER button on the key pad
within 15 seconds. If you begin a dispense
without confirming the new amount, the Preset
Amount will return to the previous confirmed
amount shown in field L.
For example, in FIG. 23 (below) the original, confirmed
Preset Dispense amount (A) was 4.0 qts. The amount
was increased to 8.0 quarts and because it has not
been confirmed, is flashing on the screen (B). After the
user confirmed this change by pressing the center
ENTER button on the key pad, the new dispense
amount no longer flashes and is now set to dispense 8.0
quarts (C).
To permanently modify the default Preset Quantities,
see To set Preset amounts: on page 15.
FIG. 23
MOBIL15W-20
2.50 QTS
4.0
XXXXXXXXXXXXXXXX
8.0
8.0
A
B
C
This manual suits for next models
1
Table of contents
Other Graco Measuring Instrument manuals

Graco
Graco SDMM8 User manual

Graco
Graco XDI12 User manual

Graco
Graco 260415 Operation manual

Graco
Graco 309455C Operation manual

Graco
Graco Pulse TLM User manual

Graco
Graco G250 Parts list manual

Graco
Graco PM8 User manual

Graco
Graco 255348 User manual

Graco
Graco Series B User manual

Graco
Graco 244023 Datasheet

Graco
Graco 215474 User manual

Graco
Graco IniFlex Operation manual

Graco
Graco 121325 Operation manual

Graco
Graco PSM User manual

Graco
Graco Pulse TEK 811 Setup guide

Graco
Graco 24J318 User manual

Graco
Graco HORIZON Operation manual

Graco
Graco G3000 User manual

Graco
Graco EFC C Series Operation manual

Graco
Graco HG6000 Technical reference manual