1. Press the ke to access to configuration.
2. Use the ke to select the desired mode.
3. Press the ke to confirm.
UF-9 Gravity Value setup
Although the balance allows the Standard Calibration, the G value setup provides a
greater accurac . This pre-calibration is performed with the switch CAL in ADJ
position. The value will be set to -00- and it will be replaced each time ou enter a new
value.
Gravit value of Distributor: configurated before the weight calibration.
Gravit value of User: configurated after the weight calibration.
The value of gravit will be denied if is greater than 9.83217 (Polar Gravit ) or less
than 9.78031 (Ecuador Gravit )
Default value: 9.79423
1. Press the ke to access to configuration.
2. The displa will show the last value for 1 second.
3. Use the ke s and to move from a digit to other and the ke to increase
the digit value.
4. Press the ke to confirm
GUARANTEE
This scale is guaranteed for one ear from the deliver date. The guarantee covers an
fabrication defect of the material.
During this period GRAM PRECISION, covers the manpower and the spare parts necessar for
the reparation of the scale.
This guarantee does not cover the failures caused b an inappropriate use or overcharge.
The guarantee does not cover the freight cost (transport) necessary to repair the scale.